SourceMed · Shelf #8857 in the ask interface
Point-of-care coagulometer INR vs laboratory INR — 4 primaries, no synthesis, spanning 2009–2023 and two device generations: the era contradiction is the headline — the 2009 head-to-head found 40% (CoaguChek S) and 43% (XS) of readings deviating more than 15% from the laboratory, while the 2016 XS validation in 304 warfarin patients found 93.9% within 0.5 units in the 2.0–3.5 target range; in liver disease the XS Pro put 93–100% of paediatric and 71% of adult readings within 15%; qLabs in the ICU reached 90% / 81% clinical agreement within ±0.3
Shelf #8857 · Heart, vessels, blood pressure & clotting
MEASUREMENT ACCURACY — precise devices, range-dependent accuracy, and an era split that is this shelf's headline: the 2009 head-to-head found good precision (CV 3.4% (S) / 2.3% (XS)) but 40% (S) and 43% (XS) of readings deviating more than 15% from the laboratory; the 2016 XS validation in 304 warfarin patients found 93.9% of readings within 0.5 units of the laboratory in the 2.0–3.5 target range with variability rising above INR 3.6; in liver disease the XS Pro put 93–100% of paediatric readings and 71% of adult readings within 15% of the laboratory; the qLabs device in critically ill patients reached 90% (finger-prick) and 81% (central-line droplet) clinical agreement for INR within ±0.3 — while its aPTT limb failed (counted nowhere). The 2009 and 2016 findings both print at full weight: device generation (S vs XS), INR range and the agreement criterion (15% relative vs 0.5 units absolute) all differ and are the likely explanation, but no anchor adjudicates — and a reading that quotes only the 2009 figure is quoting a superseded device.
Stamped: 2026-11-28.
Evidence published: 2009–2023.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients on coumarin/warfarin anticoagulation and in patients with liver disease or critical illness, how closely does point-of-care coagulometer INR agree with laboratory INR — for the CoaguChek S and XS, the CoaguChek XS in routine warfarin monitoring, the CoaguChek XS Pro in liver disease, and the qLabs portable coagulometer in the ICU?
Balance of the evidence
The composition is 4 primaries, no synthesis, spanning 2009–2023 and two device generations, across three spectra that print separately: coumarin/warfarin monitoring (Christensen, Benade), liver disease — where the INR is a liver-function measurement, not a warfarin dose check (Shadlyn) — and critical illness (Casado-Méndez). The 2009 study's base is 24 patients carrying 564 measurements; the counting unit is measurements, printed as such. Cross-point in prose only: the ICU multi-device report PMID 26126049 found good correlation for its Hemochron INR limb in 40 samples — it is carded on the POC creatinine shelf (pack #8856) and counted nowhere here. Agreement is device-specific, so devices are named exactly as the publications name them (CoaguChek S, XS, XS Pro; qLabs) under the standing product-naming exception. The homed anticoagulation ground (warfarin dosing, genotype-guided dosing) asks treatment questions, not device-agreement questions. No record carries a registry ID; the Benade record prints no MeSH.
MeSH: Aged · Anticoagulants/administration & dosage* · Coumarins/administration & dosage* · Drug Monitoring/instrumentation,methods · Female · Humans · International Normalized Ratio/instrumentation*,methods · Male · Middle Aged · Point-of-Care Systems · Self Administration/methods · Sensitivity and Specificity · Adult · Child · International Normalized Ratio · Point-of-Care Systems* · Anticoagulants · Prospective Studies · Liver Diseases* · Critical Illness · International Normalized Ratio/methods*,standards · Partial Thromboplastin Time/methods*,standards · Point-of-Care Systems/standards* · Reproducibility of Results
Cards (5)
Card 1 — Christensen head-to-head 2009 (Thrombosis and Haemostasis — prospective study, 24 coumarin-treated patients followed six weeks with weekly duplicate measurements (141 duplicate sets, 564 measurements) on CoaguChek S and XS vs the hospital laboratory): precision CV 3.4% (S) and 2.3% (XS), only the XS within the predefined 3% limit; coagulometer readings tended lower than laboratory, and 40% (S) / 43% (XS) of measurements deviated more than 15% from the laboratory value — the row's oldest and most negative limb; a 24-patient base with 564 measurements — the counting unit is measurements, printed as such
Study: Prospective agreement study (primary report; era limb 2009)
Population: 24 patients treated with coumarin, prospectively followed for six weeks; INRs were analysed weekly in duplicates on both coagulometers (CoaguChek S and CoaguChek XS) and compared with results from the hospital laboratory — 141 duplicate measurements, a total of 564 measurements.
Summary: Christensen head-to-head 2009 (Thrombosis and Haemostasis — prospective study, 24 coumarin-treated patients followed six weeks with weekly duplicate measurements (141 duplicate sets, 564 measurements) on CoaguChek S and XS vs the hospital laboratory): precision CV 3.4% (S) and 2.3% (XS), only the XS within the predefined 3% limit; coagulometer readings tended lower than laboratory, and 40% (S) / 43% (XS) of measurements deviated more than 15% from the laboratory value — the row's oldest and most negative limb; a 24-patient base with 564 measurements — the counting unit is measurements, printed as such
Result: The CoaguChek S and CoaguChek XS had a precision (CV) of 3.4% and 2.3%, respectively; only the CoaguChek XS had a precision within the predefined limit of 3%. Regarding analytical accuracy, the INR measurements tended to be lower on the coagulometers compared to the laboratory. Regarding diagnostic accuracy the CoaguChek S and CoaguChek XS deviated more than 15% from the laboratory measurements in 40% and 43% of the measurements, respectively. Whether this will have a clinical impact awaits further studies, as the authors state.
Watch out: 564 measurements from 24 patients are not 564 patients — repeated measures within the same patients and devices concentrate the evidence. The >15% deviation criterion is a relative criterion: at low INRs 15% is a small absolute gap, at high INRs a large one, and the record does not stratify the deviations by range on its face. This is a 2009, first-generation-in-part (S) result; the XS in the same study already met the precision limit.
Source: Christensen TD et al. International normalised ratio (INR) measured on the CoaguChek S and XS compared with the laboratory for determination of precision and accuracy. Thrombosis and haemostasis. 2009. PMID: 19277421. Registry ID: none in the record. MeSH: Aged · Anticoagulants/administration & dosage* · Coumarins/administration & dosage* · Drug Monitoring/instrumentation,methods · Female · Humans · International Normalized Ratio/instrumentation*,methods · Male · Middle Aged · Point-of-Care Systems · Self Administration/methods · Sensitivity and Specificity
Card 2 — Benade XS validation 2016 (South African Medical Journal — validation in 304 patients on long-term warfarin (standard target 2–3 and mechanical-valve target 2.5–3.5) at a Johannesburg academic hospital): CoaguChek XS mean INR 2.75 (SD 1.18) vs STAGO laboratory analyser 2.65 (SD 1.04); Bland-Altman bias small with imprecision within acceptable limits; within the 2.0–3.5 range 93.9% of XS readings were within 0.5 units of the laboratory result, with variability of differences increasing at INR >3.6 — the record prints no MeSH ('none in the record')
Study: Validation study (primary report; era limb 2016)
Population: 304 patients referred for routine testing on long-term warfarin therapy — standard warfarin therapy (INR 2–3) and mechanical heart valve replacements (INR 2.5–3.5) — comparing the CoaguChek XS point-of-care device with the STAGO laboratory analyser for accuracy and precision.
Summary: Benade XS validation 2016 (South African Medical Journal — validation in 304 patients on long-term warfarin (standard target 2–3 and mechanical-valve target 2.5–3.5) at a Johannesburg academic hospital): CoaguChek XS mean INR 2.75 (SD 1.18) vs STAGO laboratory analyser 2.65 (SD 1.04); Bland-Altman bias small with imprecision within acceptable limits; within the 2.0–3.5 range 93.9% of XS readings were within 0.5 units of the laboratory result, with variability of differences increasing at INR >3.6 — the record prints no MeSH ('none in the record')
Result: The mean INR value of the CoaguChek XS of 2.75 (standard deviation (SD) 1.18) was comparable to that of the STAGO (2.65 (SD 1.04)). The Bland-Altman difference plot revealed good agreement; bias between the two methods was small, and the imprecision was within acceptable limits. Within the target range (2.0–3.5), 93.9% of the CoaguChek XS INR readings were within 0.5 units of the standard laboratory method result. There was, however, an increase in the variability of the differences between the two test methods when the INR was >3.6.
Watch out: The agreement criterion differs from the 2009 card's — 0.5 absolute units within a restricted target range, vs 15% relative deviation across all ranges — so the two headlines (43% deviating vs 93.9% agreeing) are not measured on the same ruler, and neither card adjudicates the other. The >3.6 variability warning matters clinically: high INRs are where dose decisions are most consequential. The record prints no MeSH.
Source: Benade EL et al. Validation of the CoaguChek XS international normalised ratio point-of-care analyser in patients at Charlotte Maxeke Johannesburg Academic Hospital, South Africa. South African medical journal = Suid-Afrikaanse tydskrif vir geneeskunde. 2016. PMID: 26915943. Registry ID: none in the record. MeSH: none in the record
Card 3 — Shadlyn liver-disease cohort 2023 (Journal of Pediatric Gastroenterology and Nutrition — prospective cohort, 15 paediatric patients without liver disease, 13 paediatric and 17 adult patients with liver disease (45 patients)): CoaguChek XS Pro vs laboratory INR correlations r = 0.82 (paediatric, no liver disease), 0.89 (paediatric liver disease) and 0.96 (adult liver disease) — correlations printed as correlations; the proportion of POC values within 15% of the paired laboratory INR was 93% (14/15), 100% (13/13) and 71% (12/17) respectively — a non-anticoagulated, liver-disease spectrum: the INR here is a liver-function measurement, not a warfarin dose check, printed as such
Study: Prospective cohort — agreement study (primary report; liver-disease spectrum)
Population: 15 pediatric patients free of liver disease, 13 pediatric patients who had liver disease, and 17 adults with liver disease (45 patients) formed this prospective cohort, set up to test whether the CoaguChek XS Pro point-of-care device measures INR accurately and precisely against the gold standard laboratory INR.
Summary: Shadlyn liver-disease cohort 2023 (Journal of Pediatric Gastroenterology and Nutrition — prospective cohort, 15 paediatric patients without liver disease, 13 paediatric and 17 adult patients with liver disease (45 patients)): CoaguChek XS Pro vs laboratory INR correlations r = 0.82 (paediatric, no liver disease), 0.89 (paediatric liver disease) and 0.96 (adult liver disease) — correlations printed as correlations; the proportion of POC values within 15% of the paired laboratory INR was 93% (14/15), 100% (13/13) and 71% (12/17) respectively — a non-anticoagulated, liver-disease spectrum: the INR here is a liver-function measurement, not a warfarin dose check, printed as such
Result: Correlations between the device and laboratory INR were excellent in every stratum: r = 0.82 in pediatric patients without liver disease, r = 0.89 in pediatric patients with liver disease, and r = 0.96 in adult patients with liver disease. Among pediatric patients without liver disease, Fourteen (93%) POC INR values sat ≤15% from the paired laboratory value. For the pediatric liver-disease group, each of the 13 paired readings fell within ≤15%. In adults with liver disease, 12 (71%) POC INR values lay within ≤15% of the laboratory INR they were paired with.
Watch out: Small strata (13–17 patients per group): the 100% paediatric-liver figure is 13 of 13 readings, and the adult figure (71%, 12/17) is the shelf's weakest within-15% rate — in the population where INR feeds transplant-priority scores, a 29% out-of-tolerance rate is the operative number. One paired measurement per patient, as the record presents it; correlations are printed as correlations, not agreement.
Source: Shadlyn T et al. Accuracy and Precision of Point-of-Care International Normalized Ratio in Patients With Liver Disease. Journal of pediatric gastroenterology and nutrition. 2023. PMID: 36800295. Registry ID: none in the record. MeSH: Adult · Humans · Child · International Normalized Ratio · Point-of-Care Systems* · Anticoagulants · Prospective Studies · Liver Diseases*
Card 4 — Casado-Méndez ICU study 2018 (Blood Coagulation & Fibrinolysis — prospective study in critically ill patients, qLabs portable coagulometer vs central laboratory, finger-prick group n = 30 and central-venous-droplet group n = 60): INR precision ICC >0.90; INR clinical agreement (±0.3 units) 90% (finger-prick) and 81% (droplet); accuracy was better near normal values — only the INR limb is counted here; the aPTT limb (clinical agreement 55% / 68%, ICC 0.70 / 0.50) failed in the same study and is counted nowhere
Study: Prospective agreement study (primary report; INR limb only)
Population: Critically ill patients in a prospective observational study of the qLabs portable coagulometer against standard central laboratory measurement: a finger-prick group (N = 30 patients) and a blood-droplet group from central venous catheter draws (N = 60); clinical agreement was defined as ±0.3 for INR and ±10 s for aPTT.
Summary: Casado-Méndez ICU study 2018 (Blood Coagulation & Fibrinolysis — prospective study in critically ill patients, qLabs portable coagulometer vs central laboratory, finger-prick group n = 30 and central-venous-droplet group n = 60): INR precision ICC >0.90; INR clinical agreement (±0.3 units) 90% (finger-prick) and 81% (droplet); accuracy was better near normal values — only the INR limb is counted here; the aPTT limb (clinical agreement 55% / 68%, ICC 0.70 / 0.50) failed in the same study and is counted nowhere
Result: Precision of INR measurement in qLabs showed excellent intraclass correlation coefficient (ICC > 90% — as the record prints it, >0.90). Accuracy of qLabs was acceptable for INR assessment (clinical agreement 90 and 81%, for finger prick and blood droplet groups, respectively), but not for aPTT (clinical agreement 55 and 68%, respectively). Precision of aPTT measurement was less acceptable for both finger prick [ICC: 0.70] and blood droplet [ICC: 0.50] groups. Accuracy of finger prick and blood droplet measurements was better for INR and aPTT values near-to-normal (1.2 and 37 s, respectively). INR values from qLabs were consistent with the 'gold standard'.
Watch out: Critically ill patients are mostly not in the warfarin target range — agreement 'better near normal values' means this card validates the device where ICU patients actually sit, and says least about the 2.0–3.5 monitoring range the warfarin cards address. The same device failed its aPTT limb in the same patients; a device-level 'reliable' claim that merged the two limbs would misstate this record.
Source: Casado-Méndez M et al. Reliability of the portable coagulometer qLabs to accurately measure the activated thromboplastin time and international normalized ratio: a prospective study in critically ill patients. Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis. 2018. PMID: 30234544. Registry ID: none in the record. MeSH: Critical Illness · Humans · International Normalized Ratio/methods*,standards · Partial Thromboplastin Time/methods*,standards · Point-of-Care Systems/standards* · Prospective Studies · Reproducibility of Results · Sensitivity and Specificity
Card 5 — Synthesis — 4 primaries, no synthesis, two device generations: the era contradiction (43% deviate >15% in 2009 vs 93.9% within 0.5 units in 2016) is the headline and is not adjudicated by any anchor
Study: Synthesis
Population: 24 coumarin patients carrying 564 measurements (2009); 304 warfarin patients (2016); 45 liver-disease-spectrum patients (2023); 90 critically ill patients in two sampling groups (2018). Measurements, patients and sampling groups are different counting units and print per card.
Summary: Synthesis — 4 primaries, no synthesis, two device generations: the era contradiction (43% deviate >15% in 2009 vs 93.9% within 0.5 units in 2016) is the headline and is not adjudicated by any anchor
Result: Precision is good wherever measured (CV 2.3–3.4% in 2009; ICC >0.90 in the ICU). Accuracy verdicts split by era, device, range and criterion: 40–43% of readings >15% from the laboratory (2009, S and XS, all ranges, relative criterion); 93.9% within 0.5 units in the 2.0–3.5 range (2016, XS, absolute criterion), variability rising above 3.6; 93%/100%/71% within 15% across the liver-disease strata (2023, XS Pro); 90%/81% within ±0.3 units in the ICU (2018, qLabs). Cross-point in prose only: PMID 26126049's Hemochron INR limb (good correlation, 40 ICU samples) is carded on pack #8856 and counted nowhere here. No report was refused; no slate-vs-record discrepancy was found at build.
Watch out: This shelf was rowed against an earlier desk verdict of 'documented zero' for the family — the census found four fetch-verified primaries, and records govern; the contradiction is flagged for the coordinator in the part summary. Clinically: POC INR is at its best inside the target range on the later device, and least verified exactly where decisions bite — above INR 3.6 and in adult liver disease (71% within 15%).
Source: Synthesis of the reports carded above; no additional source.