Point-of-care glucose in critically ill adults (glucometers and arterial blood-gas analysers vs central laboratory) — 2 overlapping syntheses + 2 primaries: arterial blood is the fix and capillary-in-shock is the failure — arterial blood-gas analysers and arterial-sample glucometers were significantly more accurate than capillary glucometers (odds ratios for error 0.04 and 0.36, relative), arterial glucometer bias was 0.01 in the 32-study meta-analysis, and the largest primary found venous catheter samples worst of all (bias +15.1 mg/dL, limits −51.7 to +81.9)

Shelf #8859 · Critical care & sepsis

MEASUREMENT ACCURACY — arterial blood is the fix and capillary-in-shock is the failure: the 2013 systematic review (21 studies) found arterial blood-gas analysers and arterial-sample glucometers significantly more accurate than capillary glucometers (odds ratios for error 0.04 and 0.36 — relative measures, labelled as such), with error rates rising in the hypoglycaemic range (<81 mg/dL) and with oedema, vasopressors and insulin; the 2024 meta-analysis (32 studies, 5,451 patients) found no discrepancy for arterial glucometer vs laboratory (bias 0.01) but significant overestimation by gasometer and a consensus of poor capillary-glucometer agreement in shock/vasopressor use, renal failure and vitamin C treatment; the largest primary found venous catheter samples worst of all (bias +15.1 mg/dL, limits −51.7 to +81.9). The most positive primary's conclusion ('in agreement') and its limits of agreement (−63 to +54 mg/dL, the widest in the row) print together, never the conclusion alone — on the printed evidence this shelf sides with the syntheses.

Stamped: 2026-11-28.

Evidence published: 2013–2024.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In critically ill adults, how accurately do point-of-care glucose measurements — glucometers on capillary, arterial and venous samples, and arterial blood-gas analysers — agree with central-laboratory glucose, and which sample/method combinations fail?

Balance of the evidence

The composition is 2 overlapping syntheses + 2 primaries: the 2024 meta-analysis's primary pool overlaps the 2013 review's, so the two syntheses are not independent confirmations. Sample type (arterial / capillary / venous-catheter) is the row's organising variable and every figure prints with its sample attached. Glucose prints in mg/dL exactly as the records give it. Devices are named exactly as the publications name them (Accu-Chek Advantage II, Precision PCx) under the standing product-naming exception. The homed neighbour is pack #7857 (blood-glucose monitoring system accuracy under ISO 15197 for self-monitoring) — a self-monitoring population and a systems-certification question; this shelf's question is bedside measurement in critically ill adults against the central laboratory, discharged at question level and cross-pointed; the two bases must not be merged, because #7857's system evaluations say nothing about vasopressor-era capillary error. The vitamin-C interference limb inside the 2024 synthesis connects to a located primary (PMID 30230568, high-dose IV vitamin C cohort — located in census, not carded, no figure quoted). The 2024 record's databank list carries no registry ID; its OSF protocol DOI (10.17605/OSF.IO/T8KYP) is printed in the record text and noted on its card.

MeSH: Blood Gas Analysis/methods,standards · Blood Glucose/metabolism* · Critical Illness* · Humans · Monitoring, Physiologic/methods,standards* · Adult · Blood Glucose*/analysis · Reproducibility of Results · Aged · Aged, 80 and over · Blood Chemical Analysis/instrumentation,standards* · Blood Glucose/analysis* · Central Venous Catheters/standards · Critical Care/standards* · Cross-Sectional Studies · Dimensional Measurement Accuracy · Female · Intensive Care Units · Male · Middle Aged · Point-of-Care Systems/standards* · Glucose* · Critical Illness

Cards (5)

Card 1 — Inoue systematic review 2013 (Critical Care — 21 studies comparing glucometer and/or arterial blood-gas analyser glucose against central laboratory in critically ill adults; 11 studies evaluable on common accuracy criteria): accuracy with arterial blood was significantly higher for blood-gas analysers and for arterial-sample glucometers than for capillary glucometers (ORs for error 0.04, p < 0.001, and 0.36, p < 0.001 — relative measures, labelled as such); error incidence was higher in the hypoglycaemic range (<81 mg/dL) for all methods (ORs 1.86, 1.84 and 2.33); unstable haemodynamics (oedema, vasopressor use) and insulin use were associated with increased glucometer error; the authors caution that accuracy varied largely among devices — synthesis, counted as a synthesis

Study: Systematic review (synthesis)

Population: 21 clinical studies (from 879 articles screened, literature search January 2001 to August 2012) comparing blood-glucose values measured with glucose meters and/or arterial blood gas analysers with those simultaneously measured with a central laboratory machine in critically ill adult patients; 11 studies using ISO criteria, error-grid method, or percentage within 20% of the reference were selected for evaluation.

Summary: Inoue systematic review 2013 (Critical Care — 21 studies comparing glucometer and/or arterial blood-gas analyser glucose against central laboratory in critically ill adults; 11 studies evaluable on common accuracy criteria): accuracy with arterial blood was significantly higher for blood-gas analysers and for arterial-sample glucometers than for capillary glucometers (ORs for error 0.04, p < 0.001, and 0.36, p < 0.001 — relative measures, labelled as such); error incidence was higher in the hypoglycaemic range (<81 mg/dL) for all methods (ORs 1.86, 1.84 and 2.33); unstable haemodynamics (oedema, vasopressor use) and insulin use were associated with increased glucometer error; the authors caution that accuracy varied largely among devices — synthesis, counted as a synthesis

Result: The accuracy of blood-glucose measurements by arterial blood gas analysers and glucose meters by using arterial blood was significantly higher than that of measurements with glucose meters by using capillary blood (odds ratios for error: 0.04, P < 0.001; and 0.36, P < 0.001). The accuracy with arterial blood gas analysers tended to be higher than with glucose meters using arterial blood (P = 0.20). In the hypoglycemic range (defined as <81 mg/dl), the incidence of errors using these devices was higher than in the nonhypoglycemic range (odds ratios for error: arterial blood gas analyzers, 1.86, P = 0.15; glucose meters with capillary blood, 1.84, P = 0.03; glucose meters with arterial blood, 2.33, P = 0.02). Unstable hemodynamics (edema and use of a vasopressor) and use of insulin were associated with increased error of blood glucose monitoring with glucose meters.

Watch out: Odds ratios for error are relative measures pooled across studies using three different accuracy criteria — they rank sample types convincingly but state no absolute error rate for any device. The hypoglycaemic-range odds ratio for blood-gas analysers (1.86) is itself non-significant (P = 0.15) and prints with its p-value. Device variation within each method class was large, as the authors caution — the sample-type ranking is not a device guarantee.

Source: Inoue S et al. Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review. Critical care (London, England). 2013. PMID: 23506841. Registry ID: none in the record. MeSH: Blood Gas Analysis/methods,standards · Blood Glucose/metabolism* · Critical Illness* · Humans · Monitoring, Physiologic/methods,standards*

Card 2 — Arias-Rivera systematic review and meta-analysis 2024 (Enfermería Intensiva — 32 articles / 5,451 critically ill adults): arterial glucometer vs laboratory showed no discrepancy (bias 0.01, 95% CI −0.12 to 0.14), arterial gasometer significantly overestimated (bias 0.12, 0.01 to 0.24), and capillary glucometer trended to overestimation (bias 0.07, −0.02 to 0.15) — the record labels these standardised-mean-difference biases in 'mg/dL', an internally inconsistent unit: the figures print as published and the unit must not be silently repaired; the greatest cross-study consensus was poor capillary-glucometer agreement with shock and vasopressor support, renal failure or vitamin C treatment; OSF protocol DOI 10.17605/OSF.IO/T8KYP printed in the record — synthesis, counted as a synthesis; its primary pool overlaps the 2013 review's

Study: Systematic review and meta-analysis (synthesis; pool overlaps the 2013 review)

Population: 32 articles and 5,451 critically ill adult patients in a systematic review and meta-analysis (searches in PubMed, CINAHL and Embase to September 2022, plus grey literature) comparing blood samples (arterial, venous, capillary) and instruments (glucometers, gasometers, central laboratory) for glycaemia reliability, with methodological quality assessed by QUADAS-2.

Summary: Arias-Rivera systematic review and meta-analysis 2024 (Enfermería Intensiva — 32 articles / 5,451 critically ill adults): arterial glucometer vs laboratory showed no discrepancy (bias 0.01, 95% CI −0.12 to 0.14), arterial gasometer significantly overestimated (bias 0.12, 0.01 to 0.24), and capillary glucometer trended to overestimation (bias 0.07, −0.02 to 0.15) — the record labels these standardised-mean-difference biases in 'mg/dL', an internally inconsistent unit: the figures print as published and the unit must not be silently repaired; the greatest cross-study consensus was poor capillary-glucometer agreement with shock and vasopressor support, renal failure or vitamin C treatment; OSF protocol DOI 10.17605/OSF.IO/T8KYP printed in the record — synthesis, counted as a synthesis; its primary pool overlaps the 2013 review's

Result: No discrepancies were obtained between arterial glucometer vs laboratory samples [bias (95% CI): 0.01 (−0.12 to 0.14) mg/dL, as the record labels the units]. In contrast, arterial samples with a gasometer did significantly overestimate [bias (95% CI): 0.12 (0.01 to 0.24) mg/dL, as labelled]. The same trend is seen in capillaries with a glucometer, although not significantly [bias (95% CI): 0.07 (−0.02 to 0.15) mg/dL, as labelled]. There is discrepancy between studies on the effect of haematocrit and acid-base balance. The greatest consensus is on the poor agreement of glucometer with capillary vs laboratory samples in the presence of shock and vasopressor support, renal failure or during vitamin C treatment. The authors conclude the evidence recommends arterial blood with a blood glucose meter for better reliability.

Watch out: UNIT FLAG: these are standardised-mean-difference biases from the meta-analysis, and the record labels them in 'mg/dL' — a standardised difference of 0.01 is not 0.01 mg/dL, and the figures print as published with this flag rather than repaired. Read qualitatively, the ordering (arterial glucometer centred, gasometer overestimating, capillary trending over) matches the 2013 review; read quantitatively in mg/dL, the numbers would be meaningless. The two syntheses share primary studies and are never counted as independent.

Source: Arias-Rivera S et al. Blood glucose monitoring in critically ill adult patients: type of sample and method of analysis. Systematic review and meta-analysis. Enfermeria intensiva. 2024. PMID: 37474427. Registry ID: none in the record (OSF protocol DOI 10.17605/OSF.IO/T8KYP printed in the record text). MeSH: Adult · Humans · Blood Glucose*/analysis · Critical Illness* · Reproducibility of Results

Card 3 — Pereira cross-sectional study 2015 (PLOS ONE — prospective study, 145 mixed-ICU patients, simultaneous arterial, fingerstick and central-venous-catheter samples on two named glucometers vs arterial central-laboratory glucose): Accu-Chek arterial bias +10.7 mg/dL (limits −21.3 to +42.7), Precision PCx arterial bias +18.6 (limits −12.6 to +49.5), fingerstick bias +10.0 with wider limits (−31.8 to +51.8), and venous catheter samples poorest (bias +15.1, limits −51.7 to +81.9); hyperglycaemia, low haematocrit and acidosis were associated with larger errors, and vasopressors with fingerstick error; the authors conclude venous catheter sampling should not be used for glycaemic control and both glucometers' reliability was insufficient — devices named as the publication names them; agreement is device-specific

Study: Prospective cross-sectional agreement study (primary report)

Population: 145 patients, prospectively enrolled from a 41-bed adult mixed ICU at a tertiary care hospital. From each patient, samples were taken at the same time by fingerstick, from a central venous catheter, and from an indwelling arterial catheter — a single set per patient. Arterial blood was measured on the Precision PCx, while arterial, fingerstick and venous blood were measured on the Accu-Chek Advantage II, and all were compared against arterial measurements from the central laboratory.

Summary: Pereira cross-sectional study 2015 (PLOS ONE — prospective study, 145 mixed-ICU patients, simultaneous arterial, fingerstick and central-venous-catheter samples on two named glucometers vs arterial central-laboratory glucose): Accu-Chek arterial bias +10.7 mg/dL (limits −21.3 to +42.7), Precision PCx arterial bias +18.6 (limits −12.6 to +49.5), fingerstick bias +10.0 with wider limits (−31.8 to +51.8), and venous catheter samples poorest (bias +15.1, limits −51.7 to +81.9); hyperglycaemia, low haematocrit and acidosis were associated with larger errors, and vasopressors with fingerstick error; the authors conclude venous catheter sampling should not be used for glycaemic control and both glucometers' reliability was insufficient — devices named as the publication names them; agreement is device-specific

Result: Against the central lab, arterial samples on the Accu-Chek differed on average by 10.7 mg/dL (95% limits of agreement −21.3 to 42.7 mg/dL); arterial samples on the Precision PCx differed by 18.6 mg/dL (95% LA −12.6 to 49.5 mg/dL). Fingerstick Accu-Chek readings, compared with central lab arterial values, carried a similar bias (10.0 mg/dL) yet spread wider, 95% LA (−31.8 to 51.8 mg/dL). Venous samples agreed most poorly with arterial central lab values (mean bias 15.1 mg/dL; 95% LA −51.7 to 81.9). Larger differences went with hyperglycemia, low hematocrit and acidosis, and vasopressor administration went with greater error in fingerstick measurements. The authors conclude that blood from central venous catheters should not be used to control glycemia in ICU patients, and that neither of the two glucometers evaluated was reliable enough.

Watch out: Every sample type on both devices biased positive (+10.0 to +18.6 mg/dL) — a systematic over-read that, at insulin-dosing thresholds, pushes toward over-treatment; the limits (up to ±~67 mg/dL around the venous bias) are wider than most ICU glucose targets are wide. One sample set per patient, two devices, one ICU: the venous-catheter verdict is the robust finding; device rankings beyond these two meters are not in this record.

Source: Pereira AJ et al. Inaccuracy of Venous Point-of-Care Glucose Measurements in Critically Ill Patients: A Cross-Sectional Study. PloS one. 2015. PMID: 26067093. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Blood Chemical Analysis/instrumentation,standards* · Blood Glucose/analysis* · Central Venous Catheters/standards · Critical Care/standards* · Cross-Sectional Studies · Dimensional Measurement Accuracy · Female · Humans · Intensive Care Units · Male · Middle Aged · Point-of-Care Systems/standards*

Card 4 — Chandra Prabhakar & Halder cross-sectional study 2024 (Clinical Nutrition ESPEN — facility-based study, 110 critically ill patients; two POC glucometers' capillary glucose vs laboratory venous glucose (glucose-oxidase)): glucometer means 152.9 and 152.2 mg/dL vs laboratory 148.6 mg/dL; Bland-Altman bias 4 mg/dL with limits of agreement −63 to +54 mg/dL; glucometer AUROCs 0.92 and 0.93 for the classification task reported; the authors conclude capillary POC glucose was in agreement with venous laboratory glucose — the row's most positive primary, and its limits of agreement (±~60 mg/dL) are the widest of any anchor: the conclusion and the limits print together, never the conclusion alone

Study: Cross-sectional agreement study (primary report; the shelf's most positive primary)

Population: 110 critically ill patients (38 (34%) women; mean age 52.1 years, range 14–85; 14 patients (12%) known to have diabetes), comparing two point-of-care glucometers' capillary glucose with laboratory venous glucose by the glucose-oxidase method, analysed with Bland-Altman plots and the Clark error grid; 2 (2%) subjects had glucose below 70 mg/dL, 85 (77%) between 70 and 180 mg/dL and 23 (21%) above 180 mg/dL.

Summary: Chandra Prabhakar & Halder cross-sectional study 2024 (Clinical Nutrition ESPEN — facility-based study, 110 critically ill patients; two POC glucometers' capillary glucose vs laboratory venous glucose (glucose-oxidase)): glucometer means 152.9 and 152.2 mg/dL vs laboratory 148.6 mg/dL; Bland-Altman bias 4 mg/dL with limits of agreement −63 to +54 mg/dL; glucometer AUROCs 0.92 and 0.93 for the classification task reported; the authors conclude capillary POC glucose was in agreement with venous laboratory glucose — the row's most positive primary, and its limits of agreement (±~60 mg/dL) are the widest of any anchor: the conclusion and the limits print together, never the conclusion alone

Result: The mean glucose value (glucometer 1) was 152.9 mg/dL (SD 83.1), range 48–501 mg/dL; that by glucometer 2 was 152.2 mg/dL (SD 76.2), range 30–458 mg/dL; and by the laboratory was 148.6 mg/dL (SD 81.5), range 52–480 mg/dL. The Bland-Altman plot showed a bias of 4 mg% (95% CI −9.8 to +1.1, as printed) and the limits of agreement were −63 and +54 mg%. The area under the receiver operating characteristic curve for the two glucometers was 0.92 and 0.93 respectively. The authors conclude point of care glucose measured by glucometers was in agreement with the venous glucose estimation and both glucometers performed uniformly well across the wide range of blood glucose values.

Watch out: The conclusion and the data diverge inside this one record: a bias of 4 mg/dL with limits of −63 to +54 mg/dL means individual capillary readings can sit ~60 mg/dL from the laboratory value in either direction — at a hypoglycaemia threshold of 70 mg/dL that is the whole decision. Only 2 of 110 patients were hypoglycaemic, so the range where the syntheses warn loudest is nearly untested here. The AUROCs (0.92/0.93) are classification statistics for a task the abstract does not fully specify, printed as reported. This shelf sides with the limits over the conclusion.

Source: Chandra Prabhakar M & Halder P. Reliability and accuracy of bedside capillary blood glucose measurement by glucometers compared to venous blood glucose in critically ill patients: A facility based cross-sectional study. Clinical nutrition ESPEN. 2024. PMID: 38479916. Registry ID: none in the record. MeSH: Humans · Female · Middle Aged · Male · Blood Glucose*/analysis · Glucose* · Cross-Sectional Studies · Reproducibility of Results · Critical Illness

Card 5 — Synthesis — 2 overlapping syntheses + 2 primaries: arterial sampling is the convergent recommendation; capillary-in-shock is the convergent failure; the one dissenting primary dissents in its conclusion, not in its limits

Study: Synthesis

Population: 21 studies (2013 review, 11 evaluable on common criteria); 32 articles / 5,451 patients (2024 meta-analysis, pool overlapping the first); 145 ICU patients (2015 primary); 110 critically ill patients (2024 primary). Studies, articles and patients print per card and are never merged.

Summary: Synthesis — 2 overlapping syntheses + 2 primaries: arterial sampling is the convergent recommendation; capillary-in-shock is the convergent failure; the one dissenting primary dissents in its conclusion, not in its limits

Result: Sample-type ordering is consistent across all four anchors: arterial (best) > capillary > venous catheter (worst, in the one primary that tested it). Error concentrates in the hypoglycaemic range (<81 mg/dL) and under oedema, vasopressors, insulin, renal failure and vitamin C. Absolute bedside figures: arterial glucometer biases +10.7 and +18.6 mg/dL with limits spanning ~64 mg/dL; venous catheter +15.1 with limits −51.7 to +81.9; capillary (2024 primary) bias 4 with limits −63 to +54. The 2024 synthesis's bias figures carry a unit flag (standardised differences labelled 'mg/dL' in the record) and print as published. No report was refused; no slate-vs-record discrepancy was found at build beyond the flagged unit inconsistency, which is the record's own.

Watch out: ICU glucose values drive insulin infusions: an error of ±50–80 mg/dL at the limits of agreement is not a rounding issue but a dosing hazard, concentrated in exactly the patients (shock, vasopressors, hypoglycaemia) least able to tolerate it. Nothing here judges self-monitoring devices in stable outpatients — that question is homed at pack #7857 and this base must not be read onto it.

Source: Synthesis of the reports carded above; no additional source.