Corneal sensitivity: falls with age on the filament, instrument-defined thresholds, and a discordant air-pulse finding in women
Shelf #7177 · Eyes, skin, nose & allergy
PHYSIOLOGY — ESTABLISHED
Stamped: 2026-11-28.
Evidence published: 1977–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In healthy humans, how sensitive is the cornea — corneal sensitivity thresholds by Cochet–Bonnet and non-contact air-pulse esthesiometry across the cornea and conjunctiva — and how much does sensitivity fall with age, contact-lens wear and time of day?
Balance of the evidence
PHYSIOLOGY — ESTABLISHED — framed by Ganong (somatosensory / special senses chapters; thin framing for corneal esthesiometry specifically — this shelf rests chiefly on the primary measurement literature, as its slate row directs), and resting on three human studies whose core finding is concordant for the filament method: corneal touch threshold rises with age — the 1977 classic (205 healthy people) found thresholds increasing gradually throughout life, more significantly after the fifth decade, and a 2025 Cochet–Bonnet cohort (324 volunteers) found central sensitivity falling with age with correlations of about −0.78 across all quadrants, accelerating in the 50s and 60s. The air-pulse method (49 subjects, Belmonte esthesiometer) is carded per instrument and does not merge: its corneal threshold averaged 76.2 ± 26.8 mL/min, its conjunctival threshold 123.7 ± 49.1 mL/min, and its age pattern differed by sex. Cross-point: cenegermin for neurotrophic keratitis (pack #4252) is the disease state this measurement defines — not carded here.
MeSH: Cornea/physiology · Differential Threshold · Sensory Thresholds · Age Factors · Sex Factors · Reference Values
Cards (4)
Card 1 — Millodot 1977: in 205 healthy people, corneal touch threshold rose gradually throughout life, more significantly after the fifth decade
Study: Human classic psychophysical study, Invest Ophthalmol Vis Sci 1977
Population: 205 healthy people of different ages, with corneal touch threshold measured (the classic touch-threshold study); the results were compared with Boberg-Ans's earlier work and with a report claiming sensitivity rose to the fifth decade and then fell.
Summary: Millodot 1977: in 205 healthy people, corneal touch threshold rose gradually throughout life, more significantly after the fifth decade
Result: Corneal touch threshold increased gradually throughout life, although more significantly after the fifth decade of life. The results accorded with Boberg-Ans, refuting the earlier report that corneal sensitivity (threshold⁻¹) increased up to the fifth decade and declined thereafter. The record's abstract is short and gives no numerical thresholds per decade — none are printed here — so the verified content is the direction, the timing (acceleration after 50) and the cohort size.
Watch out: A short 1977 abstract with no per-decade numbers and no named instrument in the record — it anchors the direction of the age effect, and the quantitative values on this shelf come from the 2025 and 2008 cards.
Source: Millodot M. The influence of age on the sensitivity of the cornea. Investigative Ophthalmology & Visual Science. 1977. PMID: 844979. Registry ID: none in the record (pre-registration era). MeSH: Adolescent · Adult · Aged · Aging · Child · Cornea/physiology · Differential Threshold · Humans · Middle Aged · Touch
Card 2 — Eratilgan 2025: in 324 healthy Turkish volunteers, Cochet–Bonnet central corneal sensitivity averaged 50.15 ± 9.7 mm and fell with age (r ≈ −0.78), accelerating in the 5th and 6th decades
Study: Human normative study (Cochet–Bonnet), Clin Exp Optom 2025
Population: 324 healthy volunteers aged 20–84 years (right eyes; mean age 44.31 ± 15.65 years; 165 women, 159 men) attending for routine examination in Turkey, excluding any ocular disease other than refractive error, systemic disease, contact-lens use, chronic ocular or systemic drug use and previous ocular surgery; central and four-quadrant corneal sensitivity measured with a Cochet–Bonnet esthesiometer (Luneau Ophthalmology, France, named as published), in five age groups (20–29, 30–39, 40–49, 50–59, ≥60).
Summary: Eratilgan 2025: in 324 healthy Turkish volunteers, Cochet–Bonnet central corneal sensitivity averaged 50.15 ± 9.7 mm and fell with age (r ≈ −0.78), accelerating in the 5th and 6th decades
Result: Mean central, superior, inferior, nasal and temporal sensitivities were 50.15 ± 9.7, 48.62 ± 10.3, 48.59 ± 10.3, 47.73 ± 10.2 and 47.71 ± 10.3 mm respectively (filament length — longer means a lighter touch is felt). Sensitivities in all quadrants decreased significantly with age (P = 0.001 for all; correlation coefficients r = −0.783, −0.791, −0.789, −0.760, −0.775), with the decrease accelerating in groups 4 and 5 (50–59 and ≥60 years). Gender differences were significant in central and temporal quadrants.
Watch out: One Turkish population and one filament instrument, routine-examination attenders rather than a random sample — its values are the Cochet–Bonnet norm for that population, and the authors note geography, gender and ethnicity may affect values.
Source: Eratilgan NF et al. Alterations in corneal sensitivity according to age and gender in healthy population in Turkey. Clinical & Experimental Optometry. 2025. PMID: 39586826. Registry ID: none in the record. MeSH: Humans · Male · Female · Adult · Turkey/epidemiology · Middle Aged · Cornea/physiology · Aged · Young Adult · Aged, 80 and over · Healthy Volunteers · Sex Factors · Age Factors · Reference Values · Aging/physiology
Card 3 — Golebiowski 2008: with a Belmonte air-pulse esthesiometer in 49 normal subjects, corneal threshold averaged 76.2 ± 26.8 mL/min and conjunctival 123.7 ± 49.1 mL/min — thresholds were higher in men, and the age pattern differed from the filament studies
Study: Human retrospective esthesiometry study (Belmonte air-pulse), Optom Vis Sci 2008
Population: A retrospective analysis of ocular surface threshold measurements made by one examiner using the CRCERT–Belmonte esthesiometer (named as published) — multiple corneal measurements for 49 normal subjects (24 men, 25 women) and conjunctival measurements for 33 subjects (16 men, 17 women); thresholds at the corneal apex and at the inferior conjunctiva 2 mm from the limbus, measured between 09:00 and 18:00 at 20–26 °C and 52–87% humidity.
Summary: Golebiowski 2008: with a Belmonte air-pulse esthesiometer in 49 normal subjects, corneal threshold averaged 76.2 ± 26.8 mL/min and conjunctival 123.7 ± 49.1 mL/min — thresholds were higher in men, and the age pattern differed from the filament studies
Result: Mean corneal threshold was 76.2 ± 26.8 mL/min and conjunctival threshold 123.7 ± 49.1 mL/min (n = 33, P < 0.001) — the cornea is the more sensitive tissue, and the two thresholds were well correlated (r = 0.66). Thresholds were significantly higher for men than women at both cornea (82.2 ± 23.5 vs 67.6 ± 24.1 mL/min, P = 0.04) and conjunctiva (144.1 ± 40.7 vs 105.8 ± 50.2, P = 0.02). A significant reduction in corneal threshold with age was apparent for females (n = 25, r = −0.49, P = 0.01) — i.e. thresholds fell, meaning sensitivity rose with age in women on this instrument — but not males; no effect of time of day, ambient humidity or temperature was found at either site.
Watch out: Airflow thresholds in mL/min cannot be compared numerically with filament lengths in mm — the two instruments are carded separately for that reason. The female age finding runs opposite to the filament studies and rests on 25 women in a retrospective analysis; it is printed as a discordance, not smoothed over.
Source: Golebiowski B et al. Factors affecting corneal and conjunctival sensitivity measurement. Optometry and Vision Science. 2008. PMID: 18382334. Registry ID: none in the record. MeSH: Adult · Conjunctiva/physiology · Cornea/physiology · Female · Humans · Male · Middle Aged · Reference Values · Retrospective Studies · Sensitivity and Specificity · Sensory Thresholds · Sex Characteristics
Card 4 — Synthesis: corneal sensitivity falls with age on the filament — but the instrument defines the threshold, and the air-pulse record disagrees on the age direction in women
Study: Synthesis
Population: Synthesis of the three carded human records (the 1977 classic of 205 healthy people, the 2025 Cochet–Bonnet cohort of 324 volunteers, and the 2008 Belmonte air-pulse analysis of 49 normal subjects), framed by Ganong on somatosensory and special senses — thin framing for esthesiometry, as the row states.
Summary: Synthesis: corneal sensitivity falls with age on the filament — but the instrument defines the threshold, and the air-pulse record disagrees on the age direction in women
Result: By Cochet–Bonnet filament, the story is concordant across 48 years: touch threshold rises (sensitivity falls) gradually through life and faster after the fifth decade, with central sensitivity about 50 mm of filament in the modern cohort and a strong age correlation (r ≈ −0.78) in every quadrant, and small gender differences in central and temporal quadrants. By air-pulse esthesiometry the quantities are different in kind — corneal threshold about 76 mL/min of airflow, conjunctival about 124 mL/min — men need more airflow than women at both sites, time of day and ambient conditions do not move thresholds, and in women the threshold falls with age, i.e. sensitivity appears to rise. That last finding is a genuine instrument-and-sample discordance on the face of the shelf, not a synthesis error: filament and air-pulse thresholds are not interchangeable, and no single 'corneal sensitivity' number is printed without its instrument.
Watch out: No carded record here quantifies contact-lens wear's effect on sensitivity (the 2025 cohort excluded lens wearers), so the row's lens-wear limb is unquantified on this shelf and is stated as such rather than filled from the textbook.
Source: Synthesis of Millodot M. Investigative Ophthalmology & Visual Science. 1977 (PMID: 844979); Eratilgan NF et al. Clinical & Experimental Optometry. 2025 (PMID: 39586826); Golebiowski B et al. Optometry and Vision Science. 2008 (PMID: 18382334), all carded above, with Ganong textbook framing. Registry ID: none in the record. MeSH: not captured from the record