HIV self-testing accuracy — 4 primaries + 1 synthesis: supervised oral self-testing 98.7%/100% (n=2,198), unsupervised sensitivity 89.7–90% with a 15.1% invalid rate, and oral fluid costing ~2 points of sensitivity against blood in the specimen synthesis

Shelf #8604 · Critical care & sepsis

PREDICTIVE PERFORMANCE — the kit is highly accurate in supervised hands (sensitivity 98.7%, specificity 100%, n=2,198) but unsupervised field use costs sensitivity (89.7% in Kenya; 90% vs 100% supervised in the Uganda randomised comparison, non-inferiority not shown intention-to-treat) and carries a 15.1% invalid-result rate; oral fluid itself costs about 2 points of sensitivity against blood (98.03% vs 99.68% pooled), and weaker oral-fluid brands fell to 28–59% sensitivity in health-worker hands. Base: 4 primaries + 1 synthesis spanning three different accuracy questions that are never merged.

Stamped: 2026-11-28.

Evidence published: 2012–2016.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults, how accurate is HIV self-testing — oral-fluid and finger-stick whole-blood kits, supervised and unsupervised — against laboratory reference testing, and how much accuracy does unsupervised use cost?

Balance of the evidence

5 reports: 4 primaries + 1 synthesis, spanning three different accuracy questions printed separately and never merged — the kit in supervised hands (n=2,198), the kit unsupervised in the field (n=240 and n=246), and the specimen itself (a provider-use point-of-care synthesis plus a health-worker-performed home-use candidate evaluation, n=239). The invalid-result rate is treated as an accuracy figure, not a usability footnote. Cross-point only: live pack #534 asks whether making self-tests available increases uptake and linkage — a different question; its accuracy pairs are not borrowed here and this pack's pairs borrow no outcome claim from it.

MeSH: AIDS Serodiagnosis/methods* · Adolescent · Adult · Ambulatory Care Facilities · Antibodies/metabolism · Awareness · Cross-Sectional Studies · Diagnostic Self Evaluation* · Female · HIV · HIV Infections/diagnosis* · Humans · Male · Mass Screening/methods* · Middle Aged · Mouth Mucosa/immunology* · Primary Health Care · Reproducibility of Results · Rural Population* · Self Care/methods* · South Africa · Young Adult · AIDS Serodiagnosis/methods*,statistics & numerical data · HIV Antibodies/blood*,immunology · HIV Infections/blood,diagnosis*,immunology · HIV Seropositivity/blood,diagnosis*,epidemiology,immunology · Health Services Accessibility · Kenya · Prospective Studies · Reagent Kits, Diagnostic/statistics & numerical data · Reagent Kits, Diagnostic*/standards · Sensitivity and Specificity · Surveys and Questionnaires · Directive Counseling* · Feasibility Studies · Health Knowledge, Attitudes, Practice · Patient Acceptance of Health Care · Patient Education as Topic · Prevalence · Sexual Partners* · Uganda/epidemiology · Case-Control Studies · Fingers/blood supply* · HIV Antibodies/blood · HIV-1/immunology · HIV-2/immunology · Saliva/virology* · Self Administration* · HIV Antibodies/analysis,blood* · HIV Infections/blood*,diagnosis* · Point-of-Care Systems/standards* · Saliva/chemistry* · Time Factors

Cards (6)

Card 1 — Supervised oral self-testing (Martínez Pérez 2016, Tropical Medicine & International Health — primary validation, 2,198 participants in rural KwaZulu-Natal): OraQuick oral self-test sensitivity 98.7% (95% CI 96.8–99.6) / specificity 100% (99.8–100) against blood-based Determine/Unigold reference; inter-rater agreement 99.8% (kappa 0.9925); only 2 participants needed a repeat test

Study: Validation study (primary report, supervised use)

Population: 2,198 participants (66.3% women) at two government primary healthcare clinics and three Médecins Sans Frontières fixed-testing sites in uMlalazi municipality, KwaZulu-Natal, South Africa; lay counsellors demonstrated the OraQuick oral self-test and participants used it in front of the counsellor.

Summary: Supervised oral self-testing (Martínez Pérez 2016, Tropical Medicine & International Health — primary validation, 2,198 participants in rural KwaZulu-Natal): OraQuick oral self-test sensitivity 98.7% (95% CI 96.8–99.6) / specificity 100% (99.8–100) against blood-based Determine/Unigold reference; inter-rater agreement 99.8% (kappa 0.9925); only 2 participants needed a repeat test

Result: Sensitivity for the oral self-test was 98.7% (95% CI 96.8–99.6) and specificity 100% (95% CI 99.8–100). Inter-rater agreement was 99.8% (Kappa 0.9925). Only two participants had to repeat their OraQuick test.

Watch out: This is the kit in supervised hands — a counsellor demonstrated the test and watched it being used. It does not estimate unsupervised home performance, which the next two cards measure separately and lower.

Source: Martínez Pérez G et al. Supervised oral HIV self-testing is accurate in rural KwaZulu-Natal, South Africa. Tropical medicine & international health : TM & IH. 2016. PMID: 27098272. Registry ID: none in the record. MeSH: AIDS Serodiagnosis/methods* · Adolescent · Adult · Ambulatory Care Facilities · Antibodies/metabolism · Awareness · Cross-Sectional Studies · Diagnostic Self Evaluation* · Female · HIV · HIV Infections/diagnosis* · Humans · Male · Mass Screening/methods* · Middle Aged · Mouth Mucosa/immunology* · Primary Health Care · Reproducibility of Results · Rural Population* · Self Care/methods* · South Africa · Young Adult

Card 2 — Unsupervised oral self-testing (Kurth 2016, AIDS and Behavior — primary prospective validation, 240 adults in western Kenya, HIV prevalence 14.6%): sensitivity 89.7% (95% CI 73–98) / specificity 98% (89–99); 36 of 239 self-tests (15.1%) were invalid, men twice as likely as women to produce an invalid result

Study: Validation study (primary report, unsupervised use)

Population: 240 adults enrolled in a general population in western Kenya to perform unsupervised rapid oral-fluid HIV self-testing, compared against staff-administered oral-fluid and rapid fingerstick tests, with laboratory ELISA confirmation of reactive, discrepant and a proportion of negative results.

Summary: Unsupervised oral self-testing (Kurth 2016, AIDS and Behavior — primary prospective validation, 240 adults in western Kenya, HIV prevalence 14.6%): sensitivity 89.7% (95% CI 73–98) / specificity 98% (89–99); 36 of 239 self-tests (15.1%) were invalid, men twice as likely as women to produce an invalid result

Result: HIV prevalence was 14.6%. Thirty-six of the 239 HIV self-tests were invalid (15.1%; 95% CI 11.1–20.1%), with males twice as likely to have invalid results as females. HIV self-test sensitivity was 89.7% (95% CI 73–98%) and specificity was 98% (95% CI 89–99%). Acceptability was high (94%).

Watch out: The 15.1% invalid rate is an accuracy figure, not a usability footnote: roughly one in seven unsupervised tests returned no usable result at all, before sensitivity is even counted. The sensitivity interval is wide (73–98%) on this sample.

Source: Kurth AE et al. Accuracy and Acceptability of Oral Fluid HIV Self-Testing in a General Adult Population in Kenya. AIDS and behavior. 2016. PMID: 26438487. Registry ID: none in the record. MeSH: AIDS Serodiagnosis/methods*,statistics & numerical data · Adult · Female · HIV Antibodies/blood*,immunology · HIV Infections/blood,diagnosis*,immunology · HIV Seropositivity/blood,diagnosis*,immunology · Health Services Accessibility · Humans · Kenya · Male · Mass Screening · Prospective Studies · Reagent Kits, Diagnostic/statistics & numerical data · Self Care* · Sensitivity and Specificity · Surveys and Questionnaires · Young Adult

Card 3 — Randomised supervised-vs-unsupervised comparison (Asiimwe 2014, AIDS and Behavior — randomised implementation trial, 246 Ugandan fisherfolk, prevalence 22.1%): intention-to-treat sensitivity 90% unsupervised vs 100% supervised (difference −10%, 90% CI −21 to 1) — non-inferiority NOT shown; per-protocol difference −5.6% (−14.4 to 3.3) did show non-inferiority

Study: Randomized controlled trial (implementation, non-inferiority)

Population: 246 participants from selected high-HIV-risk fisherfolk in three fishing villages in Uganda (HIV prevalence 22.1%), randomised 1:1 to unsupervised or provider-supervised self-administered HIV testing, July–September 2013.

Summary: Randomised supervised-vs-unsupervised comparison (Asiimwe 2014, AIDS and Behavior — randomised implementation trial, 246 Ugandan fisherfolk, prevalence 22.1%): intention-to-treat sensitivity 90% unsupervised vs 100% supervised (difference −10%, 90% CI −21 to 1) — non-inferiority NOT shown; per-protocol difference −5.6% (−14.4 to 3.3) did show non-inferiority

Result: In an intent-to-treat analysis, sensitivity was 90% in the unsupervised arm and 100% among the provider-supervised, yielding a difference of −10% (90% CI −21, 1%); non-inferiority was not shown. In a per-protocol analysis, the difference in sensitivity was −5.6% (90% CI −14.4, 3.3%) and did show non-inferiority.

Watch out: The two analyses answer differently and both print: on the intention-to-treat analysis that counts real-world use, unsupervised testing was not shown non-inferior; the per-protocol analysis excludes exactly the failures unsupervised use produces.

Source: Asiimwe S et al. Accuracy of un-supervised versus provider-supervised self-administered HIV testing in Uganda: A randomized implementation trial. AIDS and behavior. 2014. PMID: 24691923. Registry ID: none in the record. MeSH: Adult · Directive Counseling* · Feasibility Studies · Female · HIV Seropositivity/diagnosis*,epidemiology · Health Knowledge, Attitudes, Practice · Humans · Male · Mass Screening · Patient Acceptance of Health Care · Patient Education as Topic · Prevalence · Reagent Kits, Diagnostic*/standards · Reproducibility of Results · Self Care* · Sexual Partners* · Surveys and Questionnaires · Uganda/epidemiology

Card 4 — Home-use candidate tests in health-worker hands (Jaspard 2014, PLoS ONE — primary multicentre evaluation, 179 HIV-infected patients + 60 controls, France): oral-fluid sensitivities 87.2% (OraQuick), 88.3% (Chembio DPP), 58.9% and 28% (two anonymised tests) vs 100% for the finger-stick Sure Check test — the specimen/brand limb, performed by trained health workers, NOT self-administered

Study: Evaluation study (primary report; specimen/brand limb — health-worker performed, labelled as such, never self-testing performance)

Population: 179 HIV-infected adults and 60 HIV-uninfected controls in a prospective multicentre French study; 67.6% of the HIV-infected patients had an undetectable viral load on antiretroviral therapy. Five candidate home-use tests (4 oral-fluid, 1 finger-stick) were performed by trained health workers and compared with laboratory tests.

Summary: Home-use candidate tests in health-worker hands (Jaspard 2014, PLoS ONE — primary multicentre evaluation, 179 HIV-infected patients + 60 controls, France): oral-fluid sensitivities 87.2% (OraQuick), 88.3% (Chembio DPP), 58.9% and 28% (two anonymised tests) vs 100% for the finger-stick Sure Check test — the specimen/brand limb, performed by trained health workers, NOT self-administered

Result: Sensitivities of the oral-fluid tests were 87.2%, 88.3%, 58.9% and 28% (OraQuick, DPP, test A and test B respectively) compared with 100% for the finger-stick Sure Check test. Among patients with a detectable viral load (>50 copies/mL), oral-fluid sensitivities rose to 94.8%, 96.5%, 90% and 53.1%. Specificities of the four oral-fluid tests were 98.3%, 100%, 100% and 87.5%, compared with 100% for the finger-stick test.

Watch out: Performed by health workers, not self-administered: this card measures the specimen and the brand, not home use. Antibody-test sensitivity falls in treated patients with suppressed viral load (two-thirds of this cohort) — a population self-testing programmes will meet. Two of the five tests are anonymised in the record and cannot be named.

Source: Jaspard M et al. Finger-stick whole blood HIV-1/-2 home-use tests are more sensitive than oral fluid-based in-home HIV tests. PloS one. 2014. PMID: 24971842. Registry ID: none in the record. MeSH: AIDS Serodiagnosis/methods* · Adult · Case-Control Studies · Female · Fingers/blood supply* · HIV Antibodies/blood · HIV Infections/diagnosis* · HIV-1/immunology · HIV-2/immunology · Humans · Male · Reagent Kits, Diagnostic* · Saliva/virology* · Self Administration* · Sensitivity and Specificity

Card 5 — Specimen synthesis (Pant Pai 2012, Lancet Infectious Diseases — systematic review and meta-analysis of the OraQuick rapid point-of-care test by specimen): oral pooled sensitivity 98.03% (95% CI 95.85–99.08) vs blood 99.68% (97.31–99.96); specificity 99.74% vs 99.91%; in low-prevalence settings the oral positive predictive value fell to 88.55% vs 97.65% for blood — a provider-use synthesis, counted as the specimen-accuracy limb only

Study: Systematic review and meta-analysis (synthesis; provider-use specimen limb)

Population: Adults at risk of HIV in studies directly comparing the Oraquick advance rapid HIV-1/2 test (OraSure Technologies Inc, PA, USA, as named in the record) on oral versus blood-based specimens; studies in children, in co-infected populations, with self-reported inferior reference standards or incomplete reporting were excluded.

Summary: Specimen synthesis (Pant Pai 2012, Lancet Infectious Diseases — systematic review and meta-analysis of the OraQuick rapid point-of-care test by specimen): oral pooled sensitivity 98.03% (95% CI 95.85–99.08) vs blood 99.68% (97.31–99.96); specificity 99.74% vs 99.91%; in low-prevalence settings the oral positive predictive value fell to 88.55% vs 97.65% for blood — a provider-use synthesis, counted as the specimen-accuracy limb only

Result: Pooled sensitivity was about 2% lower in oral (98.03%, 95% CI 95.85–99.08) than in blood-based specimens (99.68%, 97.31–99.96), with similar specificity (oral 99.74%, 99.47–99.88; blood 99.91%, 99.84–99.95). Negative likelihood ratios were small and similar (oral 0.019; blood 0.003); positive likelihood ratios differed (oral 383.37; blood 1105.16) — likelihood ratios labelled as such. In high-prevalence settings PPVs were similar (oral 98.65%; blood 98.50%); in low-prevalence settings PPVs were lower for oral (88.55%) than blood (97.65%) specimens.

Watch out: A provider-use point-of-care synthesis: it prices the oral-fluid specimen, not self-administration. Its 98% oral sensitivity must never be quoted as unsupervised self-test performance — the field primaries above measured 89.7–90% plus invalid results.

Source: Pant Pai N et al. Head-to-head comparison of accuracy of a rapid point-of-care HIV test with oral versus whole-blood specimens: a systematic review and meta-analysis. The Lancet. Infectious diseases. 2012. PMID: 22277215. Registry ID: none in the record. MeSH: HIV Antibodies/analysis,blood* · HIV Infections/blood*,diagnosis* · Humans · Point-of-Care Systems/standards* · Saliva/chemistry* · Time Factors

Card 6 — Synthesis — supervised use is near-laboratory accurate; unsupervised use costs about 10 points of sensitivity plus a 15.1% invalid rate; and the oral-fluid specimen itself costs sensitivity against blood, catastrophically so for weaker brands

Study: Synthesis

Population: 5 reports: 4 primaries (n=2,198 supervised; n=240 and n=246 unsupervised; n=239 health-worker candidate-test evaluation) + 1 provider-use specimen synthesis. Three different accuracy questions — supervised use, unsupervised use, and the specimen — printed separately and never merged into a single pair.

Summary: Synthesis — supervised use is near-laboratory accurate; unsupervised use costs about 10 points of sensitivity plus a 15.1% invalid rate; and the oral-fluid specimen itself costs sensitivity against blood, catastrophically so for weaker brands

Result: Supervised oral self-testing: 98.7% / 100% (n=2,198). Unsupervised: 89.7% sensitivity with 15.1% invalid results (Kenya), and 90% vs 100% against supervised use in the randomised comparison, non-inferiority not shown intention-to-treat. Specimen: oral fluid pooled 98.03% vs blood 99.68% sensitivity in provider hands, with home-use candidate oral tests at 28–88.3% in health-worker hands against 100% for the finger-stick test. No report was refused on this row; the Jaspard and Pant Pai limbs are counted as specimen/brand evidence only, labelled on their faces. Cross-point: live pack #534 (self-test availability, uptake and linkage) is a different question.

Watch out: Accuracy is not uptake or linkage: nothing here shows self-testing finds more infections or links anyone to care. Antibody tests miss early infection and lose sensitivity in virally suppressed patients; a negative self-test is not the same evidentiary object in those groups.

Source: Synthesis of the reports carded above; no additional source.