Peripheral thermometry vs core temperature in adults: 3 overlapping syntheses (labelled as syntheses, never independent cohorts) — limits of agreement beyond the acceptable ±0.5 °C, fever-detection sensitivity only 64%, and only oesophageal measurement clinically acceptable in ICU
Shelf #7781 · Heart, vessels, blood pressure & clotting
PREDICTIVE PERFORMANCE — NOT ESTABLISHED as clinically acceptable: 3 overlapping syntheses, counted as syntheses only — pooled limits of agreement outside ±0.5 °C (febrile adults −1.44 °C to 1.46 °C) and fever-detection sensitivity 64% in the largest synthesis, with only oesophageal measurement clinically acceptable in ICU; a measurement-accuracy stamp, not a treatment stamp
Stamped: 2026-11-28.
Evidence published: 2011–2021.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
At the bedside, how accurately do peripheral thermometers (tympanic membrane, temporal artery, axillary and oral) estimate core body temperature in adults — what are their pooled limits of agreement against central thermometry (pulmonary artery, bladder, oesophageal, rectal), how poor is fever detection, and does any non-invasive method reach clinically acceptable accuracy in intensive care? (Base: 3 syntheses whose primary-study sets overlap — counted as syntheses, never as independent cohorts.)
Balance of the evidence
This shelf rests on 3 syntheses whose primary-study sets overlap — the count prints as 3 syntheses, not 3 independent cohorts, everywhere. The largest (75 studies, 8,682 patients) found pooled 95% limits of agreement outside the pre-defined clinically acceptable ±0.5 °C range — −1.44 °C to 1.46 °C in febrile adults and −2.07 °C to 1.90 °C in hypothermia — with fever-detection sensitivity of only 64% beside specificity of 96%. The adult-ICU synthesis (13 studies, 632 patients, 105,375 measurements) found axillary, tympanic infrared and zero-heat-flux thermometers all underestimated intravascular temperature, only oesophageal measurement was clinically acceptable, and heterogeneity was 99–100%. The oldest, smallest synthesis (3 febrile-ICU studies) is the most favourable — mean differences within ±0.2 °C for most tympanic combinations and the one oral study — and that disagreement between syntheses is content, not noise to be averaged away. No single 'peripheral thermometers are accurate / inaccurate' verdict prints without method, population and the limits-of-agreement framing; no card implies a thermometer-guided pathway improves outcomes.
MeSH: Adult · Axilla · Child · Fever/diagnosis · Humans · Hypothermia/diagnosis · Mouth · Temporal Arteries · Thermometers/standards · Tympanic Membrane
Cards (4)
Card 1 — Peripheral thermometry synthesis 1 of 3 (Niven 2015, Annals of Internal Medicine — 75 studies, 8,682 patients, adults and children): pooled 95% limits of agreement against central thermometry fell outside the pre-defined clinically acceptable ±0.5 °C range, especially in fever (−1.44 °C to 1.46 °C in adults) and hypothermia (−2.07 °C to 1.90 °C in adults); fever-detection sensitivity was low at 64% (95% CI 55–72%) with specificity 96% (93–97%)
Study: Systematic review and meta-analysis (synthesis 1 of 3; primary-study sets overlap across the three syntheses)
Population: Prospective studies comparing peripheral thermometers (tympanic membrane, temporal artery, axillary or oral) with central thermometers (pulmonary artery catheter, urinary bladder, oesophageal or rectal) in adults and children; 75 studies, 8,682 patients. Most studies were at high or unclear risk of patient-selection bias (74%) or index-test bias (67%). One of 3 overlapping syntheses on this shelf — labelled as such, never an independent cohort.
Summary: Peripheral thermometry synthesis 1 of 3 (Niven 2015, Annals of Internal Medicine — 75 studies, 8,682 patients, adults and children): pooled 95% limits of agreement against central thermometry fell outside the pre-defined clinically acceptable ±0.5 °C range, especially in fever (−1.44 °C to 1.46 °C in adults) and hypothermia (−2.07 °C to 1.90 °C in adults); fever-detection sensitivity was low at 64% (95% CI 55–72%) with specificity 96% (93–97%)
Result: Compared with central thermometers, peripheral thermometers had pooled 95% limits of agreement outside the pre-defined clinically acceptable range (±0.5 °C), especially among patients with fever (−1.44 °C to 1.46 °C for adults; −1.49 °C to 0.43 °C for children) and hypothermia (−2.07 °C to 1.90 °C for adults; no data for children). For detection of fever, sensitivity was low (64%, 95% CI 55–72%) but specificity was high (96%, 93–97%). Only 1 study reported sensitivity and specificity for detection of hypothermia. The authors conclude peripheral thermometers do not have clinically acceptable accuracy and should not be used when accurate measurement will influence clinical decisions.
Watch out: The largest synthesis and the least favourable: limits of agreement — not mean bias — are the accuracy measure here, and they are roughly three times the acceptable width in febrile adults. Fever sensitivity of 64% means about a third of fevers are missed at the bedside by a peripheral reading.
Source: Niven DJ et al. Accuracy of peripheral thermometers for estimating temperature: a systematic review and meta-analysis. Annals of internal medicine. 2015. PMID: 26571241. Registry ID: none in the record. MeSH: Adult · Axilla · Child · Fever/diagnosis · Humans · Hypothermia/diagnosis · Mouth · Temporal Arteries · Thermometers/standards · Tympanic Membrane
Card 2 — Peripheral thermometry synthesis 2 of 3 (Cutuli 2021, adult ICU — 13 cohort studies, 632 patients, 105,375 measurements): axillary, tympanic infrared and zero-heat-flux thermometers all underestimated intravascular temperature, only oesophageal measurement showed clinically acceptable accuracy, and heterogeneity was 99–100%
Study: Systematic review and meta-analysis in adult ICU patients (synthesis 2 of 3; primary-study sets overlap across the three syntheses)
Population: Published studies reporting accuracy and precision of non-invasive peripheral thermometers (index test) in adult ICU patients against intravascular temperature measurement (reference test), searched 1966–2017; 13 cohort studies, 632 patients and 105,375 measurements; study quality assessed with QUADAS-2. One of 3 overlapping syntheses on this shelf — labelled as such, never an independent cohort. This record carries no MeSH terms (source line reads 'MeSH: none in the record').
Summary: Peripheral thermometry synthesis 2 of 3 (Cutuli 2021, adult ICU — 13 cohort studies, 632 patients, 105,375 measurements): axillary, tympanic infrared and zero-heat-flux thermometers all underestimated intravascular temperature, only oesophageal measurement showed clinically acceptable accuracy, and heterogeneity was 99–100%
Result: Axillary, tympanic infrared and zero-heat-flux thermometers all underestimated intravascular temperature. Only oesophageal measurements showed clinically acceptable accuracy. There were too few studies to assess precision for any technique. Study heterogeneity was high (99–100%), and risk of bias for the index test was unclear, mostly because of no device calibration or control for confounders. The authors conclude non-invasive peripheral thermometers have low accuracy in ICU patients and their clinical and trial-related use is unreliable and potentially misleading.
Watch out: 105,375 measurements in only 632 patients — measurement count is not patient count, and the two must never be interchanged in a summary of this shelf. Heterogeneity of 99–100% means the pooled bias figures describe an average across unlike studies, not a property of any one device.
Source: Cutuli SL et al. Accuracy of non-invasive body temperature measurement methods in adult patients admitted to the intensive care unit: a systematic review and meta-analysis. Critical care and resuscitation : journal of the Australasian Academy of Critical Care Medicine. 2021. PMID: 38046384. Registry ID: none in the record. MeSH: none in the record
Card 3 — Peripheral thermometry synthesis 3 of 3 (Jefferies 2011, febrile critically ill adults — 3 studies against pulmonary-artery core temperature): studies were too heterogeneous and too poorly reported for meta-analysis; mean differences were within ±0.2 °C in five of seven tympanic combinations and in the one oral study, but outside that range for all three rectal combinations
Study: Systematic review in febrile critically ill adults (synthesis 3 of 3; primary-study sets overlap across the three syntheses)
Population: Clinical trials comparing peripheral thermometry in critically ill adult patients with core temperatures >37.5 °C against pulmonary-artery catheter core temperature; 3 studies identified. The oldest synthesis on this shelf, and the most favourable to tympanic/oral measurement — the disagreement between syntheses is content. One of 3 overlapping syntheses — labelled as such, never an independent cohort.
Summary: Peripheral thermometry synthesis 3 of 3 (Jefferies 2011, febrile critically ill adults — 3 studies against pulmonary-artery core temperature): studies were too heterogeneous and too poorly reported for meta-analysis; mean differences were within ±0.2 °C in five of seven tympanic combinations and in the one oral study, but outside that range for all three rectal combinations
Result: The studies were heterogeneous and all failed to report appropriate measurements of variation, which prevented meta-analysis and limited interpretation. Mean differences were within ±0.2 °C in five of seven tympanic thermometer/mode/temperature combinations and in the one oral thermometer studied; all three rectal thermometer/temperature combinations reported mean differences outside this range. The authors suggest tympanic and oral thermometry provide, on average, accurate measures of core temperature within the febrile range, while calling for further studies with appropriate statistical methods.
Watch out: Near-zero mean bias in 3 small febrile-range studies is not accuracy: a thermometer can average correctly while individual readings stray widely — which is exactly what the largest synthesis's limits of agreement show. This card must never be quoted without the limits-of-agreement card beside it.
Source: Jefferies S et al. A systematic review of the accuracy of peripheral thermometry in estimating core temperatures among febrile critically ill patients. Critical care and resuscitation : journal of the Australasian Academy of Critical Care Medicine. 2011. PMID: 21880009. Registry ID: none in the record. MeSH: Adult · Body Temperature/physiology · Confounding Factors, Epidemiologic · Critical Illness · Fever/diagnosis · Humans · Mouth · Rectum · Reproducibility of Results · Temporal Arteries · Thermometers · Tympanic Membrane
Card 4 — Synthesis: 3 overlapping syntheses — counted as syntheses, never as independent cohorts — agree that peripheral thermometers underestimate or stray widely from core temperature (limits of agreement beyond ±0.5 °C; fever sensitivity 64% in the largest synthesis; only oesophageal measurement clinically acceptable in ICU), against near-zero mean bias in the small, oldest febrile-range review
Study: Synthesis
Population: All syntheses carded on this shelf: Niven 2015 (75 studies, 8,682 patients), Cutuli 2021 (13 ICU cohort studies, 632 patients, 105,375 measurements) and Jefferies 2011 (3 febrile-ICU studies). Their primary-study sets overlap; the count on this shelf is 3 syntheses, not 3 independent cohorts.
Summary: Synthesis: 3 overlapping syntheses — counted as syntheses, never as independent cohorts — agree that peripheral thermometers underestimate or stray widely from core temperature (limits of agreement beyond ±0.5 °C; fever sensitivity 64% in the largest synthesis; only oesophageal measurement clinically acceptable in ICU), against near-zero mean bias in the small, oldest febrile-range review
Result: The headline is disagreement-plus-weakness, and it must be printed by method and population: tympanic, temporal, axillary and oral readings do not reach clinically acceptable agreement with central temperature where agreement has been pooled properly, and fever detection misses about a third of fevers in the largest synthesis. No card on this shelf implies any thermometer-guided fever-management pathway improves outcomes — this is a measurement shelf.
Watch out: See the individual cards for each synthesis's limits; this synthesis adds no new data.
Source: Synthesis of the reports carded above; no additional source.