Point-of-care creatinine (and creatinine-based eGFR) vs central-laboratory creatinine — 5 positive-agreement primaries across ED, pre-contrast, HIV-clinic and stroke settings + 1 negative ICU limb (6 reports): 94.5% of 14,567 ED POC values within 0.3 mg/dL of the laboratory, fingerstick ICC 0.996 — but pre-contrast eGFR ran +11.0 mL/min/1.73 m² above the laboratory with limits of agreement −22.4 to +44.4, StatSensor eGFR bias was −4.36 with limits −19.68 to +28.40, and ICU creatinine was among the analytes that did not correlate well

Shelf #8856 · Kidney, fluids & electrolytes

MEASUREMENT ACCURACY — classification-level agreement is strong and eGFR-level agreement is not: 94.5% of ED POC creatinine values fell within 0.3 mg/dL of the central laboratory (14,567 patients with paired assays); fingerstick i-STAT creatinine showed ICC 0.996 in 246 ED patients and POCT creatinine ICC 0.980 in suspected stroke; but the pre-contrast meter's eGFR ran a mean 11.0 mL/min/1.73 m² above the laboratory value with limits of agreement −22.4 to +44.4, the StatSensor's eGFR bias was −4.36 with limits −19.68 to +28.40 despite 92.9% sensitivity / 94.7% specificity for eGFR <60, and in the ICU multi-device evaluation creatinine was among the analytes that did not correlate well. The two eGFR biases point in opposite directions (+11.0 pre-contrast vs −4.36 in HIV screening) — device- and population-bound, never pooled; a build that leads with the ICCs (0.98–0.996) without the eGFR limits misrepresents this base.

Stamped: 2026-11-28.

Evidence published: 2011–2025.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In emergency, radiology, ICU and clinic patients, how closely does point-of-care whole-blood creatinine (and creatinine-based eGFR) agree with central-laboratory serum creatinine — in a 14,567-patient ED paired-assay cohort, before contrast imaging, by fingerstick in the ED, by a named handheld device for kidney-dysfunction screening in HIV care, and in suspected stroke?

Balance of the evidence

The composition is 5 positive-agreement primaries across four settings + 1 negative ICU limb. Counting units differ (patients, paired assays, samples) and print per anchor. Two cards are shared reports, printed as shared: Dashevsky 2017 is carded for its creatinine limb only (its sodium, potassium, BUN and haematocrit limbs are counted nowhere — and its sodium agreement criterion is printed in its record as 'within 5 mg/dL', a unit sodium is not reported in; that limb is not used here and the creatinine criterion of 0.3 mg/dL is unaffected); You 2013 is carded for its creatinine limb only (its BUN limb counted nowhere); Kok 2015 is carded here for its creatinine limb — the row's negative limb — while its INR limb is cross-pointed on the POC INR shelf (pack #8857) in prose only and counted nowhere, and its blood-gas limbs are counted nowhere. Agreement is device-specific, so tested devices are named exactly as the publications name them (i-STAT, StatSensor) under the standing product-naming exception. Creatinine prints in mg/dL and eGFR in mL/min/1.73 m² exactly as the records give them. Cross-points in prose only: the cystatin-C eGFR prediction pack #6070 and the homed contrast-nephropathy ground ask different questions. No record carries a registry ID.

MeSH: Adult · Biological Assay · Biomarkers/blood · Blood Chemical Analysis/instrumentation,standards* · Blood Urea Nitrogen · Cost-Benefit Analysis · Creatinine/blood · Emergency Medical Services/methods,standards* · Female · Humans · Laboratories, Hospital/standards* · Male · Middle Aged · Point-of-Care Systems*/standards · Potassium/blood · Quality Assurance, Health Care · Reproducibility of Results · Retrospective Studies · Sodium/blood · United States · Aged · Aged, 80 and over · Contrast Media* · Creatinine/blood* · Glomerular Filtration Rate* · Image Enhancement · Kidney Diseases/blood*,diagnosis* · Magnetic Resonance Imaging · Point-of-Care Systems* · Prospective Studies · Tomography, X-Ray Computed · Creatinine*/blood · Emergency Service, Hospital* · Fingers/blood supply · HIV Infections*/complications,blood · Cross-Sectional Studies · Tanzania/epidemiology · ROC Curve · Sensitivity and Specificity · Point-of-Care Testing · Acute Kidney Injury/chemically induced,prevention & control* · Benchmarking · Clinical Laboratory Techniques · Contrast Media/adverse effects* · Creatinine/urine* · Emergency Medical Services* · Evidence-Based Medicine · Kidney Function Tests · Reference Standards · Stroke/urine · Time Factors · Australia · Clinical Laboratory Techniques* · Critical Care · Ebolavirus/isolation & purification* · Hemorrhagic Fever, Ebola/therapy* · Partial Thromboplastin Time · Point-of-Care Testing* · Tertiary Care Centers

Cards (7)

Card 1 — Dashevsky ED paired-assay cohort 2017 (Western Journal of Emergency Medicine — retrospective paired-assay cohort, 14,567 ED patients with blood sent to both POC and central laboratory): using clinical agreement criteria, creatinine agreement was 94.5% (within 0.3 mg/dL) — shared report, printed as shared: only the creatinine limb is counted here; the sodium, potassium, BUN and haematocrit limbs are counted nowhere

Study: Retrospective paired-assay cohort (shared report — creatinine limb only)

Population: 14,567 patients seen in an urban Level I emergency department (of 163,661 seen between March 2013 and October 2014) who had blood sent to both point-of-care and central laboratory analysis, with agreement measured by Bland-Altman plots against both statistical and clinical-judgment criteria.

Summary: Dashevsky ED paired-assay cohort 2017 (Western Journal of Emergency Medicine — retrospective paired-assay cohort, 14,567 ED patients with blood sent to both POC and central laboratory): using clinical agreement criteria, creatinine agreement was 94.5% (within 0.3 mg/dL) — shared report, printed as shared: only the creatinine limb is counted here; the sodium, potassium, BUN and haematocrit limbs are counted nowhere

Result: Using clinical criteria, the levels of agreement for sodium were 98.6% (within 5 mg/dL, as the record prints the criterion), for potassium 90.7% (0.5 mmol/L), for BUN 89.0% (within 5 mg/dL), for creatinine 94.5% (within 0.3 mg/dL), and for hematocrit 96.5% (within 5 g/dL, as printed). The authors conclude agreement between POC and central laboratory values is excellent and duplicate central-laboratory assays are unnecessary.

Watch out: A within-tolerance proportion (94.5% within 0.3 mg/dL) is a classification statement, not a bias or limits-of-agreement statement — at a creatinine near a contrast or dosing threshold, the 5.5% outside tolerance are the clinically operative patients. The record prints its sodium criterion in mg/dL, a unit sodium is not reported in; that limb is unused here. Retrospective, single-system, and the POC panel is a whole-blood multi-analyte device, not a creatinine-only meter.

Source: Dashevsky M et al. Agreement Between Serum Assays Performed in ED Point-of-Care and Hospital Central Laboratories. The western journal of emergency medicine. 2017. PMID: 28435491. Registry ID: none in the record. MeSH: Adult · Biological Assay · Biomarkers/blood · Blood Chemical Analysis/instrumentation,standards* · Blood Urea Nitrogen · Cost-Benefit Analysis · Creatinine/blood · Emergency Medical Services/methods,standards* · Female · Humans · Laboratories, Hospital/standards* · Male · Middle Aged · Point-of-Care Systems*/standards · Potassium/blood · Quality Assurance, Health Care · Reproducibility of Results · Retrospective Studies · Sodium/blood · United States

Card 2 — Morita pre-contrast study 2011 (Japanese Journal of Radiology — prospective study, 113 patients scheduled for contrast-enhanced CT/MRI without recent creatinine data): the POC whole-blood creatinine meter's eGFR averaged 81.2 ± 24.6 vs the laboratory's 70.2 ± 19.7 mL/min/1.73 m² (p < 0.0001), correlation r = 0.74, Bland-Altman mean difference +11.0 mL/min/1.73 m² with limits of agreement −22.4 to +44.4 — correlation printed as correlation, never as agreement; the meter systematically overestimates eGFR, the direction that under-calls renal impairment before contrast

Study: Prospective agreement study (primary report)

Population: 113 patients scheduled to undergo contrast-enhanced CT or MRI examinations who did not have creatinine data from the prescribed intervals before the examination day, whose renal function was assessed with a rapid point-of-care whole-blood creatinine meter (PCM) and central laboratory measurement, with eGFR calculated from each.

Summary: Morita pre-contrast study 2011 (Japanese Journal of Radiology — prospective study, 113 patients scheduled for contrast-enhanced CT/MRI without recent creatinine data): the POC whole-blood creatinine meter's eGFR averaged 81.2 ± 24.6 vs the laboratory's 70.2 ± 19.7 mL/min/1.73 m² (p < 0.0001), correlation r = 0.74, Bland-Altman mean difference +11.0 mL/min/1.73 m² with limits of agreement −22.4 to +44.4 — correlation printed as correlation, never as agreement; the meter systematically overestimates eGFR, the direction that under-calls renal impairment before contrast

Result: The mean eGFR measured with the PCM was slightly higher than the laboratory value (81.2 ± 24.6 vs. 70.2 ± 19.7 ml/min/1.73 m(2), P < 0.0001). They were well correlated (r = 0.74, y = 0.92x + 16.9, P < 0.0001). Bland-Altman plots (mean difference 11.0 ml/min/1.73 m(2); limits of agreement −22.4 to +44.4 ml/min/1.73 m(2)) showed a moderate agreement with some degree of dispersion.

Watch out: The bias direction is the dangerous one for this use: overestimating eGFR by a mean of 11 points, with an upper agreement limit of +44.4, moves patients across contrast-safety thresholds in the reassuring direction. A correlation of 0.74 is the weakest on this shelf and is printed as correlation only. The authors' own conclusion ('almost equally') is more generous than their limits of agreement.

Source: Morita S et al. Assessing renal function with a rapid, handy, point-of-care whole blood creatinine meter before using contrast materials. Japanese journal of radiology. 2011. PMID: 21519992. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Contrast Media* · Creatinine/blood* · Female · Glomerular Filtration Rate* · Humans · Image Enhancement · Kidney Diseases/blood*,diagnosis* · Magnetic Resonance Imaging · Male · Middle Aged · Point-of-Care Systems* · Prospective Studies · Reproducibility of Results · Tomography, X-Ray Computed

Card 3 — Öztürk fingerstick study 2025 (BMC Emergency Medicine — prospective single-centre study, 246 adult ED patients analysed (250 enrolled, 4 excluded for kit problems)): fingerstick creatinine on an i-STAT cartridge vs laboratory enzymatic serum creatinine: ICC 0.996, with strong agreement in every subgroup examined (complaint, comorbidity, age, sex, vitals, GFR band) — an ICC is a consistency statistic, printed as such; the record's abstract prints no Bland-Altman bias for this anchor and none may be quoted

Study: Prospective agreement study (primary report)

Population: 250 patients aged 18 years and over presenting to an adult emergency department who underwent serum biochemistry testing (December 2023), of whom 4 were excluded for kit problems, leaving 246 analysed (133 female, 113 male; median age 55.5); fingertip blood was scanned using the i-STAT device and a creatinine cartridge alongside laboratory enzymatic serum creatinine.

Summary: Öztürk fingerstick study 2025 (BMC Emergency Medicine — prospective single-centre study, 246 adult ED patients analysed (250 enrolled, 4 excluded for kit problems)): fingerstick creatinine on an i-STAT cartridge vs laboratory enzymatic serum creatinine: ICC 0.996, with strong agreement in every subgroup examined (complaint, comorbidity, age, sex, vitals, GFR band) — an ICC is a consistency statistic, printed as such; the record's abstract prints no Bland-Altman bias for this anchor and none may be quoted

Result: A strong agreement was found between the measured values (ICC: 0.996). When patients were analysed among themselves in terms of admission complaints, existing diseases, age, gender, admission vital signs and glomerular filtration rates, a strong agreement was observed between fingertip creatinine and serum creatinine in each group. The authors conclude the fingerstick creatinine test may be considered for early use in the ED.

Watch out: An ICC of 0.996 in a sample spanning a wide creatinine range is expected even with clinically important absolute errors at the extremes — consistency statistics rise with between-patient spread. Without printed bias and limits, this card supports 'tracks the laboratory closely in this ED sample' and nothing stronger. Single centre, single month.

Source: Öztürk SB et al. Agreement between serum creatinine and fingerstick creatinine measurements in adult patients admitted to the emergency department. BMC emergency medicine. 2025. PMID: 41214497. Registry ID: none in the record. MeSH: Humans · Female · Male · Creatinine*/blood · Emergency Service, Hospital* · Prospective Studies · Middle Aged · Adult · Aged · Fingers/blood supply

Card 4 — Msilanga StatSensor study 2025 (PLOS ONE — diagnostic-accuracy cross-sectional study, 358 people living with HIV at a Tanzanian referral HIV clinic; the named StatSensor POC creatinine device (named in the record — agreement is device-specific, the standing product-naming exception applies)): POC-eGFR vs Jaffe-method laboratory eGFR (CKD-EPI 2021) for kidney dysfunction (eGFR <60): sensitivity 92.9%, specificity 94.7%, overall accuracy 94.4%, AUC 0.938, kappa 0.805; Bland-Altman bias −4.36 mL/min/1.73 m² with limits of agreement −19.68 to +28.40 and R² = 0.813

Study: Diagnostic-accuracy cross-sectional study (primary report)

Population: 358 people living with HIV aged 18 years and over at the Temeke Regional Referral Hospital HIV Clinic, Tanzania (January–March 2025); median age 48 years, 66.2% female, 15.6% with kidney dysfunction (eGFR <60 mL/min/1.73 m² by CKD-EPI 2021 from Jaffe-method serum creatinine).

Summary: Msilanga StatSensor study 2025 (PLOS ONE — diagnostic-accuracy cross-sectional study, 358 people living with HIV at a Tanzanian referral HIV clinic; the named StatSensor POC creatinine device (named in the record — agreement is device-specific, the standing product-naming exception applies)): POC-eGFR vs Jaffe-method laboratory eGFR (CKD-EPI 2021) for kidney dysfunction (eGFR <60): sensitivity 92.9%, specificity 94.7%, overall accuracy 94.4%, AUC 0.938, kappa 0.805; Bland-Altman bias −4.36 mL/min/1.73 m² with limits of agreement −19.68 to +28.40 and R² = 0.813

Result: The StatSensor demonstrated 92.9% sensitivity, 94.7% specificity, and 94.4% overall diagnostic accuracy compared to the Jaffe method. The ROC curve (AUC = 0.938) indicated strong test performance, showing substantial agreement with a kappa value of 0.805. Bland-Altman analysis revealed a negative bias of 4.36 mL/min/1.73 m² with limits of agreement from −19.68 to 28.40 and a strong correlation (R² = 0.813) between the two methods.

Watch out: Classification at the <60 threshold is strong while the eGFR limits (−19.68 to +28.40) are roughly ±24 wide — both are true at once, because most screened patients sit far from the threshold. The record words the bias as 'a negative bias of 4.36'; it prints here as −4.36. The reference is the Jaffe method, itself not the enzymatic reference used in other anchors.

Source: Msilanga D et al. Point-of-care creatinine-based eGFR (StatSensor) in detecting kidney dysfunction (KD) among people living with HIV in Tanzania. PloS one. 2025. PMID: 40934196. Registry ID: none in the record. MeSH: Humans · Female · Creatinine*/blood · Male · HIV Infections*/complications,blood · Middle Aged · Glomerular Filtration Rate* · Adult · Cross-Sectional Studies · Tanzania/epidemiology · Point-of-Care Systems* · ROC Curve · Sensitivity and Specificity · Point-of-Care Testing

Card 5 — You suspected-stroke analysis 2013 (Emergency Medicine Journal — retrospective analysis in ED suspected-stroke patients (cohort N not stated in the record's abstract — N not documented for this limb)): POCT BUN/creatinine vs serum reference: ICC 0.980 for creatinine and 0.914 for BUN (consistency statistics, printed as such); result turnaround 11.4 vs 46.8 minutes — only the creatinine limb is counted here

Study: Retrospective agreement analysis (primary report; creatinine limb only)

Population: Suspected stroke patients presenting to an emergency department between November 2009 and November 2010 who had BUN and creatinine measured by rapid point-of-care BUN/creatinine testing and by reference laboratory tests on blood transferred to the central laboratory. The record's abstract states no cohort N — N not documented.

Summary: You suspected-stroke analysis 2013 (Emergency Medicine Journal — retrospective analysis in ED suspected-stroke patients (cohort N not stated in the record's abstract — N not documented for this limb)): POCT BUN/creatinine vs serum reference: ICC 0.980 for creatinine and 0.914 for BUN (consistency statistics, printed as such); result turnaround 11.4 vs 46.8 minutes — only the creatinine limb is counted here

Result: The mean (SD) interval from arrival at the ED to reporting of the results was 11.4 (4.9) min for POCT-BUN/Cr and 46.8 (38.5) min for the serum reference laboratory tests (p < 0.001). Intra-class correlation coefficient (ICC) analysis demonstrated a high level of agreement (the consistency agreement) between POCT and the serum reference tests for both BUN (ICC = 0.914) and Cr (ICC = 0.980).

Watch out: The turnaround advantage (11.4 vs 46.8 minutes) is this card's hardest figure; the ICC is a consistency statistic with no bias or limits printed in the record, and the cohort size is not documented, so the limb's weight cannot be checked. The clinical frame is contrast-induced-nephropathy risk triage, where the Morita card shows what an eGFR bias can do at exactly that decision.

Source: You JS et al. The usefulness of rapid point-of-care creatinine testing for the prevention of contrast-induced nephropathy in the emergency department. Emergency medicine journal : EMJ. 2013. PMID: 22833599. Registry ID: none in the record. MeSH: Acute Kidney Injury/chemically induced,prevention & control* · Benchmarking · Clinical Laboratory Techniques · Contrast Media/adverse effects* · Creatinine/urine* · Emergency Medical Services* · Evidence-Based Medicine · Humans · Kidney Function Tests · Point-of-Care Systems* · Reference Standards · Retrospective Studies · Stroke/urine · Time Factors

Card 6 — Kok ICU multi-device evaluation 2015 (Pathology — prospective multi-device ICU evaluation, 40 arterial samples from 28 critically unwell patients): across the i-STAT CHEM 8+/CG4+, Hemochron and HemoCue devices, good correlation (bias <10%) was shown for sodium, potassium, chloride, ionised calcium, glucose, urea, haemoglobin and haematocrit — but not for creatinine, anion gap, pO₂, base excess or lactate — shared report, printed as shared: this report is carded on this row for its creatinine limb (the row's negative limb); its INR limb is cross-pointed on the POC INR shelf (pack #8857) in prose only and counted nowhere; its blood-gas limbs are counted nowhere

Study: Prospective multi-device evaluation (shared report — creatinine limb; the shelf's negative limb)

Population: 40 arterial blood samples collected from 28 medical and surgical patients admitted to the intensive care unit at an Australian tertiary care hospital, measured on the i-STAT CHEM 8+ and CG4+, Hemochron Signature Elite, and HemoCue Hb 201+ and WBC Diff systems against clinical laboratory analysers (XE-5000, STA-R Evolution, Dimension Vista 1500 and ABL800 FLEX).

Summary: Kok ICU multi-device evaluation 2015 (Pathology — prospective multi-device ICU evaluation, 40 arterial samples from 28 critically unwell patients): across the i-STAT CHEM 8+/CG4+, Hemochron and HemoCue devices, good correlation (bias <10%) was shown for sodium, potassium, chloride, ionised calcium, glucose, urea, haemoglobin and haematocrit — but not for creatinine, anion gap, pO₂, base excess or lactate — shared report, printed as shared: this report is carded on this row for its creatinine limb (the row's negative limb); its INR limb is cross-pointed on the POC INR shelf (pack #8857) in prose only and counted nowhere; its blood-gas limbs are counted nowhere

Result: There was good correlation demonstrated for sodium, potassium, chloride, ionised calcium, glucose, urea, haemoglobin and haematocrit values (i-STAT Chem 8+); pH, pCO(2), bicarbonate and oxygen saturation (i-STAT CG4+); haemoglobin, white cell, neutrophil count and lymphocyte counts (Hemocue); and internationalised normal ratio (INR; Hemochron Signature Elite), but not creatinine, anion gap, pO(2), base excess, lactate, eosinophil count, prothrombin and activated partial thromboplastin time. Good correlation was defined as a bias of <10% between the POC device and the reference method.

Watch out: A negative limb from 40 samples in 28 patients — small, and 'not good correlation' under a <10% bias definition is a threshold verdict, not a measured bias; the record prints no creatinine bias figure and none may be quoted. Critically unwell patients (haemodilution, oedema, vasopressors) are exactly the spectrum where whole-blood creatinine is hardest, so this limb bounds the positive ED/clinic findings rather than refuting them.

Source: Kok J et al. Evaluation of point-of-care testing in critically unwell patients: comparison with clinical laboratory analysers and applicability to patients with Ebolavirus infection. Pathology. 2015. PMID: 26126049. Registry ID: none in the record. MeSH: Australia · Clinical Laboratory Techniques* · Critical Care · Ebolavirus/isolation & purification* · Hemorrhagic Fever, Ebola/therapy* · Humans · Partial Thromboplastin Time · Point-of-Care Testing* · Tertiary Care Centers

Card 7 — Synthesis — classification agrees, eGFR disagrees, and the bias flips sign: three layers that must print together

Study: Synthesis

Population: 14,567 ED patients (paired assays); 113 pre-contrast patients; 246 ED patients (fingerstick); 358 people living with HIV (screening); a suspected-stroke cohort of undocumented N; 40 ICU samples from 28 patients (negative limb). Counting units differ and are never merged.

Summary: Synthesis — classification agrees, eGFR disagrees, and the bias flips sign: three layers that must print together

Result: Classification layer: 94.5% of ED POC creatinine within 0.3 mg/dL of the laboratory; sensitivity 92.9% / specificity 94.7% for eGFR <60 on the StatSensor. Consistency layer: ICC 0.996 (fingerstick) and 0.980 (stroke POCT) — consistency statistics, not agreement limits. eGFR layer: bias +11.0 (limits −22.4 to +44.4) pre-contrast vs −4.36 (limits −19.68 to +28.40) in HIV screening — opposite directions, device- and population-bound. ICU layer: creatinine among the failing analytes in the multi-device evaluation. Shared-report discipline: Dashevsky and You are carded for creatinine only; Kok is carded here for creatinine only, its INR limb cross-pointed in prose on pack #8857 and counted nowhere. No slate-vs-record discrepancy was found at build.

Watch out: Creatinine decisions are threshold decisions (contrast, drug dosing, AKI staging): agreement statistics averaged over wide ranges flatter performance exactly where it matters least. The pre-contrast bias direction (+11.0, overestimating eGFR) under-calls renal impairment before contrast — the one card whose error direction is itself the finding.

Source: Synthesis of the reports carded above; no additional source.