SourceMed · Shelf #8904 in the ask interface
Wrong cuff size, wrong blood pressure — 1 randomised crossover trial (N = 195): a regular cuff on an extra-large arm overread systolic pressure by 19.5 mm Hg, and on a small arm underread it by 3.6 mm Hg
Shelf #8904 · Heart, vessels, blood pressure & clotting
POSSIBLE (benefit) — a single randomised crossover trial (N = 195): using the appropriately sized cuff avoids large, direction-predictable errors from a one-size regular cuff (systolic −3.6 mm Hg in small arms to +19.5 mm Hg in extra-large arms). Single-trial shelf — Tier C
Stamped: 2026-11-28.
Evidence published: 2023.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In adults having blood pressure measured with an automated device, how much error does using a regular-sized cuff instead of an appropriately sized cuff introduce, across arm sizes?
Balance of the evidence
POSSIBLE (benefit) — a single randomised crossover trial (N = 195): using the appropriately sized cuff avoids large, direction-predictable errors from a one-size regular cuff (systolic −3.6 mm Hg in small arms to +19.5 mm Hg in extra-large arms). Single-trial shelf —. In the Cuff(SZ) trial, among people requiring a small cuff, a regular cuff read systolic pressure 3.6 mm Hg lower (mean difference −3.6, 95% CI −5.6 to −1.7); among people requiring a large cuff it read 4.8 mm Hg higher (95% CI 3.0 to 6.6), and among people requiring an extra-large cuff 19.5 mm Hg higher (95% CI 16.1 to 22.9). Errors grew with each step of overcuffing or undercuffing and were consistent across strata of systolic pressure and body mass index. POSSIBLE (benefit): the direction and gradient are internally coherent, but one trial in one city's recruitment settings carries the shelf.
MeSH: Adult · Humans · Male · Middle Aged · Female · Blood Pressure/physiology · Cross-Over Studies · Blood Pressure Determination/methods · Hypertension/diagnosis* · Hypotension*
Cards (2)
Card 1 — Cuff(SZ) randomised crossover trial: miscuffing errors are large and arm-size-dependent (Ishigami 2023, JAMA internal medicine — 195 adults): regular cuff systolic differences −3.6 (small arm), +4.8 (large), +19.5 mm Hg (extra-large)
Study: Randomized controlled trial (crossover)
Population: 195 community-dwelling adults in Baltimore, Maryland (mean age 54 years; 34% male; 51% with hypertension), recruited 2021 via screening events, mailings and clinic referrals; each participant had 4 sets of triplicate automated BP measurements — appropriate, too-small or too-large cuffs in random order, then a final set with the appropriate cuff.
Summary: Cuff(SZ) randomised crossover trial: miscuffing errors are large and arm-size-dependent (Ishigami 2023, JAMA internal medicine — 195 adults): regular cuff systolic differences −3.6 (small arm), +4.8 (large), +19.5 mm Hg (extra-large)
Result: Among individuals requiring a small BP cuff, use of a regular cuff resulted in a lower reading (mean systolic BP difference −3.6 [95% CI −5.6 to −1.7] mm Hg). Among individuals requiring a large or extra-large cuff, a regular cuff resulted in higher readings (mean systolic differences 4.8 [95% CI 3.0–6.6] mm Hg and 19.5 [95% CI 16.1–22.9] mm Hg, respectively). For the secondary outcome, BP differences with overcuffing and undercuffing by 1 and 2 cuff sizes were greater among those requiring larger cuffs; results were consistent in analyses stratified by systolic BP (≥130 vs <130 mm Hg) and body mass index (≥30 vs <30).
Watch out: One trial, one automated device family, one city. The 19.5 mm Hg figure is for the extreme mismatch (regular cuff on an extra-large arm) — most real-world miscuffing is one size off, with smaller errors. The trial measures reading error, not downstream misdiagnosis or treatment outcomes.
Source: Ishigami J et al. Effects of Cuff Size on the Accuracy of Blood Pressure Readings: The Cuff(SZ) Randomized Crossover Trial. JAMA internal medicine. 2023. PMID: 37548984. Registry ID: NCT04610775. MeSH: Adult · Humans · Male · Middle Aged · Female · Blood Pressure/physiology · Cross-Over Studies · Blood Pressure Determination/methods · Hypertension/diagnosis* · Hypotension*
Card 2 — Synthesis — POSSIBLE (benefit): right-size the cuff; the error is mechanical, not subtle
Study: Synthesis
Population: 1 randomised crossover trial, N = 195.
Summary: Synthesis — POSSIBLE (benefit): right-size the cuff; the error is mechanical, not subtle
Result: A regular cuff systematically underreads small arms and overreads large arms, by up to 19.5 mm Hg systolic at the extreme. Because the error is produced by the measurement itself and reverses with the correct cuff in the same sitting, the finding is credible despite resting on a single trial — but it is a single trial, and the shelf is Tier C for that reason, not because the data look weak.
Watch out: No second randomised trial of automated-device miscuffing was found in the wider search; older auscultatory-era studies were not carded because they test a different measurement method.