Non-invasive haemoglobin (SpHb) by pulse CO-oximetry vs laboratory haemoglobin — 1 meta-analysis + 3 paired-measurement primaries: pooled bias 0.13 g/dL, limits of agreement spanning up to −2.25 to +3.96 g/dL

Shelf #8816 · Emergency & first hours

MEASUREMENT ACCURACY — near-zero average bias, transfusion-relevant individual disagreement: the 36-study meta-analysis (1,888 patients) pools a mean difference of 0.13 g/dL between non-invasive and invasive methods (its printed 95% CI of 0.10–0.36 alongside p > 0.05 is internally inconsistent — flagged as published, never repaired), but the primaries' 95% limits of agreement are wide — bias 0.01 g/dL with limits −1.33 to +1.34 g/dL across volume/perfusion states (210 paired readings), bias +0.86 ± 1.58 g/dL with limits −2.25 to +3.96 g/dL in liver transplantation (282 pairs, 53 patients), and bias 1.0 ± 1.4 g/dL (laboratory higher) with limits −1.7 to +3.8 g/dL in hip arthroplasty (108 pairs, 43 patients), where using SpHb instead of laboratory haemoglobin would have changed the transfusion decision in 1.9% of paired measurements.

Stamped: 2026-11-28.

Evidence published: 2017–2025.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In perioperative and surgical patients, how closely does non-invasive continuous haemoglobin by pulse CO-oximetry (SpHb) agree with laboratory haemoglobin — in the perioperative meta-analysis, in liver transplantation, during major orthopaedic bleeding and resuscitation, and in hip arthroplasty transfusion decisions?

Balance of the evidence

1 meta-analysis + 3 prospective paired-measurement primaries. Bias sign conventions differ between anchors (SpHb-minus-laboratory vs laboratory-minus-SpHb) and print as each record defines them — the apparent bias-direction disagreement is partly definitional and partly real, and no pooled bias is printed beyond the meta-analysis's own. The meta-analysis's pooled interval is internally inconsistent (flagged on its card, printed as published, never repaired). Correlation and trending figures print as correlation, never as agreement. Haemoglobin values print in g/dL exactly as the records give them. Refused at build-source stage by the slate desk and named here so it is never counted: PMID 37868333 is a protocol for a pregnancy SpHb agreement substudy reporting no results. Cross-points in prose only: transcutaneous bilirubinometry (#8438) and pulse-oximetry accuracy (#7718) are different analytes/devices. The tested device technology is named exactly as the publications name it (Masimo pulse CO-oximetry; the Radical-7 device where the record names it) because agreement is device-specific — the standing product-naming exception.

MeSH: Humans · Hemoglobins*/analysis · Perioperative Care*/methods · Oximetry*/methods · Anemia*/diagnosis,blood · Reproducibility of Results · Blood Transfusion · Sensitivity and Specificity · Hemoglobinometry/methods · Blood Gas Monitoring, Transcutaneous* · Data Accuracy · Hemoglobins/analysis* · Liver Transplantation* · Middle Aged · Monitoring, Intraoperative/methods* · Prospective Studies · Adult · Blood Volume · Female · Fluid Therapy · Hemoglobinometry/methods*,statistics & numerical data,trends · Hemorrhage/blood,therapy · Male · Monitoring, Intraoperative/methods*,statistics & numerical data,trends · Monitoring, Physiologic/methods*,statistics & numerical data,trends · Oximetry/methods,statistics & numerical data · Resuscitation · Aged · Arthroplasty, Replacement, Hip/adverse effects*,methods · Blood Transfusion/methods* · Hemoglobinometry/methods*

Cards (5)

Card 1 — Moreira meta-analysis 2025 (Journal of Clinical Monitoring and Computing — systematic review with meta-analysis, 36 studies / 1,888 patients comparing non-invasive haemoglobin with invasive reference methods in perioperative management): pooled mean difference 0.13 g/dL — with its printed 95% CI of 0.10–0.36 alongside p > 0.05, an internally inconsistent interval flagged as published and never silently repaired; I² = 0%

Study: Meta-analysis (synthesis, counted as a synthesis; defective printed interval flagged on the face)

Population: 1,888 perioperative patients in 36 studies comparing non-invasive haemoglobin measurement (pulse CO-oximetry) with invasive reference methods; the review was PROSPERO-registered (no registration number printed in the record) and QUADAS-2 assessed.

Summary: Moreira meta-analysis 2025 (Journal of Clinical Monitoring and Computing — systematic review with meta-analysis, 36 studies / 1,888 patients comparing non-invasive haemoglobin with invasive reference methods in perioperative management): pooled mean difference 0.13 g/dL — with its printed 95% CI of 0.10–0.36 alongside p > 0.05, an internally inconsistent interval flagged as published and never silently repaired; I² = 0%

Result: The meta-analysis pooled a mean difference between the non-invasive and invasive methods of 0.13 g/dL, printed in the record with a 95% confidence interval of 0.10–0.36 and a P-value > 0.05 — an interval excluding zero cannot accompany that p-value, and the point estimate sits at the interval's edge; the figures are printed flagged, exactly as published, and are never silently repaired on this shelf. Homogeneity was very good (I² = 0%). Sensitivity analyses by sampling site, Masimo finger-sensor revision, geography, bias risk, population and study type showed no statistically significant difference between methods. Trending was acceptable in a majority of studies.

Watch out: The pooled mean difference is an average bias across studies — it says nothing about how far an individual SpHb reading can sit from that patient's laboratory value, which is what the primary cards' limits of agreement measure. The defective interval means even the pooled figure's precision cannot be stated from this record.

Source: Moreira L et al. Non-invasive vs biological blood determination of haemoglobinemia for perioperative management: a systematic review with meta-analysis. Journal of clinical monitoring and computing. 2025. PMID: 40739388. Registry ID: none in the record. MeSH: Humans · Hemoglobins*/analysis · Perioperative Care*/methods · Oximetry*/methods · Anemia*/diagnosis,blood · Reproducibility of Results · Blood Transfusion · Sensitivity and Specificity · Hemoglobinometry/methods

Card 2 — Erdogan Kayhan liver-transplant study 2017 (Minerva Anestesiologica — prospective paired-measurement study in liver transplantation, 282 data pairs from 53 patients): SpHb–laboratory correlation r = 0.73, bias and precision 0.86 ± 1.58 g/dL with limits of agreement −2.25 to +3.96 g/dL; 4-quadrant sensitivity 79% — the device technology is named in the record as Masimo Corporation (Irvine, CA), printed as published because agreement is device-specific

Study: Prospective paired-measurement study (primary report)

Population: 53 patients undergoing liver transplantation; 282 paired SpHb and laboratory haemoglobin measurements obtained twice in each of the three surgical phases (pre-anhepatic, anhepatic, neohepatic). SpHb values ranged 6.9–17.7 g/dL and laboratory values 5.4–17.1 g/dL.

Summary: Erdogan Kayhan liver-transplant study 2017 (Minerva Anestesiologica — prospective paired-measurement study in liver transplantation, 282 data pairs from 53 patients): SpHb–laboratory correlation r = 0.73, bias and precision 0.86 ± 1.58 g/dL with limits of agreement −2.25 to +3.96 g/dL; 4-quadrant sensitivity 79% — the device technology is named in the record as Masimo Corporation (Irvine, CA), printed as published because agreement is device-specific

Result: The correlation coefficient between SpHb and laboratory haemoglobin was 0.73 (P<0.001), and for changes in SpHb versus changes in laboratory haemoglobin 0.76 (P<0.001). The sensitivity determined using a 4-quadrant plot was 79%. Bias and precision of SpHb to laboratory haemoglobin were 0.86 ± 1.58 g/dL, with limits of agreement −2.25 to +3.96 g/dL. Correlation remained stable across the surgical phases.

Watch out: Limits of agreement spanning more than 6 g/dL in a population where transfusion thresholds sit inside a 2 g/dL window: an individual SpHb reading in liver transplantation can be far from the laboratory value in either direction despite the small average bias. Correlation (r = 0.73) is printed as correlation, never as agreement.

Source: Erdogan Kayhan G et al. Accuracy of non-invasive hemoglobin monitoring by pulse CO-oximeter during liver transplantation. Minerva anestesiologica. 2017. PMID: 28106356. Registry ID: none in the record. MeSH: Blood Gas Monitoring, Transcutaneous* · Data Accuracy · Hemoglobins/analysis* · Humans · Liver Transplantation* · Middle Aged · Monitoring, Intraoperative/methods* · Prospective Studies

Card 3 — Adel orthopaedic study 2018 (Journal of Clinical Monitoring and Computing — prospective paired-measurement study, 70 patients undergoing major orthopaedic procedures with anticipated major blood loss, 210 time-matched readings across baseline, post-bleeding and post-resuscitation): correlation r = 0.938; mean bias 0.01 g/dL with limits of agreement −1.33 to +1.34 g/dL overall, −0.08 (−1.27 to +1.11) in fluid non-responsive and 0.09 (−1.36 to +1.54) in fluid-responsive samples — the Radical-7 device is named in the record, as published

Study: Prospective paired-measurement study (primary report)

Population: 70 patients scheduled for major orthopedic procedures with anticipated major blood loss; 210 time-matched SpHb (Radical-7 device) and laboratory haemoglobin readings at baseline, after major bleeding and after resuscitation, split into fluid-responsive (104) and fluid non-responsive (106) samples and by perfusion index.

Summary: Adel orthopaedic study 2018 (Journal of Clinical Monitoring and Computing — prospective paired-measurement study, 70 patients undergoing major orthopaedic procedures with anticipated major blood loss, 210 time-matched readings across baseline, post-bleeding and post-resuscitation): correlation r = 0.938; mean bias 0.01 g/dL with limits of agreement −1.33 to +1.34 g/dL overall, −0.08 (−1.27 to +1.11) in fluid non-responsive and 0.09 (−1.36 to +1.54) in fluid-responsive samples — the Radical-7 device is named in the record, as published

Result: Correlation between laboratory haemoglobin and SpHb was r = 0.938. Mean bias (limits of agreement) was 0.01 (−1.33 to +1.34) g/dL across all samples, −0.08 (−1.27 to +1.11) g/dL in fluid non-responsive samples, 0.09 (−1.36 to +1.54) g/dL in fluid-responsive samples, 0.01 (−1.34 to +1.31) g/dL in high-perfusion-index and 0.04 (−1.31 to +1.39) g/dL in low-perfusion-index samples. Polar-plot analysis showed good trending ability. The authors note that despite the favourable bias, the relatively wide limits might limit accuracy.

Watch out: The tightest limits on this shelf still span ±1.3 g/dL around the laboratory value — acceptable for trending, wide for a threshold decision. Volume and perfusion state did not rescue or wreck agreement in this cohort, which is a finding about this device in this setting, not about SpHb generally.

Source: Adel A et al. Accuracy and trending of non-invasive hemoglobin measurement during different volume and perfusion statuses. Journal of clinical monitoring and computing. 2018. PMID: 29335914. Registry ID: none in the record. MeSH: Adult · Blood Transfusion · Blood Volume · Female · Fluid Therapy · Hemoglobinometry/methods*,statistics & numerical data,trends · Hemorrhage/blood,therapy · Humans · Male · Middle Aged · Monitoring, Intraoperative/methods*,statistics & numerical data,trends · Monitoring, Physiologic/methods*,statistics & numerical data,trends · Oximetry/methods,statistics & numerical data · Prospective Studies · Resuscitation

Card 4 — Campos hip-arthroplasty study 2020 (Transfusion and Apheresis Science — prospective paired-measurement study in elective hip replacement, 108 paired measurements from 43 patients): Pearson r = 0.7; Bland-Altman bias 1.0 ± 1.4 g/dL with laboratory haemoglobin recurrently higher than SpHb (the opposite sign convention from the liver-transplant anchor — printed as each record defines it) and limits of agreement −1.7 to +3.8 g/dL; at a transfusion threshold of 8 g/dL the transfusion decision would have differed in 2 of 108 paired measurements (1.9%)

Study: Prospective paired-measurement study (primary report)

Population: 43 patients undergoing elective hip replacement (randomly selected); 108 paired measurements, with SpHb acquired using the Radical-7 device whenever blood was collected for laboratory haemoglobin.

Summary: Campos hip-arthroplasty study 2020 (Transfusion and Apheresis Science — prospective paired-measurement study in elective hip replacement, 108 paired measurements from 43 patients): Pearson r = 0.7; Bland-Altman bias 1.0 ± 1.4 g/dL with laboratory haemoglobin recurrently higher than SpHb (the opposite sign convention from the liver-transplant anchor — printed as each record defines it) and limits of agreement −1.7 to +3.8 g/dL; at a transfusion threshold of 8 g/dL the transfusion decision would have differed in 2 of 108 paired measurements (1.9%)

Result: The Pearson correlation between SpHb and laboratory haemoglobin was 0.7 (p < 0.001). Bland-Altman bias was 1 ± 1.4 g/dL, meaning laboratory haemoglobin was recurrently higher than SpHb, with limits of agreement −1.7 to +3.8 g/dL. Considering a red-cell transfusion threshold of 8 g/dL, in two situations (1.9% of paired measurements) the transfusion decision would have differed by measurement method. Trending ability was limited perioperatively, with a significant difference between preoperative and postoperative laboratory-minus-SpHb differences. The authors conclude the device can be a reference but cannot replace venous puncture as the gold standard.

Watch out: The smallest primary (43 patients) and the only one framed around the decision itself: 1.9% of decisions flipping sounds small, but the limits of agreement (−1.7 to +3.8 g/dL) are what a clinician faces on the next reading. Its bias sign convention (laboratory minus SpHb) is the reverse of the liver-transplant anchor's — the two biases must never be subtracted or averaged.

Source: Campos C et al. Clinical implications of using non-invasive haemoglobin monitoring for red blood cell transfusion decision in hip arthroplasty. Transfusion and apheresis science : official journal of the World Apheresis Association : official journal of the European Society for Haemapheresis. 2020. PMID: 32381421. Registry ID: none in the record. MeSH: Aged · Arthroplasty, Replacement, Hip/adverse effects*,methods · Blood Transfusion/methods* · Female · Hemoglobinometry/methods* · Hemoglobins/analysis · Humans · Male · Prospective Studies

Card 5 — Synthesis — 1 meta-analysis (36 studies / 1,888 patients) + 3 prospective paired-measurement primaries (53 + 70 + 43 patients): near-zero average bias coexisting with transfusion-relevant individual disagreement — the claim rests on the limits of agreement, not on any bias point estimate

Study: Synthesis

Population: Composition: Moreira 2025 meta-analysis (36 studies / 1,888 patients) + Erdogan Kayhan 2017 liver transplantation (282 pairs, 53 patients) + Adel 2018 major orthopaedic bleeding/resuscitation (210 readings, 70 patients) + Campos 2020 hip arthroplasty (108 pairs, 43 patients). Refused and counted nowhere: PMID 37868333, a protocol with no results.

Summary: Synthesis — 1 meta-analysis (36 studies / 1,888 patients) + 3 prospective paired-measurement primaries (53 + 70 + 43 patients): near-zero average bias coexisting with transfusion-relevant individual disagreement — the claim rests on the limits of agreement, not on any bias point estimate

Result: Average bias is small everywhere it is pooled or measured: 0.13 g/dL in the meta-analysis (its printed interval defective, flagged), 0.01 g/dL in the orthopaedic cohort, +0.86 g/dL (SpHb above laboratory) in liver transplantation and 1.0 g/dL in the opposite direction (laboratory above SpHb) in hip arthroplasty — the last partly a sign-convention difference in the record. The limits of agreement tell the clinical story: −1.33 to +1.34 g/dL at best, −2.25 to +3.96 g/dL at worst, against transfusion thresholds that turn on 1 g/dL. In the one decision-framed primary, 1.9% of transfusion decisions would have changed. The stamp is MEASUREMENT ACCURACY at trend-monitor weight: SpHb tracks haemoglobin; it does not replace the laboratory value at a threshold.

Watch out: All primaries test one manufacturer's pulse CO-oximetry technology (Masimo; Radical-7 where named) — agreement figures are device- and sensor-generation-specific and do not transfer to other non-invasive haemoglobin devices. Surgical spectra only; no emergency-department, ward or pregnancy results paper is in this base.

Source: Synthesis of the reports carded above; no additional source.