SourceMed · Shelf #8594 in the ask interface
CT-derived fractional flow reserve (FFRCT) accuracy — 5 syntheses, overlapping pools: sensitivity ~0.81–0.89, specificity 0.71–0.83 vs CCTA 0.32–0.61, accuracy >90% only outside 0.49–0.90
Shelf #8594 · Heart, vessels, blood pressure & clotting
PREDICTIVE PERFORMANCE — consistent sensitivity, value concentrated in specificity: across 5 overlapping syntheses FFRCT sensitivity runs ~0.81–0.89 with specificity 0.71–0.83 against invasive FFR ≤0.80, versus CCTA specificity of 0.32–0.61 in the same pools; agreement with invasive FFR is moderate (Spearman 0.67) and diagnostic accuracy exceeds 90% only for FFR-CT values >0.90 or <0.49 — the grey zone where most referred lesions live
Stamped: 2026-11-28.
Evidence published: 2015–2024.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with suspected or known coronary artery disease, how accurately does CT-derived fractional flow reserve (FFRCT) classify haemodynamically significant stenosis against invasive FFR ≤0.80 — per patient and per vessel — and how does its specificity compare with CT angiography alone?
Balance of the evidence
5 syntheses with heavily overlapping pools (the 2015 pool is the three programme trials inside every later pool) — printed as one literature read five ways, never as five independent confirmations. Per-patient and per-vessel pairs are never merged. One anchor (Ding 2016) prints bounds only and carries no exact pair. Li 2015's record prints two impossible intervals, reproduced flagged on its card, never repaired. Cross-points: the FFRCT-guided care strategy question (pack #2263) and post-PCI invasive FFR (#6453) are different questions.
MeSH: Aged · Computed Tomography Angiography/methods* · Coronary Angiography/methods* · Female · Fractional Flow Reserve, Myocardial/physiology* · Humans · Male · Middle Aged · Myocardial Ischemia/diagnostic imaging*,physiopathology · ROC Curve · Sensitivity and Specificity · Area Under Curve · Coronary Stenosis/complications*,diagnostic imaging*,physiopathology · Myocardial Ischemia/diagnostic imaging*,etiology*,physiopathology · Reproducibility of Results · Severity of Illness Index · Coronary Angiography · Coronary Artery Disease/diagnostic imaging*,physiopathology* · Fractional Flow Reserve, Myocardial* · Myocardial Ischemia/diagnostic imaging* · Tomography, X-Ray Computed/methods* · Fractional Flow Reserve, Myocardial*/physiology · Coronary Stenosis*/physiopathology,diagnostic imaging,diagnosis · Coronary Angiography*/methods · Computed Tomography Angiography*/methods · Predictive Value of Tests · Cardiac Catheterization · Coronary Artery Disease/physiopathology,diagnostic imaging,diagnosis · Coronary Vessels/diagnostic imaging,physiopathology · Tomography, X-Ray Computed/methods · Computed Tomography Angiography* · Coronary Artery Disease/diagnostic imaging*,physiopathology · Coronary Vessels/diagnostic imaging*,physiopathology · Hemodynamics*
Cards (6)
Card 1 — Largest early pool (Zhuang 2020, European Radiology — systematic review and meta-analysis, 16 studies / 1,852 patients / 2,731 vessels): FFRCT per-patient sensitivity 89% (95% CI 85–92) / specificity 71% (61–80), per-vessel 85% (82–88) / 82% (75–87)
Study: Systematic review and meta-analysis (synthesis, counted as a synthesis)
Population: 16 studies (2011–2019) of CT-derived fractional flow reserve against invasive FFR in 1,852 patients / 2,731 vessels; the same pool's CCTA figures are the comparator limb.
Summary: Largest early pool (Zhuang 2020, European Radiology — systematic review and meta-analysis, 16 studies / 1,852 patients / 2,731 vessels): FFRCT per-patient sensitivity 89% (95% CI 85–92) / specificity 71% (61–80), per-vessel 85% (82–88) / 82% (75–87)
Result: FFRCT pooled sensitivity/specificity were 89% (95% CI 85–92%) / 71% (61–80%) per patient and 85% (82–88%) / 82% (75–87%) per vessel. Sensitivity did not differ appreciably from CCTA (0.89 vs 0.93 per patient; 0.85 vs 0.88 per vessel), but specificity was higher for FFRCT than CCTA (0.71 vs 0.32 per patient; 0.82 vs 0.46 per vessel) — the specificity gain is the finding.
Watch out: Per-patient and per-vessel pairs are different counting units and are never merged on this pack. CCTA's per-patient specificity of 0.32 in this pool is the baseline the FFRCT gain is measured against.
Source: Zhuang B et al. Computed tomography angiography-derived fractional flow reserve (CT-FFR) for the detection of myocardial ischemia with invasive fractional flow reserve as reference: systematic review and meta-analysis. European radiology. 2020. PMID: 31696294. Registry ID: none in the record. MeSH: Aged · Computed Tomography Angiography/methods* · Coronary Angiography/methods* · Female · Fractional Flow Reserve, Myocardial/physiology* · Humans · Male · Middle Aged · Myocardial Ischemia/diagnostic imaging*,physiopathology · ROC Curve · Sensitivity and Specificity
Card 2 — Bounds-only pool (Ding 2016, Japanese Journal of Radiology — meta-analysis, 21 studies / 2,216 subjects / 2,798 vessels): FFRCT sensitivity ≥82% and specificity ≥73% per vessel and per patient — the abstract prints bounds, not exact pairs
Study: Meta-analysis (synthesis, counted as a synthesis)
Population: 21 studies of FFRCT for determining coronary artery disease severity, 2,216 subjects / 2,798 vessels, against invasive FFR.
Summary: Bounds-only pool (Ding 2016, Japanese Journal of Radiology — meta-analysis, 21 studies / 2,216 subjects / 2,798 vessels): FFRCT sensitivity ≥82% and specificity ≥73% per vessel and per patient — the abstract prints bounds, not exact pairs
Result: The abstract reports that FFRCT sensitivity per vessel and per patient was ≥82% and specificity was ≥73% for diagnosis of ischaemia, and that FFRCT had better diagnostic accuracy and discrimination than CCTA. No exact pooled pair is printed in the record, and none is printed on this card.
Watch out: A bounds-only anchor: ≥82% / ≥73% is all the record supports. Any exact pair attributed to this pool elsewhere was not read from this record.
Source: Ding A et al. Diagnostic performance of noninvasive fractional flow reserve derived from coronary computed tomography angiography in ischemia-causing coronary stenosis: a meta-analysis. Japanese journal of radiology. 2016. PMID: 27796727. Registry ID: none in the record. MeSH: Area Under Curve · Computed Tomography Angiography/methods* · Coronary Angiography/methods* · Coronary Stenosis/complications*,diagnostic imaging*,physiopathology · Fractional Flow Reserve, Myocardial/physiology* · Humans · Myocardial Ischemia/diagnostic imaging*,etiology*,physiopathology · ROC Curve · Reproducibility of Results · Sensitivity and Specificity · Severity of Illness Index
Card 3 — Programme-trials pool (Li 2015, Clinical Radiology — meta-analysis of the three multicentre programme studies only: NXT, DISCOVER-FLOW, DeFACTO; 609 patients / 1,050 vessels): FFRCT per-patient 89% (85–93) / 71% (65–75), per-vessel 83% / 78% — with two impossible intervals in the record, reproduced flagged, never repaired
Study: Meta-analysis (synthesis, counted as a synthesis)
Population: 609 patients / 1,050 vessels from the three multicentre programme studies (NXT Trial, DISCOVER-FLOW, DeFACTO), thresholds FFR ≤0.80 for ischaemia or CTA ≥50% stenosis.
Summary: Programme-trials pool (Li 2015, Clinical Radiology — meta-analysis of the three multicentre programme studies only: NXT, DISCOVER-FLOW, DeFACTO; 609 patients / 1,050 vessels): FFRCT per-patient 89% (85–93) / 71% (65–75), per-vessel 83% / 78% — with two impossible intervals in the record, reproduced flagged, never repaired
Result: Per patient, FFRCT pooled sensitivity 89% (85–93%), specificity 71% (65–75%), PPV 70%, NPV 90%, LR+ 3.31, LR− 0.16. Per vessel, sensitivity 83% (78–63%) and specificity 78% (75–81%), PPV 61%, NPV 92% (89–90%), LR+ 4.02, LR− 0.22 — DEFECTIVE FIGURES IN THE RECORD: the per-vessel sensitivity interval '78–63%' and the NPV interval '89–90%' around a 92% point estimate are impossible as printed and are reproduced verbatim, flagged, never silently repaired. Area under the summary ROC curve for FFRCT was 0.8909 per patient and 0.8865 per vessel, versus 0.7402 for CTA per patient.
Watch out: This pool is the three programme trials that sit inside every later pool on this pack — it is provenance, not an independent confirmation. Two of its printed intervals are impossible; the pack carries them flagged rather than fixing them.
Source: Li S et al. The diagnostic performance of CT-derived fractional flow reserve for evaluation of myocardial ischaemia confirmed by invasive fractional flow reserve: a meta-analysis. Clinical radiology. 2015. PMID: 25623512. Registry ID: none in the record. MeSH: Coronary Angiography · Coronary Artery Disease/diagnostic imaging*,physiopathology* · Fractional Flow Reserve, Myocardial* · Humans · Myocardial Ischemia/diagnostic imaging* · Tomography, X-Ray Computed/methods*
Card 4 — Agreement pool (Faulder 2024, Journal of the American Heart Association — systematic review and meta-analysis of agreement, 43 studies / 5,236 patients / 7,291 vessels): Spearman correlation with invasive FFR 0.67; overall accuracy 82.2%, sensitivity 80.9%, specificity 83.1%; accuracy exceeded 90% only for FFR-CT values >0.90 or <0.49
Study: Systematic review and meta-analysis (synthesis, counted as a synthesis)
Population: 43 studies reporting on 7,291 vessels from 5,236 patients, FFR-CT against invasively measured FFR (haemodynamically significant ≤0.8).
Summary: Agreement pool (Faulder 2024, Journal of the American Heart Association — systematic review and meta-analysis of agreement, 43 studies / 5,236 patients / 7,291 vessels): Spearman correlation with invasive FFR 0.67; overall accuracy 82.2%, sensitivity 80.9%, specificity 83.1%; accuracy exceeded 90% only for FFR-CT values >0.90 or <0.49
Result: A moderate positive linear relationship between FFR-CT and invasive FFR (Spearman correlation coefficient 0.67); agreement increased as invasive FFR approached 1. Overall diagnostic accuracy 82.2%, sensitivity 80.9%, specificity 83.1%. Diagnostic accuracy of 90% could be demonstrated for FFR-CT values >0.90 and <0.49. On-site tools 84.1% accuracy vs off-site 79.4%.
Watch out: The grey-zone finding prints at headline weight: accuracy is above 90% only outside 0.49–0.90, which is where most referred lesions live. A correlation of 0.67 is agreement, not interchangeability.
Source: Faulder TI et al. Agreement of Fractional Flow Reserve Estimated by Computed Tomography With Invasively Measured Fractional Flow Reserve: A Systematic Review and Meta-Analysis. Journal of the American Heart Association. 2024. PMID: 38726901. Registry ID: none in the record. MeSH: Fractional Flow Reserve, Myocardial*/physiology · Humans · Coronary Stenosis*/physiopathology,diagnostic imaging,diagnosis · Coronary Angiography*/methods · Computed Tomography Angiography*/methods · Predictive Value of Tests · Cardiac Catheterization · Reproducibility of Results · Coronary Artery Disease/physiopathology,diagnostic imaging,diagnosis · Coronary Vessels/diagnostic imaging,physiopathology · Severity of Illness Index · Tomography, X-Ray Computed/methods
Card 5 — Functional-CT pool, FFRCT limb only (Celeng 2019, JACC: Cardiovascular Imaging — meta-analysis of anatomical and functional CT, 54 articles / 5,330 patients): per-vessel FFRCT specificity 0.78 (95% CI 0.72–0.83) vs CTA 0.61 (0.54–0.68)
Study: Meta-analysis (synthesis, counted as a synthesis — FFRCT limb only)
Population: 54 articles / 5,330 patients comparing CTA, CT perfusion, FFRCT and transluminal attenuation gradient against FFR; only the FFRCT limb is counted on this pack — the CT-perfusion and attenuation-gradient limbs are cross-points.
Summary: Functional-CT pool, FFRCT limb only (Celeng 2019, JACC: Cardiovascular Imaging — meta-analysis of anatomical and functional CT, 54 articles / 5,330 patients): per-vessel FFRCT specificity 0.78 (95% CI 0.72–0.83) vs CTA 0.61 (0.54–0.68)
Result: At vessel level, pooled specificity was 0.78 (95% CI 0.72–0.83) for FFRCT vs 0.61 (0.54–0.68) for CTA (CT perfusion 0.86, attenuation gradient 0.77 in the same pool — cross-points). Adding FFRCT, CT perfusion or the gradient to CTA gave specificities of 0.80–0.92. On-site vs off-site FFRCT: sensitivity 0.84 vs 0.85, specificity 0.80 vs 0.73.
Watch out: A specificity-only limb for FFRCT on this card: no sensitivity pair is printed from this anchor here. The pool's other functional-CT limbs answer different index-test questions.
Source: Celeng C et al. Anatomical and Functional Computed Tomography for Diagnosing Hemodynamically Significant Coronary Artery Disease: A Meta-Analysis. JACC. Cardiovascular imaging. 2019. PMID: 30219398. Registry ID: none in the record. MeSH: Cardiac Catheterization · Computed Tomography Angiography* · Coronary Angiography/methods* · Coronary Artery Disease/diagnostic imaging*,physiopathology · Coronary Vessels/diagnostic imaging*,physiopathology · Fractional Flow Reserve, Myocardial* · Hemodynamics* · Humans · Predictive Value of Tests · Reproducibility of Results
Card 6 — Synthesis — 5 syntheses with heavily overlapping pools: sensitivity ~0.81–0.89 across a decade, with the value concentrated in specificity (0.71–0.83 vs CCTA's 0.32–0.61), and accuracy above 90% only outside the 0.49–0.90 grey zone
Study: Synthesis
Population: 5 syntheses: Zhuang 2020 (16 studies / 1,852 patients / 2,731 vessels), Ding 2016 (21 studies / 2,216 subjects / 2,798 vessels, bounds only), Li 2015 (the 3 programme trials, 609 patients / 1,050 vessels), Faulder 2024 (43 studies / 5,236 patients / 7,291 vessels), Celeng 2019 (54 articles / 5,330 patients, FFRCT limb only). The pools overlap heavily — the 2015 programme pool sits inside every later pool — so these are one literature read five ways, never five independent confirmations.
Summary: Synthesis — 5 syntheses with heavily overlapping pools: sensitivity ~0.81–0.89 across a decade, with the value concentrated in specificity (0.71–0.83 vs CCTA's 0.32–0.61), and accuracy above 90% only outside the 0.49–0.90 grey zone
Result: The consistent shape across a decade: FFRCT sensitivity ~0.81–0.89 with specificity 0.71–0.83, against CCTA specificity of 0.32–0.61 in the same pools — FFRCT's contribution is ruling in, not sensitivity. The 2024 agreement synthesis adds the discipline: correlation with invasive FFR 0.67, overall accuracy 82.2%, and accuracy above 90% only for FFR-CT values >0.90 or <0.49. Cross-points: pack #2263 asks whether an FFRCT-guided care strategy saves costs or improves outcomes (a strategy question — it found neither); pack #6453 is post-PCI invasive FFR optimisation, a different test. Defective figures: Li 2015's per-vessel sensitivity interval '83% (78–63%)' and NPV '92% (89–90%)' are impossible as printed in the record, reproduced flagged on its card, never repaired.
Watch out: Per-patient and per-vessel pairs are never merged on this pack. Ding 2016 prints bounds only (≥82% / ≥73%) and no exact pair may be sourced to it. On-site vs off-site computation made little difference in either pool that split it (accuracy 84.1% vs 79.4%; specificity 0.80 vs 0.73).
Source: Synthesis of the reports carded above; no additional source.