Continuous glucose monitor accuracy (MARD) — 6 primaries, no synthesis: in-clinic MARDs 7.8–9.7% for current factory-calibrated systems against YSI, a head-to-head G7 MARD of 13.6% contradicting its pivotal 8.0%, and home-use MARDs of 14.8–18.0% in the older-generation three-way study
Shelf #8605 · Diabetes, glucose & metabolism
PREDICTIVE PERFORMANCE — in-clinic, YSI-referenced MARDs are 7.8–9.7% for current factory-calibrated systems, but the one independent head-to-head measured the G7 at 13.6% against its own pivotal 8.0% (the disagreement prints unresolved), and home-use MARDs in the older-generation three-way study were 14.8–18.0% against a point-of-care comparator (different devices and a different reference — confounded, printed as confounded). Base: 6 primaries, no synthesis; manufacturer-run pivotals plus two independent comparisons.
Stamped: 2026-11-28.
Evidence published: 2017–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In people with type 1 or type 2 diabetes, how accurately do factory-calibrated continuous glucose monitors measure glucose against laboratory (YSI) reference — overall mean absolute relative difference (MARD), agreement rates, and the in-clinic vs home-use gap — for the FreeStyle Libre 2 and Libre 3 (Abbott), Dexcom G7 (Dexcom) and Guardian Sensor 3 (Medtronic) systems, as named in the records?
Balance of the evidence
6 primaries, no synthesis — the CGM accuracy literature is a set of manufacturer-run pivotal accuracy studies plus two independent comparisons, and prints as such. In-clinic figures use the YSI laboratory reference; the home-use study used twice-daily point-of-care plasma glucose, so the in-clinic vs home gap mixes a device-generation effect with a reference effect and is never decomposed. MARD is a mean and conceals hypoglycaemic-range performance, which prints separately wherever a record gives it. Cross-points only: the many CGM efficacy, closed-loop and outcome packs (#560, #950, #959, #1762, #2633, #2740, #3535, #3541, #5845, #5846, #6054, #7172) ask different questions; no efficacy claim is borrowed here.
MeSH: Adult · Aged · Algorithms · Blood Glucose · Blood Glucose Self-Monitoring · Child · Diabetes Mellitus, Type 1*/drug therapy · Diabetes Mellitus, Type 2*/drug therapy · Glucose · Humans · Pediatrics* · Reproducibility of Results · Female · Male · Middle Aged · Young Adult · Blood Glucose*/analysis · Blood Glucose Self-Monitoring*/instrumentation,methods · Diabetes Mellitus, Type 1*/blood,drug therapy · Diabetes Mellitus, Type 2*/blood,drug therapy · Hypoglycemic Agents/therapeutic use · Insulin/therapeutic use · Prospective Studies · United States · Continuous Glucose Monitoring*/instrumentation,methods · Adolescent · Blood Glucose/analysis* · Blood Glucose Self-Monitoring/instrumentation* · Calibration · Diabetes Mellitus, Type 2/blood* · Insulin Infusion Systems* · Reference Values · Blood Glucose/analysis · Continuous Glucose Monitoring*/instrumentation,standards · Diabetes Mellitus, Type 1*/blood · Diabetes Mellitus, Type 2*/blood,diagnosis · Diabetes Mellitus, Type 1*/diagnosis · Pancreas, Artificial
Cards (7)
Card 1 — FreeStyle Libre 2 pivotal (Alva 2022, Journal of Diabetes Science and Technology — primary accuracy study, 144 adults + 129 children at 7 US sites against YSI): MARD 9.2% (adults) and 9.7% (paediatric); 93.2% / 92.1% of values within ±20%/20 mg/dL (≈1.1 mmol/L); hypoglycaemic-range agreement 94.3% / 96.1% within 15 mg/dL (≈0.8 mmol/L)
Study: Accuracy study (primary report, manufacturer pivotal)
Population: 144 adults and 129 paediatric participants aged 4 years and above with type 1 or type 2 diabetes at seven US sites; participants aged 11+ underwent supervised glycaemic manipulation during in-clinic sessions across the measurement range.
Summary: FreeStyle Libre 2 pivotal (Alva 2022, Journal of Diabetes Science and Technology — primary accuracy study, 144 adults + 129 children at 7 US sites against YSI): MARD 9.2% (adults) and 9.7% (paediatric); 93.2% / 92.1% of values within ±20%/20 mg/dL (≈1.1 mmol/L); hypoglycaemic-range agreement 94.3% / 96.1% within 15 mg/dL (≈0.8 mmol/L)
Result: Percent of sensor results within ±20%/20 mg/dL (≈1.1 mmol/L) of YSI reference were 93.2% and 92.1%, and MARD was 9.2% and 9.7% for the adults and paediatric participants respectively. In the hypoglycaemic range, 94.3% of adult and 96.1% of paediatric results were within 15 mg/dL (≈0.8 mmol/L) of the YSI reference. Time lag was 2.4 ± 4.6 minutes for adults and 2.1 ± 5.0 minutes for paediatrics.
Watch out: A manufacturer-run pivotal under in-clinic conditions with deliberate glycaemic manipulation — the setting that produces the best MARD a system will ever show. Two registry IDs are printed in the record (adult and paediatric studies).
Source: Alva S et al. Accuracy of a 14-Day Factory-Calibrated Continuous Glucose Monitoring System With Advanced Algorithm in Pediatric and Adult Population With Diabetes. Journal of diabetes science and technology. 2022. PMID: 32954812. Registry ID: ClinicalTrials.gov:NCT03607448; ClinicalTrials.gov:NCT03820050. MeSH: Adult · Aged · Algorithms · Blood Glucose · Blood Glucose Self-Monitoring · Child · Diabetes Mellitus, Type 1*/drug therapy · Diabetes Mellitus, Type 2*/drug therapy · Glucose · Humans · Pediatrics* · Reproducibility of Results
Card 2 — FreeStyle Libre 3 pivotal (Alva 2023, Diabetes Therapy — primary accuracy study, 100 evaluable participants aged ≥4 years at 4 US sites, 6,845 matched pairs against YSI): overall MARD 7.8%, 93.4% within ±20%/20 mg/dL (1.1 mmol/L, as the record itself prints), stable over 14 days; ages 4–5 against a fingerstick reference, MARD 10.0%
Study: Accuracy study (primary report, manufacturer pivotal)
Population: 108 participants aged ≥4 years with type 1 or type 2 diabetes enrolled at four US sites; data from 100 participants were ultimately evaluated. Adult participants attended three in-clinic sessions and paediatric participants up to two, stratified across sensor-wear days 1–14.
Summary: FreeStyle Libre 3 pivotal (Alva 2023, Diabetes Therapy — primary accuracy study, 100 evaluable participants aged ≥4 years at 4 US sites, 6,845 matched pairs against YSI): overall MARD 7.8%, 93.4% within ±20%/20 mg/dL (1.1 mmol/L, as the record itself prints), stable over 14 days; ages 4–5 against a fingerstick reference, MARD 10.0%
Result: The overall MARD was 7.8%, and 93.4% of the CGM values were within ±20% or ±20 mg/dL (1.1 mmol/L) of the YSI reference for participants aged ≥6 years, with 6,845 CGM-YSI matched pairs. Performance was stable over the 14-day wear period. For participants aged 4–5 years, MARD was 10.0% and 88.9% of values were within 20%/20 mg/dL against a self-monitoring blood glucose reference. No serious adverse events were reported.
Watch out: The record prints no registry ID and no MeSH — both absences print as such. The youngest children's figure (ages 4–5) uses a fingerstick reference, not YSI, and is not the same measurement as the 7.8% headline.
Source: Alva S et al. Accuracy of the Third Generation of a 14-Day Continuous Glucose Monitoring System. Diabetes therapy : research, treatment and education of diabetes and related disorders. 2023. PMID: 36877403. Registry ID: none in the record. MeSH: none in the record
Card 3 — Dexcom G7 15.5-day pivotal (Garg 2025, Diabetes Technology & Therapeutics — primary accuracy study of the investigational 15-day G7 iCGM, 130 adults at 6 US sites, 20,310 matched pairs against YSI): overall MARD 8.0%; %15/15 agreement 87.7%, %20/20 94.2%, %40/40 99.8%
Study: Accuracy study (primary report, manufacturer pivotal; investigational 15.5-day device)
Population: 130 adults (mean age 43.0 ± 14.4 years, 53.1% female, 86.9% with type 1 diabetes) with type 1 or type 2 diabetes on intensive insulin therapy at six US clinical sites, with four in-clinic visits and deliberate, closely monitored glucose manipulations.
Summary: Dexcom G7 15.5-day pivotal (Garg 2025, Diabetes Technology & Therapeutics — primary accuracy study of the investigational 15-day G7 iCGM, 130 adults at 6 US sites, 20,310 matched pairs against YSI): overall MARD 8.0%; %15/15 agreement 87.7%, %20/20 94.2%, %40/40 99.8%
Result: The study analyzed 20,310 CGM-YSI matched pairs from 130 15-day G7 CGM devices. The overall MARD was 8.0% and the %15/15, %20/20, %30/30 and %40/40 agreement rates were 87.7%, 94.2%, 98.9% and 99.8% respectively. The device exceeded iCGM performance goals; no serious adverse events were reported.
Watch out: The device studied is the investigational 15.5-day G7, not the marketed 10-day G7 — and the independent head-to-head card below measured a G7 MARD of 13.6%, flatly contradicting this 8.0%. Neither figure is suppressed; protocol, wear-day composition and reference handling differ.
Source: Garg SK et al. Accuracy of the 15.5-Day G7 iCGM in Adults with Diabetes. Diabetes technology & therapeutics. 2025. PMID: 40108991. Registry ID: ClinicalTrials.gov:NCT05263258. MeSH: Adult · Female · Humans · Male · Middle Aged · Young Adult · Blood Glucose*/analysis · Blood Glucose Self-Monitoring*/instrumentation,methods · Diabetes Mellitus, Type 1*/blood,drug therapy · Diabetes Mellitus, Type 2*/blood,drug therapy · Hypoglycemic Agents/therapeutic use · Insulin/therapeutic use · Prospective Studies · Reproducibility of Results · United States · Continuous Glucose Monitoring*/instrumentation,methods
Card 4 — Guardian Sensor 3 (Christiansen 2017, Diabetes Technology & Therapeutics — primary accuracy study, 88 participants aged 14–75 against YSI): MARD 9.6% ± 9.0% (abdomen, pump-paired), 9.4% ± 9.8% (abdomen, phone-paired), 8.7% ± 8.0% (arm); ~91–93% within-20% agreement — a previous-generation, calibration-dependent limb (calibration 3–4 times daily), labelled as such
Study: Accuracy study (primary report; previous-generation calibration-dependent system)
Population: 88 subjects (14–75 years, mean 42.0 ± 19.1 years) with type 1 or type 2 diabetes wearing two abdominal sensors (paired with a MiniMed 640G pump or a Guardian Connect phone system) and a third arm sensor, with 12–14-hour in-clinic visits on days 1, 3 and 7 for YSI reference sampling.
Summary: Guardian Sensor 3 (Christiansen 2017, Diabetes Technology & Therapeutics — primary accuracy study, 88 participants aged 14–75 against YSI): MARD 9.6% ± 9.0% (abdomen, pump-paired), 9.4% ± 9.8% (abdomen, phone-paired), 8.7% ± 8.0% (arm); ~91–93% within-20% agreement — a previous-generation, calibration-dependent limb (calibration 3–4 times daily), labelled as such
Result: Overall MARD ± SD between abdomen sensor glucose and YSI was 9.6% ± 9.0% (MiniMed 640G pump) and 9.4% ± 9.8% (Guardian Connect system), and 8.7% ± 8.0% for arm sensors. The percentage of sensor values within 20% agreement of YSI (for YSI >80 mg/dL, ≈4.4 mmol/L) was 90.7%, 91.8% and 93.1% respectively. Mean functional sensor life, calibrating 3–4 times/day, was about 146–148 hours.
Watch out: This sensor required calibration 3–4 times daily — it is not a factory-calibrated system and its MARDs are a previous-generation limb, never pooled with the factory-calibrated pivotals above.
Source: Christiansen MP et al. Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor. Diabetes technology & therapeutics. 2017. PMID: 28700272. Registry ID: ClinicalTrials.gov:NCT02246582. MeSH: Adolescent · Adult · Aged · Blood Glucose/analysis* · Blood Glucose Self-Monitoring/instrumentation* · Calibration · Diabetes Mellitus, Type 2/blood* · Female · Humans · Insulin Infusion Systems* · Male · Middle Aged · Reference Values · Reproducibility of Results · Young Adult
Card 5 — Independent head-to-head (Hanson 2024, Journal of Diabetes Science and Technology — primary comparison, 55 adults against YSI and capillary reference): MARD 13.6% for Dexcom G7 vs 8.9% for FreeStyle Libre 3 (P < 0.0001); within-±20%/20 mg/dL (≈1.1 mmol/L) agreement 78.6% vs 91.4%; the G7 deficit appeared after the first 12 hours (15.1% vs 10.0%) — the pack's discordant limb, printed at headline weight against the G7 pivotal's own 8.0%
Study: Accuracy study (primary report, independent head-to-head)
Population: 55 adults (≥18 years) with diagnosed type 1 or type 2 diabetes in a multicentre, single-arm, prospective, nonsignificant-risk evaluation wearing both systems, with sensor data compared to YSI laboratory reference and capillary blood glucose values.
Summary: Independent head-to-head (Hanson 2024, Journal of Diabetes Science and Technology — primary comparison, 55 adults against YSI and capillary reference): MARD 13.6% for Dexcom G7 vs 8.9% for FreeStyle Libre 3 (P < 0.0001); within-±20%/20 mg/dL (≈1.1 mmol/L) agreement 78.6% vs 91.4%; the G7 deficit appeared after the first 12 hours (15.1% vs 10.0%) — the pack's discordant limb, printed at headline weight against the G7 pivotal's own 8.0%
Result: Analysis showed significantly lower MARD with the FreeStyle Libre 3 sensor vs the Dexcom G7 sensor (8.9% vs 13.6%, P < .0001) with a higher percentage of glucose values within ±20 mg/dL/±20 of reference (91.4% vs 78.6%). MARD values were similar during the first 12 hours; the Libre 3 MARD was notably lower than the G7 MARD during the next 12 hours (10.0% vs 15.1%, P < .0001) and throughout the study period.
Watch out: This MARD for the G7 (13.6%) flatly contradicts the G7 pivotal figure (8.0%) carded above, and the disagreement is itself a finding of this pack — it is not averaged away. The comparison covers the early wear period in which the G7 deficit appeared; wear-day composition differs from the pivotals.
Source: Hanson K et al. Comparison of Point Accuracy Between Two Widely Used Continuous Glucose Monitoring Systems. Journal of diabetes science and technology. 2024. PMID: 38189290. Registry ID: none in the record. MeSH: Adult · Aged · Female · Humans · Male · Middle Aged · Blood Glucose/analysis · Continuous Glucose Monitoring*/instrumentation,standards · Diabetes Mellitus, Type 1*/blood · Diabetes Mellitus, Type 2*/blood,diagnosis · Prospective Studies · Reproducibility of Results
Card 6 — Three-way home-use study (Jafri 2020, Diabetes Technology & Therapeutics — primary comparison, 23 participants with type 1 diabetes wearing Dexcom G5, FreeStyle Libre Pro and Senseonics Eversense concurrently for 6 weeks against twice-daily point-of-care plasma glucose, 829 three-way sets): MARDs Eversense 14.8%, G5 16.3%, Libre Pro 18.0% — all higher than the same devices' in-clinic figures; older-generation professional/blinded variants, labelled as such
Study: Accuracy study (primary report, home-use; older-generation devices)
Population: 23 subjects with type 1 diabetes in a 6-week free-living home-use bionic pancreas study, wearing all three devices concurrently; point-of-care plasma glucose values were obtained approximately twice daily by the subjects.
Summary: Three-way home-use study (Jafri 2020, Diabetes Technology & Therapeutics — primary comparison, 23 participants with type 1 diabetes wearing Dexcom G5, FreeStyle Libre Pro and Senseonics Eversense concurrently for 6 weeks against twice-daily point-of-care plasma glucose, 829 three-way sets): MARDs Eversense 14.8%, G5 16.3%, Libre Pro 18.0% — all higher than the same devices' in-clinic figures; older-generation professional/blinded variants, labelled as such
Result: In the three-way comparison (829 sets), Eversense achieved the lowest nominal MARD (14.8%), followed by Dexcom G5 (16.3%) and Libre Pro (18.0%) (Eversense vs Libre Pro P = 0.004, other comparisons not significant). In the two-way comparison (2,277 sets), MARDs were Eversense 15.1% and G5 16.9% (P = 0.008). All three systems produced higher average MARDs than during in-clinic studies.
Watch out: Two effects are confounded here and print as confounded: these are older-generation, professional/blinded device variants AND the reference is a point-of-care meter-class measurement, not YSI. The home-vs-clinic gap cannot be assigned to either cause from this study alone. n=23 is small.
Source: Jafri RZ et al. A Three-Way Accuracy Comparison of the Dexcom G5, Abbott Freestyle Libre Pro, and Senseonics Eversense Continuous Glucose Monitoring Devices in a Home-Use Study of Subjects with Type 1 Diabetes. Diabetes technology & therapeutics. 2020. PMID: 32453604. Registry ID: none in the record. MeSH: Blood Glucose/analysis · Blood Glucose Self-Monitoring/instrumentation* · Diabetes Mellitus, Type 1*/diagnosis · Humans · Pancreas, Artificial · Reproducibility of Results
Card 7 — Synthesis — current factory-calibrated systems read at MARD 7.8–9.7% in clinic against YSI, but the only independent head-to-head contradicts one pivotal outright, and home-use figures in older devices run 14.8–18.0%
Study: Synthesis
Population: 6 primaries, no synthesis: four manufacturer pivotal accuracy studies (n=273, n=100 evaluable, n=130, n=88), one independent head-to-head (n=55) and one home-use three-way comparison (n=23). Populations are never pooled — MARDs are study-level figures against differing references.
Summary: Synthesis — current factory-calibrated systems read at MARD 7.8–9.7% in clinic against YSI, but the only independent head-to-head contradicts one pivotal outright, and home-use figures in older devices run 14.8–18.0%
Result: In-clinic YSI-referenced MARDs: Libre 3 7.8%, G7 15.5-day investigational 8.0%, Libre 2 9.2–9.7%, Guardian Sensor 3 (calibration-dependent, previous generation) 8.7–9.6%. Independent head-to-head: Libre 3 8.9% vs G7 13.6%. Home use (older devices, point-of-care reference): 14.8–18.0%. No report was refused on this row. Cross-points: CGM efficacy and closed-loop packs are different questions and borrow nothing from these accuracy figures.
Watch out: MARD is a mean: it conceals hypoglycaemic-range performance (printed separately only where a record gives it — the Libre 2 card) and says nothing about outcomes. Accuracy is not benefit; no glycaemic-outcome claim may be built on these figures.
Source: Synthesis of the reports carded above; no additional source.