Corneal hysteresis and corneal resistance factor (Ocular Response Analyzer) in healthy eyes — 4 reports: population means cluster at CH ≈10–11.6 mmHg (Saudi 11.16 ± 2.11, Japanese 10.1 ± 1.5), with within-cohort spread (SD 1.5–2.1) dwarfing the age, sex and refractive effects

Shelf #8622 · Eyes, skin, nose & allergy

PHYSIOLOGY — ESTABLISHED — normal Ocular Response Analyzer values in healthy eyes centre on CH ≈10–11.6 mmHg and CRF ≈10–11.2 mmHg across four cohorts (combined N=559), printed as a wide, device-bound distribution: within-cohort standard deviations of 1.5–2.1 mmHg dwarf the systematic effects (weak negative age correlation r=−0.17 for CH; about 0.13 mmHg lower CH per dioptre of myopia), and all values are ORA values, not cornea-in-general values

Stamped: 2026-11-28.

Evidence published: 1995–2016.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In healthy eyes, what are the normal values for corneal hysteresis (CH) and corneal resistance factor (CRF) measured with the Ocular Response Analyzer — and how do they move with age, refractive error and central corneal thickness?

Balance of the evidence

The composition is 4 reports, all using the same instrument family (the Ocular Response Analyzer, named as the publications give it): a Saudi population study (215 normal adults), a Japanese age study (204 healthy volunteers), a Caucasian refractive study (95 normal adults), and a repeatability study (45 healthy volunteers, counted as a repeatability report). Nearest homed neighbours are the IOP distribution (#7454) and central corneal thickness (#7181) packs — pressure and anatomy, not biomechanics — discharged at question level and named in prose only. No registry IDs in any record.

MeSH: Adult · Aged · Aged, 80 and over · Aging/physiology* · Biomechanical Phenomena/physiology · Cornea/physiology* · Elastic Tissue/physiology* · Female · Humans · Intraocular Pressure/physiology · Male · Middle Aged · Prospective Studies · Young Adult · Biomechanical Phenomena · Cornea/pathology,physiopathology* · Intraocular Pressure/physiology* · Myopia/pathology,physiopathology* · Refractive Errors/pathology,physiopathology* · United Kingdom · Adolescent · Diagnostic Techniques, Ophthalmological/instrumentation*,standards · Elasticity/physiology* · Elasticity Imaging Techniques/instrumentation* · Reference Values · Reproducibility of Results · Tonometry, Ocular

Cards (5)

Card 1 — Saudi population study (Al-Arfaj 2016, Saudi Journal of Ophthalmology): 215 normal Saudi adults; mean CH 11.16 ± 2.11 mmHg, CRF 11.07 ± 2.31 mmHg, with sex differences in both (CH P=0.020, CRF P=0.047; higher in female subjects); CRF correlated with central corneal thickness (r=0.447, p=0.000) and CH (r=0.878); the record prints no MeSH — product named as the publication gives it: Ocular Response Analyzer (Reichert)

Study: Reference-value study (population cross-section)

Population: 215 normal Saudi adult individuals in a prospective cross-sectional observational study; CH, CRF, Goldmann-correlated and corneal-compensated IOP measured by ocular response analyzer, with Pentacam central keratometry, central corneal thickness and anterior chamber depth.

Summary: Saudi population study (Al-Arfaj 2016, Saudi Journal of Ophthalmology): 215 normal Saudi adults; mean CH 11.16 ± 2.11 mmHg, CRF 11.07 ± 2.31 mmHg, with sex differences in both (CH P=0.020, CRF P=0.047; higher in female subjects); CRF correlated with central corneal thickness (r=0.447, p=0.000) and CH (r=0.878); the record prints no MeSH — product named as the publication gives it: Ocular Response Analyzer (Reichert)

Result: Mean CH was 11.16 ± 2.11, CRF 11.07 ± 2.31 and IOP 15.12 ± 3.5. Mean central keratometry, anterior chamber depth, CH and CRF were distinct among gender (P values 0.05, 0.006, 0.020 and 0.047 respectively), with CH and CRF higher in female Saudi subjects. CRF was negatively correlated with anterior chamber depth (r=−0.146, P=0.032) and positively with spherical equivalent (r=0.176, P=0.010), central corneal thickness (r=0.447, p=0.000) and CH (r=0.878, p=0.000). CH was negatively correlated with corneal-compensated IOP (−0.433, p=0.000) and positively with central corneal thickness (r=0.412, p=0.000) and Goldmann-correlated IOP (r=0.183, p=0.007). The authors state CH and CRF were higher in Saudi subjects than in other populations, suggesting ethnic differences and the importance of population norms.

Watch out: The record prints no MeSH — stated on the face. One internal inconsistency in the record's correlation list (a 'positive' CH–anterior-chamber-depth correlation printed with a negative r) is not carded; the means and the CCT correlations are the load-bearing figures. A single-population sample: its higher means may be population, protocol, or both.

Source: Al-Arfaj K et al. Corneal biomechanics in normal Saudi individuals. Saudi journal of ophthalmology : official journal of the Saudi Ophthalmological Society. 2016. PMID: 28210179. Registry ID: none in the record. MeSH: none in the record

Card 2 — Japanese age study (Kamiya 2009, Journal of Refractive Surgery): 204 healthy Japanese volunteers aged 19–89; mean CH 10.1 ± 1.5 mmHg and CRF 10.1 ± 1.6 mmHg at mean central corneal thickness 539.1 ± 30.9 µm, with weak significant negative age correlations for CH (r=−0.17, P=.02) and CRF (r=−0.18, P=.01) and no age correlation with CCT or IOP — product named as the publication gives it: Reichert Ocular Response Analyzer

Study: Reference-value study (age effects)

Population: 204 normal eyes of 204 healthy Japanese volunteers (68 men, 136 women; mean age 46.7 ± 19.4 years, range 19–89); measurements performed 3 times and averaged; central corneal thickness by ultrasound pachymeter.

Summary: Japanese age study (Kamiya 2009, Journal of Refractive Surgery): 204 healthy Japanese volunteers aged 19–89; mean CH 10.1 ± 1.5 mmHg and CRF 10.1 ± 1.6 mmHg at mean central corneal thickness 539.1 ± 30.9 µm, with weak significant negative age correlations for CH (r=−0.17, P=.02) and CRF (r=−0.18, P=.01) and no age correlation with CCT or IOP — product named as the publication gives it: Reichert Ocular Response Analyzer

Result: Mean corneal hysteresis and corneal resistance factor were 10.1 ± 1.5 mmHg and 10.1 ± 1.6 mmHg respectively. Mean central corneal thickness was 539.1 ± 30.9 microm. A weak but significant negative correlation was found between age and corneal hysteresis (r=−0.17, P=.02) and corneal resistance factor (r=−0.18, P=.01). No significant correlation was found between age and central corneal thickness, corneal-compensated IOP, or Goldmann-correlated IOP.

Watch out: The age correlations are weak (r≈−0.17 explains about 3% of variance): age moves CH slightly, it does not define it. The authors' collagen cross-linking interpretation is their mechanism hypothesis, printed in their conclusion, not a measured quantity.

Source: Kamiya K et al. Effect of aging on corneal biomechanical parameters using the ocular response analyzer. Journal of refractive surgery (Thorofare, N.J. : 1995). 2009. PMID: 19835329. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Aging/physiology* · Biomechanical Phenomena/physiology · Cornea/physiology* · Elastic Tissue/physiology* · Female · Humans · Intraocular Pressure/physiology · Male · Middle Aged · Prospective Studies · Young Adult

Card 3 — Refractive study (Plakitsi 2011, Ophthalmic and Physiological Optics — Multicenter Study): 95 normal Caucasian adults aged 19–48; CH 10.9 ± 1.5 mmHg in emmetropes/low myopes vs 10.0 ± 1.2 mmHg in high myopes (>−6.00 D), a fall of about 0.13 mmHg per dioptre (r²=0.084), with inter-subject variation much greater than any systematic change and CRF uncorrelated with refractive error

Study: Reference-value study (refractive-error effects)

Population: 95 normal adult subjects aged 19–48 years, spherical equivalent refractive errors from +0.25 to −14.00 D, CH and CRF measured with the Ocular Response Analyser in both eyes.

Summary: Refractive study (Plakitsi 2011, Ophthalmic and Physiological Optics — Multicenter Study): 95 normal Caucasian adults aged 19–48; CH 10.9 ± 1.5 mmHg in emmetropes/low myopes vs 10.0 ± 1.2 mmHg in high myopes (>−6.00 D), a fall of about 0.13 mmHg per dioptre (r²=0.084), with inter-subject variation much greater than any systematic change and CRF uncorrelated with refractive error

Result: CH was slightly lower in high myopes (>−6.00 D, mean CH 10.0 ± 1.2 mmHg) than in moderate myopes (10.1 ± 1.4) and emmetropes/low myopes (+0.25 to −2.75 D, 10.9 ± 1.5). The decrease in CH with degree of myopia was about 0.13 mmHg per D (r²=0.084, p<0.001). Inter-subject variations were much greater than any systematic changes. CRF was not correlated with refractive error (r²=0.001, p=0.66). Refractive error was highly correlated between eyes (r²=0.89), CH and CRF less so (r²=0.68 and 0.77).

Watch out: r²=0.084: refractive error explains about 8% of CH variance in this sample — a real but small systematic shift inside a wide scatter, exactly as the authors state. A 19–48 age band cannot speak to older corneas.

Source: Plakitsi A et al. Corneal biomechanical properties measured with the Ocular Response Analyser in a myopic population. Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists). 2011. PMID: 21615446. Registry ID: none in the record. MeSH: Adult · Biomechanical Phenomena · Cornea/pathology,physiopathology* · Female · Humans · Intraocular Pressure/physiology* · Male · Middle Aged · Myopia/pathology,physiopathology* · Refractive Errors/pathology,physiopathology* · United Kingdom

Card 4 — Repeatability study (Goebels 2012, Current Eye Research): 45 healthy volunteers; mean CH 11.58 and CRF 11.21 across a five-measurement sequence, repeatability coefficients (Cronbach's α) >0.9 for both, with no significant difference between the four-shot average and the device's best-signal value — counted as a repeatability report, not a population derivation; product named as the publication gives it: ORA, Reichert Ophthalmic Instruments, Depew, NY

Study: Reference-value study (repeatability)

Population: 45 eyes from 45 healthy volunteers without ocular pathologies and with normal visual acuity; a sequence of five consecutive ORA measurements per subject, each the average of four shots, compared with the best signal value (BSV) from the device's internal processing.

Summary: Repeatability study (Goebels 2012, Current Eye Research): 45 healthy volunteers; mean CH 11.58 and CRF 11.21 across a five-measurement sequence, repeatability coefficients (Cronbach's α) >0.9 for both, with no significant difference between the four-shot average and the device's best-signal value — counted as a repeatability report, not a population derivation; product named as the publication gives it: ORA, Reichert Ophthalmic Instruments, Depew, NY

Result: Mean value for CH was 11.58 and BSV 11.55; for CRF, mean 11.21 and BSV 11.28, with no significant difference between average and BSV for either (p>0.05). Within a five-measurement sequence, the average Goldmann-correlated and corneal-compensated IOP values decreased by 1.19 and 1.05 mmHg on average. The repeatability test revealed good results for CH and CRF (all Cronbach's α higher than 0.9).

Watch out: Good within-session repeatability (α>0.9) does not narrow the between-person distribution — it certifies the instrument's consistency, not a tight norm. The IOP drift across a measurement sequence (about 1 mmHg) is a protocol caution for the pressure outputs, not for CH/CRF.

Source: Goebels SC et al. Precision of ocular response analyzer. Current eye research. 2012. PMID: 22587394. Registry ID: none in the record. MeSH: Adolescent · Adult · Biomechanical Phenomena · Cornea/physiology* · Diagnostic Techniques, Ophthalmological/instrumentation*,standards · Elasticity/physiology* · Elasticity Imaging Techniques/instrumentation* · Female · Humans · Intraocular Pressure/physiology · Male · Middle Aged · Reference Values · Reproducibility of Results · Tonometry, Ocular · Young Adult

Card 5 — Synthesis — CH and CRF normals are established as wide, device-bound distributions centred near 10–11.6 mmHg; population, age and refractive effects are all small against the within-cohort spread

Study: Synthesis

Population: 4 reports, combined N=559 (215 + 204 + 95 + 45): three population/refractive cross-sections and one repeatability study, all on the Ocular Response Analyzer family. Means are never pooled into a single combined norm — they print per population.

Summary: Synthesis — CH and CRF normals are established as wide, device-bound distributions centred near 10–11.6 mmHg; population, age and refractive effects are all small against the within-cohort spread

Result: Centres: CH 11.16 ± 2.11 (Saudi), 10.1 ± 1.5 (Japanese), 10.9 ± 1.5 emmetropes/low myopes falling to 10.0 ± 1.2 in high myopes (Caucasian), 11.58 (repeatability cohort); CRF tracks CH closely (r=0.878 in the Saudi cohort). Modifiers: weak negative age correlations (r=−0.17/−0.18), about 0.13 mmHg CH per dioptre of myopia, positive CCT correlations (r=0.41–0.45), and higher values in women in the Saudi cohort. Repeatability α>0.9 certifies measurement consistency. Cross-points: IOP distribution (#7454) and central corneal thickness (#7181) are homed questions — pressure and anatomy, not biomechanics — named in prose only.

Watch out: No registry IDs in any record. These are ORA values from one instrument family; another biomechanics device would need its own normative base. The Saudi record's internal correlation-sign inconsistency is noted on its card and no figure from that line is carded.

Source: Synthesis of the reports carded above; no additional source.