Peripheral thermometers in children — 3 meta-analyses (documented pooled N = 8,087, overlapping primary sets): infrared ear temperature reads 0.22 °C below rectal on average with 95% limits of agreement of −0.44 to +1.30 °C, yet as a fever screen peripheral thermometry pools at sensitivity 82% / specificity 88% with AUC 0.95, and new-generation tympanic devices peak at a 37.8 °C cut-off (AUC 0.97)
Shelf #8942 · Children’s health
MEASUREMENT ACCURACY — split by question: as temperature measurement, infrared ear readings in children average 0.22 °C below rectal with 95% limits of agreement of −0.44 to +1.30 °C (febrile subgroup: 0.15 °C, limits −0.32 to +1.10 °C) — too wide to substitute for rectal measurement; as fever screening, peripheral thermometry pools at sensitivity 82% / specificity 88% (labelled in words at first occurrence) with AUC 0.95 across 3,448 emergency/outpatient children, and new-generation tympanic devices perform best at a 37.8 °C cut-off (sROC AUC 0.97, Youden index 0.83, across 4,639 children; no pooled sensitivity/specificity pair is printed in that record)
Stamped: 2026-11-28.
Evidence published: 2014–2024.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In children, how accurately do peripheral thermometers (infrared tympanic/ear, and other peripheral devices used in outpatient and emergency settings) estimate body temperature against rectal or core references, and how accurately do they detect fever — what are the mean difference and limits of agreement, and the pooled sensitivity and specificity at the cut-offs used?
Balance of the evidence
MEASUREMENT ACCURACY — composition: 3 systematic reviews/meta-analyses whose primary-study sets overlap, counted as syntheses throughout (documented pooled N = 8,087 across the two that state one; the 2014 review's record states no pooled N and its card says so). This is a caution-class shelf. Cross-point in prose only: shelf #7781 homes the same question in adults against central thermometry — different population and reference standard; the paediatric question was unhomed.
MeSH: Child · Ear, External/physiology · Humans · Infrared Rays · Rectum/physiology · Thermometry/standards · Diagnostic Tests, Routine · Reproducibility of Results · Sensitivity and Specificity · Thermometers · Thermometry · Body Temperature · Emergency Service, Hospital · Fever/diagnosis · Pediatrics/methods/standards · Thermometers/standards
Cards (4)
Card 1 — Zhen ear-thermometry meta-analysis 2014 (Clinical Pediatrics — systematic review of infrared ear vs rectal thermometry in children; pooled study count and child N not stated in the record's abstract, printed as not documented): tympanic temperature was always lower than rectal, pooled mean difference 0.22 °C with 95% limits of agreement −0.44 to +1.30 °C; in febrile children the mean difference was 0.15 °C with limits −0.32 to +1.10 °C — the authors conclude ear thermometry cannot replace rectal thermometry in children
Study: Systematic review and meta-analysis (synthesis limb) — pooled N not documented in the record
Population: Children in the primary studies of a systematic review and meta-analysis (Medline, Ovid, Elsevier, Google Scholar, Cochrane Library; cross-sectional prospective designs, dual independent extraction); the record's abstract does not state the number of pooled studies or children — N not documented in the record for this limb.
Summary: Zhen ear-thermometry meta-analysis 2014 (Clinical Pediatrics — systematic review of infrared ear vs rectal thermometry in children; pooled study count and child N not stated in the record's abstract, printed as not documented): tympanic temperature was always lower than rectal, pooled mean difference 0.22 °C with 95% limits of agreement −0.44 to +1.30 °C; in febrile children the mean difference was 0.15 °C with limits −0.32 to +1.10 °C — the authors conclude ear thermometry cannot replace rectal thermometry in children
Result: Mean tympanic temperature was always lower than rectal temperature. The pooled (random-effects) mean difference between tympanic and rectal temperature was 0.22 °C (95% limits of agreement −0.44 °C to 1.30 °C), similar within rectal-device subgroups (mercury 0.21 °C, limits −0.44 to 1.27 °C; electronic 0.24 °C, limits −0.46 to 1.34 °C). In the febrile-children group, the pooled mean difference was 0.15 °C (limits −0.32 to 1.10 °C). The authors conclude the mean difference is large, the limits of agreement wide, and the accuracy of infrared ear thermometry in children poor.
Watch out: A synthesis limb, counted as a synthesis and never as a cohort: its primary studies overlap unknown amounts with the other two syntheses on this shelf. Rectal temperature is itself a peripheral reference, not core temperature — agreement with rectal is the question here, exactly as the adult shelf (#7781) judges agreement with central references.
Source: Zhen C et al. Accuracy of infrared ear thermometry in children: a meta-analysis and systematic review. Clinical pediatrics. 2014. PMID: 24879119. Registry ID: none in the record. MeSH: Child · Ear, External/physiology · Humans · Infrared Rays · Rectum/physiology · Thermometry/standards
Card 2 — Shi new-generation tympanic meta-analysis 2020 (BMC Pediatrics — 12 diagnostic studies, 4,639 children, QUADAS-2 assessed): across fever cut-offs from 37.0 to 38.0 °C, the 37.8 °C cut-off performed best (sROC AUC 0.97, Youden index 0.83) — the record prints no pooled sensitivity/specificity pair at that cut-off, and none is quoted; the authors conclude new-generation tympanic thermometry has high diagnostic accuracy for fever screening
Study: Systematic review and meta-analysis (synthesis limb)
Population: 12 diagnostic studies involving 4,639 paediatric patients, identified by systematic search of PubMed, Web of Science and Embase, quality-assessed with QUADAS-2, and synthesised for the diagnostic accuracy of new-generation infrared tympanic thermometers at cut-offs from 37.0 °C to 38.0 °C.
Summary: Shi new-generation tympanic meta-analysis 2020 (BMC Pediatrics — 12 diagnostic studies, 4,639 children, QUADAS-2 assessed): across fever cut-offs from 37.0 to 38.0 °C, the 37.8 °C cut-off performed best (sROC AUC 0.97, Youden index 0.83) — the record prints no pooled sensitivity/specificity pair at that cut-off, and none is quoted; the authors conclude new-generation tympanic thermometry has high diagnostic accuracy for fever screening
Result: The cut-off of 37.8 °C had the highest sROC area under the curve (0.97) and Youden index (0.83) and was deemed the optimal cut-off. The authors conclude that new-generation tympanic thermometry is of high diagnostic accuracy in paediatric patients and can be an alternative for fever screening in children. The record's abstract does not print the pooled sensitivity and specificity at 37.8 °C, so this shelf quotes the AUC and Youden index only.
Watch out: 'New generation' is the primary studies' device label, not a standard — and an AUC of 0.97 for fever detection coexists with the wide temperature limits of agreement on the Zhen card: a thermometer can rank febrile children well while still reading individual temperatures ±1 °C off. Screening accuracy and measurement agreement are different claims, and this shelf keeps them separate.
Source: Shi D et al. Diagnostic test accuracy of new generation tympanic thermometry in children under different cutoffs: a systematic review and meta-analysis. BMC pediatrics. 2020. PMID: 32398036. Registry ID: none in the record. MeSH: Child · Diagnostic Tests, Routine · Humans · Reproducibility of Results · Sensitivity and Specificity · Thermometers · Thermometry
Card 3 — Yuan paediatric emergency meta-analysis 2024 (International Emergency Nursing — 10 studies, 3,448 children in outpatient and emergency settings, searched to November 2023): pooled fever-detection sensitivity 82% / specificity 88% (labelled in words at first occurrence: sensitivity 82%, specificity 88%), positive likelihood ratio 11.35, negative likelihood ratio 0.20, diagnostic odds ratio 58.02, AUC 0.95 — with the infrared tympanic thermometer the most sensitive of the peripheral methods pooled
Study: Systematic review and meta-analysis (synthesis limb)
Population: 10 studies involving 3,448 children on the accuracy of peripheral thermometers (including infrared tympanic membrane devices) in outpatient and emergency-department settings, from nine databases searched to 20 November 2023, synthesised by diagnostic meta-analysis.
Summary: Yuan paediatric emergency meta-analysis 2024 (International Emergency Nursing — 10 studies, 3,448 children in outpatient and emergency settings, searched to November 2023): pooled fever-detection sensitivity 82% / specificity 88% (labelled in words at first occurrence: sensitivity 82%, specificity 88%), positive likelihood ratio 11.35, negative likelihood ratio 0.20, diagnostic odds ratio 58.02, AUC 0.95 — with the infrared tympanic thermometer the most sensitive of the peripheral methods pooled
Result: The combined sensitivity was 82% (95% CI 81% to 84%), the combined specificity 88% (95% CI 87% to 89%), the combined positive likelihood ratio 11.35 (95% CI 6.07 to 21.20), the combined negative likelihood ratio 0.20 (95% CI 0.14 to 0.29), the combined diagnostic odds ratio 58.02 (95% CI 30.66 to 109.76), and the area under the ROC curve 0.95. The authors conclude peripheral temperature measurement reflects core temperature to a degree, with the infrared tympanic membrane thermometer showing the highest sensitivity among the methods pooled.
Watch out: Sensitivity 82% means roughly 1 in 5 febrile children is missed at the pooled threshold — the shelf prints the pair with that plain consequence attached. The pooled devices are heterogeneous (the 'peripheral thermometer' umbrella), so the pair describes the class in emergency/outpatient use, not any single device.
Source: Yuan X et al. Temperature measurement accuracy of peripheral thermometer in the pediatric emergency department: A meta-analysis and systematic review. International emergency nursing. 2024. PMID: 38824868. Registry ID: none in the record. MeSH: Child · Humans · Body Temperature · Emergency Service, Hospital · Fever/diagnosis · Pediatrics/methods/standards · Sensitivity and Specificity · Thermometers/standards
Card 4 — Synthesis — peripheral thermometry in children: two different verdicts for two different questions — as a temperature measurement it reads low and wide against rectal (mean difference 0.22 °C, limits −0.44 to +1.30 °C); as a fever screen it performs well at optimised cut-offs (AUC 0.95–0.97) with pooled sensitivity 82% / specificity 88%, missing about 1 in 5 febrile children
Study: Synthesis
Population: Composition: 3 systematic reviews/meta-analyses with overlapping primary-study sets, counted as syntheses, never as independent cohorts: Zhen 2014 (pooled N not documented in the record), Shi 2020 (12 studies, 4,639 children), Yuan 2024 (10 studies, 3,448 children). Documented pooled N = 8,087 across the two syntheses that state one, with overlap between them unquantified in the records.
Summary: Synthesis — peripheral thermometry in children: two different verdicts for two different questions — as a temperature measurement it reads low and wide against rectal (mean difference 0.22 °C, limits −0.44 to +1.30 °C); as a fever screen it performs well at optimised cut-offs (AUC 0.95–0.97) with pooled sensitivity 82% / specificity 88%, missing about 1 in 5 febrile children
Result: The adult counterpart shelf (#7781) found the same agreement failure against central references in adults; the paediatric base adds the screening limb the adult shelf lacks, and the two limbs must not be merged: limits of agreement judge the number on the display, sensitivity and specificity judge the fever decision at a cut-off. A device can pass the second and fail the first — on this base, tympanic thermometry in children does exactly that. No report was refused.
Watch out: All three anchors are syntheses of small primary studies with heterogeneous devices, sites and reference standards; the primary-study overlap between the syntheses is real and unquantified, so this shelf's weight is 'three reviews agree', not '8,087 children studied'.
Source: Synthesis of the reports carded above; no additional source.