Blood-gas analyser vs central-laboratory autoanalyser for haemoglobin and electrolytes — 4 primaries, no synthesis, 43,000+ patients/pairs in aggregate (counting units differ): small average biases, interchangeability refused — concordance correlations 0.89 (chloride) to 0.98 (haemoglobin) across 9,398 ICU pairs yet only potassium interchangeable; no significant haemoglobin difference (−0.35 g/L) across five ICUs; glucose bias +6.19 mg/dL with limits −67.69 to +80.07 in 1,000 ED patients; and in 31,060 ED patients correlations of 0.59 (potassium) to 0.98 (haemoglobin) with no acceptable agreement limits for any parameter

Shelf #8860 · Critical care & sepsis

MEASUREMENT ACCURACY — small average biases, interchangeability refused: the 9,398-pair ICU study found concordance correlations of 0.89 (chloride) to 0.98 (haemoglobin) yet concluded the methods cannot be used interchangeably except for potassium (7.1% of glucose, 5.9% of haemoglobin and 5.2% of chloride readings outside US-CLIA acceptable variation); the five-ICU prospective study found no significant haemoglobin difference (−0.35 g/L) and small significant sodium (+1.49 mmol/L) and potassium (+0.19 mmol/L) biases; the 1,000-patient ED study found ICCs of 0.91–0.97 but a glucose bias of +6.19 mg/dL with limits of agreement −67.69 to +80.07 mg/dL; and the 31,060-patient ED study found correlations of 0.59 (potassium) to 0.98 (haemoglobin) with no acceptable agreement limits for any parameter — the largest study is the most negative and prints at the largest weight.

Stamped: 2026-11-28.

Evidence published: 2018–2025.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In critically ill and emergency patients, how closely do point-of-care blood-gas analyser measurements of haemoglobin and electrolytes (sodium, potassium, chloride, bicarbonate, glucose) agree with central-laboratory autoanalyser measurements on paired samples?

Balance of the evidence

The composition is 4 primaries, no synthesis, 43,000+ patients/pairs in aggregate — counting units differ (pairs, patients, paired tests) and print per anchor, never merged. The two ED anchors and the two ICU anchors split by setting as well as by verdict. Sodium and potassium print in mmol/L, numerically equal to the mEq/L of US reporting; haemoglobin prints in g/L or g/dL exactly as each record gives it — no silent unit conversion. The homed pack #8517 asks venous-vs-arterial site agreement for blood-gas values — a different question, discharged at question level and cross-pointed; one anchor here (Mittal 2025) builds arterial-vs-venous sampling into its own design, and that confound prints on its card. The ICU multi-device report PMID 26126049 carries blood-gas limbs but is carded on the POC creatinine shelf (pack #8856) and counted nowhere here. Two records print no MeSH (Prakash 2018; Mittal 2025). No record carries a registry ID.

MeSH: Automation · Blood Gas Analysis*/standards · Hemoglobins*/analysis · Humans · Point-of-Care Systems · Potassium*/blood · Prospective Studies · Sodium*/blood · Western Australia · Adult · Aged · Autoanalysis/instrumentation,standards*,statistics & numerical data · Blood Gas Analysis/instrumentation,standards*,statistics & numerical data · Female · Glucose/analysis · Hematocrit/instrumentation,standards · Hemoglobins/analysis · Male · Middle Aged · Point-of-Care Systems/standards,statistics & numerical data · Potassium/analysis · Retrospective Studies · Sodium/analysis

Cards (5)

Card 1 — Prakash ICU paired-results study 2018 (Journal of Clinical Laboratory Analysis — retrospective study, 9,398 matched blood-gas/laboratory pairs from 1,765 critically ill patients (median sampling interval 5 minutes)): concordance correlation coefficients 0.89 (chloride) to 0.98 (haemoglobin), largest bias for haemoglobin, largest limits relative to bias for sodium; readings outside US-CLIA acceptable variation: glucose 7.1%, haemoglobin 5.9%, chloride 5.2%, sodium 3.4%; the authors conclude moderate-to-substantial concordance but interchangeability only for potassium — the record prints no MeSH ('none in the record')

Study: Retrospective paired-results agreement study (primary report)

Population: 9,398 matched pairs of blood-gas analyser and laboratory autoanalyser results obtained from 1,765 critically ill patients, assessing concentration pairs for hemoglobin, glucose, sodium, potassium, chloride, and bicarbonate, with the median time interval between sampling for the two methods 5 minutes; agreement assessed by concordance correlation coefficient, Bland-Altman analysis and the US-CLIA 88 acceptable-variation rules.

Summary: Prakash ICU paired-results study 2018 (Journal of Clinical Laboratory Analysis — retrospective study, 9,398 matched blood-gas/laboratory pairs from 1,765 critically ill patients (median sampling interval 5 minutes)): concordance correlation coefficients 0.89 (chloride) to 0.98 (haemoglobin), largest bias for haemoglobin, largest limits relative to bias for sodium; readings outside US-CLIA acceptable variation: glucose 7.1%, haemoglobin 5.9%, chloride 5.2%, sodium 3.4%; the authors conclude moderate-to-substantial concordance but interchangeability only for potassium — the record prints no MeSH ('none in the record')

Result: The CCC values ranged from 0.89 (95% CI 0.89–0.90) for chloride to 0.98 (95% CI 0.98–0.99) for hemoglobin. The largest bias was for hemoglobin. The limits of agreement relative to bias were largest for sodium, with 3.4% of readings outside the US-CLIA variation rule. The number of readings outside the US-CLIA acceptable variation was highest for glucose (7.1%) followed by hemoglobin (5.9%) and chloride (5.2%). The authors conclude there is moderate to substantial concordance between the machines, however the two test methods cannot be used interchangeably, except for potassium.

Watch out: Pairs (9,398) outnumber patients (1,765) fivefold — repeated pairs within patients concentrate the evidence, and a 5-minute median sampling gap lets genuine physiology drift into the 'disagreement'. The record prints no MeSH and, in its abstract, no per-analyte bias figures beyond the ordering statements, so none may be quoted. 'Interchangeable' here is judged against US-CLIA variation rules — a regulatory criterion, stated as such.

Source: Prakash S et al. Concordance between point-of-care blood gas analysis and laboratory autoanalyzer in measurement of hemoglobin and electrolytes in critically ill patients. Journal of clinical laboratory analysis. 2018. PMID: 29500827. Registry ID: none in the record. MeSH: none in the record

Card 2 — Triplett five-ICU prospective study 2019 (Anaesthesia and Intensive Care — 5 Western Australian ICUs, 219 paired tests for haemoglobin/sodium and 215 for potassium within 1 h in the first 24 h of admission): haemoglobin mean difference −0.35 g/L (95% CI −1.20 to +0.51, p = 0.425); sodium +1.49 mmol/L (1.23–1.76) and potassium +0.19 mmol/L (0.15–0.24), both p < 0.0001 but small and magnitude-independent on Bland-Altman; the authors conclude routine concurrent laboratory testing is unwarranted when values are near the normal range

Study: Prospective multicentre agreement study (primary report)

Population: Patients in Western Australia's five intensive care units who had a blood gas, a full blood count, and urea/electrolytes/creatinine performed, each within 1 h of the others, during the first 24 h of ICU admission: paired tests numbered 219 for haemoglobin and for sodium, and 215 for potassium, comparing laboratory haematology and biochemistry auto-analysers with blood gas analysers.

Summary: Triplett five-ICU prospective study 2019 (Anaesthesia and Intensive Care — 5 Western Australian ICUs, 219 paired tests for haemoglobin/sodium and 215 for potassium within 1 h in the first 24 h of admission): haemoglobin mean difference −0.35 g/L (95% CI −1.20 to +0.51, p = 0.425); sodium +1.49 mmol/L (1.23–1.76) and potassium +0.19 mmol/L (0.15–0.24), both p < 0.0001 but small and magnitude-independent on Bland-Altman; the authors conclude routine concurrent laboratory testing is unwarranted when values are near the normal range

Result: For haemoglobin, the laboratory and blood gas auto-analysers' results did not differ statistically significantly (mean difference −0.35 g/L, 95% confidence interval (CI) −1.20 to 0.51, P = 0.425). For sodium and potassium, by contrast, the mean differences between the two methods did reach statistical significance (sodium: mean difference 1.49 mmol/L, 95% CI 1.23–1.76, P < 0.0001; potassium: mean difference 0.19 mmol/L, 95% CI 0.15–0.24, P < 0.0001), yet on the Bland-Altman plots the mean biases were small, and did not depend on the magnitude of the measurements. The authors conclude the two methods agreed under most clinical situations when values were within or close to the normal range, suggesting routine concurrent testing is unwarranted.

Watch out: The comfort is explicitly range-bound: 'within or close to normal range' — the study does not clear blood-gas values at the extremes where ICU decisions concentrate, and its sampling window (within 1 h, not simultaneous) is the loosest pairing on this shelf. Statistically significant biases (sodium, potassium) are clinically small; the record prints no limits of agreement in its abstract, so the spread around those biases cannot be quoted from this record.

Source: Triplett KE et al. Can the blood gas analyser results be believed? A prospective multicentre study comparing haemoglobin, sodium and potassium measurements by blood gas analysers and laboratory auto-analysers. Anaesthesia and intensive care. 2019. PMID: 31070468. Registry ID: none in the record. MeSH: Automation · Blood Gas Analysis*/standards · Hemoglobins*/analysis · Humans · Point-of-Care Systems · Potassium*/blood · Prospective Studies · Sodium*/blood · Western Australia

Card 3 — Mittal ED cross-sectional study 2025 (Indian Journal of Critical Care Medicine — prospective study, 1,000 adults with simultaneous arterial (blood-gas) and venous (laboratory) sampling): ICCs: sodium 0.946, potassium 0.907, haemoglobin 0.947, glucose 0.967 (consistency statistics, printed as such); despite the glucose ICC, its Bland-Altman bias was +6.19 mg/dL with limits of agreement −67.69 to +80.07 mg/dL — poor agreement; the authors advocate blood-gas sodium/potassium/haemoglobin for quick decisions but laboratory confirmation for glucose — arterial-vs-venous sampling is built into this anchor's design and prints as a confound: part of its disagreement is sample site, not device (the site-agreement question itself is homed at pack #8517); the record prints no MeSH ('none in the record')

Study: Prospective cross-sectional agreement study (primary report)

Population: 1,000 adult patients in an Emergency Medicine Department; their samples were taken at the same time: an arterial sample for blood gas analysis, a venous sample for laboratory analysis. For sodium, potassium, hemoglobin, and glucose, intraclass correlation coefficient evaluated reliability and Bland-Altman analysis, with 95% limits of agreement, evaluated agreement.

Summary: Mittal ED cross-sectional study 2025 (Indian Journal of Critical Care Medicine — prospective study, 1,000 adults with simultaneous arterial (blood-gas) and venous (laboratory) sampling): ICCs: sodium 0.946, potassium 0.907, haemoglobin 0.947, glucose 0.967 (consistency statistics, printed as such); despite the glucose ICC, its Bland-Altman bias was +6.19 mg/dL with limits of agreement −67.69 to +80.07 mg/dL — poor agreement; the authors advocate blood-gas sodium/potassium/haemoglobin for quick decisions but laboratory confirmation for glucose — arterial-vs-venous sampling is built into this anchor's design and prints as a confound: part of its disagreement is sample site, not device (the site-agreement question itself is homed at pack #8517); the record prints no MeSH ('none in the record')

Result: Reliability by intraclass correlation coefficient was good to excellent across all parameters: glucose (ICC = 0.967, 95% CI: 0.962–0.971), Hb (ICC = 0.947, 95% CI: 0.938–0.953), sodium (ICC = 0.946, 95% CI: 0.790–0.976), and potassium (ICC = 0.907, 95% CI: 0.769–0.951). On the Bland-Altman plot, agreement for sodium, potassium, and Hb levels was moderate to high, and mean bias stayed within acceptable limits. Glucose told a different story: despite an excellent ICC and high correlation, its Bland-Altman analysis showed bias that was substantial (+6.19 mg/dL) and LOA that were very wide (−67.69 to +80.07 mg/dL), indicating poor agreement. Quick decision-making, the authors advocate, can rest on blood-gas sodium, potassium and Hb, but glucose needs laboratory confirmation.

Watch out: Design confound printed at full weight: the blood-gas sample is arterial and the laboratory sample is venous, so part of every difference here is sample site, not analyser — glucose especially differs by site. This card therefore cannot be placed beside the same-sample anchors as a pure device verdict. The glucose ICC of 0.967 beside limits spanning 148 mg/dL is this shelf's cleanest demonstration that consistency statistics flatter. The record prints no MeSH.

Source: Mittal G et al. Comparison of Sodium, Potassium, Hemoglobin, and Glucose Levels by Blood Gas Analyzer and Hospital Laboratory Autoanalyzer in Emergency Department Settings: A Cross-sectional Study. Indian journal of critical care medicine : peer-reviewed, official publication of Indian Society of Critical Care Medicine. 2025. PMID: 41509697. Registry ID: none in the record. MeSH: none in the record

Card 4 — Altunok ED retrospective study 2019 (American Journal of Emergency Medicine — single-centre study, 31,060 ED patients): blood-gas vs laboratory correlations: sodium 0.725, potassium 0.593, haemoglobin 0.982, haematocrit 0.958, glucose 0.984 — correlations printed as correlations — but no parameter had acceptable agreement limits, unchanged across acidosis/normal/alkalosis pH strata; the authors conclude clinicians should be aware of the limitations of blood-gas analyser results — the row's largest and most negative anchor, printed at that weight

Study: Retrospective agreement study (primary report; largest and most negative anchor)

Population: 31,060 patients admitted to the emergency department of a tertiary care teaching hospital between May 2014 and December 2017, after exclusions (under 18 years, lacking data, any treatment before the laboratory tests, venous gas results reported more than 30 minutes after sampling, tests performed at different times, and recurrent results from a single patient).

Summary: Altunok ED retrospective study 2019 (American Journal of Emergency Medicine — single-centre study, 31,060 ED patients): blood-gas vs laboratory correlations: sodium 0.725, potassium 0.593, haemoglobin 0.982, haematocrit 0.958, glucose 0.984 — correlations printed as correlations — but no parameter had acceptable agreement limits, unchanged across acidosis/normal/alkalosis pH strata; the authors conclude clinicians should be aware of the limitations of blood-gas analyser results — the row's largest and most negative anchor, printed at that weight

Result: The correlation coefficients for sodium, potassium, hemoglobin, hematocrit, and glucose levels measured by a blood gas analyzer and a laboratory autoanalyzer were 0.725, 0.593, 0.982, 0.958, and 0.984, respectively; however, there were no good, acceptable agreement limits for any of the parameters. In addition, these results did not change according to the different pH stages (acidosis, normal pH and alkalosis). The authors conclude the two types of measurements showed moderate correlation for sodium and potassium and strong correlation for glucose, hemoglobin and hematocrit, but none of the levels had acceptable agreement limits.

Watch out: The record's abstract does not print the numeric limits of agreement or the acceptability criterion's source, so 'no acceptable limits for any parameter' is quoted as the authors' verdict without the underlying numbers — the verdict is unambiguous, its magnitude is not checkable from this record. Retrospective record pairs (sampling within 30 minutes) allow physiological drift. Potassium's correlation (0.593) is the shelf's weakest — notable against the ICU study's potassium-only interchangeability verdict, which rests on a different criterion (CLIA variation) in a different setting.

Source: Altunok İ et al. Correlation between sodium, potassium, hemoglobin, hematocrit, and glucose values as measured by a laboratory autoanalyzer and a blood gas analyzer. The American journal of emergency medicine. 2019. PMID: 30131206. Registry ID: none in the record. MeSH: Adult · Aged · Autoanalysis/instrumentation,standards*,statistics & numerical data · Blood Gas Analysis/instrumentation,standards*,statistics & numerical data · Female · Glucose/analysis · Hematocrit/instrumentation,standards · Hemoglobins/analysis · Humans · Male · Middle Aged · Point-of-Care Systems/standards,statistics & numerical data · Potassium/analysis · Retrospective Studies · Sodium/analysis

Card 5 — Synthesis — 4 primaries, no synthesis: correlation flatters and limits of agreement decide, in every anchor; the largest study refuses interchangeability for everything, the ICU studies refuse it for almost everything

Study: Synthesis

Population: 9,398 pairs from 1,765 ICU patients; 219/215 paired tests across 5 ICUs; 1,000 ED patients; 31,060 ED patients. Pairs, tests and patients are different counting units and print per card.

Summary: Synthesis — 4 primaries, no synthesis: correlation flatters and limits of agreement decide, in every anchor; the largest study refuses interchangeability for everything, the ICU studies refuse it for almost everything

Result: Concordance/correlation layer: CCC 0.89–0.98 (ICU pairs); ICC 0.907–0.967 (ED, site-confounded); correlations 0.593–0.984 (31,060 ED patients). Bias layer: haemoglobin −0.35 g/L (NS), sodium +1.49 mmol/L, potassium +0.19 mmol/L (five-ICU); glucose +6.19 mg/dL (ED). Verdict layer: interchangeable only for potassium (CLIA criterion, ICU); comfortable near the normal range (five-ICU); no acceptable agreement limits for any parameter (largest study). Setting splits the verdicts (ICU more favourable than ED) as well as the criteria (CLIA variation vs authors' acceptability limits). No report was refused; no slate-vs-record discrepancy was found at build.

Watch out: A haemoglobin correlation of 0.982 coexisting with unacceptable agreement limits is the shelf in one line: these analysers rank patients well and price individual values loosely. Treatment thresholds for potassium, sodium and haemoglobin are narrow; the biases are small enough for screening and the limits too wide — on the largest evidence — for unconfirmed dosing decisions at extremes.

Source: Synthesis of the reports carded above; no additional source.