SourceMed · Shelf #8438 in the ask interface
Transcutaneous bilirubinometry — correlation with serum bilirubin r=0.83 in preterm infants before phototherapy, degrading to 0.65–0.71 during phototherapy; the only sensitivity/specificity pair in the base is a predischarge prediction pair (100% / 88.1%) from 1 primary study of 490 newborns
Shelf #8438 · Children’s health
PREDICTIVE PERFORMANCE — correlations and Bland–Altman differences, not manufactured diagnostic pairs: pooled correlation with serum bilirubin was r=0.83 at each site in preterm infants before phototherapy (r=0.89 under 32 weeks), degrading under phototherapy to 0.71 at covered sites and 0.65 at uncovered sites in term/near-term infants, and improving only marginally after phototherapy (0.72); the only sensitivity/specificity pair in this base answers a different question — a predischarge value above the 75th-percentile track predicting later excessive hyperbilirubinaemia at sensitivity 100% / specificity 88.1% (490 newborns) — and the degradation under phototherapy prints at headline weight
Stamped: 2026-11-28.
Evidence published: 2000–2016.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In newborn infants, how accurately does transcutaneous bilirubinometry estimate serum bilirubin — in preterm infants before phototherapy, in term and near-term infants during and after phototherapy, and as a predischarge predictor of later excessive hyperbilirubinaemia?
Balance of the evidence
3 reports: 2 syntheses from the same programme (Nagar/Kumar — 2013 preterm/before-phototherapy spectrum and 2016 term/near-term during/after-phototherapy spectrum; the shared programme prints) and 1 independent primary study (490 newborns). Devices are named as the publications give them (JM-103, BiliCheck — SpectRx Inc in the primary's record) under the product-naming exception. Cross-points in prose only: visual assessment of neonatal jaundice is homed at pack #8270 (a different index test — the eye, not the device); the 2024 Cochrane review (PMID 38804265) is a screening-effectiveness review resting on one RCT, not an accuracy synthesis, and is not carded.
MeSH: Bilirubin/analysis,blood* · Equipment Design/standards* · Forehead/blood supply · Humans · Infant, Newborn · Infant, Premature/blood* · Jaundice, Neonatal/blood,diagnosis · Neonatal Screening/instrumentation,standards* · Reproducibility of Results · Bilirubin/blood* · Equipment and Supplies · Hyperbilirubinemia, Neonatal/diagnosis · Infant, Premature/blood · Jaundice, Neonatal/blood,diagnosis* · Neonatal Screening/instrumentation*,methods · Phototherapy*/adverse effects,methods · Premature Birth/blood* · Skin · Term Birth/blood* · Analysis of Variance · Asian People · Black People · Chromatography, High Pressure Liquid/methods · Cost-Benefit Analysis · Equipment Design · Fiber Optic Technology · Jaundice, Neonatal/blood,diagnosis*,ethnology*,therapy · Microspectrophotometry · Neonatal Screening/economics,instrumentation* · Phototherapy · Point-of-Care Systems · Predictive Value of Tests · Risk Assessment · Sensitivity and Specificity · United States/epidemiology · White People · Black or African American
Cards (4)
Card 1 — Preterm synthesis (Nagar 2013, Pediatrics — systematic review, 22 studies, before phototherapy): pooled correlation r=0.83 at each site, r=0.89 (95% CI 0.82–0.93) under 32 weeks; JM-103 more precise than BiliCheck at the sternum (SD of differences 24.3 vs 31.98 µmol/L, ≈1.4 vs ≈1.9 mg/dL)
Study: Systematic review (synthesis; same programme as the 2016 synthesis, different spectrum)
Population: Preterm infants before phototherapy in 22 studies comparing transcutaneous bilirubin with total serum bilirubin; 21 studies reported correlation coefficients and 13 contributed Bland–Altman plots.
Summary: Preterm synthesis (Nagar 2013, Pediatrics — systematic review, 22 studies, before phototherapy): pooled correlation r=0.83 at each site, r=0.89 (95% CI 0.82–0.93) under 32 weeks; JM-103 more precise than BiliCheck at the sternum (SD of differences 24.3 vs 31.98 µmol/L, ≈1.4 vs ≈1.9 mg/dL)
Result: Pooled correlation was r=0.83 for each site of measurement, and r=0.89 (95% CI 0.82–0.93) in infants under 32 weeks' gestation. Bland–Altman analysis showed negligible bias at forehead or sternum with either device, but the JM-103 exhibited better precision than BiliCheck (SD for transcutaneous-minus-serum differences 24.3 vs 31.98 µmol/L, ≈1.4 vs ≈1.9 mg/dL), as published.
Watch out: A correlation of 0.83 is not a sensitivity/specificity pair and none may be manufactured from it; precision differences between devices print beside the pooled correlation.
Source: Nagar G et al. Reliability of transcutaneous bilirubin devices in preterm infants: a systematic review. Pediatrics. 2013. PMID: 24127472. Registry ID: none in the record. MeSH: Bilirubin/analysis,blood* · Equipment Design/standards* · Forehead/blood supply · Humans · Infant, Newborn · Infant, Premature/blood* · Jaundice, Neonatal/blood,diagnosis · Neonatal Screening/instrumentation,standards* · Reproducibility of Results
Card 2 — Phototherapy synthesis (Nagar 2016, Neonatology — systematic review and meta-analysis, 14 studies, term and near-term infants during/after phototherapy): pooled correlations 0.71 at covered sites and 0.65 at uncovered sites during phototherapy
Study: Systematic review and meta-analysis (synthesis; same programme as the 2013 synthesis)
Population: Term and near-term infants receiving phototherapy or in the post-phototherapy phase in 14 studies comparing transcutaneous bilirubin with total serum bilirubin.
Summary: Phototherapy synthesis (Nagar 2016, Neonatology — systematic review and meta-analysis, 14 studies, term and near-term infants during/after phototherapy): pooled correlations 0.71 at covered sites and 0.65 at uncovered sites during phototherapy
Result: Pooled correlations during phototherapy were 0.71 (95% CI 0.64–0.77) at covered sites and 0.65 (0.55–0.74) at uncovered sites (forehead 0.70, sternum 0.64), improving only marginally after phototherapy (0.72 (0.64–0.78)). The two Bland–Altman studies printed mean transcutaneous-minus-serum differences of −29.2 and 30 µmol/L (≈−1.7 and ≈1.8 mg/dL), as published.
Watch out: Accuracy degrades exactly when bilirubin matters most clinically — during treatment — and the two Bland–Altman means point in opposite directions, so no single bias correction can be quoted.
Source: Nagar G et al. Effect of Phototherapy on the Reliability of Transcutaneous Bilirubin Devices in Term and Near-Term Infants: A Systematic Review and Meta-Analysis. Neonatology. 2016. PMID: 26789390. Registry ID: none in the record. MeSH: Bilirubin/blood* · Equipment and Supplies · Humans · Hyperbilirubinemia, Neonatal/diagnosis · Infant, Newborn · Infant, Premature/blood · Jaundice, Neonatal/blood,diagnosis* · Neonatal Screening/instrumentation*,methods · Phototherapy*/adverse effects,methods · Premature Birth/blood* · Reproducibility of Results · Skin · Term Birth/blood*
Card 3 — Predischarge primary study (Bhutani 2000, Pediatrics — 490 term and near-term newborns at 2 institutions, BiliCheck device (SpectRx Inc, named in the record)): correlation with HPLC serum bilirubin r=0.91; above the 75th-percentile track, sensitivity 100% / specificity 88.1% for later excessive hyperbilirubinaemia
Study: Primary study (predischarge prediction)
Population: 490 term and near-term newborns (racially diverse; 1,788 paired samples) evaluated at 12–98 hours of age with the BiliCheck device, paired with heelstick serum bilirubin by HPLC and diazo methods.
Summary: Predischarge primary study (Bhutani 2000, Pediatrics — 490 term and near-term newborns at 2 institutions, BiliCheck device (SpectRx Inc, named in the record)): correlation with HPLC serum bilirubin r=0.91; above the 75th-percentile track, sensitivity 100% / specificity 88.1% for later excessive hyperbilirubinaemia
Result: Overall correlation of serum bilirubin (HPLC) with transcutaneous bilirubin was r=0.91. In the 24–72-hour subgroup, all 23 infants whose serum values put them at high risk (above the 95th-percentile track) had BiliCheck values above the 75th-percentile track: sensitivity 100%, specificity 88.1%, negative predictive value 100%, positive predictive value 32.86% and likelihood ratio 8.43, as published.
Watch out: The 100%/88.1% pair predicts later excessive hyperbilirubinaemia from a percentile track — it is not the accuracy of the device for the current bilirubin level, and the two questions must never be conflated; positive predictive value was only 32.86%.
Source: Bhutani VK et al. Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia. Pediatrics. 2000. PMID: 10920173. Registry ID: none in the record. MeSH: Analysis of Variance · Asian People · Black People · Chromatography, High Pressure Liquid/methods · Cost-Benefit Analysis · Equipment Design · Fiber Optic Technology · Humans · Infant, Newborn · Jaundice, Neonatal/blood,diagnosis*,ethnology*,therapy · Microspectrophotometry · Neonatal Screening/economics,instrumentation* · Phototherapy · Point-of-Care Systems · Predictive Value of Tests · Reproducibility of Results · Risk Assessment · Sensitivity and Specificity · United States/epidemiology · White People · Black or African American
Card 4 — Synthesis: a good estimator before phototherapy, a weaker one during it — 2 syntheses (one programme, two spectra) + 1 primary study (490 newborns)
Study: Synthesis
Population: 3 reports: Nagar 2013 (22 studies), Nagar 2016 (14 studies) — same programme, different spectra — and Bhutani 2000 (490 newborns).
Summary: Synthesis: a good estimator before phototherapy, a weaker one during it — 2 syntheses (one programme, two spectra) + 1 primary study (490 newborns)
Result: Correlations print as correlations; the only diagnostic pair is the primary's prediction pair. The phototherapy degradation (0.83 → 0.65–0.71) is headline material. Refused as an accuracy anchor: the 2024 Cochrane screening-effectiveness review (38804265), named in prose only. Cross-point: visual jaundice assessment at pack #8270.
Watch out: Do not convert any correlation here into a sensitivity or specificity, and do not use transcutaneous values during phototherapy as if they carried pre-phototherapy precision.
Source: Synthesis of the reports carded above; no additional source.