HHIE-S: ten-item hearing-handicap screen whose accuracy moves with the audiometric definition, the cut-off (>8 vs >11) and the spectrum — 4 reports
Shelf #8114 · Eyes, skin, nose & allergy
SCREENING ACCURACY — definition- and spectrum-dependent: sensitivity 53–72% / specificity 70–84% across five audiometric definitions at >8 (derivation, n = 178); 89.1% / 75.0% in an ENT-complaint clinic (n = 138); 86.4% / 72.4% at a migrated cut-off of >11 in Italian outpatients (n = 167); population-spectrum sensitivity only 34% at >8 (cross-row, #8115); composition 4 reports; cross-points #8115 and #7813; a screening-accuracy stamp, not a hearing-treatment stamp
Stamped: 2026-11-28.
Evidence published: 1988–2023.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately does the HHIE-S (Hearing Handicap Inventory for the Elderly — Screening version: ten self-rated social / emotional hearing-handicap items, scored 0–40, conventionally positive above 8) identify hearing impairment against pure-tone audiometry in older adults — what sensitivity and specificity did its original diagnostic-performance study find across definitions of hearing loss, what accuracy did the Brazilian study find, how does it compare head-to-head with a single hearing question in a Japanese community sample, and how did the Italian version perform?
Balance of the evidence
This is a screening-accuracy shelf. The audiometric definition moves the figures materially (the derivation's ranges span five definitions of hearing loss) and every pair prints with its definition and cut-off — the cut-off itself migrates (>8 in the original study vs >11 as the Italian optimum). The HHIE-S measures perceived handicap, not thresholds. Cross-row caution, printed as directed: in a population sample (Beaver Dam, PMID 9776206 — carded on the single-question shelf #8115 for its single-question limb, where its HHIE-S limb is a comparator) the HHIE-S's sensitivity at >8 was only 34% — clinic and complaint spectra (89.1% sensitivity in the Brazilian ENT-complaint study) and population spectra give very different sensitivities for the same instrument. PMID 22833152 anchors THIS shelf only: its single-question limb is a comparator here; the single question's own shelf is #8115, on different reports — neither report is an anchor on both packs. The three hearing rows cross-point: this shelf, #8115 (single self-report question) and pack #7813 (whispered voice test — an examiner-administered bedside manoeuvre).
MeSH: Aged · Aged, 80 and over · Audiometry, Pure-Tone · Female · Hearing Disorders/classification,diagnosis,psychology · Humans · Male · Psychological Tests/methods · Depression/diagnosis,psychology · Geriatric Assessment/methods · Hearing Loss/diagnosis,psychology · Japan · Logistic Models · Loneliness/psychology · Middle Aged · Personality Inventory · Quality of Life · Reproducibility of Results · Self Report · Sensitivity and Specificity · Surveys and Questionnaires · Hearing · Deafness · Psychological Well-Being · Psychometrics
Cards (5)
Card 1 — Original diagnostic-performance study (Lichtenstein 1988, Ear and hearing — 178 elderly primary-care subjects against pure-tone audiometry under five definitions of hearing loss): at a cut-off of >8 sensitivity ranged 53–72% and specificity 70–84% across definitions, with similar ROC / likelihood-ratio behaviour regardless of definition
Study: Diagnostic-performance study (primary report) — derivation-spectrum ranges across five definitions
Population: 178 elderly subjects screened in primary care; hearing loss assessed by pure-tone audiometry under five definitions.
Summary: Original diagnostic-performance study (Lichtenstein 1988, Ear and hearing — 178 elderly primary-care subjects against pure-tone audiometry under five definitions of hearing loss): at a cut-off of >8 sensitivity ranged 53–72% and specificity 70–84% across definitions, with similar ROC / likelihood-ratio behaviour regardless of definition
Result: Using a score of greater than 8 as the cut point, the HHIE-S had sensitivities ranging from 53 to 72% and specificities ranging from 70 to 84% across the different definitions of hearing loss. The HHIE-S receiver-operating characteristics and likelihood ratios were similar regardless of the hearing-loss definition used. The authors conclude the HHIE-S is a valid, robust test for identifying hearing-impaired elderly irrespective of the audiometric definition used, as published.
Watch out: Ranges, not a pair: sensitivity 53–72% and specificity 70–84% span five audiometric definitions — no single pair may be quoted from this study without its definition. A primary-care screened spectrum from 1988.
Source: Lichtenstein et al. Diagnostic performance of the hearing handicap inventory for the elderly (screening version) against differing definitions of hearing loss. Ear and hearing. 1988. PMID: 3169401. Registry ID: none in the record. MeSH: Aged · Aged, 80 and over · Audiometry, Pure-Tone · Female · Hearing Disorders/classification,diagnosis,psychology · Humans · Male · Psychological Tests/methods
Card 2 — Brazilian accuracy study (Servidoni 2018, International archives of otorhinolaryngology — 138 Brazilians aged >60 with ENT complaints): HHIE-S diagnostic accuracy was 86.2%, sensitivity 89.1% and specificity 75.0% against audiometry
Study: Cross-sectional accuracy study (primary report)
Population: 138 individuals aged over 60 with otorhinolaryngological complaints, recruited at a university otorhinolaryngology and speech-therapy clinic in Marília, Brazil; patients already in auditory rehabilitation and those without the oral understanding needed for interview were excluded. Audiometric prevalence of hearing loss was 79.7%; questionnaire prevalence was 76.1%.
Summary: Brazilian accuracy study (Servidoni 2018, International archives of otorhinolaryngology — 138 Brazilians aged >60 with ENT complaints): HHIE-S diagnostic accuracy was 86.2%, sensitivity 89.1% and specificity 75.0% against audiometry
Result: The diagnostic accuracy of the HHIE-S was 86.2%, with sensitivity of 89.1% and specificity of 75.0%, regardless of gender. The authors conclude the standardised questionnaire is suitable for screening hearing loss in the elderly given its high accuracy and user-friendly quality, as published.
Watch out: A complaint spectrum: everyone had ENT complaints and audiometric prevalence was 79.7% — the 89.1% sensitivity is a clinic figure. In a population spectrum the same instrument at the same cut-off reached only 34% sensitivity (Beaver Dam, carded on #8115); the two figures print as a pair of spectra, never as one performance claim. This record carries no MeSH terms (printed as none in the record on its source line).
Source: Servidoni et al. Hearing Loss in the Elderly: Is the Hearing Handicap Inventory for the Elderly - Screening Version Effective in Diagnosis When Compared to the Audiometric Test?. International archives of otorhinolaryngology. 2018. PMID: 29371892. Registry ID: none in the record. MeSH: none in the record
Card 3 — Japanese community head-to-head (Tomioka 2013, Quality of life research : an international journal of quality of life aspects of treatment, care and rehabilitation — 1,731 questionnaire respondents; 197 with audiometry / test–retest): HHIE-S α 0.91, ICC 0.85; the HHIE-S had significantly lower sensitivity for >25 dB impairment but significantly higher specificity for >40 dB impairment than the single question, and stronger associations with quality-of-life measures — its single-question limb is a comparator here; the single question's own row is #8115, on different reports; this report anchors #8114 only
Study: Head-to-head accuracy / reliability study (primary report) — anchors this shelf only; single-question limb comparator-only here
Population: Community elderly in Japan (781 males and 950 females) completing a self-report questionnaire including the HHIE-S and a single question (SQ); among them, 97 males and 100 females also underwent test–retest and auditory testing, with criterion validity tested against pure-tone averages.
Summary: Japanese community head-to-head (Tomioka 2013, Quality of life research : an international journal of quality of life aspects of treatment, care and rehabilitation — 1,731 questionnaire respondents; 197 with audiometry / test–retest): HHIE-S α 0.91, ICC 0.85; the HHIE-S had significantly lower sensitivity for >25 dB impairment but significantly higher specificity for >40 dB impairment than the single question, and stronger associations with quality-of-life measures — its single-question limb is a comparator here; the single question's own row is #8115, on different reports; this report anchors #8114 only
Result: For the HHIE-S, Cronbach's alpha was 0.91, the Spearman-Brown coefficient 0.90 and the intra-class correlation coefficient 0.85 (the SQ's test–retest kappa was 0.65). The HHIE-S had significantly lower sensitivity in identifying >25 dB hearing impairment, but significantly higher specificity for detection of >40 dB impairment, than the SQ. The HHIE-S had stronger associations with quality-of-life measures (subjective well-being, depressive symptoms, loneliness and physical functioning) than the SQ. The authors conclude the HHIE-S is the more effective instrument for assessing hearing impairment and quality-of-life research in the elderly community, as published.
Watch out: Anchor discipline: this report is carded here for its HHIE-S limb; its single-question limb is comparator-only on this shelf and is not an anchor on #8115. The comparison is severity-dependent — the HHIE-S wins on specificity at >40 dB and loses on sensitivity at >25 dB; neither direction may be printed without its threshold.
Source: Tomioka et al. The Hearing Handicap Inventory for Elderly-Screening (HHIE-S) versus a single question: reliability, validity, and relations with quality of life measures in the elderly community, Japan. Quality of life research : an international journal of quality of life aspects of treatment, care and rehabilitation. 2013. PMID: 22833152. Registry ID: none in the record. MeSH: Aged · Aged, 80 and over · Audiometry, Pure-Tone · Depression/diagnosis,psychology · Female · Geriatric Assessment/methods · Hearing Loss/diagnosis,psychology · Humans · Japan · Logistic Models · Loneliness/psychology · Male · Middle Aged · Personality Inventory · Quality of Life · Reproducibility of Results · Self Report · Sensitivity and Specificity · Surveys and Questionnaires
Card 4 — Italian validation (Apa 2023, Acta otorhinolaryngologica Italica : organo ufficiale della Societa italiana di otorinolaringologia e chirurgia cervico-facciale — 167 elderly outpatients): α 0.908, significant test–retest reliability, moderate-to-high correlations with better-ear pure-tone average, and at the best cut-off of >11 sensitivity was 86.4%, specificity 72.4%, PPV 52.8% and NPV 93.7%
Study: Cross-cultural validation (primary report)
Population: 167 elderly outpatients administered the Italian HHIE-S (HHIE-S-It) together with the Psychological General Well-Being Index, after translation.
Summary: Italian validation (Apa 2023, Acta otorhinolaryngologica Italica : organo ufficiale della Societa italiana di otorinolaringologia e chirurgia cervico-facciale — 167 elderly outpatients): α 0.908, significant test–retest reliability, moderate-to-high correlations with better-ear pure-tone average, and at the best cut-off of >11 sensitivity was 86.4%, specificity 72.4%, PPV 52.8% and NPV 93.7%
Result: Cronbach's alpha was 0.908 for the total score (0.832 and 0.816 for the two subscales). Significant test–retest reliability was observed (p < 0.001). Moderate-to-high correlations were found between HHIE-S-It and pure-tone average in the better ear (p < 0.001), and scores differed significantly across hearing-loss degree groups (p < 0.001). The criterion HHIE-S-It >11 was the best cut-off, with sensitivity 86.4%, specificity 72.4%, positive predictive value 52.8%, negative predictive value 93.7% and likelihood ratios 3.12 and 0.19. The authors conclude its adoption is justified for clinical and research purposes and as a screening tool for age-related hearing loss, as published.
Watch out: Cut-off migration prints: the Italian optimum is >11, not the original >8 — a cut-off quoted without its study is meaningless. At that cut-off the PPV was only 52.8% in this outpatient spectrum: about half of screen-positives did not have the target impairment.
Source: Apa et al. Italian validation of the Hearing Handicap Inventory for Elderly - Screening version (HHIE-S-It). Acta otorhinolaryngologica Italica : organo ufficiale della Societa italiana di otorinolaringologia e chirurgia cervico-facciale. 2023. PMID: 37488990. Registry ID: none in the record. MeSH: Aged · Humans · Reproducibility of Results · Hearing · Deafness · Psychological Well-Being · Psychometrics
Card 5 — Synthesis (4 reports): the HHIE-S identifies audiometric hearing impairment with definition- and spectrum-dependent accuracy — sensitivity 53–72% across five definitions at >8 in its derivation, 89.1% in an ENT-complaint clinic, 86.4% at a migrated >11 cut-off in Italian outpatients — and only 34% in a population sample (cross-row figure from #8115); cross-points #8115 and #7813
Study: Synthesis
Population: Synthesis of the 4 reports carded above (derivation n = 178; Brazilian clinic study n = 138; Japanese community study 1,731 respondents / 197 tested; Italian validation n = 167). Cross-points: the single self-report hearing question is shelved at #8115 and the whispered voice test at pack #7813; PMID 22833152 is an anchor here only, its single-question limb comparator-only.
Summary: Synthesis (4 reports): the HHIE-S identifies audiometric hearing impairment with definition- and spectrum-dependent accuracy — sensitivity 53–72% across five definitions at >8 in its derivation, 89.1% in an ENT-complaint clinic, 86.4% at a migrated >11 cut-off in Italian outpatients — and only 34% in a population sample (cross-row figure from #8115); cross-points #8115 and #7813
Result: Across four spectra the HHIE-S is reliable (α 0.91–0.908; ICC 0.85) and its accuracy tracks the audiometric definition and the population tested: derivation ranges of 53–72% sensitivity and 70–84% specificity across five definitions at >8; 89.1% / 75.0% where four in five patients had audiometric loss; a severity-dependent trade against the single question in the community (less sensitive at >25 dB, more specific at >40 dB); and 86.4% / 72.4% at a best cut-off of >11 in Italian outpatients. The population-spectrum caution from the Beaver Dam study (sensitivity 34% at >8, carded on #8115) bounds all of these clinic figures.
Watch out: The HHIE-S measures perceived hearing handicap: a high score is a handicap report, not a threshold measurement, and a low score does not exclude audiometric loss — least of all in unselected populations.
Source: Synthesis of the reports carded above; no additional source.