Doppler echocardiography for pulmonary artery systolic pressure vs right heart catheterisation — meta-analysis sensitivity 88% / specificity 56%, near-zero mean bias (+0.07 and +2.4 mm Hg) with 95% limits of agreement of roughly ±20–32 mm Hg: the wide limits of agreement are the headline, not the mean bias
Shelf #8766 · Heart, vessels, blood pressure & clotting
MEASUREMENT ACCURACY — unbiased on average, imprecise in the individual, and the wide limits of agreement are the headline, not the mean bias: the meta-analysis pools sensitivity 88% / specificity 56% / accuracy 63% for the diagnosis of pulmonary hypertension (correlations 0.65–0.97 across 9 studies); the paired cohorts find near-zero mean bias (+0.07 mm Hg in one, +2.4 mm Hg in another) with 95% limits of agreement of roughly ±20–32 mm Hg, and the KARUM database concludes echocardiography accurately represents invasive pressures for population studies but may mislead for individual diagnosis; in ventilated ICU patients measured simultaneously, a Doppler PASP >39 mm Hg predicted pulmonary hypertension with 100% sensitivity and specificity in 40 patients — this pack's smallest and most selected limb, whose estimate partly incorporates an invasive central venous pressure.
Stamped: 2026-11-28.
Evidence published: 2013–2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with suspected or established pulmonary hypertension, how accurately does Doppler echocardiography estimate pulmonary artery systolic pressure (PASP) against right heart catheterisation — in the diagnostic meta-analysis, in the simultaneous-measurement ICU study, and in the large paired-measurement cohorts — and how often is an individual patient's estimate wrong by a clinically important margin?
Balance of the evidence
4 reports = 1 meta-analysis (9 articles, 1984–2009) + 3 paired-measurement primaries (218 patients retrospectively paired within 7 days; 419 subjects in a dual-centre database paired within 3 hours; 40 ventilated ICU patients measured simultaneously). Sensitivity/specificity pairs (a classification question at a pressure threshold) and bias/limits-of-agreement figures (a measurement question) are different quantities and print labelled, never merged. The pulmonary-hypertension definition in every anchor is the older mean PAP ≥25 mm Hg threshold — transport to the revised >20 mm Hg definition is not in this base. Tricuspid signal absence excludes a substantial minority of real-world patients from the technique entirely, a spectrum limit the KARUM anchor's exclusions embody and the face carries.
MeSH: Adult · Blood Pressure Determination/methods*,statistics & numerical data* · Echocardiography, Doppler/methods*,statistics & numerical data* · Humans · Hypertension, Pulmonary/diagnostic imaging*,epidemiology,physiopathology* · Male · Middle Aged · Pulmonary Wedge Pressure* · Reproducibility of Results · Sensitivity and Specificity · Blood Pressure · Cardiac Catheterization/methods · Cross-Sectional Studies · Echocardiography, Doppler/methods · Hypertension, Pulmonary*/diagnostic imaging · Pulmonary Artery/diagnostic imaging · Retrospective Studies · Tricuspid Valve Insufficiency* · Cardiac Catheterization · Echocardiography · Echocardiography, Doppler · Hemodynamics · Predictive Value of Tests · Blood Flow Velocity/physiology · Blood Pressure/physiology · Central Venous Pressure/physiology · Echocardiography, Doppler* · Female · Hypertension, Pulmonary/diagnosis* · Intensive Care Units · Prospective Studies · Pulmonary Artery/diagnostic imaging* · Respiration, Artificial* · Systole/physiology · Tricuspid Valve Insufficiency/diagnostic imaging,physiopathology
Cards (5)
Card 1 — Diagnostic meta-analysis (Taleb 2013, Echocardiography — 9 articles (1984–2009) comparing Doppler echocardiography with right heart catheterisation): correlations between estimated and invasive PASP ranged r = 0.65 to 0.97; pooled sensitivity, specificity and accuracy for the diagnosis of pulmonary hypertension were 88% (95% CI 84–92), 56% (46–66) and 63% (53–73) respectively
Study: Meta-analysis (synthesis, counted as a synthesis)
Population: Patients in 9 articles (May 1984–January 2009, English and non-English) that compared Doppler echocardiography with right heart catheterisation in the assessment of PASP, identified from MEDLINE and EMBASE; the record prints no single pooled patient total.
Summary: Diagnostic meta-analysis (Taleb 2013, Echocardiography — 9 articles (1984–2009) comparing Doppler echocardiography with right heart catheterisation): correlations between estimated and invasive PASP ranged r = 0.65 to 0.97; pooled sensitivity, specificity and accuracy for the diagnosis of pulmonary hypertension were 88% (95% CI 84–92), 56% (46–66) and 63% (53–73) respectively
Result: The correlation between PASP estimated by Doppler echocardiography and by right heart catheterisation ranged from r = 0.65 (P < 0.001) to r = 0.97 (P < 0.001). The pooled sensitivity, specificity and accuracy of Doppler echocardiography for the diagnosis of pulmonary hypertension were 88% (95% CI 84–92%), 56% (95% CI 46–66%) and 63% (95% CI 53–73%) respectively. The authors conclude Doppler echocardiography is useful to screen for pulmonary hypertension and can reliably determine whether PASP is normal, mildly elevated or markedly elevated, but abnormal results need confirmation by catheterisation.
Watch out: A classification pair (88%/56%) from studies spanning 1984–2009 — older machines and the older ≥25 mm Hg definition throughout. Specificity of 56% means nearly half of non-PH patients screen positive; this pool supports screening, not diagnosis, exactly as its authors state. Correlation (up to 0.97) is not agreement, as the paired cohorts below show.
Source: Taleb M et al. The diagnostic accuracy of Doppler echocardiography in assessment of pulmonary artery systolic pressure: a meta-analysis. Echocardiography (Mount Kisco, N.Y.). 2013. PMID: 23227919. Registry ID: none in the record. MeSH: Adult · Blood Pressure Determination/methods*,statistics & numerical data* · Echocardiography, Doppler/methods*,statistics & numerical data* · Humans · Hypertension, Pulmonary/diagnostic imaging*,epidemiology,physiopathology* · Male · Middle Aged · Pulmonary Wedge Pressure* · Reproducibility of Results · Sensitivity and Specificity
Card 2 — Retrospective paired cohort (Lv 2022, BMC Medical Imaging — 218 patients with highly suspected pulmonary hypertension undergoing catheterisation and echocardiography within 7 days): the tricuspid-regurgitation-derived systolic estimate had the highest correlation (r = 0.781), an AUC of 0.98 against a catheter mean PAP ≥25 mm Hg criterion, and the lowest bias (mean +0.07 mm Hg; 95% limits of agreement −32.08 to +32.22 mm Hg); tricuspid signal quality, invasive pressure level and pulmonary artery wedge pressure significantly affected accuracy
Study: Retrospective cross-sectional study (paired-measurement primary)
Population: 218 patients with highly suspected pulmonary hypertension who underwent right heart catheterisation and echocardiography within 7 days; five tricuspid-regurgitation-related methods were compared (TR Vmax, TR pressure gradient, TR mean pressure gradient, estimated mean PAP, and estimated systolic PAP).
Summary: Retrospective paired cohort (Lv 2022, BMC Medical Imaging — 218 patients with highly suspected pulmonary hypertension undergoing catheterisation and echocardiography within 7 days): the tricuspid-regurgitation-derived systolic estimate had the highest correlation (r = 0.781), an AUC of 0.98 against a catheter mean PAP ≥25 mm Hg criterion, and the lowest bias (mean +0.07 mm Hg; 95% limits of agreement −32.08 to +32.22 mm Hg); tricuspid signal quality, invasive pressure level and pulmonary artery wedge pressure significantly affected accuracy
Result: The systolic estimate (sPAPECHO) had the highest correlation coefficient (r = 0.781, P < 0.001), the best diagnostic efficiency (AUC = 0.98, against a catheter mean PAP ≥25 mm Hg criterion) and the lowest bias (mean bias = 0.07 mm Hg; 95% limits of agreement −32.08 to +32.22 mm Hg) compared with the other TR-related methods. Ordinal regression showed TR signal quality, invasive systolic pressure level and pulmonary artery wedge pressure affected accuracy (P < 0.05): relative to good signal quality, the odds ratios for medium and poor signal quality were 0.26 (95% CI 0.14–0.48) and 0.23 (0.07–0.73) — relative measures; compared with a high invasive pressure level, low and medium levels had odds ratios of 21.56 (9.57–48.55) and 5.13 (2.55–10.32) — relative; the wedge-pressure odds ratio was 0.94 (0.89–0.99) — relative. TR severity and right ventricular systolic function had no significant effect.
Watch out: The pack's clearest single picture: a mean bias of +0.07 mm Hg beside limits of agreement spanning 64 mm Hg. An individual estimate from this technique can be 30 mm Hg out in either direction — the limits, not the bias, are the headline. Retrospective pairing within 7 days (not simultaneous) adds physiological drift to the disagreement. Odds ratios print labelled relative, as published.
Source: Lv GJ et al. The accuracy and influencing factors of Doppler echocardiography in estimating pulmonary artery systolic pressure: comparison with right heart catheterization: a retrospective cross-sectional study. BMC medical imaging. 2022. PMID: 35578318. Registry ID: none in the record. MeSH: Blood Pressure · Cardiac Catheterization/methods · Cross-Sectional Studies · Echocardiography, Doppler/methods · Humans · Hypertension, Pulmonary*/diagnostic imaging · Pulmonary Artery/diagnostic imaging · Retrospective Studies · Tricuspid Valve Insufficiency*
Card 3 — KARUM haemodynamic database (Venkateshvaran 2021, International Journal of Cardiovascular Imaging — 419 subjects undergoing echocardiography within 3 hours of catheterisation): echocardiographic systolic and mean pulmonary pressures showed minimal bias (+2.4 and +1.9 mm Hg) but wide limits of agreement (−20 to +25 and −14 to +18 mm Hg) and frequent misclassification; guideline-based right-atrial-pressure estimation was falsely elevated in 32% of cases, and substituting a fixed median right atrial pressure reduced bias to +1.4 mm Hg
Study: Database study (paired-measurement primary, dual-centre)
Population: 419 consecutive pulmonary-hypertension referrals at two centres who underwent comprehensive echocardiography within 3 hours of clinically indicated right heart catheterisation; subjects with absent or severe, free-flowing tricuspid regurgitation were excluded.
Summary: KARUM haemodynamic database (Venkateshvaran 2021, International Journal of Cardiovascular Imaging — 419 subjects undergoing echocardiography within 3 hours of catheterisation): echocardiographic systolic and mean pulmonary pressures showed minimal bias (+2.4 and +1.9 mm Hg) but wide limits of agreement (−20 to +25 and −14 to +18 mm Hg) and frequent misclassification; guideline-based right-atrial-pressure estimation was falsely elevated in 32% of cases, and substituting a fixed median right atrial pressure reduced bias to +1.4 mm Hg
Result: Echocardiographic PA systolic and mean pressures demonstrated minimal bias against invasive measurements (+2.4 and +1.9 mm Hg respectively) but wide limits of agreement (−20 to +25 and −14 to +18 mm Hg respectively) and frequently misclassified subjects; accuracy for individual diagnosis was defined as an estimate within ±10 mm Hg of the invasive measurement. Recommendation-based right atrial pressure demonstrated poor precision and was falsely elevated in 32% of individual cases. Applying a fixed, median right atrial pressure gave lower bias for PA systolic (+1.4 mm Hg; 95% limits of agreement +25 to −22 mm Hg) and PA mean pressures (+1.4 mm Hg; +19 to −16 mm Hg). The authors conclude echocardiography accurately represents invasive PA pressures for population studies but may be misleading for individual diagnosis owing to modest precision and frequent misclassification.
Watch out: The exclusions define the spectrum: patients with absent or severe free-flowing tricuspid regurgitation — a substantial real-world minority — are not in this cohort at all, so even these wide limits flatter unselected practice. The right-atrial-pressure component, not the Doppler gradient, is identified as a major error source, and the fixed-RAP fix narrows bias without narrowing the limits materially.
Source: Venkateshvaran A et al. Accuracy of echocardiographic estimates of pulmonary artery pressures in pulmonary hypertension: insights from the KARUM hemodynamic database. The international journal of cardiovascular imaging. 2021. PMID: 34146206. Registry ID: none in the record. MeSH: Cardiac Catheterization · Echocardiography · Echocardiography, Doppler · Hemodynamics · Humans · Hypertension, Pulmonary*/diagnostic imaging · Predictive Value of Tests · Pulmonary Artery/diagnostic imaging
Card 4 — Simultaneous-measurement ICU study (Mercado 2019, Critical Care Medicine — prospective, 40 mechanically ventilated ICU patients): the tricuspid-regurgitation gradient plus invasive central venous pressure correlated best with invasive systolic PAP (r = 0.87, bias −3 mm Hg, precision 9 mm Hg), and a Doppler systolic PAP >39 mm Hg predicted pulmonary hypertension (mean PAP ≥25 mm Hg) with 100% sensitivity and specificity in this sample — this pack's smallest limb (n = 40), and its pressure estimate partly incorporates an invasive measurement (central venous pressure), printed as such
Study: Prospective observational study (simultaneous paired-measurement primary; smallest limb, n = 40)
Population: 40 ICU patients receiving mechanical ventilation at Amiens ICU, France, in whom Doppler echocardiography variables and invasive central venous, systolic, diastolic and mean pulmonary artery pressures were recorded simultaneously.
Summary: Simultaneous-measurement ICU study (Mercado 2019, Critical Care Medicine — prospective, 40 mechanically ventilated ICU patients): the tricuspid-regurgitation gradient plus invasive central venous pressure correlated best with invasive systolic PAP (r = 0.87, bias −3 mm Hg, precision 9 mm Hg), and a Doppler systolic PAP >39 mm Hg predicted pulmonary hypertension (mean PAP ≥25 mm Hg) with 100% sensitivity and specificity in this sample — this pack's smallest limb (n = 40), and its pressure estimate partly incorporates an invasive measurement (central venous pressure), printed as such
Result: Systolic PAP assessed from the tricuspid regurgitation maximal pressure gradient added to the central venous pressure showed the best correlation with invasive systolic PAP (r = 0.87), with a small bias (−3 mm Hg) and a precision of 9 mm Hg. A Doppler systolic PAP greater than 39 mm Hg predicted pulmonary hypertension (mean PAP ≥25 mm Hg) with 100% sensitivity and specificity. Tricuspid regurgitation maximal velocity greater than 2.82 m/s and pressure gradient greater than 32 mm Hg also predicted pulmonary hypertension. Pulmonary regurgitation was recorded in only 10 patients (25%), and no correlation was found between pulmonary regurgitation velocities and mean or diastolic PAP. Pulmonary acceleration time less than 57 ms and isovolumic relaxation time less than 40 ms each predicted pulmonary hypertension 100% of the time with a 100% negative predictive value.
Watch out: 100% sensitivity and specificity in 40 selected, simultaneously measured, ventilated patients is a small-sample ceiling, not a transportable property — and the best-performing estimate adds an invasive central venous pressure to the Doppler gradient, so it is not a purely non-invasive measurement. Pulmonary regurgitation signals were obtainable in only a quarter of patients.
Source: Mercado P et al. Reassessment of the Accuracy of Cardiac Doppler Pulmonary Artery Pressure Measurements in Ventilated ICU Patients: A Simultaneous Doppler-Catheterization Study. Critical care medicine. 2019. PMID: 30379666. Registry ID: none in the record. MeSH: Blood Flow Velocity/physiology · Blood Pressure/physiology · Central Venous Pressure/physiology · Echocardiography, Doppler* · Female · Humans · Hypertension, Pulmonary/diagnosis* · Intensive Care Units · Male · Middle Aged · Prospective Studies · Pulmonary Artery/diagnostic imaging* · Respiration, Artificial* · Systole/physiology · Tricuspid Valve Insufficiency/diagnostic imaging,physiopathology
Card 5 — Synthesis — Doppler echo PASP vs catheterisation: 1 meta-analysis + 3 paired-measurement primaries (218 + 419 + 40 patients), one coherent bias/imprecision story
Study: Synthesis
Population: Meta-analysis: 9 articles (1984–2009). Primaries: 677 paired-measurement patients in total (218 retrospective within 7 days; 419 database within 3 hours; 40 simultaneous ICU). Classification figures and agreement figures are kept apart throughout.
Summary: Synthesis — Doppler echo PASP vs catheterisation: 1 meta-analysis + 3 paired-measurement primaries (218 + 419 + 40 patients), one coherent bias/imprecision story
Result: Every layer agrees on the shape: the average error is near zero (+0.07 and +2.4 mm Hg bias) and the individual error is large (95% limits of agreement roughly ±20–32 mm Hg), with pooled classification performance of 88% sensitive / 56% specific for pulmonary hypertension under the older ≥25 mm Hg definition. MEASUREMENT ACCURACY — unbiased on average, imprecise in the individual — is the stamp, with the wide limits of agreement as its headline rather than the mean bias. The ICU study's perfect classification (n = 40, partly invasive estimate) prints at small-limb weight and does not soften the stamp.
Watch out: Transport to the revised pulmonary-hypertension definition (mean PAP >20 mm Hg) is not evidenced anywhere in this base and no figure may be re-based to it. Patients without a usable tricuspid signal are excluded from the technique itself — a spectrum limit no anchor quantifies for unselected practice.
Source: Synthesis of the reports carded above; no additional source.