MUNIX / MUSIX reference values — 5 reports, 362 documented healthy subjects plus one N-undocumented cohort: five-muscle adult means 152.1–191.0, and a ≥20% sum-score change is the usable significance threshold
Shelf #8556 · Brain, stroke & mind
PHYSIOLOGY — ESTABLISHED — the adult five-muscle MUNIX scale rests on one 150-subject reference table (means 152.1–191.0 across muscles, falling with age in every muscle), and the reproducibility story prints at equal weight: single-muscle MUNIX is operator- and occasion-sensitive (coefficients of variation up to ~21–22%), so the 118-subject variability study's threshold — a MUNIX sum-score variation of ≥20% reads as a significant change in muscle innervation — is the pack's clinically usable figure. The family's flagship multi-centre reference report (PMID 21689981) is named in prose only: its record has a 0-character abstract and can never be an anchor. This pack carries the #7081 caveat on its face (see balance).
Stamped: 2026-11-28.
Evidence published: 2010–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In healthy humans, what are the reference values for the motor unit number index (MUNIX) and motor unit size index (MUSIX) — what values did 150 Chinese adults give in five muscles, how reproducible is the index in 34 healthy subjects and in 118 healthy subjects measured twice, and how do children's values grow with age in 60 control children?
Balance of the evidence
5 reports, 362 documented healthy subjects (34 + 150 + 118 + 60 children) plus one cohort whose abstract states no sample size (N not documented, orbicularis oculi). Only healthy-subject limbs are counted; the principle report's ALS serial limb is context, not normative data. Standing caveat, printed pack #7081 homes direct motor/sensory conduction velocity and amplitude in median, ulnar and sural nerves — this pack asks a different measurand that #7081 does not ask about, and the two packs cross-point. Developer ruling recorded as open. FENCE: PMID 21689981 (Neuwirth et al. 2011, Clinical Neurophysiology — multi-centre MUNIX reference values for five muscles) was fetched; its abstract is 0 characters, so under the standing fence it is named here in prose only and is NOT carded — this pack stands on the five abstract-present reports below. Distinct from pack #7081 and from the quantitative needle-EMG pack #8560: MUNIX is a surface-EMG index, not a conduction velocity and not a needle MUP distribution.
MeSH: Adult · Aged · Algorithms · Amyotrophic Lateral Sclerosis/pathology · Cohort Studies · Electrodiagnosis · Electromyography · Electrophysiology · Female · Humans · Isometric Contraction/physiology · Male · Middle Aged · Models, Statistical · Motor Neuron Disease/pathology* · Motor Neurons/pathology* · Muscle Fibers, Skeletal/pathology* · Muscle, Skeletal/innervation*,pathology* · Regression Analysis · Reproducibility of Results · Action Potentials* · Arm · Isometric Contraction* · Leg · Motor Neurons/physiology* · Muscle, Skeletal/innervation*,physiology · Neuromuscular Diseases/diagnosis,physiopathology · Reference Values · Superficial Back Muscles/innervation,physiology · Young Adult · Electromyography/methods · Healthy Volunteers · Muscle, Skeletal/physiology* · Recruitment, Neurophysiological/physiology* · Child · Adolescent · Child, Preschool · Muscle, Skeletal*/physiology · Action Potentials*/physiology · Motor Neurons*/physiology · Age Factors · Action Potentials/physiology · Electric Stimulation · Facial Muscles/physiology* · Muscle Contraction/physiology*
Cards (6)
Card 1 — Method principle report (Nandedkar et al. 2010, Muscle & Nerve — 34 healthy subjects, hypothenar muscles): good reproducibility; no change in CMAP amplitude or MUNIX over serial study in 4 healthy subjects
Study: Method / principle report (programme-labelled: the method developers' own report); healthy limb only
Population: 34 healthy subjects studied to define the normal range and reproducibility of MUNIX/MUSIX in the hypothenar muscles; 4 healthy subjects (and 13 ALS patients) studied serially over 1 year — only the healthy limb is counted here.
Summary: Method principle report (Nandedkar et al. 2010, Muscle & Nerve — 34 healthy subjects, hypothenar muscles): good reproducibility; no change in CMAP amplitude or MUNIX over serial study in 4 healthy subjects
Result: In healthy subjects, MUNIX showed good reproducibility. In serial studies, healthy subjects showed no change in CMAP amplitude and MUNIX, while ALS patients with minimal CMAP change had a significant drop in MUNIX and increase in MUSIX; when CMAP changed significantly (>30%) in 1 year, CMAP and MUNIX decreased in parallel.
Watch out: Programme-labelled — the developers' own report, and its healthy N (34) is small. The ALS limb shows what the index is for; it is not normative data and is counted nowhere in this pack's totals.
Source: Nandedkar SD et al. Motor unit number index (MUNIX): principle, method, and findings in healthy subjects and in patients with motor neuron disease. Muscle & nerve. 2010. PMID: 20976783. Registry ID: none in the record. MeSH: Adult · Aged · Algorithms · Amyotrophic Lateral Sclerosis/pathology · Cohort Studies · Electrodiagnosis · Electromyography · Electrophysiology · Female · Humans · Isometric Contraction/physiology · Male · Middle Aged · Models, Statistical · Motor Neuron Disease/pathology* · Motor Neurons/pathology* · Muscle Fibers, Skeletal/pathology* · Muscle, Skeletal/innervation*,pathology* · Regression Analysis · Reproducibility of Results
Card 2 — Five-muscle reference values (Cao et al. 2020, Muscle & Nerve — 150 healthy Chinese adults, right-sided muscles): mean MUNIX 191.0 ± 43.6 (abductor pollicis brevis), 179.1 ± 38.8 (abductor digiti minimi), 179.7 ± 36.2 (biceps brachii), 152.1 ± 38.8 (tibialis anterior), 166.1 ± 40.7 (trapezius)
Study: Normative-data study (primary cohort report)
Population: 150 healthy participants (76 men, 74 women); MUNIX and MUSIX measured in the right abductor pollicis brevis, abductor digiti minimi, biceps brachii, tibialis anterior and trapezius.
Summary: Five-muscle reference values (Cao et al. 2020, Muscle & Nerve — 150 healthy Chinese adults, right-sided muscles): mean MUNIX 191.0 ± 43.6 (abductor pollicis brevis), 179.1 ± 38.8 (abductor digiti minimi), 179.7 ± 36.2 (biceps brachii), 152.1 ± 38.8 (tibialis anterior), 166.1 ± 40.7 (trapezius)
Result: Mean MUSIX for the same muscles were 60.4 ± 12.5, 59.3 ± 13.7, 43.8 ± 11.5, 41.3 ± 10.7 and 49.1 ± 15.2 respectively. MUNIX in all five muscles was inversely related to age.
Watch out: The pack's only full reference table — one population (Chinese adults), right-sided muscles only. Because MUNIX falls with age in every muscle, no age-free 'normal MUNIX' may be printed.
Source: Cao B et al. Reference values for the motor unit number index and the motor unit size index in five muscles. Muscle & nerve. 2020. PMID: 32068896. Registry ID: none in the record. MeSH: Action Potentials* · Adult · Aged · Arm · Electromyography · Female · Humans · Isometric Contraction* · Leg · Male · Middle Aged · Motor Neurons/physiology* · Muscle, Skeletal/innervation*,physiology · Neuromuscular Diseases/diagnosis,physiopathology · Reference Values · Superficial Back Muscles/innervation,physiology · Young Adult
Card 3 — Intra-individual variability (Delmont et al. 2020, Clinical Neurophysiology — 118 healthy subjects at 11 centres, measured twice 90–360 days apart): 95th-percentile coefficients of variability 22% / 18% / 15% for 2-, 3- and 4-muscle sum scores (experienced operators); ≥20% sum-score variation is a significant change
Study: Reproducibility study (primary report)
Population: 118 healthy subjects at 11 hospital centres; MUNIX assessed twice, at least 3 months apart, in tibialis anterior, abductor pollicis brevis, abductor digiti minimi and deltoid.
Summary: Intra-individual variability (Delmont et al. 2020, Clinical Neurophysiology — 118 healthy subjects at 11 centres, measured twice 90–360 days apart): 95th-percentile coefficients of variability 22% / 18% / 15% for 2-, 3- and 4-muscle sum scores (experienced operators); ≥20% sum-score variation is a significant change
Result: Repeatability was better for sum scores than for single-muscle recordings. Variability was independent of sex, age and measurement interval, and was lower for experienced than non-experienced operators. The 95th percentile of the coefficient of variability for the 2-, 3- and 4-muscle sum scores was 22%, 18% and 15% respectively for experienced operators. The authors conclude a variation of the MUNIX sum score ≥20% can be interpreted as a significant change of muscle innervation.
Watch out: The ≥20% threshold belongs to experienced operators using sum scores — single-muscle MUNIX in inexperienced hands varies more, and a smaller 'change' in that setting is noise, not progression.
Source: Delmont E et al. Motor unit number index as an individual biomarker: Reference limits of intra-individual variability over time in healthy subjects. Clinical neurophysiology : official journal of the International Federation of Clinical Neurophysiology. 2020. PMID: 32707479. Registry ID: none in the record. MeSH: Adult · Aged · Electromyography/methods · Female · Healthy Volunteers · Humans · Male · Middle Aged · Motor Neurons/physiology* · Muscle, Skeletal/physiology* · Neuromuscular Diseases/diagnosis,physiopathology · Recruitment, Neurophysiological/physiology* · Reproducibility of Results · Young Adult
Card 4 — Control children (Boulay et al. 2025, Muscle & Nerve — 60 healthy children in four age bands): abductor pollicis brevis MUNIX rose 6.5 per year of age univariably, but was not independently associated with age, height or weight multivariably
Study: Normative-data study — paediatric limb (printed as such)
Population: 60 control children (28 girls): under 5 years n = 16; 5–10 n = 15; 10–15 n = 20; 15–17 n = 9; abductor digiti minimi and abductor pollicis brevis measured by an experienced and an inexperienced investigator.
Summary: Control children (Boulay et al. 2025, Muscle & Nerve — 60 healthy children in four age bands): abductor pollicis brevis MUNIX rose 6.5 per year of age univariably, but was not independently associated with age, height or weight multivariably
Result: Abductor pollicis brevis MUNIX increased by 6.5 (95% CI 3.2 to 9.7) per year of age, 1.0 per cm and 1.7 per kg (p < 0.001 each) in univariable analyses, but was not independently associated with any of the three in multivariable analysis. Reproducibility was high for the experienced investigator and lower between investigators.
Watch out: PAEDIATRIC LIMB: children's values are growth-bound and are never merged with the adult table. The univariable growth slopes did not survive multivariable adjustment — printed as published, not smoothed.
Source: Boulay C et al. Motor Unit Number Index (MUNIX) in Control Children: Reference Values and Reliability. Muscle & nerve. 2025. PMID: 40686349. Registry ID: none in the record. MeSH: Humans · Female · Male · Child · Adolescent · Child, Preschool · Reference Values · Reproducibility of Results · Electromyography · Muscle, Skeletal*/physiology · Action Potentials*/physiology · Motor Neurons*/physiology · Age Factors
Card 5 — Orbicularis oculi MUNIX (Ahn et al. 2015, Muscle & Nerve — healthy subjects, N not documented, bilateral): test–retest correlation 0.803 for MUNIX and 0.592 for MUSIX; coefficients of variation 20.9% and 8.5%
Study: Normative / reproducibility study — facial-muscle limb; N not documented
Population: N not documented — the abstract states no sample size; healthy subjects, bilateral orbicularis oculi muscles.
Summary: Orbicularis oculi MUNIX (Ahn et al. 2015, Muscle & Nerve — healthy subjects, N not documented, bilateral): test–retest correlation 0.803 for MUNIX and 0.592 for MUSIX; coefficients of variation 20.9% and 8.5%
Result: MUNIX was applied successfully to the orbicularis oculi muscles and showed good reproducibility: correlation coefficients 0.803 for MUNIX and 0.592 for MUSIX, and coefficients of variation 20.9% and 8.5% respectively.
Watch out: FACIAL-MUSCLE LIMB, printed as such, with N not documented: a feasibility-plus-reproducibility report whose 20.9% MUNIX coefficient of variation underlines the pack's variability warning; no reference mean is printed from it.
Source: Ahn SW et al. Motor unit number index (MUNIX) in the orbicularis oculi muscle of healthy subjects. Muscle & nerve. 2015. PMID: 24865812. Registry ID: none in the record. MeSH: Action Potentials/physiology · Adult · Electric Stimulation · Electromyography · Facial Muscles/physiology* · Female · Healthy Volunteers · Humans · Male · Motor Neurons/physiology* · Muscle Contraction/physiology* · Young Adult
Card 6 — Synthesis — an established adult scale with a variability warning at its centre; the flagship reference report is fence-excluded and named in prose only; the #7081 caveat stands on this pack
Study: Synthesis
Population: 5 reports: 362 documented healthy subjects (34 + 150 + 118 + 60 children) plus one N-undocumented cohort; adult reference table from a single 150-subject cohort.
Summary: Synthesis — an established adult scale with a variability warning at its centre; the flagship reference report is fence-excluded and named in prose only; the #7081 caveat stands on this pack
Result: The adult scale is established in shape: five-muscle means 152.1–191.0 with an inverse age relation in every muscle. The reproducibility limbs govern its use: sum scores beat single muscles, experienced operators beat inexperienced, and a ≥20% sum-score change is the threshold for a significant change. Children's values grow through childhood. Refused as an anchor: PMID 21689981 (2011, multi-centre reference values for five muscles) — its fetched record has a 0-character abstract, and an empty-abstract record can never be an anchor; it is named in prose only, here and in the balance. The stamp is PHYSIOLOGY — ESTABLISHED for the adult scale. Caveat restated: if the developer judges pack #7081 to have taken electrodiagnostic norms as a whole, this pack falls to cross-point status inside it.
Watch out: A laboratory adopting the 150-subject table without the variability discipline will over-read small changes; the index measures innervation change over time in one patient better than it classifies a single value.