SourceMed · Shelf #7758 in the ask interface
Reticulocyte-profile reference intervals in healthy adults: intervals are platform-specific — reticulocyte haemoglobin content 22.84–29.70 pg on the Mindray BC-6200 vs 29.95–35.12 pg in Chengdu and 30.9–35.7 pg on the BC-6800, across 4 independent populations
Shelf #7758 · Heart, vessels, blood pressure & clotting
PHYSIOLOGY — ESTABLISHED — reticulocyte-profile reference intervals are established in 4 independent populations and are platform-specific: reticulocyte haemoglobin content runs 29.95–35.12 pg (Chengdu, men) and 30.9–35.7 pg (Mindray BC-6800) but only 22.84–29.70 pg on the Mindray BC-6200 (N = 60), and sex partitioning is needed only for selected derived parameters (DxH 900 dataset)
Stamped: 2026-11-28.
Evidence published: 2017–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In healthy adults, what are the reference intervals for the reticulocyte profile — absolute reticulocyte count, reticulocyte percentage, immature reticulocyte fraction, reticulocyte haemoglobin content / equivalent (CHr / Ret-He / RHE) and mean reticulocyte volume — how far apart do the intervals sit between analyser platforms (Beckman Coulter DxH, Sysmex, Mindray) and populations, and which parameters need sex partitioning? (Textbook framing: Guyton & Hall, red-cell / erythropoiesis sections — framing only; the shelf rests chiefly on the primary datasets carded.)
Balance of the evidence
PHYSIOLOGY — reticulocyte-profile reference intervals rest on 4 independent populations, and platform is the content: no haemoglobin-content figure prints without its analyser. Textbook framing: Guyton & Hall, red-cell / erythropoiesis sections — framing only; this shelf rests chiefly on the primary reference literature, and says so. The Madrid dataset (175 healthy adults, Beckman Coulter DxH 900) established intervals for 14 reticulocyte-profile parameters per CLSI EP28-A3c, with sex partitioning required for two derived parameters (low haemoglobin density, microcytic anaemia factor) but not the core reticulocyte measures; its abstract does not print the numeric intervals, and none are invented here. The Chengdu dataset (691 healthy Han adults) gives reticulocyte haemoglobin content 29.95–35.12 pg (men) and 29.77–34.52 pg (women), mean reticulocyte volume 99.76–115.97 fL (men) and 98.72–113.83 fL (women), and reticulocyte percentage 0.485–1.504%. The Mindray BC-6800 dataset (184 healthy adults) gives reticulocytes 23.2–93.2 ×10⁹/L and haemoglobin equivalent 30.9–35.7 pg, with the Sysmex XE-5000 comparison showing a different upper count limit (101.3 vs 93.2 ×10⁹/L). The Thai BC-6200 dataset — the smallest, N = 60, printed on the face — gives a haemoglobin-content interval of 22.84–29.70 pg, well below the Chengdu and BC-6800 intervals on other platforms. Counts print in ×10⁹/L as in the records.
MeSH: Humans · Adult · Reticulocyte Count · Male · Female · Reference Values · Reticulocytes · Middle Aged · Adolescent · Young Adult · Cohort Studies
Cards (5)
Card 1 — Dataset 1 of 4 (Posada-Franco 2024, Madrid, N = 175 healthy adults aged 18–62, Beckman Coulter DxH 900): reference intervals established for 14 reticulocyte-profile parameters per CLSI EP28-A3c, with sex partitioning required for two derived parameters (low haemoglobin density and microcytic anaemia factor) but not for the core reticulocyte measures — the abstract does not print the numeric intervals, and none are claimed here
Study: Primary human physiology dataset — reticulocyte-profile reference intervals, Beckman Coulter DxH 900 (CLSI EP28-A3c)
Population: 175 healthy adults (18 to 62 years) recruited from the blood donation centre of Hospital Clínico San Carlos, Madrid, Spain, with whole blood assayed for 14 haematological parameters of the reticulocyte profile on the Beckman Coulter DxH 900 analyser; intervals estimated by non-parametric, Harrell-Davis bootstrap and robust methods, with partitioning judged by Lahti's method.
Summary: Dataset 1 of 4 (Posada-Franco 2024, Madrid, N = 175 healthy adults aged 18–62, Beckman Coulter DxH 900): reference intervals established for 14 reticulocyte-profile parameters per CLSI EP28-A3c, with sex partitioning required for two derived parameters (low haemoglobin density and microcytic anaemia factor) but not for the core reticulocyte measures — the abstract does not print the numeric intervals, and none are claimed here
Result: Intervals estimated by the non-parametric and Harrell-Davis bootstrap methods were very similar; robust-method intervals were narrower. Gender partitioning was required for two parameters — low haemoglobin density (LHD) and microcytic anaemia factor (MAF) — while the rest of the parameters did not need partitioning. The authors' stated rationale is that, absent standardisation, each laboratory should establish reference ranges for its own methodology and instrumentation.
Watch out: This card carries the partitioning finding and the platform, not numeric intervals — its abstract reports that 14 intervals were established without printing them, and no interval is reconstructed here. Do not borrow another platform's numbers to fill it in.
Source: Posada-Franco et al. Reference intervals for reticulocyte count and derived reticulocyte parameters in a cohort of healthy adults. International journal of laboratory hematology. 2024. PMID: 38924378. Registry ID: none in the record. MeSH: Humans · Adult · Reticulocyte Count · Male · Female · Reference Values · Reticulocytes · Middle Aged · Adolescent · Young Adult · Cohort Studies
Card 2 — Dataset 2 of 4 (Zhen 2020, Chengdu, N = 691 healthy Han adults aged 20–90): reticulocyte haemoglobin content (RHE) 29.95–35.12 pg in men and 29.77–34.52 pg in women, mean reticulocyte volume 99.76–115.97 fL in men and 98.72–113.83 fL in women, and reticulocyte percentage 0.485–1.504% (intervals defined as mean ± 1.96 SD)
Study: Primary human physiology dataset — reticulocyte haemoglobin content, mean reticulocyte volume and percentage intervals, healthy Han adults
Population: 691 healthy Han adults without iron deficiency, aged 20 to 90 years, in Chengdu, China; reference intervals for reticulocyte haemoglobin content, mean reticulocyte volume and reticulocyte count percentage defined as mean ± 1.96 SD.
Summary: Dataset 2 of 4 (Zhen 2020, Chengdu, N = 691 healthy Han adults aged 20–90): reticulocyte haemoglobin content (RHE) 29.95–35.12 pg in men and 29.77–34.52 pg in women, mean reticulocyte volume 99.76–115.97 fL in men and 98.72–113.83 fL in women, and reticulocyte percentage 0.485–1.504% (intervals defined as mean ± 1.96 SD)
Result: Reference intervals: RHE 29.95–35.12 pg (males) and 29.77–34.52 pg (females); mean reticulocyte volume 99.76–115.97 fL (males) and 98.72–113.83 fL (females); reticulocyte percentage 0.485–1.504%. The authors report these as the first RHE, MRV and RET intervals established for the Han population of Chengdu. The abstract does not name the analyser platform, and none is attributed here.
Watch out: Sex-partitioned intervals print as published (men / women) and are not collapsed. The platform is not named in the abstract — so these figures must not be assigned to a named analyser, and the platform contrast on this shelf is carried by the datasets that do name theirs.
Source: Zhen et al. Reference Intervals for Reticulocyte Hemoglobin Content, Mean Reticulocyte Volume and Reticulocyte Count Percentage in Healthy Han Ethnic Adults of Chengdu. Clinical laboratory. 2020. PMID: 32902211. Registry ID: none in the record. MeSH: Adult · Anemia, Iron-Deficiency · Ethnicity · Female · Hemoglobins · Humans · Male · Reference Values · Reticulocyte Count · Reticulocytes
Card 3 — Dataset 3 of 4 (Buttarello 2017, N = 184 healthy adults, Mindray BC-6800): reticulocytes 23.2–93.2 ×10⁹/L, immature reticulocyte fraction 0.015–0.14, mean reticulocyte haemoglobin equivalent 30.9–35.7 pg and mean reticulocyte volume 97.8–118 fL — with the Sysmex XE-5000 comparison showing a different upper count limit (101.3 vs 93.2 ×10⁹/L)
Study: Primary human physiology dataset — reticulocyte reference intervals, Mindray BC-6800 with Sysmex XE-5000 comparison
Population: 184 healthy adults of both sexes for the reference intervals (with a further 368 subjects with various pathological conditions used for the analyser's diagnostic-efficiency evaluation, not for the intervals); reticulocytes and extended parameters quantified on the Mindray BC-6800 and compared with the Sysmex XE-5000.
Summary: Dataset 3 of 4 (Buttarello 2017, N = 184 healthy adults, Mindray BC-6800): reticulocytes 23.2–93.2 ×10⁹/L, immature reticulocyte fraction 0.015–0.14, mean reticulocyte haemoglobin equivalent 30.9–35.7 pg and mean reticulocyte volume 97.8–118 fL — with the Sysmex XE-5000 comparison showing a different upper count limit (101.3 vs 93.2 ×10⁹/L)
Result: Reference intervals on the BC-6800: reticulocytes 23.2–93.2 ×10⁹/L; immature reticulocyte fraction 0.015–0.14; mean reticulocyte haemoglobin equivalent (RHE) 30.9–35.7 pg; mean reticulocyte volume 97.8–118 fL. Imprecision was close to the analytic goal, carryover (2.3%) was negligible and time-stability excellent up to 8 hours. Against the XE-5000, overall correlation on counting was good despite the difference in the upper reference limit (93.2 vs 101.3 ×10⁹/L), and total agreement of RHE was 91.6%.
Watch out: The immature reticulocyte fraction here is a fraction (0.015–0.14), while the Thai dataset reports IRF as a percentage (2.71–15.28%) — same quantity, different expression, never mixed. The 368 pathological subjects belong to the efficiency limb only and are not part of the healthy denominator.
Source: Buttarello et al. Reticulocyte count and extended reticulocyte parameters by Mindray BC-6800: Reference intervals and comparison with Sysmex XE-5000. International journal of laboratory hematology. 2017. PMID: 28675649. Registry ID: none in the record. MeSH: Adult · Female · Humans · Male · Reticulocyte Count
Card 4 — Dataset 4 of 4 (Aphaiwiwat 2025, Thailand, N = 60 healthy adults — the smallest dataset, printed on the face — Mindray BC-6200): reticulocytes 34.88–118.50 ×10⁹/L, immature reticulocyte fraction 2.71–15.28%, and reticulocyte haemoglobin content 22.84–29.70 pg — well below the Chengdu (29.95–35.12 pg) and BC-6800 (30.9–35.7 pg) intervals — with counts correlating r > 0.9785 against flow cytometry and manual microscopy
Study: Primary human physiology dataset — reticulocyte reference intervals, Mindray BC-6200, Thai adults (small N = 60)
Population: 60 healthy Thai adults of both sexes for the reference intervals (with 182 leftover samples from individuals with pathological conditions used for method comparison), with reticulocyte counts on the Mindray BC-6200 compared against manual microscopy and flow cytometry as the reference method.
Summary: Dataset 4 of 4 (Aphaiwiwat 2025, Thailand, N = 60 healthy adults — the smallest dataset, printed on the face — Mindray BC-6200): reticulocytes 34.88–118.50 ×10⁹/L, immature reticulocyte fraction 2.71–15.28%, and reticulocyte haemoglobin content 22.84–29.70 pg — well below the Chengdu (29.95–35.12 pg) and BC-6800 (30.9–35.7 pg) intervals — with counts correlating r > 0.9785 against flow cytometry and manual microscopy
Result: Reference intervals on the BC-6200: reticulocytes 34.88–118.50 ×10⁹/L; immature reticulocyte fraction 2.71–15.28%; reticulocyte haemoglobin content (RHE) 22.84–29.70 pg. Imprecision was below the manufacturer's claim, carryover was <0.001% and time-stability excellent up to 24 hours. Reticulocyte counts correlated strongly between the analyser, manual microscopy and flow cytometry (r > 0.9785). The authors concluded the BC-6200 is a reliable alternative for reticulocyte counting while noting further studies are needed on clinical utility in diagnosing haematological conditions.
Watch out: N = 60 exceeds the ~10-human reference floor but is small for an interval study, and its N prints here rather than being rounded up by the larger datasets. Its haemoglobin-content interval (22.84–29.70 pg) barely overlaps the other platforms' — the clearest single demonstration on this shelf that the interval belongs to the analyser, not to the analyte alone.
Source: Aphaiwiwat et al. Reference Intervals and Comparative Analysis of Reticulocyte Counts Using the Mindray BC-6200, Flow Cytometry, and Manual Microscopy. International journal of laboratory hematology. 2025. PMID: 39915923. Registry ID: none in the record. MeSH: Humans · Flow Cytometry · Reticulocyte Count · Male · Female · Adult · Reference Values · Reticulocytes · Middle Aged · Microscopy · Young Adult
Card 5 — Synthesis: 4 independent populations establish reticulocyte-profile reference intervals per platform — DxH 900 intervals for 14 parameters with sex partitioning needed only for two derived measures (numeric intervals not printed in that abstract), Chengdu RHE 29.95–35.12 pg (men) with reticulocyte percentage 0.485–1.504%, BC-6800 reticulocytes 23.2–93.2 ×10⁹/L with RHE 30.9–35.7 pg, and BC-6200 reticulocytes 34.88–118.50 ×10⁹/L with RHE 22.84–29.70 pg — so no reticulocyte haemoglobin figure is meaningful without its analyser named
Study: Synthesis
Population: All datasets carded on this shelf: Madrid (N = 175, Beckman Coulter DxH 900), Chengdu (N = 691, healthy Han adults), BC-6800 series (N = 184 healthy adults) and Thai BC-6200 series (N = 60 healthy adults) — four independent populations, intervals kept per platform and per population. Textbook framing (Guyton & Hall, red-cell / erythropoiesis sections) is framing only; the shelf rests on these datasets. A fifth candidate dataset was excluded at verification as veterinary (dogs) and is not counted.
Summary: Synthesis: 4 independent populations establish reticulocyte-profile reference intervals per platform — DxH 900 intervals for 14 parameters with sex partitioning needed only for two derived measures (numeric intervals not printed in that abstract), Chengdu RHE 29.95–35.12 pg (men) with reticulocyte percentage 0.485–1.504%, BC-6800 reticulocytes 23.2–93.2 ×10⁹/L with RHE 30.9–35.7 pg, and BC-6200 reticulocytes 34.88–118.50 ×10⁹/L with RHE 22.84–29.70 pg — so no reticulocyte haemoglobin figure is meaningful without its analyser named
Result: Sex partitioning is parameter-specific (required for low haemoglobin density and microcytic anaemia factor in Madrid; published as sex-specific intervals in Chengdu), and count intervals differ by platform at both ends (upper limit 93.2 vs 118.50 ×10⁹/L across the Mindray datasets, with the Sysmex XE-5000 at 101.3). Cross-point: full-blood-count reference intervals (pack #7321) cover haemoglobin, white cells and platelets, not reticulocyte parameters.
Watch out: See the individual cards for each dataset's limits; this synthesis adds no new data.
Source: Synthesis of the reports carded above; no additional source.