DPOAE norms by device: ILO92 database (101 ears), emission thresholds within ±13 dB at f2/f1 = 1.21, and Titan vs Eclipse agreeing on amplitude but not on SNR
Shelf #7249 · Eyes, skin, nose & allergy
PHYSIOLOGY — EMERGING
Stamped: 2026-11-28.
Evidence published: 1996–2024.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In normal-hearing adults, what are distortion-product otoacoustic emission (DPOAE) levels and signal-to-noise ratios by frequency (the 2f1-f2 DP-gram) — their means, lower reference limits and test-retest behaviour — and how do device, primary-tone levels, sex and age shift them?
Balance of the evidence
Textbook framing: Ganong, hearing (cochlear mechanics) — the framing is brief, and this shelf rests chiefly on the primary test-norming literature, and says so. Three adult normative datasets span three device generations, carded PER DEVICE / PROTOCOL as the measurement demands: the 1996 ILO92 database (101 normal ears), a 2007 research protocol deriving an emission-threshold reference standard in 20 excellent-hearing young adults (optimal at f2/f1 = 1.21, 1-6 kHz, 95% within ±13 dB), and a 2024 head-to-head of Titan vs Eclipse in 63 normal-hearing adults (amplitudes agree; noise and SNR differ; both pass ≥6 dB SNR across frequencies, with unreliable responses at the edge frequencies). DPOAE dissociates from the pure-tone audiogram — that dissociation, against the built pure-tone threshold shelf, is core content; the TEOAE companion is shelf #7250. The stamp is graded down because the total adult base is small and every number is device-bound. PHYSIOLOGY — EMERGING.
MeSH: Acoustic Stimulation · Adult · Audiometry, Pure-Tone · Auditory Threshold · Cochlea/physiology · Electric Stimulation · Female · Humans · Male · Random Allocation · Adolescent · Audiometry/instrumentation,methods · Hearing Disorders/diagnosis · Hearing Loss, Sensorineural/diagnosis · Hearing Tests/instrumentation,methods · Otoacoustic Emissions, Spontaneous/physiology · Pressure · Reference Standards · Reproducibility of Results · Research Design · Sound
Cards (4)
Card 1 — First normative database on the ILO92 analyser: DP-grams and DP-growth functions from 101 normal ears of 101 healthy young adults (Vinck 1996)
Study: Primary human physiology study (DPOAE normative database, ILO92 analyser)
Population: 101 normal ears of 101 healthy young adults; distortion-product otoacoustic emissions (2f1-f2) measured automatically on the commercially available Otodynamics ILO92 analyser using two data-collection protocols, in the form of DP-grams and DP-growth (input/output) functions, statistically processed into a normative database.
Summary: First normative database on the ILO92 analyser: DP-grams and DP-growth functions from 101 normal ears of 101 healthy young adults (Vinck 1996)
Result: The study's product is the normative database itself: at the time, the ILO92 let centres perform DPOAE measurements in different populations but with no relationship to normative data, and these 101-ear data were assembled to serve as the basis for research on DPOAE testing in ear pathology. The fetched record states the database's construction and scope; it does not print numeric DP-gram levels, and this card claims none.
Watch out: Device- and protocol-bound by construction: ILO92 DP-gram norms from 1996 are the first-generation reference, not a universal DPOAE limit — later cards use different stimulus rules and devices.
Source: Vinck BM et al. Distortion product otoacoustic emissions: a normative study. Audiology : official organ of the International Society of Audiology. 1996. PMID: 8937656. Registry ID: none in the record. MeSH: Acoustic Stimulation · Adult · Audiometry, Pure-Tone · Auditory Threshold · Cochlea/physiology · Electric Stimulation · Female · Humans · Male · Random Allocation
Card 2 — Emission-threshold reference standard (20 young adults, 40 ears): the optimal diagnostic measure is the DPOAE threshold at f2/f1 = 1.21 for f2 1-6 kHz, 95% within ±13 dB of the mean (Mills 2007)
Study: Primary human physiology study (normative DPOAE emission thresholds, research protocol)
Population: 20 young adults aged 18-24 years (40 ears; equal men and women) with excellent hearing — thresholds of 10 dB HL or better at 0.5-8 kHz and tympanometric peak within ±30 daPa of ambient in both ears. Two primary tones varied in 5-dB steps with L1 always 10 dB above L2; emission isoresponse threshold defined as the stimulus level giving an emission amplitude of -10 dB SPL; f2 at audiometric frequencies 1-8 kHz at ratios f2/f1 = 1.21 and 1.28.
Summary: Emission-threshold reference standard (20 young adults, 40 ears): the optimal diagnostic measure is the DPOAE threshold at f2/f1 = 1.21 for f2 1-6 kHz, 95% within ±13 dB of the mean (Mills 2007)
Result: Reference standards for emission threshold level (an HL-like scale for emissions) were derived. For diagnosis, the optimal measure was the emission threshold at ratio 1.21 for f2 frequencies 1-6 kHz; the 8 kHz emission had a relatively high mean threshold and seemed less useful. Variability was adequately small: 95% of emission thresholds fell within ±13 dB of the mean at each frequency. Even within the 10 dB HL inclusion limit, emission thresholds rose slightly with auditory threshold — importantly only at 8 kHz, strongest in men.
Watch out: A deliberately super-normal reference group (≤10 dB HL) of only 20 subjects: the ±13 dB band describes excellent ears under this exact L1/L2 rule, and the authors state the specific numerical results for distinguishing cochlear disorders remain to be firmly established.
Source: Mills DM et al. Evaluation of cochlear hearing disorders: normative distortion product otoacoustic emission measurements. Ear and hearing. 2007. PMID: 17982366. Registry ID: none in the record. MeSH: Adolescent · Adult · Audiometry/instrumentation,methods · Female · Hearing Disorders/diagnosis · Hearing Loss, Sensorineural/diagnosis · Hearing Tests/instrumentation,methods · Humans · Male · Otoacoustic Emissions, Spontaneous/physiology · Pressure · Reference Standards · Reproducibility of Results · Research Design · Sound
Card 3 — Two current clinical devices compared head-to-head (63 normal-hearing adults): Titan and Eclipse give the same DPOAE amplitudes but different noise and SNR — both pass the ≥6 dB SNR criterion across frequencies (Cañete 2024)
Study: Primary human physiology study (per-device DPOAE normative comparison, Titan vs Eclipse)
Population: 63 participants with normal hearing (mean age 23.2, SD 2.0 years); DPOAEs measured on two commercial Interacoustics clinical devices, Titan and Eclipse, at 31 frequencies from 500 to 10,000 Hz; signal and noise amplitudes and signal-to-noise ratios (SNRs) analysed. This record carries no MeSH terms (see source line).
Summary: Two current clinical devices compared head-to-head (63 normal-hearing adults): Titan and Eclipse give the same DPOAE amplitudes but different noise and SNR — both pass the ≥6 dB SNR criterion across frequencies (Cañete 2024)
Result: DPOAE amplitude differed across frequencies, with a high number of unreliable responses at the edge frequencies (<828 Hz and >6,072 Hz). There were no significant differences between the two devices in DPOAE amplitudes, but recorded noise levels differed significantly, and SNRs — frequency-dependent, larger at higher frequencies — differed significantly between devices. Despite this, OAEs from both devices met the pass criterion (SNR ≥ 6 dB) consistently across frequencies, so the differences do not compromise test outcomes.
Watch out: The device lesson is in the noise floor, not the signal: SNR-based pass/fail is device-dependent even when amplitudes agree, and edge frequencies (<828 Hz, >6,072 Hz) yield many unreliable responses in normal ears.
Source: Cañete OM et al. Comparison of Two Clinical Devices for the Measurement of Distortion Product Otoacoustic Emissions in Normal-Hearing Adults. Journal of audiology & otology. 2024. PMID: 38382519. Registry ID: none in the record. MeSH: none in the record
Card 4 — Synthesis: DPOAE norms are real but device- and protocol-specific — three adult datasets across three device generations agree on the DP-gram's shape and a ≥6 dB SNR pass, on a small total base (101 ears; 20 adults; 63 adults)
Study: Synthesis
Population: Synthesis of the three primary reports carded above.
Summary: Synthesis: DPOAE norms are real but device- and protocol-specific — three adult datasets across three device generations agree on the DP-gram's shape and a ≥6 dB SNR pass, on a small total base (101 ears; 20 adults; 63 adults)
Result: Framed by Ganong (hearing — cochlear mechanics; the framing is brief, and this shelf rests chiefly on the primary test-norming literature). The first-generation ILO92 normative database covers 101 normal ears (Vinck 1996); the research-protocol reference standard defines an emission threshold (the level giving -10 dB SPL) in 20 excellent-hearing young adults, optimal at f2/f1 = 1.21 for 1-6 kHz with 95% of thresholds within ±13 dB of the mean (Mills 2007); and the current-device comparison in 63 adults shows Titan and Eclipse agree on amplitude but differ on noise and SNR while both meet the ≥6 dB SNR pass across frequencies (Cañete 2024). DPOAE is an objective cochlear measure that dissociates from the audiogram — the built pure-tone hearing-threshold physiology shelf is the cross-point, not re-carded here; the transient-evoked companion shelf is #7250. PHYSIOLOGY — EMERGING: the adult normative base is thin (no dataset exceeds 101 ears) and every value is bound to its device and primary-tone protocol, as printed on each card.
Watch out: Never present one device's DP-gram limits as universal: primary-tone levels (L1/L2) and ratio dominate the numbers, and 8 kHz emissions behave differently from the 1-6 kHz range.
Source: Synthesis of the reports carded above; no additional source.