ROX index for HFNC failure: AUROC 0.68→0.76 over 12 h in pneumonia (≥4.88 = success), but only 0.57–0.61 at 2 h in COVID-19 validations

Shelf #7544 · Critical care & sepsis

PREDICTIVE PERFORMANCE — modest and time-dependent: ROX AUROC rose from 0.679 (2 h) to 0.759 (12 h) at derivation in pneumonia with time-point-specific cut-offs (≥4.88 success; <2.85 / <3.47 / <3.85 failure), but COVID-19 validations found weak 2-hour discrimination (0.57 in Japan; 0.612 in Thai EDs, below the plain SpO2/FiO2 ratio's 0.651); a prognostic-accuracy stamp, not a treatment stamp

Stamped: 2026-11-28.

Evidence published: 2019–2023.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults with acute hypoxaemic respiratory failure treated with high-flow nasal cannula (HFNC), how well does the ROX index (oxygen saturation / inspired oxygen fraction divided by respiratory rate), measured at 2, 6 and 12 hours, predict HFNC failure (intubation) — what cut-offs separate success from failure at each time point, and does the index hold up in external validation, including in COVID-19 pneumonia and emergency-department cohorts?

Balance of the evidence

PREDICTIVE PERFORMANCE — time-point- and population-dependent. In the pre-COVID pneumonia derivation/validation cohort, the ROX index (SpO2/FiO2 divided by respiratory rate) discriminated modestly and better with time: AUROC 0.679 at 2 hours, 0.703 at 6 hours and 0.759 at 12 hours, with ROX ≥4.88 at any of those points associated with a lower risk of intubation and failure cut-offs of <2.85, <3.47 and <3.85 respectively. The two COVID-19 validations are materially weaker early: in 300 Japanese patients ROX AUROC was 0.57 at 2 hours, rising to 0.75 only at 48 hours; and in 173 Thai emergency-department patients the plain SpO2/FiO2 ratio (AUROC 0.651) discriminated better than ROX (0.612) at 2 hours. Cut-offs are time-point-specific and are never a single number. This is a prognostic-accuracy shelf: no randomised ROX-guided strategy trial is carded, and no card claims that acting on a ROX threshold improves outcomes. Two of the three reports are COVID-19 cohorts; the pre-COVID pneumonia cohort is the generalisable limb.

MeSH: Aged · Blood Gas Analysis · Catheterization · Cohort Studies · Data Accuracy · Diagnostic Techniques and Procedures · Extracorporeal Membrane Oxygenation · Female · Humans · Male · Middle Aged · Noninvasive Ventilation · Oxygen Inhalation Therapy · Pneumonia · Practice Guidelines as Topic · Prospective Studies · Respiratory Rate · Adult · Oxygen · Cannula · Oxygen Saturation · Retrospective Studies · COVID-19 · Respiratory Insufficiency

Cards (4)

Card 1 — Derivation/validation (Roca 2019): ROX discrimination rose with time — AUROC 0.679 at 2 h, 0.703 at 6 h, 0.759 at 12 h; ROX ≥4.88 at any point marked lower intubation risk, with failure cut-offs <2.85, <3.47 and <3.85

Study: Multicentre prospective cohort — derivation and validation of the ROX index (pneumonia, pre-COVID)

Population: Multicentre prospective observational cohort of patients with pneumonia and acute hypoxaemic respiratory failure treated with high-flow nasal cannula (HFNC) over 2 years; in the 191-patient validation cohort, 68 patients (35.6%) required intubation.

Summary: Derivation/validation (Roca 2019): ROX discrimination rose with time — AUROC 0.679 at 2 h, 0.703 at 6 h, 0.759 at 12 h; ROX ≥4.88 at any point marked lower intubation risk, with failure cut-offs <2.85, <3.47 and <3.85

Result: The prediction accuracy of the ROX index (oxygen saturation by pulse oximetry / FiO2, divided by respiratory rate) increased over time: AUROC 0.679 at 2 hours, 0.703 at 6 hours and 0.759 at 12 hours after HFNC initiation. ROX ≥4.88 measured at 2, 6 or 12 hours was consistently associated with a lower risk of intubation (hazard ratios 0.434, 95% CI 0.264 to 0.715; 0.304, 95% CI 0.182 to 0.509; and 0.291, 95% CI 0.161 to 0.524 respectively — relative measures, printed beside the cohort's 35.6% intubation rate). ROX below 2.85 at 2 hours, below 3.47 at 6 hours and below 3.85 at 12 hours predicted HFNC failure. Of the index's components, the SpO2/FiO2 ratio carried greater weight than respiratory rate.

Watch out: A pneumonia cohort from before COVID-19: its early (2-hour) discrimination was already modest (AUROC 0.679), and the COVID-19 validations below are weaker still at 2 hours — the ≥4.88 threshold is a 2/6/12-hour rule, not a single early reading.

Source: Roca O et al. An Index Combining Respiratory Rate and Oxygenation to Predict Outcome of Nasal High-Flow Therapy. American journal of respiratory and critical care medicine. 2019. PMID: 30576221. Registry ID: NCT02845128. MeSH: Aged · Blood Gas Analysis · Catheterization · Cohort Studies · Data Accuracy · Diagnostic Techniques and Procedures · Extracorporeal Membrane Oxygenation · Female · Humans · Male · Middle Aged · Noninvasive Ventilation · Oxygen Inhalation Therapy · Pneumonia · Practice Guidelines as Topic · Prospective Studies · Respiratory Rate

Card 2 — COVID-19 external validation, Japan (Okano 2024): ROX AUROC only 0.57 at 2 hours in 300 patients, rising to 0.75 at 48 hours; at the published <2.85 cut-off, 2-hour sensitivity was 3% with 100% specificity

Study: Multicentre retrospective external validation (COVID-19 pneumonia, Japan) — ROX index and HACOR score

Population: 300 adults (median age 60 years) with COVID-19 pneumonia treated with HFNC at nine Japanese institutions, January 2020 to March 2022 (retrospective multicentre study); treatment failure — intubation or death within 7 days — occurred in 127 patients (42%). The comparator HACOR score was validated alongside ROX. This record carries no MeSH terms (see source line).

Summary: COVID-19 external validation, Japan (Okano 2024): ROX AUROC only 0.57 at 2 hours in 300 patients, rising to 0.75 at 48 hours; at the published <2.85 cut-off, 2-hour sensitivity was 3% with 100% specificity

Result: At the primary 2-hour time point, ROX AUROC was 0.57 and HACOR 0.63 (P = 0.24) — the authors judged 2-hour predictive performance potentially inadequate. ROX AUROC rose with time: 0.62 at 6 hours, 0.68 at 12 hours, 0.69 at 24 hours and 0.75 at 48 hours (HACOR 0.58, 0.70, 0.68 and 0.75 at the same points). At the previously reported 2-hour cut-offs, sensitivity and specificity were 3% and 100% for ROX <2.85, and 18% and 91% for HACOR >5. Visual calibration assessment found the HACOR score well calibrated but not the ROX index.

Watch out: Retrospective, COVID-19-only, and its failure outcome includes death within 7 days as well as intubation — a broader outcome than the derivation cohort's intubation endpoint; the 2-hour ROX performance here (0.57) is close to chance.

Source: Okano H et al. External validation of the HACOR score and ROX index for predicting treatment failure in patients with coronavirus disease 2019 pneumonia managed on high-flow nasal cannula therapy: a multicenter retrospective observational study in Japan. Journal of intensive care. 2024. PMID: 38360681. Registry ID: UMIN000050024. MeSH: none in the record

Card 3 — COVID-19 emergency-department validation, Thailand (Ruangsomboon 2023): at 2 hours the plain SpO2/FiO2 ratio (AUROC 0.651) discriminated better than ROX (0.612) and modified ROX (0.606) in 173 patients

Study: Multicentre retrospective validation (COVID-19, emergency department) — SF ratio vs ROX vs modified ROX

Population: 173 adults with COVID-19 treated with HFNC in five Thai emergency departments, January to December 2021 (multicentre retrospective study); HFNC success — no mechanical ventilation required at HFNC termination — occurred in 55 patients (31.8%). The SpO2/FiO2 (SF) ratio, the ROX index and the modified ROX index (ROX divided by heart rate) were recorded at 0 and 2 hours.

Summary: COVID-19 emergency-department validation, Thailand (Ruangsomboon 2023): at 2 hours the plain SpO2/FiO2 ratio (AUROC 0.651) discriminated better than ROX (0.612) and modified ROX (0.606) in 173 patients

Result: The 2-hour SF ratio had the highest discrimination (AUROC 0.651, 95% CI 0.558 to 0.744), ahead of the 2-hour ROX index (0.612) and modified ROX index (0.606), and also had the best calibration and overall model performance. At its optimal cut-point of 128.19, the SF ratio gave a sensitivity of 65.3% and specificity of 61.8%; an SF ratio ≥128.19 at 2 hours was independently associated with HFNC success (adjusted odds ratio 0.29 for failure, 95% CI 0.13 to 0.65; P = 0.003 — a relative measure).

Watch out: Retrospective ED cohort in COVID-19 only, with success defined at HFNC termination rather than a fixed horizon: it questions ROX's added value over its own SpO2/FiO2 component in this setting, not ROX's performance in the derivation population.

Source: Ruangsomboon O et al. Ratio of Oxygen Saturation to Inspired Oxygen, ROX Index, Modified ROX Index to Predict High Flow Cannula Success in COVID-19 Patients: Multicenter Validation Study. The western journal of emergency medicine. 2023. PMID: 37278775. Registry ID: none in the record. MeSH: Adult · Humans · Oxygen · Cannula · Oxygen Saturation · Retrospective Studies · Respiratory Rate · COVID-19 · Oxygen Inhalation Therapy · Respiratory Insufficiency

Estimate type: adjusted estimate, as stated in the card.

Card 4 — Synthesis: ROX predicts HFNC failure modestly and increasingly with time in pneumonia (AUROC 0.759 by 12 h), but its early discrimination in COVID-19 validations was weak (0.57–0.61 at 2 h) — and at 2 hours in the ED its own SpO2/FiO2 component did better

Study: Synthesis

Population: All reports carded on this shelf.

Summary: Synthesis: ROX predicts HFNC failure modestly and increasingly with time in pneumonia (AUROC 0.759 by 12 h), but its early discrimination in COVID-19 validations was weak (0.57–0.61 at 2 h) — and at 2 hours in the ED its own SpO2/FiO2 component did better

Result: The derivation cohort (Roca 2019, pneumonia, pre-COVID) set time-point-specific cut-offs — success at ROX ≥4.88, failure below 2.85 / 3.47 / 3.85 at 2 / 6 / 12 hours — with discrimination rising from 0.679 to 0.759 over 12 hours. Both COVID-19 external validations found materially weaker early performance: AUROC 0.57 at 2 hours rising to 0.75 at 48 hours in Japan (Okano 2024), and 0.612 at 2 hours in Thai emergency departments, where the simpler SpO2/FiO2 ratio reached 0.651 (Ruangsomboon 2023). The honest reading: ROX is a serial, time-point-specific index whose later readings carry the information; a single 2-hour value — especially in COVID-19 — discriminates weakly. No carded report tests a ROX-guided management strategy.

Watch out: See the individual cards for each cohort's limits; this synthesis adds no new data.

Source: Synthesis of the reports carded above; no additional source.