End-tidal CO₂ as an estimate of arterial PCO₂ — 4 agreement studies (total N = 518): mainstream sampling under anaesthesia held the arterial-to-end-tidal gradient at 0.64 kPa (≈4.8 mmHg); in ED dyspnoea, mainstream bias was 0.5 mmHg with limits of agreement −10.5 to +9.5 mmHg, while microstream sidestream sampling showed a mean difference of 8 mmHg with limits of −10 to +26 mmHg and 38% of readings ≥10 mmHg out — correlations of 0.79–0.91 throughout

Shelf #8940 · Emergency & first hours

MEASUREMENT ACCURACY — method-dependent, and correlation flatters it: correlations with arterial PCO₂ run 0.79–0.91 across all four studies, but agreement splits by sampling method — mainstream under anaesthesia: gradient 0.64 (0.16) kPa (≈4.8 mmHg), limits of agreement 0.32–0.96 kPa; mainstream in ED dyspnoea (n = 162): bias 0.5 ± 5 mmHg, limits −10.5 to +9.5 mmHg, 80% of readings within ±5 mmHg; microstream sidestream in ED dyspnoea (n = 43): mean difference 8 mmHg, limits −10 to +26 mmHg, 38% of readings differing by ≥10 mmHg. End-tidal CO₂ tracks arterial PCO₂; at sidestream precision it does not replace it

Stamped: 2026-11-28.

Evidence published: 2003–2012.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In non-intubated emergency-department patients with dyspnoea, and in ventilated anaesthetised patients, how closely does end-tidal carbon dioxide (mainstream, sidestream and microstream capnography) estimate arterial PCO₂ — what are the bias and limits of agreement, and how much does the sampling method matter?

Balance of the evidence

MEASUREMENT ACCURACY — composition: 4 primary agreement studies, total N = 518 patients (73 + 43 + 162 + 240; the anaesthesia study contributes 1,007 paired measurements). One study (Chan 2003) carries a TYPE-FLAG — typed Randomized Controlled Trial in its record for a randomisation between measurement methods; counted on substance, printed on its card. This is a caution-class shelf: agreement with a reference measurement only; no diagnostic or outcome claim is made.

MeSH: Adolescent · Adult · Age Factors · Aged · Aged, 80 and over · Body Temperature · Breath Tests/methods · Capnography · Carbon Dioxide/analysis/blood · Emergencies · Emergency Service, Hospital · Female · Humans · Lung Diseases/diagnosis · Male · Middle Aged · Monitoring, Physiologic/methods · Partial Pressure · Prospective Studies · Reproducibility of Results · Single-Blind Method · Blood Gas Analysis · Capnography/instrumentation · Dyspnea/blood/etiology · Feasibility Studies · Hypercapnia/blood/diagnosis/etiology · Lung Diseases, Obstructive/blood/complications/diagnosis · Tidal Volume · Acute Disease · Capnography/methods · Carbon Dioxide/blood · Dyspnea/blood/physiopathology · Predictive Value of Tests · Craniotomy · Linear Models · Monitoring, Intraoperative/methods · Supine Position

Cards (5)

Card 1 — Yosefy sidestream study 2004 (Emergency Medicine Journal — 73 ED patients in respiratory distress, sidestream end-tidal CO₂ vs arterial PCO₂): correlation r = 0.792 for the whole group; age under 50 deflected the correlation curve left (a wider arterial-to-end-tidal gradient) and temperature above 37.6 °C deflected it right

Study: Agreement study (primary report)

Population: 73 patients (47 men, 26 women) referred to the emergency department for respiratory distress, with simultaneous arterial blood gas pressures and sidestream end-tidal CO₂ measurements, plus age, temperature, respiratory rate, blood pressure, pulse rate and diagnosis recorded.

Summary: Yosefy sidestream study 2004 (Emergency Medicine Journal — 73 ED patients in respiratory distress, sidestream end-tidal CO₂ vs arterial PCO₂): correlation r = 0.792 for the whole group; age under 50 deflected the correlation curve left (a wider arterial-to-end-tidal gradient) and temperature above 37.6 °C deflected it right

Result: A significant correlation was found between sidestream end-tidal CO₂ and arterial PCO₂ (r = 0.792). Compared with the whole-group correlation curve, age under 50 years deflected the curve to the left, while temperature above 37.6 °C deflected it to the right; the remaining parameters had no clear influence. The authors conclude there is a good correlation in the ED setting, with young age possibly increasing and raised temperature possibly decreasing the arterial-to-end-tidal gradient.

Watch out: Correlation only — this study prints no bias or limits of agreement, and r = 0.792 is compatible with clinically large individual errors (as the next card shows). Carded as the sidestream limb and for its age/temperature modifiers, never as agreement evidence.

Source: Yosefy C et al. End tidal carbon dioxide as a predictor of the arterial PCO2 in the emergency department setting. Emergency medicine journal : EMJ. 2004. PMID: 15333528. Registry ID: none in the record. MeSH: Adolescent · Adult · Age Factors · Aged · Aged, 80 and over · Body Temperature · Breath Tests/methods · Capnography · Carbon Dioxide/analysis/blood · Emergencies · Emergency Service, Hospital · Female · Humans · Lung Diseases/diagnosis · Male · Middle Aged · Monitoring, Physiologic/methods · Partial Pressure · Prospective Studies · Reproducibility of Results · Single-Blind Method

Card 2 — Delerme microstream study 2010 (American Journal of Emergency Medicine — 43 ED patients with acute dyspnoea, 48 paired measurements, microstream capnography vs arterial blood gas): correlation was good (R = 0.82) but the mean arterial-to-end-tidal difference was 8 mmHg, limits of agreement ran −10 to +26 mmHg, and 38% of measurements differed by ≥10 mmHg

Study: Agreement study (primary report)

Population: 43 non-intubated adults (48 measurements) admitted to a teaching-hospital ED with acute dyspnoea who required arterial blood gases — mainly pneumonia (n = 12), acute cardiac failure (n = 8), asthma (n = 7) and COPD exacerbation (n = 6); end-tidal CO₂ by a microstream capnometer device for spontaneously breathing patients.

Summary: Delerme microstream study 2010 (American Journal of Emergency Medicine — 43 ED patients with acute dyspnoea, 48 paired measurements, microstream capnography vs arterial blood gas): correlation was good (R = 0.82) but the mean arterial-to-end-tidal difference was 8 mmHg, limits of agreement ran −10 to +26 mmHg, and 38% of measurements differed by ≥10 mmHg

Result: Correlation between end-tidal CO₂ and PaCO₂ was good (R = 0.82). However, 18 measurements (38%) had a difference between end-tidal CO₂ and PaCO₂ of 10 mmHg or more. The mean difference between PaCO₂ and end-tidal CO₂ was 8 mmHg, and Bland–Altman limits of agreement were −10 to +26 mmHg. The authors conclude end-tidal CO₂ by the microstream method does not seem to accurately predict PaCO₂ in ED patients with acute dyspnoea.

Watch out: The shelf's cautionary centre: a 'good correlation' headline sitting on limits of agreement 36 mmHg wide. Forty-three patients at one centre, unblinded — small, but its agreement statistics are the honest kind (bias and limits, not correlation alone).

Source: Delerme S et al. Concordance between capnography and capnia in adults admitted for acute dyspnea in an ED. The American journal of emergency medicine. 2010. PMID: 20637388. Registry ID: none in the record. MeSH: Adult · Aged · Blood Gas Analysis · Capnography/instrumentation · Dyspnea/blood/etiology · Emergency Service, Hospital · Feasibility Studies · Female · Humans · Hypercapnia/blood/diagnosis/etiology · Lung Diseases, Obstructive/blood/complications/diagnosis · Male · Middle Aged · Prospective Studies · Tidal Volume

Card 3 — Cinar mainstream study 2012 (American Journal of Emergency Medicine — 162 ED patients with acute dyspnoea, mainstream capnometry vs arterial blood gas): correlation r = 0.911, mean bias 0.5 ± 5 mmHg, limits of agreement −10.5 to +9.5 mmHg, and 80% of end-tidal values fell within ±5 mmHg of arterial

Study: Agreement study (primary report)

Population: 162 non-intubated adults presenting to a university-hospital ED (over 110,000 attendances/year) with acute dyspnoea requiring arterial blood gas analysis over 6 months in 2010, breathing through an airway adapter attached to a mainstream capnometer, with simultaneous arterial sampling.

Summary: Cinar mainstream study 2012 (American Journal of Emergency Medicine — 162 ED patients with acute dyspnoea, mainstream capnometry vs arterial blood gas): correlation r = 0.911, mean bias 0.5 ± 5 mmHg, limits of agreement −10.5 to +9.5 mmHg, and 80% of end-tidal values fell within ±5 mmHg of arterial

Result: Mean end-tidal CO₂ was 39.47 ± 10.84 mmHg (range 19–82) and mean PaCO₂ 38.95 ± 12.27 mmHg (range 16–94). Correlation was positive, strong and significant (r = 0.911, P < .001). Bland–Altman mean bias was 0.5 ± 5 mmHg, with limits of agreement of −10.5 and +9.5 mmHg. Eighty percent (129) of end-tidal measurements were within ±5 mmHg of arterial. The authors conclude mainstream end-tidal measurement accurately predicts arterial PaCO₂ in ED acute dyspnoea.

Watch out: Same clinical question as Delerme, opposite conclusion — the difference is the sampling method (mainstream at the airway vs microstream sidestream) and possibly case mix. Even on this favourable reading, 1 in 5 values missed arterial by more than 5 mmHg and the limits span 20 mmHg: 'accurately predicts' is the authors' phrase, printed with its own numbers beside it.

Source: Cinar O et al. Can mainstream end-tidal carbon dioxide measurement accurately predict the arterial carbon dioxide level of patients with acute dyspnea in ED. The American journal of emergency medicine. 2012. PMID: 21277140. Registry ID: none in the record. MeSH: Acute Disease · Capnography/methods · Carbon Dioxide/blood · Dyspnea/blood/physiopathology · Emergency Service, Hospital · Female · Humans · Male · Middle Aged · Predictive Value of Tests · Prospective Studies · Tidal Volume

Card 4 — Chan mainstream-vs-sidestream study 2003 (Anaesthesia — 240 anaesthetised neurosurgical patients, 1,007 paired measurements; record typed as a randomised trial of measurement method — TYPE-FLAG): the arterial-to-end-tidal gradient was 0.64 (0.16) kPa (≈4.8 mmHg) with mainstream sampling versus 0.99 (0.40) kPa (≈7.4 mmHg) with sidestream, and limits of agreement were narrower for mainstream (0.32–0.96 kPa ≈ 2.4–7.2 mmHg) than sidestream (0.19–1.79 kPa ≈ 1.4–13.4 mmHg)

Study: Method-comparison study (primary report) — TYPE-FLAG: the record is typed Randomized Controlled Trial for a randomisation between measurement methods; counted on substance

Population: 240 adult neurosurgical patients undergoing supine craniotomy under propofol anaesthesia, randomly assigned to end-tidal measurement by mainstream (503 paired measurements) or sidestream (504 paired measurements) infrared capnometry, with arterial blood gases after induction, before dural incision, during surgery and before wound closure (1,007 pairs total).

Summary: Chan mainstream-vs-sidestream study 2003 (Anaesthesia — 240 anaesthetised neurosurgical patients, 1,007 paired measurements; record typed as a randomised trial of measurement method — TYPE-FLAG): the arterial-to-end-tidal gradient was 0.64 (0.16) kPa (≈4.8 mmHg) with mainstream sampling versus 0.99 (0.40) kPa (≈7.4 mmHg) with sidestream, and limits of agreement were narrower for mainstream (0.32–0.96 kPa ≈ 2.4–7.2 mmHg) than sidestream (0.19–1.79 kPa ≈ 1.4–13.4 mmHg)

Result: For the 1,007 paired measurements, the mean (SD) arterial-to-end-tidal difference was 0.64 (0.16) kPa for mainstream versus 0.99 (0.40) kPa for sidestream (p < 0.001) — approximately 4.8 mmHg versus 7.4 mmHg at 7.5 mmHg per kPa. Limits of agreement were 0.32–0.96 kPa (mainstream) versus 0.19–1.79 kPa (sidestream) (p < 0.001) — approximately 2.4–7.2 versus 1.4–13.4 mmHg. The gradient did not change with time in either group, but within-patient variation was greater with sidestream. The authors conclude mainstream end-tidal CO₂ estimates arterial PCO₂ more accurately than sidestream.

Watch out: Ventilated, anaesthetised, supine surgical patients — the controlled end of the spectrum, where end-tidal sampling is at its best. Its gradient (~5–7 mmHg) is the floor the ED studies should be judged against, not a figure transferable to a spontaneously breathing dyspnoeic patient.

Source: Chan KL et al. Mainstream vs. sidestream capnometry for prediction of arterial carbon dioxide tension during supine craniotomy. Anaesthesia. 2003. PMID: 12562411. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Capnography/methods · Carbon Dioxide/blood · Craniotomy · Female · Humans · Linear Models · Male · Middle Aged · Monitoring, Intraoperative/methods · Partial Pressure · Reproducibility of Results · Supine Position

Card 5 — Synthesis — end-tidal CO₂ vs arterial PCO₂: the gradient is real and method-dependent — about 5 mmHg with mainstream sampling under anaesthesia, near-zero mean bias but ±10 mmHg limits with mainstream sampling in ED dyspnoea, and an 8 mmHg mean difference with limits of −10 to +26 mmHg with microstream sidestream sampling; correlation coefficients (0.79–0.91) flatter every method

Study: Synthesis

Population: 4 agreement studies, total N = 518 patients (plus 1,007 paired measurements in the anaesthesia study): Yosefy 2004 (n = 73), Delerme 2010 (n = 43 patients, 48 measurements), Cinar 2012 (n = 162), Chan 2003 (n = 240).

Summary: Synthesis — end-tidal CO₂ vs arterial PCO₂: the gradient is real and method-dependent — about 5 mmHg with mainstream sampling under anaesthesia, near-zero mean bias but ±10 mmHg limits with mainstream sampling in ED dyspnoea, and an 8 mmHg mean difference with limits of −10 to +26 mmHg with microstream sidestream sampling; correlation coefficients (0.79–0.91) flatter every method

Result: Read by agreement statistics rather than correlations, the base is coherent: mainstream sampling beats sidestream in the head-to-head study; the two ED mainstream/microstream studies bracket the spontaneously-breathing performance (bias 0.5 mmHg, limits −10.5 to +9.5, mainstream; mean difference 8 mmHg, limits −10 to +26, microstream); and the sidestream correlation study (r = 0.792) shows why correlation alone cannot settle the question. No report was refused.

Watch out: MEASUREMENT ACCURACY here is a property of the sampling method and the patient's physiology (ventilation–perfusion matching in dyspnoea widens the gradient), not of 'capnography' as a class — any single-number summary of end-tidal accuracy would misstate this base.

Source: Synthesis of the reports carded above; no additional source.