SourceMed · Shelf #8621 in the ask interface
Full-field ERG reference limits (ISCEV standard) — 6 reports: there is no electrode-free ERG number — silver-thread amplitudes are 55–65% of gold-foil and skin 35–38% of silver-thread, while peak times transfer almost unchanged; the age gradient is rod-dominant in every cohort that looked
Shelf #8621 · Eyes, skin, nose & allergy
PHYSIOLOGY — ESTABLISHED — ISCEV-standard full-field ERG reference limits are electrode-bound: amplitude limits must travel with their electrode (silver-thread amplitudes 55–65% of gold-foil; skin-electrode amplitudes 35–38% of silver-thread), while peak times agree closely across electrodes (LA 30 Hz bias −0.2 ms, 95% limits of agreement −1.1 to 0.7 ms); rod-driven responses show the strongest age attenuation and delay in every cohort that looked; 6 reports, combined documented N=624 plus one component limb with N not documented
Stamped: 2026-11-28.
Evidence published: 2015–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In healthy humans, what are the reference values for the ISCEV-standard full-field electroretinogram (ERG) — dark-adapted 0.01/3/10, light-adapted 3 and 30 Hz flicker amplitudes and peak times — how do they fall and slow with age, and how far do the limits move with the recording electrode (silver thread, gold foil, skin, contact-lens and handheld systems)?
Balance of the evidence
The composition is 6 reports: a multicentre transference/validation study building a combined reference sample of 407 healthy individuals across 8 decades; an RM contact-lens electrode normative study (40 volunteers); an extended-protocol age study (73 participants); a skin-electrode normative study (50 young volunteers); a Romanian age-variability study (54 valid recordings); and a photopic negative response (PhNR) component limb from a handheld mini-Ganzfeld study whose record prints no N — a component limb of the light-adapted protocol, never a sixth independent full-protocol cohort. Two anchors print their age effects qualitatively at abstract level and no numeric limit is manufactured from them. Nearest homed neighbours are structural retinal-imaging packs (#7179, #7180, #7258) and the VEP row — different signals, discharged at question level. No registry IDs in any record.
MeSH: Humans · Male · Female · Adult · Middle Aged · Electrodes · Reference Values · Electroretinography*/instrumentation,standards · Young Adult · Healthy Volunteers · Aged · Adolescent · Retina*/physiology · Silver · Reproducibility of Results · Gold · Aged, 80 and over · Electroretinography*/instrumentation,methods · Aging*/physiology · Dark Adaptation/physiology · Age Factors · Electroretinography*/methods · Dark Adaptation*/physiology · Photic Stimulation* · Healthy Volunteers* · Retinal Cone Photoreceptor Cells*/physiology · Retinal Rod Photoreceptor Cells/physiology · Dark Adaptation · Electroretinography/instrumentation,methods* · Galvanic Skin Response/physiology* · Retina/physiology* · Visual Fields/physiology* · Color Vision/physiology* · Electroretinography/methods* · Microelectrodes · Observer Variation · Photic Stimulation · Retinal Ganglion Cells/physiology
Cards (7)
Card 1 — Multicentre transference study (Baker 2025, Documenta Ophthalmologica — Multicenter Study, Validation Study): combined reference sample of 407 healthy individuals across 8 decades; silver-thread data combinable across two centres, gold-foil data transformable into the silver-thread range; silver-thread amplitudes were 55–65% of gold-foil amplitudes and skin-electrode amplitudes 35–38% of silver-thread, while peak times agreed closely (LA 30 Hz bias −0.2 ms, 95% limits of agreement −1.1 to 0.7 ms)
Study: Reference-value study (multicentre transference/validation)
Population: Combined reference sample of 407 healthy individuals: silver-thread reference data from two centres verified for combination (n=251), gold-foil reference data transformed for inclusion (n=156), with paired electrode comparisons in reference subjects and patients (n=53 silver thread vs gold foil; n=41 skin).
Summary: Multicentre transference study (Baker 2025, Documenta Ophthalmologica — Multicenter Study, Validation Study): combined reference sample of 407 healthy individuals across 8 decades; silver-thread data combinable across two centres, gold-foil data transformable into the silver-thread range; silver-thread amplitudes were 55–65% of gold-foil amplitudes and skin-electrode amplitudes 35–38% of silver-thread, while peak times agreed closely (LA 30 Hz bias −0.2 ms, 95% limits of agreement −1.1 to 0.7 ms)
Result: Silver thread data were sufficiently similar across two centres to justify combination. Peak times for gold foil and silver thread were closely comparable (r² 0.75–0.98, Bland-Altman bias ≤1.6 ms for all components), with LA 30 Hz peak time showing the highest agreement (bias −0.2 ms, 95% limits of agreement −1.1 to 0.7 ms). There was a clinically significant amplitude difference: silver thread ERGs were 55–65% of gold foil amplitudes (limits of agreement 29–90%) and skin ERGs 35–38% of silver thread (limits 18–54%). Age-adjusted reference limits were derived for the main DA 0.01, DA 3, DA 10, LA 30 Hz and LA 3 components.
Watch out: The combined sample mixes directly recorded and mathematically transformed data — legitimate by the authors' verification, but the transformed limb is not 156 fresh silver-thread recordings. Amplitude limits from this sample apply to silver-thread recording; applying them to skin-electrode traces would misclassify most healthy people.
Source: Baker RA et al. ISCEV standard full-field ERG reference limits from 407 healthy subjects, derived from transference and validation of reference data between electrode types and centres. Documenta ophthalmologica. Advances in ophthalmology. 2025. PMID: 40164784. Registry ID: none in the record. MeSH: Humans · Male · Female · Adult · Middle Aged · Electrodes · Reference Values · Electroretinography*/instrumentation,standards · Young Adult · Healthy Volunteers · Aged · Adolescent · Retina*/physiology · Silver · Reproducibility of Results · Gold · Aged, 80 and over
Estimate type: adjusted estimate, as stated in the card.
Card 2 — RM electrode normative study (Jeffrey 2025, Translational Vision Science & Technology — Comparative Study): 40 healthy volunteers, mean age 39.4 years; normative reference ranges for the single-use RM contact-lens electrode across ISCEV standard and four extended protocols; RM dark-adapted a-wave amplitudes 18% larger and light-adapted b-wave 20% larger than Burian-Allen (P=0.02 and P<0.001), scotopic b-wave not different, and 30 Hz recordings more stable on RM (median %CV 12.3% vs 16.7%, P=0.001)
Study: Reference-value study (new-electrode normative ranges)
Population: 40 (19 female) healthy volunteers aged 39.4 ± 15.7 years, recorded with ISCEV standard plus four extended protocols (dark-adapted red flash, photopic negative response, ON-OFF, S-cones), with RM vs Burian-Allen comparison.
Summary: RM electrode normative study (Jeffrey 2025, Translational Vision Science & Technology — Comparative Study): 40 healthy volunteers, mean age 39.4 years; normative reference ranges for the single-use RM contact-lens electrode across ISCEV standard and four extended protocols; RM dark-adapted a-wave amplitudes 18% larger and light-adapted b-wave 20% larger than Burian-Allen (P=0.02 and P<0.001), scotopic b-wave not different, and 30 Hz recordings more stable on RM (median %CV 12.3% vs 16.7%, P=0.001)
Result: All amplitudes and implicit times followed a log-normal distribution (except photopic b-wave and ON-OFF amplitudes); most parameters varied with age. RM electrode dark-adapted a-wave and light-adapted b-wave amplitudes were 18% (P=0.02) and 20% (P<0.001) larger than the Burian-Allen electrode; scotopic b-wave amplitudes did not differ. The 30-Hz ERGs were more stable for RM (median %CV 12.3% [IQR 7.6–17.6]) than Burian-Allen (16.7% [12.5–26.8], P=0.001).
Watch out: A 40-volunteer sample: its reference ranges are adoption-enabling for a new disposable electrode, not population limits. Its electrode differences (18–20%) are smaller than the thread/foil/skin differences in card 1 but still electrode-sized.
Source: Jeffrey BG et al. Electroretinography With the RM Electrode: Normative Reference Ranges, Variation With Age, and Comparison With the Burian-Allen Electrode. Translational vision science & technology. 2025. PMID: 40478591. Registry ID: none in the record. MeSH: Humans · Electroretinography*/instrumentation,methods · Female · Male · Adult · Reference Values · Middle Aged · Electrodes · Young Adult · Aged · Healthy Volunteers · Aging*/physiology · Adolescent · Dark Adaptation/physiology · Age Factors
Card 3 — Age study (Jung 2024, Documenta Ophthalmologica): 73 healthy participants aged 14–73, extended ISCEV protocol; dark-adapted rod-driven responses showed the strongest age effects — b-wave attenuation and delay greatest at the dimmest (0.001 cd·s/m²) flash, a-wave reduction strongest at DA 3 — while cone flash/flicker responses changed comparatively little and S-cone function tended to diminish early without reaching significance; this record prints its age effects qualitatively and no numeric limit may be manufactured from it
Study: Reference-value study (age-dependency, extended protocol)
Population: 73 healthy participants aged 14–73 years, full-field ERG with an extended ISCEV standard protocol including an additional 9 Hz flicker stimulus and isolated On-Off and S-cone paradigms.
Summary: Age study (Jung 2024, Documenta Ophthalmologica): 73 healthy participants aged 14–73, extended ISCEV protocol; dark-adapted rod-driven responses showed the strongest age effects — b-wave attenuation and delay greatest at the dimmest (0.001 cd·s/m²) flash, a-wave reduction strongest at DA 3 — while cone flash/flicker responses changed comparatively little and S-cone function tended to diminish early without reaching significance; this record prints its age effects qualitatively and no numeric limit may be manufactured from it
Result: Dark-adapted ERGs displayed notable age-related alterations. The attenuation and delay of the b-wave with higher age were most significant in the dark-adapted, rod-driven 0.001 cd·s/m² flash ERG. The age-dependent reduction of a-wave amplitude was strongest in the standard dark-adapted 3 cd·s/m² flash condition. Cone-driven, light-adapted responses to flash or flicker displayed comparatively small alterations at higher age. S-cone function tended to diminish at an early age, but the effect was not significant in the whole population. Response amplitudes displayed wide variability across the whole sample.
Watch out: Qualitative at abstract level: directions and ordering of age effects only. The wide amplitude variability the authors report is itself the caution against tight age-banded amplitude limits from samples of this size.
Source: Jung R et al. Age-dependencies of the electroretinogram in healthy subjects. Documenta ophthalmologica. Advances in ophthalmology. 2024. PMID: 39251480. Registry ID: none in the record. MeSH: Humans · Electroretinography*/methods · Adult · Male · Female · Adolescent · Young Adult · Middle Aged · Aged · Dark Adaptation*/physiology · Photic Stimulation* · Healthy Volunteers* · Aging*/physiology · Retinal Cone Photoreceptor Cells*/physiology · Reference Values · Retinal Rod Photoreceptor Cells/physiology
Card 4 — Skin-electrode normative study (Fernandes 2016, Arquivos Brasileiros de Oftalmologia — Comparative Study): 50 healthy volunteers aged 17–26; skin-electrode normative values derived against microfiber electrodes, with peak-to-peak amplitude reductions of 61.4% (rod), 61.5% (maximal), 46.2% (oscillatory potentials), 57.4% (cone) and 54.4% (30 Hz flicker) and comparable implicit times
Study: Reference-value study (skin-electrode normative values)
Population: 50 healthy volunteers (17–26 years; mean 20.63 ± 2.01) recorded simultaneously with a gold disk skin electrode on the lower orbital rim of one eye and a microfiber electrode in the contralateral eye, under the ISCEV protocol.
Summary: Skin-electrode normative study (Fernandes 2016, Arquivos Brasileiros de Oftalmologia — Comparative Study): 50 healthy volunteers aged 17–26; skin-electrode normative values derived against microfiber electrodes, with peak-to-peak amplitude reductions of 61.4% (rod), 61.5% (maximal), 46.2% (oscillatory potentials), 57.4% (cone) and 54.4% (30 Hz flicker) and comparable implicit times
Result: Both electrode types showed similar wave morphologies and comparable implicit times. On peak-to-peak amplitude, skin-electrode recordings showed reductions of 61.4% for rod responses, 61.5% for maximal responses, 46.2% for oscillatory potentials, 57.4% for cone responses, and 54.4% for 30 Hz-flicker responses versus microfiber electrodes. Normative values for peak-to-peak amplitude and implicit time for skin-electrode recording were determined.
Watch out: A young-only sample (17–26 years): its skin-electrode norms cannot be age-extended. Its reduction percentages are against microfiber electrodes, a different comparator chain than card 1's thread/foil ratios — the two are consistent in direction, never arithmetically merged.
Source: Fernandes AG et al. Full-field electroretinogram recorded with skin electrodes in normal adults. Arquivos brasileiros de oftalmologia. 2016. PMID: 28076567. Registry ID: none in the record. MeSH: Adolescent · Adult · Dark Adaptation · Electrodes · Electroretinography/instrumentation,methods* · Female · Galvanic Skin Response/physiology* · Humans · Male · Retina/physiology* · Visual Fields/physiology* · Young Adult
Card 5 — Age-variability study (Corîci 2015, Current Health Sciences Journal): 54 healthy subjects aged 20–80 with valid recordings (59 enrolled); b-wave latencies larger above age 50 for DA 0.01, DA 3.0 and LA 3.0, oscillatory-potential latencies not different, and amplitude age differences not statistically provable because of variability; the record prints no MeSH
Study: Reference-value study (age variability)
Population: 59 healthy Romanian subjects aged 20–80 enrolled prospectively; valid ERG recordings for all tests from 54 subjects, recorded with a Metrovision MonPack One system under ISCEV protocols.
Summary: Age-variability study (Corîci 2015, Current Health Sciences Journal): 54 healthy subjects aged 20–80 with valid recordings (59 enrolled); b-wave latencies larger above age 50 for DA 0.01, DA 3.0 and LA 3.0, oscillatory-potential latencies not different, and amplitude age differences not statistically provable because of variability; the record prints no MeSH
Result: ERG latency values were bigger in subjects above 50 years than in younger subjects for the b wave in dark-adapted 0.01, dark-adapted 3.0 and light-adapted 3.0, and for the a wave in dark-adapted 0.01 and 3.0. There was no significant difference in latency for dark-adapted oscillatory potentials between young and old subjects. Because of increased variability, the observed amplitude differences could not be proved statistically significant. The authors conclude there is a major loss in retinal activity with aging, most of it from the rod cells' delayed response.
Watch out: The record prints no MeSH — stated on the face. A null amplitude limb (not provable because of variability) prints as a null: latency slowing with age is supported here, amplitude decline is not, in this cohort.
Source: Corîci AC et al. Variability of Normal Values of Electroretinogram Parameters Due to Aging in Healthy Individuals. Current health sciences journal. 2015. PMID: 30151247. Registry ID: none in the record. MeSH: none in the record
Card 6 — PhNR component limb (Berezovsky 2021, Documenta Ophthalmologica — handheld mini-Ganzfeld; healthy adults, N not documented in the record): normative limits (5th percentile) for photopic negative response amplitude at 1/5/7 cd·s/m² of 20.81 / 18.06 / 19.60 µV (baseline-to-trough) and 69.11 / 77.98 / 76.51 µV (peak-to-trough), peak-time limits 65.98 / 78.20 / 77.96 ms, intra-session coefficients of variation 1.35–10.28% — a component limb of the light-adapted protocol from a single centre, never a sixth independent full-protocol cohort
Study: Reference-value study (component limb — photopic negative response, handheld system)
Population: Healthy adults with no visual complaints, visual acuity 20/20 or better and negative family history for visual diseases, recorded from both eyes with a DTL-type fiber electrode and a handheld LED mini-Ganzfeld stimulator. N not documented in the record.
Summary: PhNR component limb (Berezovsky 2021, Documenta Ophthalmologica — handheld mini-Ganzfeld; healthy adults, N not documented in the record): normative limits (5th percentile) for photopic negative response amplitude at 1/5/7 cd·s/m² of 20.81 / 18.06 / 19.60 µV (baseline-to-trough) and 69.11 / 77.98 / 76.51 µV (peak-to-trough), peak-time limits 65.98 / 78.20 / 77.96 ms, intra-session coefficients of variation 1.35–10.28% — a component limb of the light-adapted protocol from a single centre, never a sixth independent full-protocol cohort
Result: Normative limits for PhNR amplitude (µV) using 1, 5 and 7 cd·s/m² stimuli were respectively 20.81, 18.06 and 19.60 for baseline-to-trough and 69.11, 77.98 and 76.51 for peak-to-trough. Peak-time normative limits were 65.98, 78.20 and 77.96 ms. Intra-session variability by coefficient of variation ranged from 1.35 to 10.28%. Inter-session variability showed significant intraclass correlation for all PhNR parameters only for the 1 cd·s/m² stimulus. The tested stimulator is named as the publication gives it: ColorBurst (Diagnosys LLC, Lowell, MA, USA).
Watch out: Single centre, N not documented in the record, and a ganglion-cell component of the light-adapted protocol — its limits belong to the PhNR on this handheld system only and support no claim about the full ISCEV protocol.
Source: Berezovsky A et al. Photopic negative response using a handheld mini-ganzfeld stimulator in healthy adults: normative values, intra- and inter-session variability. Documenta ophthalmologica. Advances in ophthalmology. 2021. PMID: 32681419. Registry ID: none in the record. MeSH: Adolescent · Adult · Color Vision/physiology* · Electroretinography/methods* · Female · Healthy Volunteers · Humans · Male · Microelectrodes · Middle Aged · Observer Variation · Photic Stimulation · Reference Values · Retina/physiology* · Retinal Ganglion Cells/physiology · Young Adult
Card 7 — Synthesis — ERG reference limits are established physiology, but only as electrode-specific, age-aware ranges: amplitudes do not transfer across electrodes, peak times largely do, and rods age faster than cones
Study: Synthesis
Population: 6 reports, combined documented N=624 (407 + 40 + 73 + 50 + 54) plus the PhNR limb with N not documented. Counting units are reference individuals/volunteers per report; the flagship's 407 includes 156 transformed gold-foil records, printed as such on its card.
Summary: Synthesis — ERG reference limits are established physiology, but only as electrode-specific, age-aware ranges: amplitudes do not transfer across electrodes, peak times largely do, and rods age faster than cones
Result: Electrode effects: silver-thread amplitudes 55–65% of gold-foil; skin 35–38% of silver-thread; skin vs microfiber reductions 46–62% by component; RM contact-lens 18–20% above Burian-Allen on two components. Peak times: near-identical across thread/foil (LA 30 Hz bias −0.2 ms) and comparable skin vs microfiber. Age: rod-driven dark-adapted responses attenuate and delay most (greatest at the dimmest flash), cone responses change little, latencies lengthen above 50, and one cohort could not prove amplitude decline amid variability — printed as its null. The PERG normative family verified at only 1 cohort in the slate census and is not stretched into this pack.
Watch out: No registry IDs in any record. Every limit in this pack is system-bound (electrode, stimulator, protocol version); a laboratory adopting another centre's limits must verify transference for its own electrode, exactly as the flagship study did.
Source: Synthesis of the reports carded above; no additional source.