SourceMed · Shelf #8685 in the ask interface
Retinal vessel oxygen saturation by fundus oximetry — 4 reports, N=476 subjects across at least three instrument families: venous saturation agrees (55.3% vs 52.8%) while arterial saturation disagrees (90.4% vs 97.0%) with opposite age directions — a device-bound distribution
Shelf #8685 · Eyes, skin, nose & allergy
PHYSIOLOGY — EMERGING — retinal oximetry values are device property, not a single normal range: the two numeric normative cohorts agree on venous saturation (55.3 ± 7.1% in 61 US subjects; 52.8 ± 8.3% in 252 Japanese subjects, Oxymap T1) but disagree on arterial saturation (90.4 ± 4.3% vs 97.0 ± 6.9%) and on its age direction (falling vs rising). Replicated modifiers: temporal quadrants read lower, children under 10 read lower than adults, and higher fundus pigmentation lowers local readings. Two of the four reports print no numeric saturations in their records, so no merged normal range is printed.
Stamped: 2026-11-28.
Evidence published: 2014–2024.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In healthy people, what are the normative values for retinal vessel oxygen saturation measured by fundus oximetry — arterial, venous and arteriovenous-difference saturations — and how do age, quadrant, fundus pigmentation and the oximeter itself move them?
Balance of the evidence
4 reports, N = 476 subjects (61 + 252 + 104 + 59; the age-structure report's unit in its subgroups is eyes — 200 eyes of 104 subjects). At least three instrument families share the shelf (Oxymap T1 named in the Japanese report; Retinal Vessel Analyser (IMEDOS) in the age-structure report; Zilia Ocular in the modifier report), named as the publications give them. Only two reports print numeric saturations; the other two are carried as an age-structure limb and a modifier limb with no figures taken from them. No registry IDs in any record. Cross-points in prose only: retinal vessel calibre and optic-nerve-sheath diameter shelves home different measurands.
MeSH: Adult · Aged · Aging/physiology · Blood Pressure/physiology · Female · Healthy Volunteers · Humans · Intraocular Pressure/physiology · Male · Middle Aged · Oximetry/methods · Oxygen/blood* · Oxygen Consumption · Reference Values · Retinal Artery/physiology* · Retinal Vein/physiology* · Young Adult · Aged, 80 and over · Japan · Oximetry/methods* · Oxygen/metabolism* · Oxygen Consumption/physiology* · Prospective Studies · Reproducibility of Results · Retina/physiology* · Software · Adolescent · Aging/physiology* · Arterioles · Child · Child, Preschool · Retinal Vessels/physiology* · Venules
Cards (5)
Card 1 — US multiethnic cohort (Jani 2014, Retina — 61 normal healthy subjects, both eyes): average venous saturation 55.3 ± 7.1% and arterial 90.4 ± 4.3%; age lowered both (venous β = −0.19, arterial β = −0.11 per year as printed), with no sex or ethnic-group difference in the averages
Study: Normative-value cohort (primary report)
Population: 61 normal healthy subjects in a multiethnic group; retinal oximetry performed in both eyes, with global and quadrant venous (SvO2) and arterial (SaO2) saturation, arteriovenous difference and vessel widths measured.
Summary: US multiethnic cohort (Jani 2014, Retina — 61 normal healthy subjects, both eyes): average venous saturation 55.3 ± 7.1% and arterial 90.4 ± 4.3%; age lowered both (venous β = −0.19, arterial β = −0.11 per year as printed), with no sex or ethnic-group difference in the averages
Result: Average SvO2 was 55.3 ± 7.1% and average SaO2 was 90.4 ± 4.3%. All average measurements were comparable in both eyes, both genders and among ethnic groups; inferonasal quadrant SaO2 was higher in Asians. Age was associated with decreased SvO2 (β = −0.19; P = 0.001) and SaO2 (β = −0.11; P = 0.003). A history of controlled systemic hypertension was associated with an increase in arteriovenous difference (β = 3.99; P = 0.013).
Watch out: One of only two anchors on this shelf that print numeric saturations, and its arterial figure (90.4%) disagrees with the Japanese cohort's 97.0% — the numbers are device-and-software property and must not be merged into a single normal range. The hypertension limb is a subgroup association in 61 subjects.
Source: Jani PD et al. Normative values and predictors of retinal oxygen saturation. Retina (Philadelphia, Pa.). 2014. PMID: 23842102. Registry ID: none in the record. MeSH: Adult · Aged · Aging/physiology · Blood Pressure/physiology · Female · Healthy Volunteers · Humans · Intraocular Pressure/physiology · Male · Middle Aged · Oximetry/methods · Oxygen/blood* · Oxygen Consumption · Reference Values · Retinal Artery/physiology* · Retinal Vein/physiology* · Young Adult
Card 2 — Japanese cohort (Nakano 2016, PLoS ONE — 252 healthy subjects, one eye each): mean arterial saturation 97.0 ± 6.9%, venous 52.8 ± 8.3%, arteriovenous difference 44.2 ± 9.2%; arterial saturation rose with age (+0.67% per 10 years, r = 0.18) while venous did not correlate with age; reproducibility ICCs 0.936–0.979 within a session — product named as the publication gives it: Oxymap T1 with Oxymap Analyzer software
Study: Normative-value cohort (primary report)
Population: 252 eyes of 252 healthy Japanese subjects; oxygen saturation measured in the major retinal vessels with Oxymap T1 and processed with Oxymap Analyzer software. Reproducibility was investigated in 20 eyes of 20 healthy subjects.
Summary: Japanese cohort (Nakano 2016, PLoS ONE — 252 healthy subjects, one eye each): mean arterial saturation 97.0 ± 6.9%, venous 52.8 ± 8.3%, arteriovenous difference 44.2 ± 9.2%; arterial saturation rose with age (+0.67% per 10 years, r = 0.18) while venous did not correlate with age; reproducibility ICCs 0.936–0.979 within a session — product named as the publication gives it: Oxymap T1 with Oxymap Analyzer software
Result: Mean retinal oxygen saturation across 4 quadrants was 97.0 ± 6.9% in arteries and 52.8 ± 8.3% in veins; the mean arteriovenous difference was 44.2 ± 9.2%. Arterial and venous saturations were significantly lower on the temporal side, especially in the temporal-inferior vessels, while the arteriovenous difference was limited across quadrants. Interphotograph, intervisit and interevaluator intraclass correlation coefficients were 0.936–0.979, 0.809–0.837 and 0.732–0.947 respectively. In major retinal arteries, saturation increased with age (r = 0.18, p < 0.01) at 0.67% per 10 years; venous saturation showed no correlation with age.
Watch out: The record is typed Clinical Trial for a normative-database study; counted on substance as a cohort. Its arterial age direction (rising) is opposite to the US cohort's (falling) — both print, unreconciled, because the instruments differ.
Source: Nakano Y et al. Retinal Oximetry in a Healthy Japanese Population. PloS one. 2016. PMID: 27434373. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Female · Humans · Intraocular Pressure/physiology · Japan · Male · Middle Aged · Oximetry/methods* · Oxygen/metabolism* · Oxygen Consumption/physiology* · Prospective Studies · Reference Values · Reproducibility of Results · Retina/physiology* · Retinal Artery/physiology* · Retinal Vein/physiology* · Software
Card 3 — Age-structure limb (Waizel 2018, Acta Ophthalmologica — 104 healthy subjects, 200 eyes; children vs five adult age bands; Retinal Vessel Analyser, IMEDOS): children under 10 had significantly lower arterial saturation and arteriovenous difference than adult subgroups — the record prints no global numeric saturations at abstract level and none is taken from it
Study: Normative-value report (age-structure limb, no abstract-level numeric saturations)
Population: 200 eyes of 104 healthy subjects: 20 eyes of children aged <10 years and 62 eyes of children aged 10–19 years, compared with adult controls in five age groups (20–29, 30–39, 40–49, 50–59 and 60–80 years; eye counts as printed in the record). Oxygen saturation was estimated with the Retinal Vessel Analyser (RVA; IMEDOS Systems UG, Jena, Germany).
Summary: Age-structure limb (Waizel 2018, Acta Ophthalmologica — 104 healthy subjects, 200 eyes; children vs five adult age bands; Retinal Vessel Analyser, IMEDOS): children under 10 had significantly lower arterial saturation and arteriovenous difference than adult subgroups — the record prints no global numeric saturations at abstract level and none is taken from it
Result: Younger children (<10 years) presented a statistically significantly lower arterial saturation (A-SO2) and arteriovenous difference (A-V SO2) than adult subgroups. Arterial diameters were significantly lower in children aged 10+ than in the other groups, while venular diameters showed no significant differences. The authors conclude retinal oxygen metabolism changes throughout life and that normative data for different age groups are mandatory. No global numeric saturation values are printed in the record.
Watch out: Figure-light at abstract level by the record's own content: this card establishes the child–adult difference and the instrument, nothing numeric. Eye counts, not subject counts, are the record's unit for its subgroups (200 eyes, 104 subjects).
Source: Waizel M et al. Normative values of retinal vessel oximetry in healthy children against adults. Acta ophthalmologica. 2018. PMID: 30187646. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Aging/physiology* · Arterioles · Child · Child, Preschool · Female · Healthy Volunteers · Humans · Male · Middle Aged · Oximetry/methods* · Oxygen/blood* · Oxygen Consumption · Prospective Studies · Reference Values · Retinal Vessels/physiology* · Venules · Young Adult
Card 4 — Modifier limb (Bisignano 2024, Clinical Optometry — 59 healthy subjects aged 22–69, racially diverse): local macular saturation at four perimacular locations with a different instrument (Zilia Ocular, named as the publication gives it); higher pigmentation was associated with lower oximetry values (P < 0.01), nasal locations read higher than temporal, age associated after race adjustment; sex, refractive error, blood pressure and prediabetic status showed no association — no numeric saturation values are printed in the record and none is manufactured
Study: Normative-value report (modifier limb, local macular oximetry, no abstract-level numeric saturations)
Population: 59 subjects aged 22–69 years (38.8 ± 14.7), racially diverse with equal gender distribution and no eye disease; oxygen saturation taken with the Zilia Ocular at 4 locations around the macula, 3.1 degrees from the fovea.
Summary: Modifier limb (Bisignano 2024, Clinical Optometry — 59 healthy subjects aged 22–69, racially diverse): local macular saturation at four perimacular locations with a different instrument (Zilia Ocular, named as the publication gives it); higher pigmentation was associated with lower oximetry values (P < 0.01), nasal locations read higher than temporal, age associated after race adjustment; sex, refractive error, blood pressure and prediabetic status showed no association — no numeric saturation values are printed in the record and none is manufactured
Result: Oximetry measures varied significantly by race, with higher pigmentation levels associated with lower oximetry values (p < 0.01). There was no relationship with sex (p = 0.34), refractive error (p = 0.67), blood pressure (systolic p = 0.61, diastolic p = 0.71) or prediabetic status (p = 0.87). Oximetry was associated with age when controlling for race (P < 0.002). Nasal oximetry was higher than temporal measures (P < 0.01). No numeric saturation values are printed in the record.
Watch out: A different measurement (local perimacular saturation) on a different instrument family — its modifiers (pigmentation, location) transfer as cautions, its values do not transfer at all. Modifier limb only.
Source: Bisignano KK et al. Variations in Retinal Oxygen Saturation in a Diverse Healthy Population. Clinical optometry. 2024. PMID: 39045010. Registry ID: none in the record. MeSH: none in the record
Card 5 — Synthesis — retinal oximetry normative values: 4 reports, N = 476 subjects (61 + 252 + 104 + 59), at least three instrument families; only 2 reports print numeric saturations and they disagree on the arterial limb
Study: Synthesis
Population: 4 reports: two numeric normative cohorts (US multiethnic, n = 61, both eyes; Japanese, n = 252, one eye each, Oxymap T1), one age-structure report (n = 104 subjects, 200 eyes, Retinal Vessel Analyser) and one local-macular modifier report (n = 59, Zilia Ocular), the last two printing no numeric saturations in their records.
Summary: Synthesis — retinal oximetry normative values: 4 reports, N = 476 subjects (61 + 252 + 104 + 59), at least three instrument families; only 2 reports print numeric saturations and they disagree on the arterial limb
Result: The venous values agree reasonably (55.3 ± 7.1% vs 52.8 ± 8.3%); the arterial values do not (90.4 ± 4.3% vs 97.0 ± 6.9%), and the two numeric cohorts give opposite age directions for arterial saturation (falling, β = −0.11/year, vs rising, +0.67% per 10 years) while agreeing that quadrant matters (temporal lower) and that children under 10 read lower than adults. Pigmentation lowers readings on the local instrument. The shelf's finding is therefore that no device-free retinal saturation exists: a value is interpretable only against its own instrument's normative data. No report was refused; retinal vessel calibre (widths, not saturations) and optic-nerve-sheath diameter are homed on their own shelves and are different measurands.
Watch out: PHYSIOLOGY — EMERGING, not ESTABLISHED: this is the weakest composition on its slate — half the anchors print no numbers, and the numeric pair contradicts itself on the arterial limb. The stamp claims the replicated parts only (venous range, quadrant and pigmentation modifiers, child–adult difference) and prints the arterial disagreement at headline weight.
Source: Synthesis of the reports carded above; no additional source.