Smartphone hearing tests versus pure-tone audiometry — 1 systematic review (17 studies) plus 3 validation studies (documented N = 462): the studied apps report sensitivity and specificity between 75% and 100% for hearing loss, hearTest screening exceeded sensitivity 0.90 / specificity 0.90 for disabling hearing loss in self-test mode, but accuracy depends on the app, the earphones and the setting
Shelf #9141 · Eyes, skin, nose & allergy
MEASUREMENT ACCURACY — ESTABLISHED for screening with named, validated apps used as specified, NOT ESTABLISHED as a replacement for diagnostic audiometry: the systematic review (17 studies) found the most-studied apps (Digits-in-Noise, uHear, hearScreen, hearTest) reporting sensitivity and specificity values between 75% and 100% against pure-tone audiometry, with only 4 of 17 studies meeting all methodological-quality criteria; hearTest screening in 301 tested participants gave sensitivity and specificity above 0.90 for disabling hearing loss in self-test mode, but identifying any level of hearing loss in adults was mode-dependent (low specificity in test-operator mode); in a 139-participant multicentre study only 2 of 4 apps produced thresholds within ±10 dB of audiometry at home, and in a 22-participant study the same app was accurate with one earphone type and not others
Stamped: 2026-11-28.
Evidence published: 2018–2024.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do smartphone-based hearing tests (automated pure-tone apps and digits-in-noise screening) identify hearing loss against conventional pure-tone audiometry — and how much do the app, the earphones, and testing at home versus in a sound-treated room change the result?
Balance of the evidence
MEASUREMENT ACCURACY — established for screening, not for diagnosis — composition: 1 systematic review (17 studies) plus 3 validation studies (n = 301 completing all procedures of 340 enrolled; 139; 22 — total documented N = 462 across the primary studies, counting the hearTest study at the 301 who completed all test procedures). The signal is real: against the audiometry reference, validated apps in self-test mode detect disabling hearing loss at sensitivity and specificity above 0.90 (labelled in words at first occurrence), and the review's range across apps and studies is 75–100% for both measures. The cautions are equally concrete: only 4 of the review's 17 studies met all quality criteria; two of four apps in the multicentre study failed to track audiometry at home; one app's accuracy flipped with the earphones used; and mild hearing loss in adults was where specificity broke down. PROPOSED MEASUREMENT ACCURACY: a named, validated app, used with the earphones and mode its validation specified, is a legitimate screening instrument that sorts people into 'test properly' and 'probably fine' — it does not produce a diagnostic audiogram, and an unvalidated app or an unspecified earphone inherits none of these numbers.
MeSH: Audiometry, Pure-Tone · Hearing · Hearing Loss/diagnosis · Humans · Noise · Sensitivity and Specificity · Smartphone
Cards (5)
Card 1 — Systematic review: 17 studies, sensitivity and specificity between 75% and 100% across apps — but only 4 studies met all quality criteria (Melo 2022, CoDAS — systematic review)
Study: Systematic review (17 diagnostic accuracy studies)
Population: A PRISMA-DTA systematic review of smartphone applications to identify hearing loss against pure-tone audiometry as the reference standard, across four databases and grey literature: 1,395 articles screened, 104 read in full, 17 included (all published in English, 2015–2020, all ages and both sexes), with risk of bias assessed by QUADAS-2. The most-studied applications were the Digits-in-Noise Test (5 articles), uHear (4), HearScreen (2), hearTest (2) and Hearing Test (2); the named applications are the subject of the claim.
Summary: Systematic review: 17 studies, sensitivity and specificity between 75% and 100% across apps — but only 4 studies met all quality criteria (Melo 2022, CoDAS — systematic review)
Result: All the named applications showed sensitivity and specificity values between 75% and 100% (labelled in words at first occurrence). The reviewers single out uHear, the Digits-in-Noise Test, hearTest and HearScreen as having shown significant sensitivity and specificity values and being the most accurate methods for screening hearing impairment. Only 4 of the 17 included studies met all criteria for methodological quality.
Watch out: A range across different apps, thresholds and populations is not a pooled estimate: 75% and 100% are both inside it, and which app, which hearing-loss criterion and which setting produced which end of the range is study-specific — the primary cards below carry the usable detail.
Source: Melo IMM et al. Accuracy of smartphone-based hearing screening tests: a systematic review. CoDAS. 2022. PMID: 35239828. Registry ID: none in the record. MeSH: Audiometry, Pure-Tone · Hearing · Hearing Loss/diagnosis · Humans · Noise · Sensitivity and Specificity · Smartphone
Card 2 — hearTest screening: sensitivity and specificity above 90% for disabling hearing loss in self-test mode, in adults and children (Corona 2020, International Journal of Audiology — diagnostic accuracy study, 340 enrolled, 301 completing all procedures)
Study: Diagnostic accuracy study (smartphone screening versus conventional audiometry)
Population: 340 individuals aged 5–92 years (301 undertook all test procedures: 273 adults and 28 children) screened with the hearTest smartphone-based audiometry application (the named application is the subject of this card's claim) in two response modes — self-test and test-operator — against conventional audiometry, for two criteria: disabling hearing loss and any level of hearing loss; test duration was also compared between modes.
Summary: hearTest screening: sensitivity and specificity above 90% for disabling hearing loss in self-test mode, in adults and children (Corona 2020, International Journal of Audiology — diagnostic accuracy study, 340 enrolled, 301 completing all procedures)
Result: Sensitivity and specificity were both above 90% (sensitivity 0.90 / specificity 0.90 as the labelled floor the record states) for hearTest screening to identify disabling hearing loss, for both response modes, in adults and children. Sensitivity for identifying any level of hearing loss was similar for both modes in children, with specificity above 80%, and for the self-test mode in adults; but specificity was low when identifying any level of hearing loss in adults using the test-operator mode. Test duration differed significantly between modes in adults. The authors conclude hearTest screening is accurate for disabling hearing loss and, in self-test mode, for any level of hearing loss.
Watch out: The record prints the accuracy thresholds as 'above 90%' and 'above 80%' rather than exact point estimates, and the shelf prints them the same way — no exact pair is quoted. The failure mode is specific and worth keeping: mild loss, adults, operator mode.
Source: Corona AP et al. Validity of hearing screening using hearTest smartphone-based audiometry: performance evaluation of different response modes. International journal of audiology. 2020. PMID: 32134341. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Audiometry · Audiometry, Pure-Tone · Child · Child, Preschool · Hearing · Hearing Loss/diagnosis · Humans · Mass Screening · Middle Aged · Sensitivity and Specificity · Smartphone · Young Adult
Card 3 — Four apps taken home: only two of four tracked audiometry within ±10 dB, in clinic and at home (Swords 2024, The Laryngoscope — multicentre validity study, 139 participants)
Study: Multicentre validity and feasibility study (4 apps; n = 139)
Population: A multicentre, prospective randomised study of patients undergoing pure-tone audiometry in clinics, each allocated to one of four publicly available smartphone applications designed to measure pure-tone thresholds (the record's abstract does not name the four applications, and no names are printed on this shelf); each participant used the app once in optimal sound-treated conditions and three times at home, with ear- and frequency-specific thresholds and pure-tone averages compared with audiometry, and acceptability surveyed.
Summary: Four apps taken home: only two of four tracked audiometry within ±10 dB, in clinic and at home (Swords 2024, The Laryngoscope — multicentre validity study, 139 participants)
Result: 139 participants submitted data. The results of two at-home automated smartphone apps correlated strongly or very strongly with the audiometry pure-tone average, and their frequency-specific medians were within ±10 dB accuracy; app results in sound-treated and home conditions correlated very strongly with each other. 90% of participants rated the apps easy or very easy to use, and 90% would be happy or very happy to use an app to monitor their hearing. The authors conclude that judicious use of self-performed smartphone audiometry was valid and feasible for two of the four apps.
Watch out: Half the tested products failed the ±10 dB criterion — the study is as much a warning about unvalidated apps as a validation of the two that passed, and because the record does not name the apps, this card cannot tell a reader which two passed; the named-app evidence is on the other cards.
Source: Swords C et al. A Multicenter Validity Study of Four Smartphone Hearing Test Apps in Optimized and Home Environments. The Laryngoscope. 2024. PMID: 38214403. Registry ID: none in the record. MeSH: Adult · Aged · Female · Humans · Male · Middle Aged · Young Adult · Audiometry, Pure-Tone/instrumentation,methods · Auditory Threshold/physiology · Feasibility Studies · Hearing Loss/diagnosis · Mobile Applications · Prospective Studies · Reproducibility of Results · Smartphone
Card 4 — The earphone is part of the test: uHear was accurate with the standard earbuds at 1000–6000 Hz and not interchangeable across transducers (Barczik 2018, American Journal of Audiology — 22 adults, 44 ears)
Study: Validation study (2 apps × 3 transducers versus conventional audiometry; n = 22)
Population: 22 adult participants (10 with normal hearing, 12 with sensorineural hearing loss; 44 ears) who underwent conventional audiometry and performed 6 self-administered hearing tests using two iPhone-based applications — uHear (Unitron) and uHearingTest (WooFu Tech) — each with 3 transducer styles (earbud earphones, supra-aural headphones, circumaural headphones). The named applications and transducers are the subject of this card's claim.
Summary: The earphone is part of the test: uHear was accurate with the standard earbuds at 1000–6000 Hz and not interchangeable across transducers (Barczik 2018, American Journal of Audiology — 22 adults, 44 ears)
Result: Accuracy differed between apps, across frequencies and across earphone styles. The uHear app using the iPhone standard EarPod earbuds was accurate to conventional thresholds (p > .002 with Bonferroni correction) at 1000, 2000, 4000 and 6000 Hz, and was valid (81–100% across sensitivity, specificity, positive and negative predictive values) for screening mild-or-greater hearing loss (above 25 dB HL) at 500, 1000, 2000, 4000 and 6000 Hz. The uHearingTest app was accurate and valid for screening mild-or-greater loss using supra-aural headphones at 2000, 4000 and 8000 Hz. The authors conclude self-hearing-test apps can be accurate when using appropriate transducers, and that manufacturers should specify earphone models to users.
Watch out: Twenty-two participants — N printed on the face of this card. Its weight is the interaction it demonstrates (app × earphone × frequency), not the point estimates: an app validated with one earphone has not been validated with another.
Source: Barczik J et al. Accuracy of Smartphone Self-Hearing Test Applications Across Frequencies and Earphone Styles in Adults. American journal of audiology. 2018. PMID: 30242342. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Audiometry, Pure-Tone · Auditory Threshold · Case-Control Studies · Female · Hearing Loss, Sensorineural/diagnosis · Hearing Tests/instrumentation,methods · Humans · Male · Mass Screening · Middle Aged · Mobile Applications · Smartphone · Young Adult
Card 5 — Synthesis — MEASUREMENT ACCURACY: real screening performance, conditional on the app, the earphones and the mode — none of it transfers to an unvalidated product
Study: Synthesis
Population: 1 systematic review (17 studies) plus 3 validation studies (n = 301 completing all procedures; 139; 22 — total documented primary N = 462).
Summary: Synthesis — MEASUREMENT ACCURACY: real screening performance, conditional on the app, the earphones and the mode — none of it transfers to an unvalidated product
Result: The favourable evidence is specific, and its specificity is the message: named apps (hearTest in self-test mode; uHear with the specified earbuds; the Digits-in-Noise family in the review) detect disabling or mild-or-greater hearing loss at sensitivity and specificity from the high 80s to above 0.90 (labelled in words) against audiometry. Against that: only 4 of 17 reviewed studies met all quality criteria, 2 of 4 apps failed at-home threshold tracking in the multicentre study, accuracy flipped with transducer choice in the smallest study, and mild loss in adults in operator mode is where specificity failed. A screening result from a validated configuration sorts; it does not diagnose, fit a hearing aid, or replace the audiogram it is screened against. No report was refused.
Watch out: Most validation was performed or supervised in clinical settings with clinic-recruited participants; performance in unsupervised home use by the general public is inferred from one multicentre study's at-home arm, not measured across the base.
Source: Synthesis of the reports carded above; no additional source.