SourceMed · Shelf #7757 in the ask interface
Viscoelastic test reference ranges (TEG / ROTEM) in healthy adults: population-derived ranges sit far from manufacturers' ranges (up to 74% of healthy volunteers outside on at least one parameter), with sex, age and native-vs-citrated differences throughout
Shelf #7757 · Heart, vessels, blood pressure & clotting
PHYSIOLOGY — ESTABLISHED — whole-blood viscoelastic reference ranges are established in 6 independent populations (4 TEG, 2 ROTEM, counted per modality and never merged), and are population-, sex- and sample-handling-specific: manufacturers' ranges misclassify healthy volunteers (8.5% diagnosis-level in Brazil; 74% any-parameter in North India), and recalcification alone shifts ROTEM values (TRUE-NATEM CT 872–1595 s vs NATEM 757–1327 s; that dataset's N is not documented)
Stamped: 2026-11-28.
Evidence published: 2009–2023.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In healthy adults, what are the reference ranges for whole-blood viscoelastic clotting tests — thromboelastography (TEG) and rotational thromboelastometry (ROTEM) — how far do population-derived ranges sit from the manufacturers' ranges, how do the ranges differ by sex, age and population, and how much do native vs citrated-recalcified samples move the ROTEM values? (Textbook framing: Guyton & Hall / Boron & Boulpaep haemostasis sections — framing only; the shelf rests chiefly on the primary datasets carded.)
Balance of the evidence
PHYSIOLOGY — reference ranges for whole-blood viscoelastic tests rest on 6 independent populations (4 TEG, 2 ROTEM), counted per modality and never merged. Textbook framing: Guyton & Hall and Boron & Boulpaep, haemostasis sections — framing only; this shelf rests chiefly on the primary reference literature, and says so. Population-derived TEG ranges sit materially apart from the manufacturers' ranges: the manufacturer's ranges would have labelled 8.5% of 118 healthy Brazilian volunteers coagulopathic (specificity 81%), 74% of 200 North Indian volunteers had at least one value outside the manufacturer's range, and about 40% of a second Indian sample (120 volunteers, including 33 children) fell outside at least one manufacturer range — three different comparisons (diagnosis-level vs any-parameter), each printed with its own definition. Sex differences recur (women more hypercoagulable in the Brazilian and Beijing datasets). For ROTEM, native-blood (NATEM) clotting time, α-angle and maximum clot firmness correlated with age in 132 healthy adults; and recalcification itself moves the values — TRUE-NATEM clotting time 872–1595 s vs NATEM 757–1327 s in a dataset whose volunteer N is not documented in its abstract, labelled as such. Activator and sample handling (citrated kaolin vs native) print on every card.
MeSH: Adult · Blood Coagulation · Blood Group Antigens · Female · Humans · Male · Racial Groups · Reference Values · Thrombelastography
Cards (7)
Card 1 — TEG dataset 1 of 4 (Scarpelini 2009, Brazil, N = 118 healthy volunteers, citrated kaolin TEG on a Haemoscope 5000): 95% ranges R 3.8–9.8 min, K 0.7–3.4 min, α-angle 47.8–77.7°, MA 49.7–72.7 mm, LY30 −2.3 to 5.77%, coagulation index −5.1 to 3.6 — and the manufacturer's ranges would have diagnosed coagulopathy in 10 of these healthy volunteers (8.5%; specificity 81%), with no disease found on extensive investigation
Study: Primary human physiology dataset — TEG reference ranges, healthy volunteers (citrated kaolin, Haemoscope 5000)
Population: 118 healthy adult volunteers (pregnant people, those on anticoagulants and those with a bleeding disorder excluded), tested with a standardised citrated-blood kaolin technique; volunteers outside the manufacturer's normal range underwent extensive coagulation investigations. Women were significantly more hypercoagulable than men; ageing was not associated with hypercoagulability.
Summary: TEG dataset 1 of 4 (Scarpelini 2009, Brazil, N = 118 healthy volunteers, citrated kaolin TEG on a Haemoscope 5000): 95% ranges R 3.8–9.8 min, K 0.7–3.4 min, α-angle 47.8–77.7°, MA 49.7–72.7 mm, LY30 −2.3 to 5.77%, coagulation index −5.1 to 3.6 — and the manufacturer's ranges would have diagnosed coagulopathy in 10 of these healthy volunteers (8.5%; specificity 81%), with no disease found on extensive investigation
Result: Reference ranges for 95% of volunteers: R 3.8–9.8 min, K 0.7–3.4 min, α-angle 47.8–77.7 degrees, MA 49.7–72.7 mm, LY30 −2.3 to 5.77%, coagulation index −5.1 to 3.6. Most values differed significantly from the manufacturer's, whose ranges had 81% specificity in this population and would incorrectly identify 8.5% of healthy volunteers as coagulopathic. The authors supported the manufacturer's own recommendation that each institution determine its own normal values.
Watch out: The 8.5% figure is a diagnosis-level misclassification (labelled coagulopathic, then cleared by investigation) — a different comparison from the any-parameter out-of-range rates in the Indian datasets, and never averaged with them. Method is citrated kaolin on one device; native-blood values would differ.
Source: Scarpelini et al. Normal range values for thromboelastography in healthy adult volunteers. Brazilian journal of medical and biological research = Revista brasileira de pesquisas medicas e biologicas. 2009. PMID: 19882085. Registry ID: none in the record. MeSH: Adult · Blood Coagulation · Blood Group Antigens · Female · Humans · Male · Racial Groups · Reference Values · Thrombelastography
Card 2 — TEG dataset 2 of 4 (Subramanian 2014, North India, N = 200 healthy volunteers — 100 men, 100 women, aged 20–50): 95% ranges r-time 1.8–14.2 min, k-time 0.7–7.3 min, α-angle 27.3–72.3°, MA 32.1–87.9 mm — and 74% (149/200) had at least one value outside the manufacturer's reference range, against 14.5% (29/200) outside this population's own derived range
Study: Primary human physiology dataset — TEG reference ranges, healthy North Indian volunteers (TEM-A, non-additive whole blood)
Population: 200 healthy North Indian volunteers (100 males, 100 females, aged 20–50 years) enrolled over 2011–2012; thromboelastometry on a TEM-A automated thromboelastometer (Framar Biomedica) using whole blood non-additive (360 µL), with prothrombin time, activated partial thromboplastin time and platelet count for all volunteers.
Summary: TEG dataset 2 of 4 (Subramanian 2014, North India, N = 200 healthy volunteers — 100 men, 100 women, aged 20–50): 95% ranges r-time 1.8–14.2 min, k-time 0.7–7.3 min, α-angle 27.3–72.3°, MA 32.1–87.9 mm — and 74% (149/200) had at least one value outside the manufacturer's reference range, against 14.5% (29/200) outside this population's own derived range
Result: The 95% reference ranges (mean ± 1.96 SD) were r-time 1.8–14.2 min, k-time 0.7–7.3 min, α-angle 27.3–72.3° and MA 32.1–87.9 mm. Maximum clot strength was higher in women than men, not statistically significantly. Overall 14.5% (29/200) had at least one abnormal parameter on the population's own range while 74% (149/200) had deranged values on the manufacturer's range, with statistically significant variation between the ranges on every parameter (P<0.001). The authors recommended each laboratory determine normal values for its target population and cautioned that TEG cannot replace conventional coagulation tests.
Watch out: 74% is an any-parameter out-of-range rate, not a coagulopathy diagnosis rate — the comparison differs from the Brazilian dataset's 8.5% and the two are never merged. The device and sample handling (TEM-A, non-additive whole blood) also differ from the citrated-kaolin datasets; ranges print per method.
Source: Subramanian et al. Establishing a normal reference range for thromboelastography in North Indian healthy volunteers. Indian journal of pathology & microbiology. 2014. PMID: 24739830. Registry ID: none in the record. MeSH: Adult · Blood Coagulation · Blood Coagulation Disorders · Female · Healthy Volunteers · Humans · India · Male · Middle Aged · Reference Values · Thrombelastography · Young Adult
Card 3 — TEG dataset 3 of 4 (Ji 2011, Beijing, N = 137 healthy volunteers, citrated kaolin, Haemoscope 5000): 95% ranges R 3.8–8.4 min, K 0.8–3.3 min, α-angle 46.2–76.2°, MA 50.0–70.8 mm, LY30 −3.3 to 4.0%, CI −3.8 to 2.9 — 24.1% (33/137) had at least one abnormal parameter on the population range and 7.3% (10/137) would have been considered coagulopathic on the manufacturer's values (specificity 76.0%), with significant sex differences on all parameters and lower fibrinogen function than the manufacturer's Western reference population
Study: Primary human physiology dataset — TEG reference ranges, healthy Beijing volunteers (citrated kaolin, Haemoscope 5000)
Population: 137 healthy adult volunteers resident in Beijing for over three years (enrolled June–August 2010), tested with citrated blood and kaolin activator on a Haemoscope 5000, with prothrombin time, APTT, thrombin time and plasma fibrinogen measured on the same sample.
Summary: TEG dataset 3 of 4 (Ji 2011, Beijing, N = 137 healthy volunteers, citrated kaolin, Haemoscope 5000): 95% ranges R 3.8–8.4 min, K 0.8–3.3 min, α-angle 46.2–76.2°, MA 50.0–70.8 mm, LY30 −3.3 to 4.0%, CI −3.8 to 2.9 — 24.1% (33/137) had at least one abnormal parameter on the population range and 7.3% (10/137) would have been considered coagulopathic on the manufacturer's values (specificity 76.0%), with significant sex differences on all parameters and lower fibrinogen function than the manufacturer's Western reference population
Result: Reference ranges for 95% of volunteers: R 3.8–8.4 min, K 0.8–3.3 min, α-Angle 46.2–76.2°, MA 50.0–70.8 mm, LY30 −3.3% to 4.0% and CI −3.8 to 2.9. Overall 24.1% (33/137) had at least one abnormal parameter while 7.3% (10/137) would have been considered coagulopathy on the manufacturer's values (test specificity 76.0%). Chinese healthy volunteers had lower fibrinogen function than the manufacturer's Western reference population, and R, K, α-Angle, MA and CI differed significantly between men and women (all P<0.01). The authors supported each institute determining its own normal reference values.
Watch out: Two different out-of-range rates appear in this one report — 24.1% any-parameter on the population range and 7.3% diagnosis-level on the manufacturer's — and each prints with its definition. Same device family and activator as the Brazilian dataset, yet the ranges and sex pattern are population-specific.
Source: Ji et al. [Establishment of normal reference values for thromboelastography on Chinese population in Beijing]. Zhonghua yi xue za zhi. 2011. PMID: 21609551. Registry ID: none in the record. MeSH: Adult · Asian People · China · Female · Humans · Male · Middle Aged · Reference Values · Thrombelastography
Card 4 — TEG dataset 4 of 4 (Ahammad 2020, India, N = 120 healthy volunteers — 87 adults and 33 children, composition printed on the face): kaolin-activated TEG ranges R-time 3.8–10.6, K-time 1.2–3.1, angle 44.9–72.0, MA 41.2–64.5, LY30 0–9.9, CI −3.7 to 3.4 — about 40% had at least one abnormal parameter on the manufacturer's range, falling to 12.5% on the derived ranges, with significant age- and sex-group differences in R-time, K-time and angle
Study: Primary human physiology dataset — TEG reference ranges, healthy Indian population including children (kaolin-activated, citrated whole blood)
Population: 120 healthy volunteers (55 adult males, 32 adult females and 33 children) screened for bleeding history and drugs affecting coagulation; kaolin-activated TEG on citrated whole blood, analysing R-time, K-time, angle, MA, LY30 and CI. The adult subset is 87; the 33 children are part of the printed composition and are not silently counted as adults.
Summary: TEG dataset 4 of 4 (Ahammad 2020, India, N = 120 healthy volunteers — 87 adults and 33 children, composition printed on the face): kaolin-activated TEG ranges R-time 3.8–10.6, K-time 1.2–3.1, angle 44.9–72.0, MA 41.2–64.5, LY30 0–9.9, CI −3.7 to 3.4 — about 40% had at least one abnormal parameter on the manufacturer's range, falling to 12.5% on the derived ranges, with significant age- and sex-group differences in R-time, K-time and angle
Result: Derived reference ranges for all volunteers irrespective of age and sex: R-time 3.8–10.6, K-time 1.2–3.1, angle 44.9–72.0, MA 41.2–64.5, LY30 0–9.9 and CI −3.7 to 3.4. Statistically significant differences were observed between age and sex groups for R-time, K-time and angle. About 40% of volunteers had at least one abnormal parameter on the manufacturer's reference range, decreasing to 12.5% on the derived ranges. The authors concluded population-specific ranges should be derived before TEG is used clinically.
Watch out: This dataset mixes 87 adults with 33 children and its headline ranges are pooled across them — it is carded with that composition on its face and its ranges are not quoted as adult-only intervals. Its ~40% figure is again an any-parameter rate, defined as in its own report.
Source: Ahammad et al. Age- and gender-related reference ranges for thromboelastography from a healthy Indian population. International journal of laboratory hematology. 2020. PMID: 31889401. Registry ID: none in the record. MeSH: Adolescent · Adult · Age Factors · Female · Humans · Male · Middle Aged · Reference Values · Sex Factors · Thrombelastography · Wounds and Injuries
Card 5 — ROTEM dataset 1 of 2 (Schneider 2015, N = 132 healthy adults, median age 48): native-blood ROTEM (NATEM) clotting time fell with age (P = 0.039) while α-angle and maximum clot firmness rose with age (P = 0.001), and ROTEM variables correlated with calibrated thrombin generation — this dataset's thrombin-generation limb belongs with the thrombin-generation family (pack #6965); only its ROTEM limb is carded here
Study: Primary human physiology dataset — ROTEM (NATEM) in healthy adults, with calibrated thrombin generation (ROTEM limb carded here)
Population: 132 healthy adults (72 men, 60 women; median age 48.0 years), with ROTEM performed without activator (NATEM) — clotting time, α-angle, clot formation time and maximum clot firmness — and thrombin generation measured in platelet-poor plasma by calibrated automated thrombogram at 1 and 5 pmol/L tissue factor.
Summary: ROTEM dataset 1 of 2 (Schneider 2015, N = 132 healthy adults, median age 48): native-blood ROTEM (NATEM) clotting time fell with age (P = 0.039) while α-angle and maximum clot firmness rose with age (P = 0.001), and ROTEM variables correlated with calibrated thrombin generation — this dataset's thrombin-generation limb belongs with the thrombin-generation family (pack #6965); only its ROTEM limb is carded here
Result: For ROTEM, age correlated positively with maximum clot firmness and α-angle (P = 0.001) and negatively with clotting time (P = 0.039). Thrombin peak and endogenous thrombin potential at 5 pmol/L tissue factor correlated significantly with α-angle and maximum clot firmness. Age-related changes in both assays were not linear. This report publishes correlations and age patterns for NATEM rather than a full reference-interval table — carded as such, with no interval invented for it.
Watch out: No reference intervals are printed from this dataset because its abstract reports correlations, not intervals — the interval-bearing ROTEM dataset is the next card. Its thrombin-generation results are cross-pointed, not carded, to avoid double-homing one report's two limbs.
Source: Schneider et al. Thrombin generation and rotational thromboelastometry in the healthy adult population. Hamostaseologie. 2015. PMID: 25529462. Registry ID: none in the record. MeSH: Aging · Blood Coagulation · Female · Humans · Male · Middle Aged · Reference Values · Reproducibility of Results · Sensitivity and Specificity · Thrombelastography · Thrombin
Card 6 — ROTEM dataset 2 of 2 (Yoon 2023 — volunteer N not documented in the abstract, labelled as such; healthy adult volunteers aged 18–65): native whole blood (TRUE-NATEM) reference values CT 872–1595 s, CFT 314–839 s, MCF 34–55 mm, α-angle 17–40° vs recalcified citrated blood (NATEM) CT 757–1327 s, CFT 219–615 s, MCF 37–61 mm, α-angle 24–51° — recalcification itself significantly activates coagulation, shortening CT / CFT and raising MCF / α-angle
Study: Primary human physiology dataset — ROTEM Delta reference values, native vs recalcified citrated blood (volunteer N not documented)
Population: Healthy adult volunteers aged 18–65 years (any medical condition or medication affecting coagulation excluded), tested on a ROTEM Delta with fresh native whole blood (TRUE-NATEM) and with recalcified citrated blood (NATEM). N not documented — the abstract does not state the number of volunteers, and no number is claimed here.
Summary: ROTEM dataset 2 of 2 (Yoon 2023 — volunteer N not documented in the abstract, labelled as such; healthy adult volunteers aged 18–65): native whole blood (TRUE-NATEM) reference values CT 872–1595 s, CFT 314–839 s, MCF 34–55 mm, α-angle 17–40° vs recalcified citrated blood (NATEM) CT 757–1327 s, CFT 219–615 s, MCF 37–61 mm, α-angle 24–51° — recalcification itself significantly activates coagulation, shortening CT / CFT and raising MCF / α-angle
Result: TRUE-NATEM reference values: CT 872–1595 s, A10 14–34 mm, A20 26–48 mm, CFT 314–839 s, MCF 34–55 mm, α-angle 17–40°. NATEM reference values: CT 757–1327 s, A10 19–43 mm, A20 33–55 mm, CFT 219–615 s, MCF 37–61 mm, α-angle 24–51°, ML 0–3%. Comparing NATEM with TRUE-NATEM, CT and CFT decreased and MCF and α-angle increased — the recalcification process itself shows significant activation of coagulation. Female volunteers had enhanced coagulation when activators were used; ethnicity showed no difference. The authors underlined native whole blood as the gold-standard reference.
Watch out: 'N not documented' applies to this dataset per the standing labelling rule — it is carded for its paired native-vs-citrated comparison, not for any sample-size weight. The native/citrated gap is the content: a NATEM interval must never be applied to a native-blood result, or vice versa.
Source: Yoon et al. Native Whole Blood (TRUE-NATEM) and Recalcified Citrated Blood (NATEM) Reference Value Validation with ROTEM Delta. Seminars in cardiothoracic and vascular anesthesia. 2023. PMID: 36631390. Registry ID: none in the record. MeSH: Adult · Humans · Female · Reference Values · Blood Coagulation · Thrombelastography · Citric Acid · Citrates · Piperidones · Benzeneacetamides
Card 7 — Synthesis: 6 independent populations — 4 TEG and 2 ROTEM, counted per modality and never merged — establish that viscoelastic reference ranges are population-, sex-, device- and sample-handling-specific: population-derived TEG ranges repeatedly sit apart from manufacturers' ranges (8.5% diagnosis-level misclassification in Brazil; 74% any-parameter out-of-range in North India; ~40% in the second Indian dataset), and ROTEM values move materially with recalcification alone (TRUE-NATEM CT 872–1595 s vs NATEM 757–1327 s)
Study: Synthesis
Population: All datasets carded on this shelf: TEG — Brazil (N = 118), North India (N = 200), Beijing (N = 137), India (N = 120, including 33 children); ROTEM — healthy adults (N = 132) and native-vs-citrated volunteers (N not documented). Textbook framing (Guyton & Hall / Boron & Boulpaep haemostasis sections) is framing only; the shelf rests on these primary datasets.
Summary: Synthesis: 6 independent populations — 4 TEG and 2 ROTEM, counted per modality and never merged — establish that viscoelastic reference ranges are population-, sex-, device- and sample-handling-specific: population-derived TEG ranges repeatedly sit apart from manufacturers' ranges (8.5% diagnosis-level misclassification in Brazil; 74% any-parameter out-of-range in North India; ~40% in the second Indian dataset), and ROTEM values move materially with recalcification alone (TRUE-NATEM CT 872–1595 s vs NATEM 757–1327 s)
Result: The consistent finding across datasets is that a laboratory's own population range classifies healthy people very differently from the manufacturer's range, and that sex, age, activator and citrate handling move the values. Pregnancy-specific viscoelastic ranges are a different population and are not carded here. Cross-points: plasma coagulation screening tests (pack #7393) and calibrated thrombin generation (pack #6965) are different measurement families.
Watch out: See the individual cards for each dataset's limits; this synthesis adds no new data.
Source: Synthesis of the reports carded above; no additional source.