Nasal nitric oxide reference values by technique: healthy controls 304.6 ± 118.8 nl/min (PCD cut-off 77 nl/min) vs 426.76 ± 143.27 ppb by tidal breathing — never pooled
Shelf #7691 · Lungs, breathing & sleep
PHYSIOLOGY — EMERGING — nasal nitric oxide reference values rest on three independent datasets that agree healthy values sit far above the primary-ciliary-dyskinesia cut-off (healthy controls 304.6 ± 118.8 nl/min vs cut-off 77 nl/min), but the values are technique-bound — nl/min velum-closure aspiration and ppb tidal-breathing figures are different quantities, never converted or pooled — the one dedicated derivation prints no numeric values in its record, and the tidal-breathing dataset is small (N = 32) with a wide interval
Stamped: 2026-11-28.
Evidence published: 2013–2017.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In healthy adults, what are the reference values for nasal nitric oxide — the mean, spread and upper range by measurement technique (breath-hold / velum-closure sampling in nl/min vs tidal-breathing NIOX MINO measurement in ppb) — which subject factors (sex, age, body size, lung function, blood eosinophils) move the value, and how far below the healthy range does the primary-ciliary-dyskinesia cut-off (77 nl/min) sit?
Balance of the evidence
Textbook framing: Guyton & Hall and Boron & Boulpaep have no nasal-nitric-oxide reference chapter — this shelf rests on the primary reference literature alone, and says so. Three independent datasets, carded per technique because technique is destiny in this literature: velum-closure aspiration values in nl/min and tidal-breathing NIOX MINO values in ppb are different quantities and are never converted or pooled. The dedicated reference study (Antosova 2017; 141 healthy adults, NIOX MINO per ATS/ERS recommendations) derived reference values and studied the affecting factors, but its fetched record prints no numeric nasal NO values, and none is claimed from it. The numeric anchor in nl/min comes from the healthy-control arm of the primary-ciliary-dyskinesia standardisation study (Leigh 2013; 78 healthy controls, mean 304.6 ± 118.8 nl/min, range 125.5–867.0) — a control group inside a diagnostic study, not a dedicated interval derivation, printed as such — against which the PCD cut-off of 77 nl/min (sensitivity 0.98, specificity >0.999) sits roughly fourfold below the healthy mean. The tidal-breathing dataset (Gelardi 2016; 32 healthy subjects) gives 426.76 ± 143.27 ppb with a wide interval, its small N and width printing beside any range claim. Fractional EXHALED nitric oxide (FeNO, the lower-airway measurement, pack #7154) is a cross-point only — the two are never merged. PHYSIOLOGY — EMERGING (see proposedStamp).
MeSH: Adolescent · Adult · Female · Humans · Male · Nasal Mucosa/chemistry,metabolism · Nitric Oxide/analysis,metabolism · Reference Values · Respiratory Function Tests/methods · Young Adult · Aged · Asthma/diagnosis · Axonemal Dyneins/genetics · Breath Tests/methods · Case-Control Studies · Child · Child, Preschool · Cilia/ultrastructure · Cystic Fibrosis/diagnosis · Kartagener Syndrome/diagnosis,genetics,pathology · Microscopy, Electron, Transmission · Middle Aged · Nasal Mucosa/cytology,ultrastructure · Nitric Oxide/analysis · Prospective Studies · Pulmonary Disease, Chronic Obstructive/diagnosis · Sensitivity and Specificity · Breath Tests/instrumentation,methods · Electrochemical Techniques/instrumentation · Mouth Breathing · Nasal Cavity · Palate, Soft/physiology · Reference Standards · Reproducibility of Results
Cards (4)
Card 1 — Dedicated reference-values study (Antosova 2017): 141 healthy adults measured by NIOX MINO per ATS/ERS recommendations, affecting factors studied — the record prints no numeric nasal NO values
Study: Primary human physiology dataset (dedicated reference-values derivation, NIOX MINO)
Population: 141 healthy adult volunteers; nasal nitric oxide measured by NIOX MINO (Aerocrine, Sweden) according to ATS & ERS recommendations. Studied as potential variables influencing nasal NO levels: sex, age, height, body weight, waist-to-hip ratio, FEV1/FVC, PEF, and numbers of leukocytes, eosinophils, basophils and monocytes.
Summary: Dedicated reference-values study (Antosova 2017): 141 healthy adults measured by NIOX MINO per ATS/ERS recommendations, affecting factors studied — the record prints no numeric nasal NO values
Result: The study derived reference values for nasal NO in this healthy population and identified the individual parameters that modulate levels, creating a homogeneous group for nasal NO monitoring; its data are offered as reference data for the given variables. There was a significant correlation between nasal NO and exhaled NO, which the authors read as supporting the 'one airway — one disease' concept. The fetched record prints no numeric nasal NO mean, spread or interval, and this card claims none — the numeric values on this shelf come from the two datasets carded below, each bound to its own technique.
Watch out: The only dedicated reference derivation on the shelf is also the one whose record carries no numbers: it establishes that reference values and affecting-factor data exist for a defined young healthy population under an ATS/ERS-conformant protocol, and it must not be quoted for any value.
Source: Antosova M et al. Nasal nitric oxide in healthy adults - reference values and affecting factors. Physiological research. 2017. PMID: 28937239. Registry ID: none in the record. MeSH: Adolescent · Adult · Female · Humans · Male · Nasal Mucosa/chemistry,metabolism · Nitric Oxide/analysis,metabolism · Reference Values · Respiratory Function Tests/methods · Young Adult
Card 2 — PCD standardisation study's healthy controls (Leigh 2013): 78 healthy controls at 304.6 ± 118.8 nl/min (range 125.5–867.0) — the PCD cut-off of 77 nl/min sits far below, at sensitivity 0.98 and specificity >0.999
Study: Primary human physiology dataset (healthy-control arm of a multicentre diagnostic standardisation study; nl/min)
Population: Multicentre standardisation study (lead site plus six validation sites). At the lead site, nasal NO was measured prospectively in 78 healthy control subjects and 146 disease controls (asthma n = 37, cystic fibrosis n = 77, COPD n = 32), against 149 confirmed primary ciliary dyskinesia (PCD) patients (143 by ciliary ultrastructural defects, 6 by DNAH11 mutations). The healthy arm is a control group inside a diagnostic standardisation study, not a dedicated interval derivation, and prints as such.
Summary: PCD standardisation study's healthy controls (Leigh 2013): 78 healthy controls at 304.6 ± 118.8 nl/min (range 125.5–867.0) — the PCD cut-off of 77 nl/min sits far below, at sensitivity 0.98 and specificity >0.999
Result: Healthy-control nasal NO was 304.6 ± 118.8 nl/min (range 125.5–867.0), against PCD at 20.7 ± 24.1 nl/min (range 1.5–207.3), asthma at 267.8 ± 103.2, COPD at 223.7 ± 87.1 and cystic fibrosis at 134.0 ± 73.5 nl/min — the healthy range overlapped the disease-control ranges but only rarely the PCD range. The disease-specific cut-off was defined at 77 nl/min (sensitivity 0.98; specificity >0.999), roughly a quarter of the healthy mean; at the six other sites this cut-off identified 70 of 71 (98.6%) participants with confirmed PCD. All values are in nl/min under the study's standardised velum-closure protocol.
Watch out: The 77 nl/min figure is a disease-specific diagnostic cut-off for PCD derived against these controls, not a lower reference limit for healthy adults; and the healthy mean comes from a control arm assembled for comparison, measured under one standardised protocol — centres using other protocols or units cannot adopt it directly.
Source: Leigh MW et al. Standardizing nasal nitric oxide measurement as a test for primary ciliary dyskinesia. Annals of the American Thoracic Society. 2013. PMID: 24024753. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Asthma/diagnosis · Axonemal Dyneins/genetics · Breath Tests/methods · Case-Control Studies · Child · Child, Preschool · Cilia/ultrastructure · Cystic Fibrosis/diagnosis · Female · Humans · Kartagener Syndrome/diagnosis,genetics,pathology · Male · Microscopy, Electron, Transmission · Middle Aged · Nasal Mucosa/cytology,ultrastructure · Nitric Oxide/analysis · Prospective Studies · Pulmonary Disease, Chronic Obstructive/diagnosis · Reference Values · Sensitivity and Specificity · Young Adult
Card 3 — Tidal-breathing NIOX MINO dataset (Gelardi 2016): 32 healthy subjects at 426.76 ± 143.27 ppb (95% CI 160.22–733.30, as published) — men 446.76, women 403.80 ppb
Study: Primary human physiology dataset (dedicated normal-range derivation, tidal-breathing technique; ppb)
Population: 32 healthy subjects measured with the NIOX MINO Nasal device (Aerocrine AB, Solna, Sweden) using the tidal-breathing technique with velum closure — a different measurement quantity (ppb) from the nl/min aspiration values of the Leigh dataset, and never converted or pooled with them.
Summary: Tidal-breathing NIOX MINO dataset (Gelardi 2016): 32 healthy subjects at 426.76 ± 143.27 ppb (95% CI 160.22–733.30, as published) — men 446.76, women 403.80 ppb
Result: The measured normal nasal NO values were 426.76 ± 143.27 ppb, with a 95% confidence interval of 160.22–733.30 as published. Men averaged 446.76 ± 133.63 ppb (interval 178.64–714.02 as published) and women 403.80 ± 154.90 ppb (interval 94.00–713.60 as published). The authors conclude the study establishes the normal range of nitric oxide produced in the nasal/sinus cavities of healthy individuals using this device and the tidal-breathing velum-closure manoeuvre, consistent with values in the international literature.
Watch out: Small (N = 32) and the published interval is very wide (160.22–733.30 ppb around a mean of 426.76): the dataset exceeds the ~10-human reference floor but its range claim carries that width, and its ppb values answer a different measurement question from the nl/min values on the card above.
Source: Gelardi M et al. Standardization procedure for the nasal nitric oxide measurement method using Niox MINO® and the tidal-breathing technique with velum-closure. Journal of biological regulators and homeostatic agents. 2016. PMID: 27655510. Registry ID: none in the record. MeSH: Adult · Breath Tests/instrumentation,methods · Electrochemical Techniques/instrumentation · Female · Humans · Male · Mouth Breathing · Nasal Cavity · Nitric Oxide/analysis · Palate, Soft/physiology · Reference Standards · Reference Values · Reproducibility of Results · Young Adult
Card 4 — Synthesis: healthy nasal NO sits far above the 77 nl/min PCD cut-off in nl/min terms (healthy mean 304.6), but reference values are technique-bound — nl/min (velum-closure aspiration) and ppb (tidal breathing) are different quantities and are never pooled
Study: Synthesis
Population: Synthesis of the three primary datasets carded above.
Summary: Synthesis: healthy nasal NO sits far above the 77 nl/min PCD cut-off in nl/min terms (healthy mean 304.6), but reference values are technique-bound — nl/min (velum-closure aspiration) and ppb (tidal breathing) are different quantities and are never pooled
Result: Framed by the primary reference literature alone — neither Guyton & Hall nor Boron & Boulpaep carries a nasal-nitric-oxide reference chapter. Antosova 2017 (141 healthy adults, NIOX MINO per ATS/ERS) derived dedicated reference values and studied sex, age, body size, lung function and leukocyte subsets as affecting factors, with a significant nasal–exhaled NO correlation; its record prints no numeric values. Leigh 2013 supplies the nl/min anchor from 78 healthy controls (304.6 ± 118.8 nl/min, range 125.5–867.0) inside the PCD standardisation study, against a disease cut-off of 77 nl/min (sensitivity 0.98, specificity >0.999). Gelardi 2016 supplies the ppb anchor from 32 healthy subjects (426.76 ± 143.27 ppb) under tidal-breathing velum-closure measurement. PHYSIOLOGY — EMERGING: the three datasets are independent and mutually consistent in direction (healthy values far above the PCD range), but each is bound to its device and manoeuvre, one prints no numbers in its record, one is a diagnostic study's control arm, and the smallest has N = 32 with a very wide interval. Fractional exhaled NO (FeNO) is the lower-airway measurement, built at pack #7154 — cross-point only, never merged with nasal NO.
Watch out: Never convert ppb to nl/min or average across techniques: sampling flow and manoeuvre define the quantity. A value below 77 nl/min is a PCD-test result under the Leigh protocol, not a statement that any low nasal NO value, on any device, carries the same meaning.
Source: Synthesis of the reports carded above; no additional source.