Stress echocardiography accuracy for coronary artery disease — 3 syntheses (two from one programme): stressors interchangeable at modern protocols (87% vs 84%), 35-year pooled sensitivity 0.81 / specificity 0.85
Shelf #8596 · Heart, vessels, blood pressure & clotting
PREDICTIVE PERFORMANCE — good pooled performance, stressors interchangeable: the 35-year pool puts stress echocardiography at sensitivity 0.81 / specificity 0.85, and head-to-head pools show dobutamine and dipyridamole within noise of each other once protocols are modern (accuracy 87% vs 84%, P = 0.48), the older larger pool's small opposed gradients (dobutamine +6 sensitivity, dipyridamole +6 specificity) printing as the earlier reading; significant publication bias across the 2025 pool caps every pair
Stamped: 2026-11-28.
Evidence published: 1990–2008.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with suspected coronary artery disease, how accurate is stress echocardiography — exercise, dobutamine and dipyridamole/vasodilator protocols — against coronary angiography, and do the pharmacological stressors differ?
Balance of the evidence
3 syntheses: two from one programme (Picano 2000 and 2008, nested pools — counted as two reports of one programme, not independent confirmations) plus one 35-year pool (Sonaglioni 2025, stress-echo limb only, TYPE-FLAG: typed Review only, a meta-analysis on substance, no MeSH in the record). The 2025 pool's publication-bias finding prints at weight. Exercise and vasodilator protocols are limbs of one question here; the pool's exercise-testing, SPECT and CMR limbs are cross-points to packs #8595 and #8597.
MeSH: Cardiotonic Agents · Coronary Disease/diagnosis*,diagnostic imaging · Dipyridamole* · Dobutamine* · Dose-Response Relationship, Drug · Echocardiography* · Exercise Test/methods · Female · Humans · Male · Probability · Sensitivity and Specificity · Vasodilator Agents · Aged · Coronary Angiography/methods · Coronary Artery Disease/diagnostic imaging* · Echocardiography, Stress/methods* · Middle Aged · Prognosis · Risk Assessment · Severity of Illness Index
Cards (4)
Card 1 — Head-to-head pool (Picano 2000, Coronary Artery Disease — meta-analysis of 12 head-to-head studies / 818 patients with angiography): dipyridamole accuracy 77% vs dobutamine 80% (NS); sensitivities 71% (403/568) vs 77% (437/568) (P < 0.05), specificities 93% (232/250) vs 87% (217/250) (P < 0.05)
Study: Meta-analysis (synthesis, counted as a synthesis)
Population: 818 patients in 12 head-to-head studies (same population) of high-dose dipyridamole (0.84 mg/kg) vs high-dose dobutamine (up to 40 micrograms/kg per min) stress echocardiography with angiographic information; 568 with and 250 without disease in the paired analysis.
Summary: Head-to-head pool (Picano 2000, Coronary Artery Disease — meta-analysis of 12 head-to-head studies / 818 patients with angiography): dipyridamole accuracy 77% vs dobutamine 80% (NS); sensitivities 71% (403/568) vs 77% (437/568) (P < 0.05), specificities 93% (232/250) vs 87% (217/250) (P < 0.05)
Result: Diagnostic accuracies were similar (631/818, 77%, dipyridamole vs 654/818, 80%, dobutamine, NS). Overall sensitivities 403/568 (71%) vs 437/568 (77%) (P < 0.05); single-vessel disease 177/275 (64%) vs 203/275 (74%) (P < 0.05); multivessel 162/203 (80%) vs 163/203 (80%) (NS). Specificities 232/250 (93%) vs 217/250 (87%) (P < 0.05). The unpaired pools (26 dipyridamole studies, n = 2,038; 47 dobutamine studies, n = 4,264) gave accuracies of 80% and 82%.
Watch out: Small opposed gradients in the older, larger pool (dobutamine +6 sensitivity, dipyridamole +6 specificity) — both print, and the protocol-restricted 2008 pool below erases the sensitivity gap. Same senior programme as the 2008 anchor: two reports of one programme, not two independent confirmations.
Source: Picano E et al. The comparable diagnostic accuracies of dobutamine-stress and dipyridamole-stress echocardiographies: a meta-analysis. Coronary artery disease. 2000. PMID: 10758817. Registry ID: none in the record. MeSH: Cardiotonic Agents · Coronary Disease/diagnosis*,diagnostic imaging · Dipyridamole* · Dobutamine* · Dose-Response Relationship, Drug · Echocardiography* · Exercise Test/methods · Female · Humans · Male · Probability · Sensitivity and Specificity · Vasodilator Agents
Card 2 — Protocol-restricted pool (Picano 2008, Cardiovascular Ultrasound — meta-analysis restricted to the 5 head-to-head papers using state-of-the-art protocols, 435 patients, 299 with CAD): dipyridamole vs dobutamine accuracy 87% (95% CI 83–90) vs 84% (80–88) (P = 0.48), sensitivity 85% vs 86%, specificity 89% vs 86%
Study: Meta-analysis (synthesis, counted as a synthesis — same programme as the 2000 anchor, nested pool)
Population: 435 patients (299 with and 136 without angiographically assessed CAD, stenosis >50%) in the 5 head-to-head papers using state-of-the-art protocols (dobutamine 40 mcg/kg/min ± atropine; dipyridamole 0.84 mg/kg in 10 min plus atropine, or in 6 min without).
Summary: Protocol-restricted pool (Picano 2008, Cardiovascular Ultrasound — meta-analysis restricted to the 5 head-to-head papers using state-of-the-art protocols, 435 patients, 299 with CAD): dipyridamole vs dobutamine accuracy 87% (95% CI 83–90) vs 84% (80–88) (P = 0.48), sensitivity 85% vs 86%, specificity 89% vs 86%
Result: Dipyridamole and dobutamine showed similar accuracy (87%, 95% CI 83–90, vs 84%, CI 80–88, P = 0.48), sensitivity (85%, CI 80–89, vs 86%, CI 78–91, P = 0.81) and specificity (89%, CI 82–94, vs 86%, CI 75–89, P = 0.15). The authors concluded the European recommendations — that dobutamine and vasodilators at appropriately high doses are equally potent ischaemic stressors — are evidence-based.
Watch out: Only 5 papers and 435 patients once protocols are restricted to state-of-the-art — the cleanest comparison is also the smallest. Its pool is nested inside the programme's wider literature; it does not independently confirm the 2000 pool.
Source: Picano E et al. The diagnostic accuracy of pharmacological stress echocardiography for the assessment of coronary artery disease: a meta-analysis. Cardiovascular ultrasound. 2008. PMID: 18565214. Registry ID: none in the record. MeSH: Aged · Coronary Angiography/methods · Coronary Artery Disease/diagnostic imaging* · Echocardiography, Stress/methods* · Female · Humans · Male · Middle Aged · Probability · Prognosis · Risk Assessment · Sensitivity and Specificity · Severity of Illness Index
Card 3 — 35-year pool, stress-echo limb only (Sonaglioni 2025, Journal of Clinical Medicine — systematic reviews and meta-analyses of 104 studies / 16,824 symptomatic individuals, 1990–2025): pooled stress-echocardiography sensitivity 0.81 (95% CI 0.79–0.83) / specificity 0.85 (0.82–0.87) — TYPE-FLAG: typed Review only at PubMed, a meta-analysis on substance; the record prints no MeSH
Study: Systematic review and meta-analysis (synthesis, counted as a synthesis — stress-echocardiography limb only) — TYPE-FLAG: the record is typed Journal Article / Review only although a meta-analysis on substance; counted on substance
Population: 104 studies comprising 16,824 symptomatic individuals with suspected or confirmed CAD across four techniques (exercise stress testing, stress echocardiography, stress myocardial SPECT, stress CMR), 1990–2025; only the stress-echocardiography limb is counted here.
Summary: 35-year pool, stress-echo limb only (Sonaglioni 2025, Journal of Clinical Medicine — systematic reviews and meta-analyses of 104 studies / 16,824 symptomatic individuals, 1990–2025): pooled stress-echocardiography sensitivity 0.81 (95% CI 0.79–0.83) / specificity 0.85 (0.82–0.87) — TYPE-FLAG: typed Review only at PubMed, a meta-analysis on substance; the record prints no MeSH
Result: Pooled stress-echocardiography sensitivity 0.81 (95% CI 0.79–0.83) / specificity 0.85 (0.82–0.87). In the same pool: exercise stress testing 0.66 / 0.61, stress SPECT 0.82 / 0.74, stress CMR 0.83 / 0.89 — cross-points to the exercise-ECG and perfusion packs. Heterogeneity I² ranged 82.5–92.5%, and Egger's test showed statistically significant publication bias (P < 0.05) for all methods, which the authors attribute to smaller studies reporting more favourable results.
Watch out: Significant publication bias for every method in this pool (smaller studies more favourable) caps every pair in this family and prints at weight. The record prints no MeSH. A 35-year pool mixes protocol eras that the Picano cards show matter.
Source: Sonaglioni A et al. Diagnostic Accuracy of Exercise Stress Testing, Stress Echocardiography, Myocardial Scintigraphy, and Cardiac Magnetic Resonance for Obstructive Coronary Artery Disease: Systematic Reviews and Meta-Analyses of 104 Studies Published from 1990 to 2025. Journal of clinical medicine. 2025. PMID: 40943997. Registry ID: none in the record. MeSH: none in the record
Card 4 — Synthesis — 3 syntheses, two from one programme (Picano) with nested pools plus one 35-year pool: dobutamine and dipyridamole are interchangeable within noise once protocols are modern (accuracy 87% vs 84%), and the modern all-protocol stress-echo pair is 0.81 / 0.85
Study: Synthesis
Population: 3 syntheses: Picano 2000 (12 head-to-head studies / 818 patients, plus unpaired pools of 2,038 and 4,264), Picano 2008 (5 protocol-restricted head-to-head papers / 435 patients — same senior programme, nested pool, counted as a second report of one programme, not an independent confirmation), Sonaglioni 2025 (104 studies / 16,824 individuals, stress-echo limb only; TYPE-FLAG typed Review only).
Summary: Synthesis — 3 syntheses, two from one programme (Picano) with nested pools plus one 35-year pool: dobutamine and dipyridamole are interchangeable within noise once protocols are modern (accuracy 87% vs 84%), and the modern all-protocol stress-echo pair is 0.81 / 0.85
Result: The stressor question answers itself consistently: in the 2000 head-to-head pool dobutamine was slightly more sensitive (77% vs 71%) and dipyridamole slightly more specific (93% vs 87%) at equal overall accuracy (80% vs 77%); restricted to state-of-the-art protocols (2008) the gap disappears (accuracy 87% vs 84%, sensitivity 85% vs 86%, specificity 89% vs 86%). The 35-year all-protocol pool puts stress echocardiography at 0.81 / 0.85, between exercise testing (0.66 / 0.61) and CMR (0.83 / 0.89) in the same analysis — those limbs are cross-points to the exercise-ECG and perfusion packs.
Watch out: The 2025 pool's publication-bias finding (smaller studies more favourable, all modalities) caps every pair in this family. Protocol era is a real variable here: the same programme's numbers move when protocols are restricted, so a single timeless stress-echo pair should not be quoted from this pack.
Source: Synthesis of the reports carded above; no additional source.