CGM profiles without diabetes: mean ~100 mg/dL (5.6 mmol/L), 96% of time at 70–140 mg/dL (3.9–7.8 mmol/L), and excursions are normal

Shelf #7172 · Diabetes, glucose & metabolism

PHYSIOLOGY — ESTABLISHED

Stamped: 2026-11-28.

Evidence published: 2010–2022.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In healthy humans without diabetes, what do continuous glucose monitoring profiles look like — mean glucose, time above 140 mg/dL (7.8 mmol/L), glycaemic variability and the 'glucotype' patterns — by age, BMI and time of day?

Balance of the evidence

PHYSIOLOGY — ESTABLISHED — framed by Guyton & Hall ch. 79 (insulin, glucagon and glucose regulation), and resting on concordant current-generation sensor data: in 153 healthy non-diabetic people aged 7–80, mean glucose was 98–99 mg/dL (5.4–5.5 mmol/L), rising to 104 mg/dL (5.8 mmol/L) over age 60, with a median 96% of time between 70 and 140 mg/dL (3.9–7.8 mmol/L) and a within-person coefficient of variation of 17 ± 3%; 39 healthy children aged 1–6 read almost identically (mean 103 mg/dL (5.7 mmol/L), median 96% in range). Two qualifications are part of the physiology, not footnotes: people normoglycaemic by standard measures still spend time in prediabetic and diabetic sensor ranges (the glucotype finding), and in the 12-week ADAG monitoring 93% of non-diabetic participants exceeded 7.8 mmol/L (140 mg/dL) at some point. All CGM homes in diabetes ask a different (management) question; this shelf is the non-diabetic norm only.

MeSH: Blood Glucose/analysis · Blood Glucose Self-Monitoring/methods · Reference Values · Circadian Rhythm · Healthy Volunteers

Cards (5)

Card 1 — Shah 2019: 153 healthy non-diabetic people aged 7–80 on a Dexcom G6 — mean glucose 98–99 mg/dL (5.4–5.5 mmol/L), median 96% of time at 70–140 mg/dL (3.9–7.8 mmol/L)

Study: Human multicentre prospective CGM normative study, J Clin Endocrinol Metab 2019

Population: 153 non-pregnant, healthy, non-diabetic children and adults (age 7–80 years) with non-obese BMI at 12 centres in the T1D Exchange Clinic Network, each wearing a blinded Dexcom G6 CGM (named as published) with once-daily calibration for up to 10 days, in a multicentre prospective study.

Summary: Shah 2019: 153 healthy non-diabetic people aged 7–80 on a Dexcom G6 — mean glucose 98–99 mg/dL (5.4–5.5 mmol/L), median 96% of time at 70–140 mg/dL (3.9–7.8 mmol/L)

Result: Mean average glucose was 98 to 99 mg/dL (5.4 to 5.5 mmol/L) for all age groups except those over 60 years, in whom it was 104 mg/dL (5.8 mmol/L). Median time between 70 and 140 mg/dL (3.9 to 7.8 mmol/L) was 96% (interquartile range 93–98). Mean within-individual coefficient of variation was 17 ± 3%. Median time above 140 mg/dL (7.8 mmol/L) was 2.1% (30 min/day) and median time below 70 mg/dL (3.9 mmol/L) was 1.1% (15 min/day).

Watch out: Up to 10 days of blinded sensor wear in non-obese volunteers: sensor values are interstitial, not blood, values, and a different sensor generation will not reproduce these figures exactly — the device is named on the face for that reason.

Source: Shah VN et al. Continuous Glucose Monitoring Profiles in Healthy Nondiabetic Participants: A Multicenter Prospective Study. Journal of Clinical Endocrinology and Metabolism. 2019. PMID: 31127824. Registry ID: none in the record. MeSH: Adolescent · Adult · Age Factors · Aged · Benchmarking · Blood Glucose/analysis · Blood Glucose Self-Monitoring/methods · Child · Circadian Rhythm · Female · Healthy Volunteers · Humans · Hyperglycemia/diagnosis · Hypoglycemia/diagnosis · Male · Middle Aged · Monitoring, Physiologic/methods · Prospective Studies · Reference Values · Sex Factors · Young Adult

Card 2 — Hall 2018: 'glucotypes' — people normoglycaemic by standard measures still spent 15% of time in prediabetic and 2% in diabetic sensor ranges (N not documented in the record)

Study: Human CGM phenotyping study (glucotypes), PLoS Biol 2018

Population: Participants phenotyped with continuous glucose monitoring, standardised nutrient challenges, and measures of insulin resistance and secretion, grouped by an analytical framework into patterns of glycaemic response ('glucotypes'). The record's abstract does not state the participant number — N not documented in the record, printed here as required rather than supplied from memory.

Summary: Hall 2018: 'glucotypes' — people normoglycaemic by standard measures still spent 15% of time in prediabetic and 2% in diabetic sensor ranges (N not documented in the record)

Result: Even individuals considered normoglycaemic by standard measures exhibited high glucose variability on CGM, with glucose levels reaching prediabetic ranges 15% of the time and diabetic ranges 2% of the time. Glycaemic responses to identical meals varied widely between individuals and reflected variable underlying physiology (insulin resistance and secretion), which the authors use to argue that glucose dysregulation is more prevalent and heterogeneous than single-time-point measures suggest.

Watch out: The cohort mixes glucose-regulation categories by design — it is a dysregulation-pattern study, not a healthy-normative cohort — so its 15%/2% figures describe what CGM reveals inside 'normoglycaemic' classifications, and must not be read as the healthy population's time-above-range (that is the Shah card's 2.1% median above 140 mg/dL (7.8 mmol/L)).

Source: Hall H et al. Glucotypes reveal new patterns of glucose dysregulation. PLoS Biology. 2018. PMID: 30040822. Registry ID: none in the record. MeSH: Adult · Aged · Blood Glucose/metabolism · Blood Glucose Self-Monitoring · Carbohydrate Metabolism · Cohort Studies · Diabetes Mellitus, Type 2/blood,diagnosis · Female · Homeostasis · Humans · Hyperglycemia/blood,diagnosis · Insulin/metabolism · Internet · Male · Meals · Middle Aged

Card 3 — DuBose 2022: healthy children aged 1–6 years (n = 39) on a Dexcom G6 Pro — mean glucose 103 mg/dL (5.7 mmol/L), median 96% of time at 70–140 mg/dL (3.9–7.8 mmol/L)

Study: Human prospective CGM normative study (young children), J Endocr Soc 2022

Population: 39 healthy, non-diabetic children aged 1–6 years with normal BMI in an institutional prospective observational study, each wearing a blinded Dexcom G6 Pro CGM (named as published) for approximately 10 days.

Summary: DuBose 2022: healthy children aged 1–6 years (n = 39) on a Dexcom G6 Pro — mean glucose 103 mg/dL (5.7 mmol/L), median 96% of time at 70–140 mg/dL (3.9–7.8 mmol/L)

Result: Mean average glucose was 103 mg/dL (5.7 mmol/L). Median time between 70 and 140 mg/dL (3.9–7.8 mmol/L) was 96% (interquartile range 92–97%), mean within-individual coefficient of variation was 17 ± 3%, median time above 140 mg/dL (7.8 mmol/L) was 3.4% (49 min/day), and median time below 70 mg/dL (3.9 mmol/L) was 0.4% (6 min/day) — essentially the adult pattern of the Shah card, in toddlers and pre-schoolers.

Watch out: Thirty-nine children at one institution, from the same research network as the Shah study — a distinct age cohort, not an independent replication in a different population; very young children's sensor wear and calibration differ from adults'.

Source: DuBose SN et al. Continuous Glucose Monitoring Profiles in Healthy, Nondiabetic Young Children. Journal of the Endocrine Society. 2022. PMID: 35506147. Registry ID: none in the record. MeSH: not captured from the record

Card 4 — Borg 2010 (ADAG): over 12 weeks of monitoring, 93% of 80 people without diabetes exceeded 7.8 mmol/L (140 mg/dL) — median 26 min/day above it — and 9% reached diabetic-range values above 11.1 mmol/L (200 mg/dL)

Study: Human intensive-monitoring cohort (ADAG, non-diabetic), Diabetologia 2010

Population: 80 participants without diabetes in the A1C-Derived Average Glucose (ADAG) study, selected on a very low baseline fasting plasma glucose with normal HbA1c, completing an intensive glucose monitoring period of 12 weeks, with time above plasma-glucose thresholds, postprandial indices and variability derived from the monitoring.

Summary: Borg 2010 (ADAG): over 12 weeks of monitoring, 93% of 80 people without diabetes exceeded 7.8 mmol/L (140 mg/dL) — median 26 min/day above it — and 9% reached diabetic-range values above 11.1 mmol/L (200 mg/dL)

Result: 93% of participants reached glucose concentrations above the impaired-glucose-tolerance threshold of 7.8 mmol/L (140 mg/dL) and spent a median of 26 min/day above this level during continuous monitoring. Eight individuals (10%) spent more than 2 hours in the IGT range — they had higher HbA1c, fasting glucose, age and BMI — and seven participants (9%) reached concentrations above 11.1 mmol/L (200 mg/dL). Brief excursions into 'prediabetic' ranges are therefore part of monitored non-diabetic life, not proof of disease.

Watch out: An older monitoring era and a cohort deliberately selected for very low fasting glucose — if anything that selection biases excursions downward, yet excursions were near-universal; the 12-week window also captures far more excursions than a 10-day sensor wear, so its percentages are not directly comparable with the Shah/DuBose medians.

Source: Borg R et al. Real-life glycaemic profiles in non-diabetic individuals with low fasting glucose and normal HbA1c: the A1C-Derived Average Glucose (ADAG) study. Diabetologia. 2010. PMID: 20396998. Registry ID: none in the record. MeSH: Adult · Blood Glucose/analysis,metabolism · Fasting/blood · Female · Glycated Hemoglobin/analysis,metabolism · Humans · Male · Middle Aged · Monitoring, Ambulatory · Reference Values

Card 5 — Synthesis: without diabetes, glucose sits at a mean of about 100 mg/dL (5.6 mmol/L) with ~96% of time at 70–140 mg/dL (3.9–7.8 mmol/L) — and brief excursions above range are normal

Study: Synthesis

Population: Synthesis of the four carded human records (153 people aged 7–80 and 39 children aged 1–6 on current-generation Dexcom sensors; the glucotype phenotyping cohort, N not documented in its record; and 80 ADAG participants monitored for 12 weeks), framed by Guyton & Hall ch. 79 on insulin, glucagon and glucose regulation.

Summary: Synthesis: without diabetes, glucose sits at a mean of about 100 mg/dL (5.6 mmol/L) with ~96% of time at 70–140 mg/dL (3.9–7.8 mmol/L) — and brief excursions above range are normal

Result: The concordant norm: mean sensor glucose 98–104 mg/dL (5.4–5.8 mmol/L) across the lifespan (slightly higher over 60), a median 96% of time between 70 and 140 mg/dL (3.9–7.8 mmol/L), within-person variability (CV) of about 17%, and only minutes per day below 70 mg/dL (3.9 mmol/L). The equally concordant qualification: excursions are universal — a median 2.1% of time above 140 mg/dL (7.8 mmol/L) over 10 days, and over 12 weeks 93% of non-diabetic people exceed 7.8 mmol/L (140 mg/dL) and 9% exceed 11.1 mmol/L (200 mg/dL). 'Normal' CGM is therefore a distribution with tails, and the glucotype work shows the tails' patterns differ by underlying physiology even when standard measures call everyone normoglycaemic.

Watch out: Sensor generation moves values: the two Dexcom G6 cohorts anchor the current-generation norm, the ADAG limb is an earlier monitoring era, and no carded record here stratifies the healthy norm by BMI category beyond the non-obese/normal-BMI inclusion of the Dexcom cohorts.

Source: Synthesis of Shah VN et al. Journal of Clinical Endocrinology and Metabolism. 2019 (PMID: 31127824); Hall H et al. PLoS Biology. 2018 (PMID: 30040822); DuBose SN et al. Journal of the Endocrine Society. 2022 (PMID: 35506147); Borg R et al. Diabetologia. 2010 (PMID: 20396998), all carded above, with Guyton & Hall textbook framing. Registry ID: none in the record. MeSH: not captured from the record