SourceMed · Shelf #9139 in the ask interface

Smartphone skin-cancer apps — 2 systematic reviews plus 2 prospective validations (785 and 1,550 lesions): the algorithm apps missed melanomas in the early evidence (sensitivities from 7% to 73% in the Cochrane review), and the best prospective market-approved app reached sensitivity 86.9% / specificity 70.4% on clinician-taken photographs of pre-selected lesions

Shelf #9139 · Eyes, skin, nose & allergy

MEASUREMENT ACCURACY — NOT ESTABLISHED for self-triage of skin lesions: the Cochrane review found artificial-intelligence app sensitivities ranging from 7% to 73% (specificities 37% to 94%) across 332 lesions with 86 melanomas, with 2–18% of images unevaluable and at least one melanoma classed unevaluable in three of four evaluations; the BMJ review's best-studied app (SkinVision, the named product, in 252 lesions) achieved sensitivity 80% / specificity 78% for malignant or premalignant lesions, while another CE-marked app scored sensitivity 0% in its only study (15 lesions, 5 melanomas); the largest prospective validation of a market-approved app (785 lesions) reached sensitivity 86.9% / specificity 70.4% — on clinician-selected lesions photographed in clinics, not on the public's own photographs

Stamped: 2026-11-28.

Evidence published: 2018–2022.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do algorithm-based smartphone applications that assess photographs of skin lesions detect melanoma and other skin cancers against histopathology — in the studies that exist, whose photographs were mostly taken by clinicians of lesions already selected in clinics?

Balance of the evidence

MEASUREMENT ACCURACY — NOT ESTABLISHED for self-triage — composition: 2 systematic reviews (Cochrane 2018, 2 studies, 332 lesions; BMJ 2020, 9 studies of 6 apps, 725 lesions verified by histology or follow-up plus 407 against expert recommendation) plus 2 prospective validations (785 lesions; 514 patients, 1,550 images). The reviews agree the underlying studies are small, selectively recruited, at high risk of bias, and used clinician-taken photographs of clinic-selected lesions — so even the favourable numbers are ceilings, not expectations for home use. The strongest prospective result (sensitivity 86.9% / specificity 70.4%, labelled in words at first occurrence) still misses about 1 in 8 premalignant or malignant lesions and flags about 3 in 10 benign lesions; on suspicious lesions alone its specificity was 45.5%. A clinician-facing dermoscopic algorithm (a different use case, carded separately) performed at specialist level on selected lesions — which shows the failure is in the consumer self-triage pathway, not in image algorithms as such. MEASUREMENT ACCURACY — NOT ESTABLISHED: no app on this base can be relied on to rule out melanoma, and a 'low risk' app result is the dangerous output, because it is the one that delays the consultation that would have caught the cancer.

MeSH: Adult · Algorithms · Diagnostic Errors/statistics & numerical data · Early Detection of Cancer/instrumentation,methods · Humans · Melanoma/diagnostic imaging · Mobile Applications · Sensitivity and Specificity · Skin Neoplasms/diagnostic imaging · Smartphone · Triage/methods · Cutaneous Malignant Melanoma

Cards (5)

Card 1 — Cochrane review: app sensitivities ranged from 7% to 73%, and melanomas were lost as 'unevaluable' (Chuchu 2018, Cochrane Database of Systematic Reviews — 2 studies, 332 lesions, 86 melanomas)

Study: Systematic review (Cochrane; 2 studies, no meta-analysis performed)

Population: A Cochrane diagnostic-accuracy review of smartphone applications intended for people in the community with lesions suspicious for melanoma or atypical intraepidermal melanocytic variants, against histological confirmation, clinical follow-up or expert opinion; data were too scarce and studies too poor for meta-analysis, so estimates are plotted per application. Five applications and 332 suspicious lesions with 86 melanomas were reported across the two included studies, both at high risk of bias (selective recruitment, high rates of non-evaluable images, lesions already selected for excision, photographs taken by clinicians).

Summary: Cochrane review: app sensitivities ranged from 7% to 73%, and melanomas were lost as 'unevaluable' (Chuchu 2018, Cochrane Database of Systematic Reviews — 2 studies, 332 lesions, 86 melanomas)

Result: Across the four artificial-intelligence applications, sensitivities ranged from 7% (95% CI 2% to 16%) to 73% (95% CI 52% to 88%) and specificities from 37% (95% CI 29% to 46%) to 94% (95% CI 87% to 97%) (sensitivity and specificity ranges, labelled in words at first occurrence). The single store-and-forward application (photographs reviewed by a dermatologist) had sensitivity 98% (95% CI 90% to 100%) and specificity 30% (95% CI 22% to 40%). Test failures — images the applications classed as unevaluable and the studies excluded — ranged from 3 to 31 per evaluation (2% to 18% of lesions), the store-and-forward application had a 15% failure rate, and at least one melanoma was classed as unevaluable in three of the four application evaluations. The reviewers conclude the AI applications have not demonstrated sufficient accuracy and carry a high likelihood of missing melanomas.

Watch out: The human-in-the-loop store-and-forward result (sensitivity 98% / specificity 30%) is a teledermatology service, not an algorithm — carded as the review reports it, and it should not be quoted as app accuracy.

Source: Chuchu N et al. Smartphone applications for triaging adults with skin lesions that are suspicious for melanoma. The Cochrane database of systematic reviews. 2018. PMID: 30521685. Registry ID: none in the record. MeSH: Adult · Algorithms · Diagnostic Errors/statistics & numerical data · Early Detection of Cancer/instrumentation,methods · Humans · Melanoma/diagnostic imaging · Mobile Applications · Sensitivity and Specificity · Skin Neoplasms/diagnostic imaging · Smartphone · Triage/methods · Cutaneous Malignant Melanoma

Card 2 — BMJ review of 6 apps: the best-studied reached sensitivity 80% / specificity 78%; another CE-marked app scored 0% sensitivity in its only study (Freeman 2020, BMJ — systematic review, 9 studies)

Study: Systematic review of diagnostic accuracy studies (9 studies, 6 apps)

Population: A systematic review of diagnostic-accuracy studies of algorithm-based smartphone applications assessing images of skin lesions suspicious for skin cancer, with histology or follow-up as reference in 6 studies (725 lesions) and expert recommendation in 3 (407 lesions); study validity assessed with QUADAS-2. The named applications are the subject of the claim and are named as the review names them.

Summary: BMJ review of 6 apps: the best-studied reached sensitivity 80% / specificity 78%; another CE-marked app scored 0% sensitivity in its only study (Freeman 2020, BMJ — systematic review, 9 studies)

Result: Two CE-marked apps were available for download. SkinScan was evaluated in a single study (15 lesions, 5 melanomas) with 0% sensitivity and 100% specificity for melanoma detection (sensitivity 0.00 / specificity 1.00 in the labelled form used on this shelf). SkinVision was evaluated in two studies (252 lesions, 61 malignant or premalignant) and achieved sensitivity 80% (95% CI 63% to 92%) and specificity 78% (67% to 87%) for malignant or premalignant lesions; its accuracy against expert recommendations was poor in three further studies. The reviewers conclude current apps cannot be relied on to detect all cases of melanoma or other skin cancers, that performance is likely poorer in clinically relevant populations and with intended users, and that the CE-marking process for such apps does not adequately protect the public. PROSPERO CRD42016033595.

Watch out: The SkinScan figure rests on 15 lesions — it proves the app was marketed on near-no evidence, not that its true sensitivity is exactly zero. Both reviews' studies predate current-generation models; the shelf judges the evidence as published, and the prospective cards below are the newest evidence in the base.

Source: Freeman K et al. Algorithm based smartphone apps to assess risk of skin cancer in adults: systematic review of diagnostic accuracy studies. BMJ (Clinical research ed.). 2020. PMID: 32041693. Registry ID: none in the record. MeSH: Algorithms · Biopsy · Dermoscopy/instrumentation,methods · False Negative Reactions · False Positive Reactions · Humans · Melanoma/diagnosis,pathology · Mobile Applications · Reproducibility of Results · Risk Assessment/methods · Skin/diagnostic imaging,pathology · Skin Neoplasms/diagnosis,pathology · Smartphone

Card 3 — Largest prospective validation of a market-approved app: sensitivity 86.9% / specificity 70.4% overall, but specificity 45.5% on suspicious lesions (Sangers 2022, Dermatology — prospective multicentre study, 785 lesions in 372 patients)

Study: Prospective multicentre diagnostic accuracy study (785 lesions)

Population: A prospective multicentre diagnostic-accuracy study (January–August 2020) of a CE-marked mobile-health app integrating a convolutional neural network, in 372 adults (50.8% women, median age 71 years) with at least one suspicious skin lesion: 785 lesions in total (418 suspicious, 367 benign controls), assessed by the app on iOS or Android after clinical diagnosis and before histopathology, against histopathological diagnosis or, where unavailable, a dermatologist's clinical diagnosis. The app is not named in the record's abstract, and no name is printed on this shelf.

Summary: Largest prospective validation of a market-approved app: sensitivity 86.9% / specificity 70.4% overall, but specificity 45.5% on suspicious lesions (Sangers 2022, Dermatology — prospective multicentre study, 785 lesions in 372 patients)

Result: The app performed at an overall sensitivity of 86.9% (95% CI 82.3 to 90.7) and specificity of 70.4% (95% CI 66.2 to 74.3) (sensitivity 0.87 / specificity 0.70, labelled in words) for premalignant and malignant lesions. Sensitivity was higher on iOS than Android devices (91.0% versus 83.0%; p = 0.02). Specificity on benign control lesions (80.1%) was significantly higher than on suspicious lesions (45.5%; p < 0.001). The authors conclude the app's accuracy is far from perfect but potentially promising for patient self-assessment, and that further prospective validation, particularly for suspicious pigmented lesions, and implementation studies in the lay population are needed.

Watch out: Photographs were taken in clinics of lesions already selected by clinicians, and benign controls flatter the specificity — the 45.5% specificity on suspicious lesions is the figure closest to real self-triage use, and the lay-population performance the authors call for is still unmeasured.

Source: Sangers T et al. Validation of a Market-Approved Artificial Intelligence Mobile Health App for Skin Cancer Screening: A Prospective Multicenter Diagnostic Accuracy Study. Dermatology (Basel, Switzerland). 2022. PMID: 35124665. Registry ID: none in the record. MeSH: Adult · Aged · Artificial Intelligence · Early Detection of Cancer · Female · Humans · Male · Melanoma/pathology · Mobile Applications · Skin Diseases · Skin Neoplasms/diagnosis,pathology · Telemedicine

Card 4 — The clinician-facing contrast: a dermoscopic algorithm matched specialists on selected lesions (AUROC 90.1% on iPhone images) (Phillips 2019, JAMA Network Open — prospective multicentre trial, 514 patients, 1,550 images)

Study: Prospective diagnostic accuracy trial (clinician-acquired dermoscopic images)

Population: A prospective, multicentre, single-arm, masked diagnostic trial in dermatology and plastic-surgery clinics at 7 UK hospitals: dermoscopic images of suspicious and control skin lesions from 514 patients (55.7% women; 96.8% white; mean age 52.1 years) captured on three cameras (iPhone 6s, Galaxy S6, and a digital single-lens reflex camera), assessed by clinicians and by the Deep Ensemble for Recognition of Malignancy algorithm (the named algorithm is the subject of this card's claim) against histopathology. Of 1,550 images, 286 (18.6%) were used to train the algorithm and 849 (54.8%) were missing or unsuitable for analysis; 551 lesions were biopsied, of which 125 (22.7%) were melanoma and 77 entered the primary analysis.

Summary: The clinician-facing contrast: a dermoscopic algorithm matched specialists on selected lesions (AUROC 90.1% on iPhone images) (Phillips 2019, JAMA Network Open — prospective multicentre trial, 514 patients, 1,550 images)

Result: The algorithm achieved an area under the receiver operating characteristic curve (AUROC) of 90.1% (95% CI 86.3% to 94.0%) for biopsied lesions and 95.8% (94.1% to 97.6%) for all lesions using iPhone 6s images (Galaxy S6: 85.8% and 93.8%; DSLR: 86.9% and 91.8%). At 100% sensitivity, the algorithm achieved a specificity of 64.8% with iPhone 6s images. Specialists achieved an AUROC of 77.8% (95% CI 72.5% to 81.9%) and a specificity of 69.9%. The authors conclude the algorithm identified melanoma from dermoscopic images of selected lesions with accuracy similar to specialists.

Watch out: A different use case from the consumer cards: dermoscopic attachments, clinician photography, clinic-selected lesions, over half the images excluded as missing or unsuitable, and a study population 96.8% white. It shows what image algorithms can do inside a clinical pathway — it does not rehabilitate self-triage apps, and its numbers must not be borrowed for them.

Source: Phillips M et al. Assessment of Accuracy of an Artificial Intelligence Algorithm to Detect Melanoma in Images of Skin Lesions. JAMA network open. 2019. PMID: 31617929. Registry ID: none in the record. MeSH: Adult · Aged · Area Under Curve · Biopsy · Deep Learning · Dermoscopy/instrumentation · Female · Humans · Male · Melanoma/diagnostic imaging,pathology · Middle Aged · Photography/instrumentation · Prospective Studies · ROC Curve · Skin Neoplasms/diagnostic imaging,pathology · Smartphone

Card 5 — Synthesis — MEASUREMENT ACCURACY — NOT ESTABLISHED for self-triage: the misses are melanomas, and the evidence was gathered under easier conditions than home use

Study: Synthesis

Population: 2 systematic reviews (332 and 725-plus-407 lesions) plus 2 prospective validations (785 lesions; 514 patients / 1,550 images).

Summary: Synthesis — MEASUREMENT ACCURACY — NOT ESTABLISHED for self-triage: the misses are melanomas, and the evidence was gathered under easier conditions than home use

Result: Every layer points the same way. The early consumer apps missed between a quarter and nearly all melanomas in formal evaluation (sensitivities 7% to 73%), lost further melanomas as unevaluable images, and were marketed on studies as small as 15 lesions. The best prospective evidence for a current market-approved app — sensitivity 86.9% / specificity 70.4% (labelled in words), falling to specificity 45.5% on suspicious lesions — was gathered on clinician-taken photographs of clinic-selected lesions in patients with a median age of 71. The one specialist-level performance on this shelf required dermoscopy, clinician photography and exclusion of over half the images. No report was refused.

Watch out: NOT ESTABLISHED is a verdict on the self-triage claim, not on skin-image algorithms generally: the same technology performs materially better inside a clinical pathway with controlled image acquisition, which is where the evidence supports developing it.

Source: Synthesis of the reports carded above; no additional source.