Bedside chest ultrasonography for haemothorax in trauma — pooled sensitivity 0.67 (vs radiography 0.54) with specificity 0.99, against primaries reporting 92–96%; the gated pneumothorax synthesis is a cross-point only
Shelf #8521 · Emergency & first hours
PREDICTIVE PERFORMANCE — more sensitive than supine radiography at the pooled level, with near-perfect specificity, but pooled sensitivity is only moderate and sits far below the primaries: pooled ultrasonography sensitivity 0.67 (0.41 to 0.86) / specificity 0.99 against radiography 0.54 / 0.99 across 12 studies / 7,361 patients (heterogeneity I² 68–99%), while the two primaries report 96.2% (25 of 26) and 92% sensitivity at 100% specificity — the gap prints at headline weight and is not averaged away.
Stamped: 2026-11-28.
Evidence published: 1997–2016.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In trauma patients, how accurately does bedside chest ultrasonography detect haemothorax against CT, thoracostomy or composite reference standards — and how does it compare with supine chest radiography?
Balance of the evidence
4 reports: 1 systematic review and meta-analysis (Rahimi-Movaghar 2016, 12 studies / 7,361 patients; TYPE-FLAG — typed Review only at PubMed although a meta-analysis on substance), 2 primaries (Ma & Mateer 1997, 240 patients analysed, identical counts reported for ultrasound and radiograph; Brooks 2004, 61 patients, FAST extension) and 1 best-evidence-topic review (McEwan & Thompson 2007, 6 best papers, no pooled pair in its record). The larger pneumothorax+haemothorax synthesis (PMID 29433802) is GATED inside pack #8414 and, per adjudication 4, is a cross-point in prose only; pack #8414 (pneumothorax ultrasound) is the neighbour question, cross-pointed.
MeSH: Adult · Hemothorax/diagnostic imaging* · Humans · Pleural Effusion/diagnostic imaging · Reproducibility of Results · Retrospective Studies · Sensitivity and Specificity · Thoracic Injuries/diagnostic imaging* · Thoracostomy · Tomography, X-Ray Computed · Ultrasonography · Acute Disease · Emergencies · Prospective Studies · Wounds, Nonpenetrating/diagnostic imaging* · Adolescent · Aged · Aged, 80 and over · Early Diagnosis · Emergency Medicine/instrumentation,methods · Hemothorax/diagnostic imaging*,etiology* · Middle Aged · Radiography · Thoracic Injuries/complications*
Cards (5)
Card 1 — Systematic review and meta-analysis (Rahimi-Movaghar 2016, Emergency (Tehran) — 12 studies, 7,361 patients): pooled ultrasonography sensitivity 0.67 (0.41–0.86) / specificity 0.99 (0.95–1.0), against radiography 0.54 (0.33–0.75) / 0.99 (0.94–1.0); heterogeneity I² 68–99% across the four pools — typed Review only at PubMed although a meta-analysis on substance
Study: Systematic review and meta-analysis (synthesis) — TYPE-FLAG: typed 'Review' only at PubMed although a meta-analysis on substance; counted on substance
Population: 12 studies with 7,361 trauma patients (77.1% male) comparing chest ultrasonography and radiography for haemothorax detection. Synthesis, counted as a synthesis.
Summary: Systematic review and meta-analysis (Rahimi-Movaghar 2016, Emergency (Tehran) — 12 studies, 7,361 patients): pooled ultrasonography sensitivity 0.67 (0.41–0.86) / specificity 0.99 (0.95–1.0), against radiography 0.54 (0.33–0.75) / 0.99 (0.94–1.0); heterogeneity I² 68–99% across the four pools — typed Review only at PubMed although a meta-analysis on substance
Result: Pooled sensitivity and specificity of ultrasonography for haemothorax were 0.67 (95% CI 0.41 to 0.86; I² = 68.38) and 0.99 (0.95 to 1.0; I² = 88.16); for radiography, 0.54 (0.33 to 0.75; I² = 92.85) and 0.99 (0.94 to 1.0; I² = 99.22). Subgroup analysis found the ultrasound operator, transducer frequency and sample size to be important sources of heterogeneity. The authors conclude ultrasonography's sensitivity is relatively higher than radiography's but still moderate, with both modalities' specificity excellent, and screening characteristics influenced by operator and transducer frequency.
Watch out: I² of 68–99% across all four pools means the pooled sensitivities are averages over deeply heterogeneous studies. The record's conclusion renders the pooled sensitivity as '0.67%' — read as the proportion 0.67, as its results section prints it.
Source: Rahimi-Movaghar et al. Application of Ultrasonography and Radiography in Detection of Hemothorax; a Systematic Review and Meta-Analysis. Emergency (Tehran, Iran). 2016. PMID: 27299139. Registry ID: none in the record. MeSH: none in the record
Card 2 — Primary study (Ma & Mateer 1997, Annals of Emergency Medicine — 240 trauma patients analysed of 245 enrolled, 26 with haemothorax): both ultrasound and the initial chest radiograph recorded 25 true positives, 1 false negative, 0 false positives and 214 true negatives (both 96.2% sensitive, 100% specific, 99.6% accurate as the record states) — the record reports identical counts for the two tests, and the counts are what are printed from this anchor
Study: Primary diagnostic comparison (retrospective analysis of a prospective study's data)
Population: A convenience sample of adult patients with major blunt or penetrating torso trauma at a Level I trauma centre over 17 months; 5 of 245 enrolled were excluded because tube thoracostomy preceded the ultrasound; haemothorax confirmed by tube thoracostomy or CT in 26 of 240.
Summary: Primary study (Ma & Mateer 1997, Annals of Emergency Medicine — 240 trauma patients analysed of 245 enrolled, 26 with haemothorax): both ultrasound and the initial chest radiograph recorded 25 true positives, 1 false negative, 0 false positives and 214 true negatives (both 96.2% sensitive, 100% specific, 99.6% accurate as the record states) — the record reports identical counts for the two tests, and the counts are what are printed from this anchor
Result: Both ultrasound examination and the initial chest radiograph resulted in 0 false-positive, 1 false-negative, 25 true-positive and 214 true-negative findings; overall both modalities were 96.2% sensitive, 100% specific and 99.6% accurate for haemothorax. Ultrasound interpretations were recorded before other diagnostic tests and were not used in management decisions; radiograph and CT interpretations were by attending radiologists not blinded to patient outcome. The authors conclude ultrasonography is comparable to the initial chest radiograph and may expedite diagnosis and treatment.
Watch out: The identical counts for ultrasound and radiograph sit in visible tension with the pooled comparison (0.67 vs 0.54 sensitivity) and print as reported, unresolved. Retrospective analysis; radiologists unblinded to outcome.
Source: Ma et al. Trauma ultrasound examination versus chest radiography in the detection of hemothorax. Annals of emergency medicine. 1997. PMID: 9055768. Registry ID: none in the record. MeSH: Adult · Hemothorax/diagnostic imaging* · Humans · Pleural Effusion/diagnostic imaging · Reproducibility of Results · Retrospective Studies · Sensitivity and Specificity · Thoracic Injuries/diagnostic imaging* · Thoracostomy · Tomography, X-Ray Computed · Ultrasonography
Card 3 — Primary FAST-extension study (Brooks 2004, Emergency Medicine Journal — 61 torso-trauma patients): 12 true positives, 48 true negatives, 0 false positives, 1 false negative; sensitivity 92%, specificity 100%, PPV 100%, NPV 98%; 4 haemothoraces seen on ultrasound were not apparent on the trauma chest radiograph
Study: Primary prospective diagnostic accuracy study (FAST extension)
Population: 61 adult patients with torso trauma (89% blunt) undergoing thoracic ultrasound as an extension of the FAST protocol by one of two non-radiologists trained in FAST, against supine chest radiography, intercostal drain, CT or thoracotomy at the attending physician's discretion.
Summary: Primary FAST-extension study (Brooks 2004, Emergency Medicine Journal — 61 torso-trauma patients): 12 true positives, 48 true negatives, 0 false positives, 1 false negative; sensitivity 92%, specificity 100%, PPV 100%, NPV 98%; 4 haemothoraces seen on ultrasound were not apparent on the trauma chest radiograph
Result: Twelve patients had a haemothorax detected by ultrasound and confirmed by CT or tube thoracostomy; four of these were not apparent on trauma chest radiography. There were 12 true positives, 48 true negatives, no false positives and one false negative scan, giving sensitivity 92%, specificity 100%, positive predictive value 100% and negative predictive value 98% for haemothorax after trauma. The authors conclude emergency chest ultrasound in the primary survey can rapidly and accurately diagnose haemothorax.
Watch out: 61 patients and 13 haemothoraces at one centre with two trained operators — the 92% sensitivity is a small, expert-operator figure and prints beside the pooled 0.67, never averaged with it.
Source: Brooks et al. Emergency ultrasound in the acute assessment of haemothorax. Emergency medicine journal : EMJ. 2004. PMID: 14734374. Registry ID: none in the record. MeSH: Acute Disease · Emergencies · Hemothorax/diagnostic imaging* · Humans · Prospective Studies · Sensitivity and Specificity · Ultrasonography · Wounds, Nonpenetrating/diagnostic imaging*
Card 4 — Best-evidence-topic review (McEwan & Thompson 2007, Emergency Medicine Journal — 6 best papers from 39 found): concluded ultrasonography is more sensitive and as specific as chest radiography or CT for haemothorax after chest trauma — the record prints no pooled pair, and none is printed from this anchor
Study: Best-evidence-topic (short-cut) review — qualitative, no pooled pair in the record
Population: Six best papers selected from 39 found by the reported searches, in patients with chest trauma, comparing ultrasonography with chest radiography or CT. Review, counted as a review.
Summary: Best-evidence-topic review (McEwan & Thompson 2007, Emergency Medicine Journal — 6 best papers from 39 found): concluded ultrasonography is more sensitive and as specific as chest radiography or CT for haemothorax after chest trauma — the record prints no pooled pair, and none is printed from this anchor
Result: The review's tabulated best evidence led to the conclusion that ultrasonography is more sensitive and as specific as chest x ray at detecting haemothorax in patients with chest trauma. No pooled sensitivity or specificity is printed in the record.
Watch out: A short-cut review is a weaker synthesis form than the 2016 meta-analysis and is never load-bearing for a figure; it corroborates direction only.
Source: McEwan et al. Ultrasound to detect haemothorax after chest injury. Emergency medicine journal : EMJ. 2007. PMID: 17652688. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Early Diagnosis · Emergency Medicine/instrumentation,methods · Hemothorax/diagnostic imaging*,etiology* · Humans · Middle Aged · Radiography · Sensitivity and Specificity · Thoracic Injuries/complications* · Ultrasonography
Card 5 — Synthesis — specificity is uniformly near-perfect; sensitivity is where the argument lives, and the pooled figure sits far below the primaries
Study: Synthesis
Population: 4 reports: 1 meta-analysis (12 studies, 7,361 patients), 2 primaries (240 and 61 patients) and 1 best-evidence-topic review (6 best papers, no pooled pair). The larger pneumothorax-plus-haemothorax synthesis (PMID 29433802) is gated, carded inside pack #8414 for its pneumothorax limb, and is a cross-point in prose only under adjudication 4 — never an anchor here; pack #8414 (bedside chest ultrasound for pneumothorax) is the neighbour question.
Summary: Synthesis — specificity is uniformly near-perfect; sensitivity is where the argument lives, and the pooled figure sits far below the primaries
Result: Pooled ultrasonography sensitivity is 0.67 against radiography's 0.54, with specificity 0.99 for both — yet the two primaries report 96.2% (25/26) and 92% sensitivity. That gap — small, single-centre, expert-operator primaries against a heterogeneous 12-study pool with I² up to 99% — prints at headline weight rather than being averaged away. The test's argument is sensitivity against supine radiography, not rule-in power: specificity is ~0.99–1.0 everywhere.
Watch out: The 1997 primary's identical counts for the two modalities remain in visible tension with the pooled comparison and are printed as reported. Operator and transducer frequency were identified sources of heterogeneity in the synthesis.
Source: Synthesis of the reports carded above; no additional source.