Blood-glucose monitoring system accuracy under ISO 15197 — multi-system evaluations, 2003 to 2015 criteria

Shelf #7857 · Diabetes, glucose & metabolism

PREDICTIVE PERFORMANCE — system accuracy variable and verdict-specific: 16/27 systems met ISO 15197:2003, 14/18 met EN ISO 15197:2015 criteria, lay-user use cost one of five systems its pass; every verdict is system-plus-lot-plus-criteria-version and no verdict generalises across a manufacturer's range. A device-accuracy stamp, not a treatment stamp

Stamped: 2026-11-28.

Evidence published: 2003–2015.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurate are blood-glucose monitoring systems for self-monitoring when evaluated as systems (meter plus strip lot) against laboratory comparison methods under the ISO 15197 standard — what proportion of systems met the standard in the 27-system evaluation under the 2003 criteria and in the 18-system evaluation under the tightened 2015 criteria, how do results differ between lay users and study personnel for the same systems, and how much does the choice of comparison method (glucose oxidase vs hexokinase) move the verdict?

Balance of the evidence

Device accuracy with product naming binding the build: each evaluation names its systems and strip lots, and every verdict attaches to system-plus-lot-plus-criteria-version (2003 limits are looser than 2013/2015) — never generalised to a manufacturer's other systems or later lots. Only 16 of 27 systems met ISO 15197:2003 and 14 of 18 met the tightened 2015 criteria; three of five anchors are one programme (Ulm), printed as a programme, with an independent four-system study as counterweight; the comparison method (glucose oxidase vs hexokinase) moves individual-lot verdicts and prints with every figure. Not a treatment shelf.

MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Blood Glucose/analysis · Blood Glucose Self-Monitoring/instrumentation/standards · Diabetes Mellitus, Type 1/blood · Diabetes Mellitus, Type 2/blood · Female · Germany · Humans · Male · Middle Aged · Reagent Strips/chemistry · Self Care/instrumentation · Young Adult · Biomarkers/analysis · Blood Glucose Self-Monitoring/instrumentation/methods/standards · Diabetes Mellitus, Type 1/blood/epidemiology · Diabetes Mellitus, Type 2/blood/epidemiology · Europe/epidemiology · Follow-Up Studies · Patient Acceptance of Health Care · Point-of-Care Systems/standards · Prognosis · Reference Standards · Reproducibility of Results · Research Design · Diabetes Mellitus/blood · Glucose Oxidase/analysis · Hexokinase/analysis · Reagent Strips · Biomarkers/blood · Blood Glucose/metabolism · Case-Control Studies · Diabetes Mellitus, Type 1/blood/diagnosis · Diabetes Mellitus, Type 2/blood/diagnosis · Equipment Design · Materials Testing · Predictive Value of Tests · Blood Glucose Self-Monitoring/methods · Insulin/administration & dosage

Cards (6)

Card 1 — 27-system evaluation under ISO 15197:2003 (Ulm programme, programme-labelled) — only 16 of 27 named systems from 18 manufacturers met the minimum accuracy requirements (≥95% of results within ±15 mg/dL below 75 mg/dL and ±20% at or above); mean proportion of acceptable results 95.2 ± 5.2% (range 80.0–100.0%)

Study: Multi-system accuracy evaluation, 27 systems (Ulm evaluation programme)

Population: 27 blood glucose monitoring systems evaluated as systems (meter plus strip lot) with duplicate measurements of 100 blood samples from ≥100 subjects, against a glucose-oxidase comparison (24 systems) or hexokinase comparison (3 systems). Every system is named in the source; verdicts attach to system-plus-lot-plus-criteria-version and are not generalised here to any manufacturer's other systems or later lots.

Summary: 27-system evaluation under ISO 15197:2003 (Ulm programme, programme-labelled) — only 16 of 27 named systems from 18 manufacturers met the minimum accuracy requirements (≥95% of results within ±15 mg/dL below 75 mg/dL and ±20% at or above); mean proportion of acceptable results 95.2 ± 5.2% (range 80.0–100.0%)

Result: More than 40% of CE-labelled systems did NOT fulfil DIN EN ISO 15197:2003. The authors' clinical-risk framing is the source's own: inaccurate systems bear the risk of false treatment decisions and possible severe health injury.

Watch out: The 2003 criteria are looser than the 2013/2015 criteria on the later cards — a 'met the standard' verdict is meaningless without its criteria version.

Source: Freckmann G et al. System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197. Diabetes Technology & Therapeutics. 2010. PMID: 20151773. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Blood Glucose/analysis · Blood Glucose Self-Monitoring/instrumentation/standards · Diabetes Mellitus, Type 1/blood · Diabetes Mellitus, Type 2/blood · Female · Germany · Humans · Male · Middle Aged · Reagent Strips/chemistry · Self Care/instrumentation · Young Adult

Card 2 — 18-system evaluation under EN ISO 15197:2015 (Ulm programme) — 14 of 18 named CE-marked current-generation systems had ≥95% of results within ±15 mg/dL (0.83 mmol/L) / ±15%, 3 reached the ±10 band, and the smallest deviation containing ≥95% of results ranged from ±7.7 to ±19.7 mg/dL / %

Study: Multi-system accuracy evaluation, 18 systems (Ulm evaluation programme)

Population: 18 named systems (ABRA, Accu-Chek Guide, AURUM, CareSens Dual, CERA-CHEK 1CODE, ContourNext One, eBsensor, FreeStyle Freedom Lite, GL50 evo, GlucoCheck GOLD, GlucoMen areo 2K, GluNEO, MyStar DoseCoach, OneTouch Verio Flex, Pic GlucoTest, Rightest GM700S, TRUEyou, WaveSense JAZZ Wireless — named in the source and named here per the standing product rule), one reagent lot each, capillary blood from 100 subjects, criteria applied at 100 mg/dL (5.55 mmol/L).

Summary: 18-system evaluation under EN ISO 15197:2015 (Ulm programme) — 14 of 18 named CE-marked current-generation systems had ≥95% of results within ±15 mg/dL (0.83 mmol/L) / ±15%, 3 reached the ±10 band, and the smallest deviation containing ≥95% of results ranged from ±7.7 to ±19.7 mg/dL / %

Result: Four of 18 current-generation systems did not reach ≥95% within ±15 mg/dL (0.83 mmol/L) / ±15%, and even among passing systems the ≥95% deviation band varied more than twofold (±7.7 to ±19.7). The authors conclude not all CE-labelled systems fulfil ISO 15197 reliably.

Watch out: One lot per system: a pass or fail here is a verdict on that system with that lot under 2015 criteria, not on the brand.

Source: Pleus S et al. System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015. BMJ Open Diabetes Research & Care. 2020. PMID: 31958308. Registry ID: NCT03737188. MeSH: Adult · Aged · Aged, 80 and over · Biomarkers/analysis · Blood Glucose/analysis · Blood Glucose Self-Monitoring/instrumentation/methods/standards · Diabetes Mellitus, Type 1/blood/epidemiology · Diabetes Mellitus, Type 2/blood/epidemiology · Europe/epidemiology · Female · Follow-Up Studies · Humans · Male · Middle Aged · Patient Acceptance of Health Care · Point-of-Care Systems/standards · Prognosis · Reference Standards · Reproducibility of Results · Research Design · Young Adult

Card 3 — Multi-system evaluation under ISO 15197:2013 with THREE reagent lots and TWO comparison methods in parallel (Ulm programme) — 7 systems met the 2013 criteria with all lots under BOTH methods, and individual-lot verdicts differed slightly between the glucose-oxidase and hexokinase comparisons, so the comparison method prints with every figure

Study: Multi-system accuracy evaluation, two comparison methods (Ulm evaluation programme)

Population: European-market systems for self-monitoring evaluated per ISO 15197:2013: ≥95% of results within ±15% or ±0.83 mmol/L (15 mg/dL) of the comparison result (threshold 5.55 mmol/L, i.e. 100 mg/dL), plus consensus-error-grid zones A+B.

Summary: Multi-system evaluation under ISO 15197:2013 with THREE reagent lots and TWO comparison methods in parallel (Ulm programme) — 7 systems met the 2013 criteria with all lots under BOTH methods, and individual-lot verdicts differed slightly between the glucose-oxidase and hexokinase comparisons, so the comparison method prints with every figure

Result: Seven systems complied with all lots under both comparison methods, with 100% of results in error-grid zones A and B; eight systems met the older 2003 criteria with all lots and one with two lots. The authors conclude the comparison method itself can have a considerable impact on the accuracy data obtained for a system.

Watch out: Glucose values print in both units as the source pairs them (0.83 mmol/L = 15 mg/dL; 5.55 mmol/L = 100 mg/dL) per the US-units rule.

Source: Freckmann G et al. System Accuracy Evaluation of Different Blood Glucose Monitoring Systems Following ISO 15197:2013 by Using Two Different Comparison Methods. Diabetes Technology & Therapeutics. 2015. PMID: 26110670. Registry ID: NCT01909687. MeSH: Blood Glucose/analysis · Blood Glucose Self-Monitoring/instrumentation/standards · Diabetes Mellitus/blood · Glucose Oxidase/analysis · Hexokinase/analysis · Humans · Reagent Strips · Reference Standards · Reproducibility of Results

Card 4 — Independent four-system evaluation (the counterweight to the Ulm programme) — four named systems (Gold-Accu Meter, Nova StatStrip Xpress Glucose Hospital Meter, FreeStyle Optium Neo H, Accu-Chek Performa; eight meter variants) ALL met both the 2003 and 2013 criteria, with all results in consensus-error-grid zones A and B

Study: Independent multi-system accuracy evaluation, 4 systems / 8 variants

Population: Four named systems evaluated in China against ISO 15197:2003 and ISO 15197:2013 criteria, registered ChiCTR-DDD-16008436.

Summary: Independent four-system evaluation (the counterweight to the Ulm programme) — four named systems (Gold-Accu Meter, Nova StatStrip Xpress Glucose Hospital Meter, FreeStyle Optium Neo H, Accu-Chek Performa; eight meter variants) ALL met both the 2003 and 2013 criteria, with all results in consensus-error-grid zones A and B

Result: All four brands (eight variants) met both standards — a cleaner result than the larger European evaluations, on a much smaller system sample; the contrast is a sampling point, not a contradiction, and prints as such.

Watch out: Four systems cannot estimate a market-wide pass rate; this card is the programme-external counterweight, weighted accordingly.

Source: Chen H et al. The accuracy evaluation of four blood glucose monitoring systems according to ISO 15197:2003 and ISO 15197:2013 criteria. Primary Care Diabetes. 2019. PMID: 30770203. Registry ID: ChiCTR-DDD-16008436. MeSH: Adult · Aged · Biomarkers/blood · Blood Glucose/metabolism · Blood Glucose Self-Monitoring/instrumentation/standards · Case-Control Studies · Diabetes Mellitus, Type 1/blood/diagnosis · Diabetes Mellitus, Type 2/blood/diagnosis · Equipment Design · Female · Humans · Male · Materials Testing · Middle Aged · Predictive Value of Tests · Reproducibility of Results

Card 5 — Five named systems in the hands of LAY USERS vs study personnel (ISO 15197:2013) — 4 of 5 systems met the criteria with lay users (≥95% within limits) and all performed at ≥99.5% with personnel; modelled median insulin-dosing errors were −0.8 to +0.6 units (lay users), printed as modelled; frequent lay-user handling errors are catalogued

Study: Multi-system accuracy evaluation in lay-user and professional hands

Population: Five named systems (Accu-Chek Aviva Connect, Contour Next One, FreeStyle Freedom Lite, GlucoMen areo, OneTouch Verio — as lettered A–E in the source), one test-strip lot each, used by adult lay users and by study personnel.

Summary: Five named systems in the hands of LAY USERS vs study personnel (ISO 15197:2013) — 4 of 5 systems met the criteria with lay users (≥95% within limits) and all performed at ≥99.5% with personnel; modelled median insulin-dosing errors were −0.8 to +0.6 units (lay users), printed as modelled; frequent lay-user handling errors are catalogued

Result: User hands cost accuracy: one system that passed with personnel failed the ≥95% criterion with lay users. All five systems' results stayed within clinically acceptable error-grid zones. Frequent lay-user errors were not checking strip expiry, applying blood incorrectly and handling the device incorrectly.

Watch out: Modelled dosing errors are modelled — they print with that word attached, never as observed patient harm.

Source: Jendrike N et al. Accuracy of five systems for self-monitoring of blood glucose in the hands of adult lay-users and professionals applying ISO 15197:2013 accuracy criteria and potential insulin dosing errors. Current Medical Research and Opinion. 2019. PMID: 29927663. Registry ID: NCT03033849. MeSH: Adult · Blood Glucose/analysis · Blood Glucose Self-Monitoring/methods · Humans · Insulin/administration & dosage · Reproducibility of Results

Card 6 — Synthesis: a device-accuracy shelf, not a treatment shelf — no card implies any system's use improves glycaemic outcomes. Verdicts attach to system-plus-lot-plus-criteria-version (2003 vs 2013/2015 limits differ) and never generalise to a manufacturer's other systems or later lots; each evaluation names its systems in its source and they are named on these cards per the standing product rule. Three of five anchors come from one evaluation programme (Ulm) and print as a programme, with the independent four-system study as counterweight. The pass picture: 16/27 under the 2003 criteria, 14/18 under the tightened 2015 criteria, 7 systems clean under both comparison methods in 2013 testing, and a lay-user penalty in the five-system study. The clinical-risk framing is the sources' own (inaccurate results risk false insulin-dosing decisions; modelled errors −0.8 to +0.6 units).

Study: Synthesis

Population: Synthesis of the reports carded above; no additional population.

Summary: Synthesis: a device-accuracy shelf, not a treatment shelf — no card implies any system's use improves glycaemic outcomes. Verdicts attach to system-plus-lot-plus-criteria-version (2003 vs 2013/2015 limits differ) and never generalise to a manufacturer's other systems or later lots; each evaluation names its systems in its source and they are named on these cards per the standing product rule. Three of five anchors come from one evaluation programme (Ulm) and print as a programme, with the independent four-system study as counterweight. The pass picture: 16/27 under the 2003 criteria, 14/18 under the tightened 2015 criteria, 7 systems clean under both comparison methods in 2013 testing, and a lay-user penalty in the five-system study. The clinical-risk framing is the sources' own (inaccurate results risk false insulin-dosing decisions; modelled errors −0.8 to +0.6 units).

Result: Balance: CE marking does not guarantee ISO accuracy for a given system and lot; accuracy is an evaluated property, not a label property.

Source: Synthesis of the reports carded above; no additional source.