Oscillometric blood-pressure measurement in atrial fibrillation — 2 invasive/beat-to-beat comparison studies (documented N = 221): average bias stays small when three readings are averaged, but within-subject variability is significantly higher in atrial fibrillation and individual error can be large (systolic SD ±14.8 mmHg against beat-to-beat reference)

Shelf #9041 · Heart, vessels, blood pressure & clotting

MEASUREMENT ACCURACY — unbiased on average with repeated readings, imprecise in the individual: against invasive arterial pressure (102 patients), mean systolic and diastolic values did not differ significantly between atrial fibrillation and sinus rhythm for either an upper-arm or a wrist device when three measurements were averaged, but within-subject variability was higher in atrial fibrillation (P < 0.01); against a beat-to-beat reference (96 patients plus 23 auxiliary), systolic deviation in atrial fibrillation was +0.1 (±14.8) mmHg and diastolic +6.3 (±9.2) mmHg, and accuracy fell further with high beat-to-beat variability or high heart rate

Stamped: 2026-11-28.

Evidence published: 1979–2018.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with atrial fibrillation, how accurately do automated oscillometric blood-pressure devices measure systolic and diastolic pressure against invasive arterial pressure or a beat-to-beat reference, and how much does averaging repeated readings help?

Balance of the evidence

MEASUREMENT ACCURACY — unbiased on average with repeated readings, imprecise in the individual — Composition: 2 comparison studies, documented N = 221 (102 + 96 + 23 auxiliary-device patients). Pagonas 2013: in 50 patients with atrial fibrillation and 52 in sinus rhythm, systolic and diastolic pressure by an automated oscillometric upper-arm and a wrist device did not differ significantly from invasive arterial pressure in either rhythm when three consecutive measurements were averaged, but within-subject variability of the oscillometric measurements was higher in atrial fibrillation (P < 0.01). Feenstra 2018: in 58 patients with atrial fibrillation, device systolic pressure deviated +0.1 (±14.8) mmHg and diastolic +6.3 (±9.2) mmHg from a beat-to-beat volume-clamp reference. MEASUREMENT ACCURACY — average bias is not the problem in atrial fibrillation; spread is. A single oscillometric reading in atrial fibrillation can sit far from the true pressure, and only averaged repeated readings recover group-level accuracy.

MeSH: Adult · Aged · Aged, 80 and over · Atrial Fibrillation/physiopathology · Blood Pressure/physiology · Blood Pressure Determination/instrumentation,methods · Blood Pressure Monitors* · Cross-Sectional Studies · Female · Humans · Male · Middle Aged · Oscillometry*

Cards (3)

Card 1 — Against invasive pressure: no bias penalty for atrial fibrillation — if three readings are averaged (Pagonas 2013, Hypertension — cross-sectional comparison, 102 patients): variability, not bias, is what atrial fibrillation adds (P < 0.01)

Study: Cross-sectional measurement-comparison study (invasive reference; n = 102)

Population: 102 patients (52 sinus rhythm, 50 atrial fibrillation); accuracy of an automated validated oscillometric upper-arm device and a wrist device (Omron M5 Professional and R5 Professional — named because device identity is part of the measurement claim) assessed against invasively measured arterial pressure; blood pressure calculated as the mean of 3 consecutive measurements.

Summary: Against invasive pressure: no bias penalty for atrial fibrillation — if three readings are averaged (Pagonas 2013, Hypertension — cross-sectional comparison, 102 patients): variability, not bias, is what atrial fibrillation adds (P < 0.01)

Result: Systolic and diastolic blood pressure did not differ significantly between patients in sinus rhythm and atrial fibrillation, independent of the method of measurement (P > 0.05 each). The within-subject variability of the oscillometric measurements was higher in atrial fibrillation than in sinus rhythm (P < 0.01 each). In Bland–Altman analysis, the biases of systolic and diastolic pressure did not differ significantly in the presence or absence of atrial fibrillation (P > 0.05 each). The authors conclude atrial fibrillation did not significantly affect the accuracy of oscillometric measurement when 3 repeated measurements were performed.

Watch out: The reassuring result is conditional on the protocol: three consecutive measurements averaged, in a supervised study setting. It does not validate the single casual reading taken in a busy clinic.

Source: Pagonas N et al. Impact of atrial fibrillation on the accuracy of oscillometric blood pressure monitoring. Hypertension (Dallas, Tex. : 1979). 2013. PMID: 23897073. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Atrial Fibrillation/physiopathology · Blood Pressure/physiology · Blood Pressure Determination/instrumentation,methods · Blood Pressure Monitors* · Cross-Sectional Studies · Female · Humans · Male · Middle Aged · Oscillometry*

Card 2 — Against beat-to-beat pressure: mean deviations look fine, the spread does not (Feenstra 2018, Blood Pressure Monitoring — comparison study, 96 patients plus 23 auxiliary): systolic +0.1 (±14.8) mmHg, diastolic +6.3 (±9.2) mmHg in atrial fibrillation

Study: Measurement-comparison study (beat-to-beat reference; n = 119 including auxiliary)

Population: Main study: 58 outpatients with atrial fibrillation and 38 in sinus rhythm, systolic and diastolic pressure measured by automated oscillometry (Welch Allyn Spot Vital Signs) and by beat-to-beat volume-clamp method; auxiliary study: 23 patients with atrial fibrillation measured with a Philips IntelliVue monitor (devices named because device identity is part of the measurement claim). Accuracy also analysed by tertiles of beat-to-beat variability and heart frequency.

Summary: Against beat-to-beat pressure: mean deviations look fine, the spread does not (Feenstra 2018, Blood Pressure Monitoring — comparison study, 96 patients plus 23 auxiliary): systolic +0.1 (±14.8) mmHg, diastolic +6.3 (±9.2) mmHg in atrial fibrillation

Result: In atrial fibrillation, oscillometric systolic pressure deviated by +0.1 (±14.8) mmHg and in sinus rhythm by −7.9 (±15.7) mmHg from the volume-clamp reference; diastolic pressure read higher than reference in both (atrial fibrillation +6.3 (±9.2) mmHg; sinus rhythm +5.0 (±7.7) mmHg). High beat-to-beat variability and high heart frequency decreased accuracy for both systolic and diastolic pressure. With the auxiliary device, deviations in atrial fibrillation were −6.8 (±13.2) mmHg systolic and +9.4 (±8.1) mmHg diastolic. The authors conclude oscillometric accuracy in atrial fibrillation is limited and individual inaccuracy may be considerable.

Watch out: The volume-clamp reference is itself an estimate, and even in sinus rhythm the systolic deviation was −7.9 mmHg — part of the spread is the method and the reference, not the arrhythmia alone. The atrial-fibrillation-specific penalty is clearest for diastolic pressure and at high heart rates.

Source: Feenstra RK et al. Accuracy of oscillometric blood pressure measurement in atrial fibrillation. Blood pressure monitoring. 2018. PMID: 29256921. Registry ID: none in the record. MeSH: Aged · Atrial Fibrillation/diagnosis,physiopathology · Blood Pressure* · Blood Pressure Determination/methods · Female · Humans · Hypertension/diagnosis,physiopathology · Male · Middle Aged · Oscillometry/methods

Card 3 — Synthesis — MEASUREMENT ACCURACY: in atrial fibrillation the average is honest and the single reading is not

Study: Synthesis

Population: 2 comparison studies: Pagonas 2013 (n = 102, invasive reference), Feenstra 2018 (n = 96 main + 23 auxiliary, beat-to-beat reference). Documented N = 221.

Summary: Synthesis — MEASUREMENT ACCURACY: in atrial fibrillation the average is honest and the single reading is not

Result: Both studies agree once their references are understood. Oscillometry in atrial fibrillation is not systematically biased: averaged over three readings, mean pressures match invasive values as well as they do in sinus rhythm. What atrial fibrillation adds is variance — significantly higher within-subject variability, systolic standard deviations around ±14.8 mmHg against a beat-to-beat reference, and worse accuracy when beat-to-beat variability or heart rate is high. The practical measurement rule the evidence supports: in atrial fibrillation, take repeated readings and average them, and treat any single oscillometric value as a rough estimate.

Watch out: Neither study validates the home self-measurement setting, where cuff placement and rest periods are uncontrolled; accuracy there will be worse than in these supervised comparisons.

Source: Synthesis of the reports carded above; no additional source.