Mobile visual-acuity apps versus standard charts — 1 meta-analysis (22 studies reviewed; 6 studies, 24,284 participants, meta-analysed) plus 3 validation studies (documented N = 826): Peek Acuity agreed with the ETDRS chart to a mean difference of 0.07 logMAR with test–retest variability of ±0.033 logMAR, and pooled app sensitivity ranged 0.85–0.87 with specificity 0.78–0.98 across age groups
Shelf #9142 · Eyes, skin, nose & allergy
MEASUREMENT ACCURACY — ESTABLISHED for supervised screening and field measurement with validated tumbling-E apps, with age-dependent screening performance: the meta-analysis pooled sensitivity 0.87 / specificity 0.78 in children aged 3–5 years, sensitivity 0.86 / specificity 0.91 at ages 6–22, and sensitivity 0.85 / specificity 0.98 at 55 years and older (labelled in words at first occurrence); the Peek Acuity validation (n = 300) found a mean difference from the ETDRS chart of 0.07 logMAR (95% CI 0.05 to 0.09) with test–retest variability of ±0.033 logMAR — as repeatable as the chart itself — and the near-vision version (n = 483) differed from the near chart by a mean of 0.008 logMAR with 92.9% screening agreement (kappa 0.85)
Stamped: 2026-11-28.
Evidence published: 2015–2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do smartphone and tablet applications measure visual acuity against standard clinical charts (ETDRS logMAR, Snellen, near-vision charts) — in agreement, in test–retest repeatability, and in detecting visual impairment for screening?
Balance of the evidence
MEASUREMENT ACCURACY — established for validated apps in supervised use — composition: 1 systematic review and meta-analysis (22 studies reviewed; 6 studies with 24,284 participants meta-analysed) plus 3 primary validations (n = 300; 483; 43 — total documented primary N = 826). The strongest evidence is agreement, not just detection: Peek Acuity's mean difference from the reference ETDRS chart (0.07 logMAR) is under one line of acuity, its test–retest variability (±0.033 logMAR) matches published chart repeatability, and community health workers administered it as fast as a Snellen test (77 versus 82 seconds) after minimal training. The near-vision version agreed with its chart to 0.008 logMAR on average. The screening numbers carry the caution: in the youngest children the pooled specificity was 0.78, meaning roughly 1 in 5 unimpaired children screen positive, and the iPad study found the app's letter charts (Snellen, Arabic figures) agreed with ETDRS while its tumbling-E and Landolt-C charts read significantly worse — the optotype set is part of the instrument. The Peek validation study was also summarised in a second journal report (PMID 26720028) — the same study, counted once on this shelf.
MeSH: Adolescent · Adult · Child · Child, Preschool · Computers, Handheld · Humans · Middle Aged · Mobile Applications · Reproducibility of Results · Smartphone · Visual Acuity · Young Adult
Cards (5)
Card 1 — Meta-analysis: pooled sensitivity 0.85–0.87 across age groups, specificity weakest in the youngest children at 0.78 (Suo 2022, JMIR mHealth and uHealth — systematic review and meta-analysis, 22 studies)
Study: Systematic review and meta-analysis (22 studies reviewed; 6 meta-analysed, 24,284 participants)
Population: A systematic review of mobile-app visual-acuity assessment (studies published 2008–2020) including 22 studies of tablet or smartphone apps, most rated high quality on QUADAS-2, with a meta-analysis of 6 studies involving 24,284 participants, subgrouped by age group (3–5 years, 6–22 years, 55 years and older), examiner (professional versus non-professional) and device (smartphone versus iPad).
Summary: Meta-analysis: pooled sensitivity 0.85–0.87 across age groups, specificity weakest in the youngest children at 0.78 (Suo 2022, JMIR mHealth and uHealth — systematic review and meta-analysis, 22 studies)
Result: In children aged 3–5 years, pooled sensitivity was 0.87 (95% CI 0.79 to 0.93) and pooled specificity 0.78 (95% CI 0.70 to 0.85) (sensitivity and specificity labelled in words at first occurrence). At ages 6–22, sensitivity was 0.86 (95% CI 0.84 to 0.87) and specificity 0.91 (95% CI 0.90 to 0.91). At 55 years and older, sensitivity was 0.85 (95% CI 0.55 to 0.98) and specificity 0.98 (95% CI 0.95 to 0.99). Non-professional examiners achieved higher accuracy (area under the curve 0.93) than professional examiners (0.87); the smartphone subgroup pooled sensitivity 0.86 / specificity 0.91 and the iPad subgroup sensitivity 0.86 / specificity 0.79. The authors conclude app-based acuity tests can be useful for on-demand detection of visual impairment.
Watch out: Pooled screening accuracy across many different apps and charts — the age gradient (specificity 0.78 in preschool children) is the actionable figure, and the wide confidence interval on sensitivity in the oldest group (0.55 to 0.98) marks how thin that subgroup's evidence is.
Source: Suo L et al. Use of Mobile Apps for Visual Acuity Assessment: Systematic Review and Meta-analysis. JMIR mHealth and uHealth. 2022. PMID: 35156935. Registry ID: none in the record. MeSH: Adolescent · Adult · Child · Child, Preschool · Computers, Handheld · Humans · Middle Aged · Mobile Applications · Reproducibility of Results · Smartphone · Visual Acuity · Young Adult
Card 2 — Peek Acuity: mean difference from the ETDRS chart 0.07 logMAR, test–retest ±0.033 logMAR, administered by community health workers in homes (Bastawrous 2015, JAMA Ophthalmology — validation study, 300 adults in Kenya)
Study: Validation study (smartphone test versus ETDRS and Snellen; n = 300)
Population: 300 adults aged 55 years and older, recruited consecutively within the Nakuru Eye Disease Cohort follow-up in central Kenya (December 2013 to March 2014), tested monocularly with the smartphone-based Peek Acuity test (the named application is the subject of this card's claim; a tumbling-E test not dependent on letter familiarity), the Snellen chart and the ETDRS logMAR chart (reference standard), in participants' homes and temporary clinic settings, including test–retest measurement and timing.
Summary: Peek Acuity: mean difference from the ETDRS chart 0.07 logMAR, test–retest ±0.033 logMAR, administered by community health workers in homes (Bastawrous 2015, JAMA Ophthalmology — validation study, 300 adults in Kenya)
Result: The 95% confidence limits for test–retest variability of the smartphone test were ±0.033 logMAR. Mean differences were 0.07 logMAR (95% CI 0.05 to 0.09) between the smartphone test and the ETDRS chart and 0.08 logMAR (95% CI 0.06 to 0.10) between the smartphone test and Snellen; the agreement of Peek Acuity with the ETDRS chart was greater than the agreement of Snellen with the ETDRS chart (95% CI 0.05–0.10; P = .08). Community health workers readily accepted the test, it required minimal training, and it took no longer than the Snellen test (77 seconds versus 82 seconds). The authors conclude the smartphone test gives accurate and repeatable acuity measurements consistent with published test–retest data for retroilluminated logMAR charts.
Watch out: One app, one population (adults 55 and older in rural Kenya), supervised administration — the agreement figures belong to that configuration. This study was also reported in a second journal summary (PMID 26720028); it is the same study and is counted once on this shelf.
Source: Bastawrous A et al. Development and Validation of a Smartphone-Based Visual Acuity Test (Peek Acuity) for Clinical Practice and Community-Based Fieldwork. JAMA ophthalmology. 2015. PMID: 26022921. Registry ID: none in the record. MeSH: Aged · Aged, 80 and over · Cell Phone · Community Health Workers · Female · Humans · Kenya · Male · Middle Aged · Professional Practice · Reproducibility of Results · Vision Disorders/diagnosis · Vision Tests/instrumentation · Visual Acuity/physiology
Card 3 — Peek near vision: mean difference 0.008 logMAR from the near chart, screening agreement 92.9% (kappa 0.85), and faster than the chart (Katibeh 2022, Translational Vision Science & Technology — validation study, 483 participants in Nepal)
Study: Validation study (smartphone near-vision test versus near chart; n = 483)
Population: 483 participants at an eye hospital in Lahan, Nepal, tested with the Peek Near Vision smartphone test (PeekNV, the named application is the subject of this card's claim) in its two forms — quantitative near-acuity measurement and a binary near-vision-impairment screening test — against a conventional tumbling-E near point vision chart as reference, with Bland–Altman limits of agreement for agreement and repeatability and Cohen's kappa, sensitivity and specificity for screening.
Summary: Peek near vision: mean difference 0.008 logMAR from the near chart, screening agreement 92.9% (kappa 0.85), and faster than the chart (Katibeh 2022, Translational Vision Science & Technology — validation study, 483 participants in Nepal)
Result: The mean difference between PeekNV and chart near acuity was 0.008 logMAR units (95% CI −0.005 to 0.021) in right-eye data, with 95% limits of agreement between 0.235 and −0.218 logMAR. As a screening tool, overall agreement between the tests was 92.9% (kappa = 0.85), with a positive predictive value of 93.2% (95% CI 89.6% to 96.9%) and a negative predictive value of 92.7% (95% CI 88.9% to 96.4%). PeekNV screening was faster than the chart (11.6 seconds, 95% CI 10.5 to 12.6, versus 14.9 seconds, 95% CI 13.5 to 16.2; P < 0.001). The authors conclude the smartphone test produces rapid near-acuity results comparable to an accepted near-vision test. The record reports predictive values for the screening form rather than a sensitivity/specificity pair, and the shelf prints them as reported.
Watch out: Hospital-based participants in Nepal, tested supervised — and note the spread behind the tiny mean difference: limits of agreement of roughly ±0.23 logMAR mean individual readings can differ from the chart by over two lines even when the average difference is near zero.
Source: Katibeh M et al. Development and Validation of a Digital (Peek) Near Visual Acuity Test for Clinical Practice, Community-Based Survey, and Research. Translational vision science & technology. 2022. PMID: 36583912. Registry ID: none in the record. MeSH: Humans · Visual Acuity · Vision Tests
Card 4 — Not every chart in the app is equal: the iPad Snellen and Arabic-figure charts agreed with ETDRS, the tumbling-E and Landolt-C charts read significantly worse (Rhiu 2016, Telemedicine and e-Health — agreement study, 43 normally sighted adults)
Study: Agreement study (tablet application versus ETDRS; n = 43)
Population: 43 normally sighted adults tested with the ETDRS chart and with an iPad visual-acuity application using a mirroring technique (randomly presented letters categorised by cognoscibility), containing four charts — Snellen, tumbling E, Landolt C, and Arabic figures — with all tests repeated after a 10-min break; agreement assessed by 95% limits of agreement (±1.96 standard deviations of the differences). The application is described in the record by device and design, not by a product name, and none is printed on this shelf.
Summary: Not every chart in the app is equal: the iPad Snellen and Arabic-figure charts agreed with ETDRS, the tumbling-E and Landolt-C charts read significantly worse (Rhiu 2016, Telemedicine and e-Health — agreement study, 43 normally sighted adults)
Result: LogMAR acuity showed no significant difference between the ETDRS chart and the iPad Snellen chart (p = 0.66) or the iPad Arabic-figure chart (p = 0.29), but ETDRS acuity was significantly better than the iPad tumbling-E chart (p < 0.01) and the iPad Landolt-C chart (p < 0.01). Subjects showed letter memory on retest for the ETDRS, tumbling-E and Landolt-C charts but not for the iPad Snellen or Arabic-figure charts. The authors conclude the iPad application showed similar repeatability and may be a rapid and convenient alternative to some existing measures.
Watch out: Normally sighted adults only — an agreement study at the good-acuity end, where the screening question (detecting impairment) is not tested. Its finding cuts against a lazy reading of the Peek cards: the same delivery device with a different optotype set performed measurably worse.
Source: Rhiu S et al. Visual Acuity Testing Using a Random Method Visual Acuity Application. Telemedicine journal and e-health : the official journal of the American Telemedicine Association. 2016. PMID: 26280910. Registry ID: none in the record. MeSH: Adult · Cohort Studies · Diabetic Retinopathy/diagnosis · Female · Humans · Male · Middle Aged · Mobile Applications/statistics & numerical data · Republic of Korea · Sensitivity and Specificity · Vision Tests/instrumentation,methods · Visual Acuity/physiology
Card 5 — Synthesis — MEASUREMENT ACCURACY: chart-level agreement for validated apps in trained hands; screening specificity is the weak point in young children
Study: Synthesis
Population: 1 systematic review and meta-analysis (22 studies; 6 meta-analysed, 24,284 participants) plus 3 primary validations (n = 300; 483; 43 — total documented primary N = 826). One further report (PMID 26720028) is a second summary of the Peek validation study and is counted once, not as separate evidence.
Summary: Synthesis — MEASUREMENT ACCURACY: chart-level agreement for validated apps in trained hands; screening specificity is the weak point in young children
Result: The agreement evidence is the strongest on this shelf: mean differences from reference charts of 0.07 logMAR (distance) and 0.008 logMAR (near) — under one line — with test–retest variability matching the charts themselves, delivered by community health workers at chart speed. The screening evidence is age-graded: pooled specificity of 0.78 in children aged 3–5 years means about one in five unimpaired preschoolers would be referred on, against 0.91–0.98 in older groups (sensitivity and specificity labelled in words). And the instrument is the whole configuration — app, optotype set, device and administrator — as the iPad study's split result between its own charts shows. No report was refused.
Watch out: Every primary validation here was supervised; unsupervised self-testing at home, with uncontrolled distance and lighting, is outside this base, and no acuity app on this shelf examines the eye — a normal app acuity does not exclude eye disease, and an abnormal one is a referral, not a diagnosis.
Source: Synthesis of the reports carded above; no additional source.