SourceMed · Shelf #9164 in the ask interface

Silver-coated endotracheal tube: in the one large randomised trial, microbiologically confirmed ventilator-associated pneumonia fell from 7.5% to 4.8% in patients intubated at least 24 hours — with no difference in mortality, intubation duration, or length of stay

Shelf #9164 · Critical care & sepsis

POSSIBLE — a silver-coated endotracheal tube reduced microbiologically confirmed ventilator-associated pneumonia in the one large randomised trial (4.8% versus 7.5% among patients intubated for 24 hours or longer; 2,003 patients randomised), with no difference in mortality or length of stay; the only other randomised study (149 patients) measured bacterial colonisation, not pneumonia, so the pneumonia claim rests on a single trial

Stamped: 2026-11-28.

Evidence published: 2006–2008.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults expected to need mechanical ventilation for at least 24 hours, does intubation with a silver-coated endotracheal tube reduce ventilator-associated pneumonia compared with an otherwise identical uncoated tube?

Balance of the evidence

POSSIBLE — single-trial benefit claim, documented N = 2,152 randomised across two trials. NASCENT (2,003 patients randomised in 54 North American centres) found microbiologically confirmed ventilator-associated pneumonia in 4.8% of the silver-coated group versus 7.5% of the uncoated group among patients intubated for 24 hours or longer (P =.03), a 35.9% relative risk reduction with the absolute rates printed here; among all intubated patients the rates were 3.8% versus 5.8% (P =.04), and pneumonia occurred later with the coated tube (P =.005). There were no statistically significant differences in duration of intubation, intensive care or hospital stay, mortality, or adverse events. The earlier Rello feasibility trial (149 patients) found delayed and reduced bacterial colonisation of the tube and airways but was not designed or powered to measure pneumonia. Neither record names the tube's brand or manufacturer — both describe it only as a silver-coated endotracheal tube (Rello's as a respiratory infection control device) — so no specific marketed product can be credited from these records. POSSIBLE: the direction is consistent (less colonisation, less pneumonia) but a single outcome trial, with no mortality or stay benefit, caps the stamp; the shelf is Tier C because the benefit claim rests on one trial.

MeSH: Adult · Aged · Aged, 80 and over · Anti-Infective Agents, Local* · Biofilms* · Bronchoalveolar Lavage Fluid/microbiology · Equipment Contamination/prevention & control* · Female · Humans · Incidence · Intubation, Intratracheal/adverse effects/instrumentation* · Male · Middle Aged · Pneumonia, Ventilator-Associated/epidemiology/microbiology/prevention & control* · Prospective Studies · Respiration, Artificial/instrumentation · Risk Factors · Silver Compounds* · Single-Blind Method · Time Factors

Cards (3)

Card 1 — NASCENT: silver-coated tube cut confirmed ventilator-associated pneumonia from 7.5% to 4.8% (Kollef 2008, JAMA — randomised trial, 2,003 patients randomised): a 35.9% relative risk reduction, with no difference in mortality or length of stay

Study: Randomized controlled trial (single-blind, multicentre)

Population: 2,003 adult patients expected to require mechanical ventilation for 24 hours or longer, randomised at 54 centres in North America (9,417 screened) to intubation with a silver-coated high-volume, low-pressure endotracheal tube or an otherwise similar uncoated tube; the record describes the tube by its silver coating only and names no brand or manufacturer. The primary outcome was microbiologically confirmed ventilator-associated pneumonia (quantitative bronchoalveolar lavage culture at 10⁴ colony-forming units/mL or greater) in patients intubated for 24 hours or longer.

Summary: NASCENT: silver-coated tube cut confirmed ventilator-associated pneumonia from 7.5% to 4.8% (Kollef 2008, JAMA — randomised trial, 2,003 patients randomised): a 35.9% relative risk reduction, with no difference in mortality or length of stay

Result: Among patients intubated for 24 hours or longer, rates of microbiologically confirmed ventilator-associated pneumonia were 4.8% (37/766 patients; 95% CI, 3.4%-6.6%) with the silver-coated tube and 7.5% (56/743; 95% CI, 5.7%-9.7%) (P = .03) with the uncoated tube (all intubated patients, 3.8% [37/968] versus 5.8% [56/964] [P = .04]), a relative risk reduction of 35.9% (95% CI, 3.6%-69.0%). The silver-coated tube was associated with delayed occurrence of ventilator-associated pneumonia (P = .005). No statistically significant between-group differences were observed in durations of intubation, intensive care unit stay, and hospital stay; mortality; and frequency and severity of adverse events.

Watch out: The 35.9% figure is a relative reduction on a low absolute base — 2.7 percentage points (7.5% to 4.8%) — and the trial's own secondary outcomes (mortality, stay, intubation duration) all showed no difference, so the tube prevents some pneumonias without any shown effect on how patients fare overall.

Source: Kollef MH et al. Silver-coated endotracheal tubes and incidence of ventilator-associated pneumonia: the NASCENT randomized trial. JAMA. 2008. PMID: 18714060. Registry ID: NCT00148642. MeSH: Adult · Aged · Aged, 80 and over · Anti-Infective Agents, Local* · Biofilms* · Bronchoalveolar Lavage Fluid/microbiology · Equipment Contamination/prevention & control* · Female · Humans · Incidence · Intubation, Intratracheal/adverse effects/instrumentation* · Male · Middle Aged · Pneumonia, Ventilator-Associated/epidemiology/microbiology/prevention & control* · Prospective Studies · Respiration, Artificial/instrumentation · Risk Factors · Silver Compounds* · Single-Blind Method · Time Factors

Card 2 — Feasibility trial: the silver-coated tube delayed and reduced bacterial colonisation of tube and airways (Rello 2006, Critical Care Medicine — randomised feasibility study, 149 patients analysed for safety; 121 intubated ≥24 hours)

Study: Randomized controlled trial (feasibility, single-blind, multicentre)

Population: 149 eligible adults who required mechanical ventilation for 24 hours or longer and did not have respiratory infections, intubated at three hospitals in Spain and one in the United States with the silver-coated respiratory infection control device or a control endotracheal tube; 121 were intubated 24 hours or longer and analysed for tube colonisation, and 67 with negative baseline quantitative endotracheal aspirates were analysed for airway colonisation. The record names no brand or manufacturer.

Summary: Feasibility trial: the silver-coated tube delayed and reduced bacterial colonisation of tube and airways (Rello 2006, Critical Care Medicine — randomised feasibility study, 149 patients analysed for safety; 121 intubated ≥24 hours)

Result: The silver-coated device was associated with delayed colonization on the tube compared with the control device at the thresholds reported (p = .02, log-rank test) and in quantitative endotracheal aspirates, with a reduced colonization rate by days on the tube (p = .04) and lower maximal bacterial burden in tracheal aspirates for 7 days (mean log-transformed burden, 4.2 +/- 2.3 vs. 5.5 +/- 1.7 log colony-forming units/mL; p = .02). Other between-group differences favored the coated device but were not significant. Five adverse events were considered device related, including three events in the coated group and two events in the control group. The authors conclude larger studies were needed to determine whether the colonisation effects would decrease ventilator-associated pneumonia — the question NASCENT then tested.

Watch out: Colonisation is a surrogate: this trial shows the coating changes airway microbiology, not that patients get less pneumonia — that inference comes only from NASCENT, and the two trials' tubes are described generically in both records, so they cannot be confirmed as the same marketed product from the records alone.

Source: Rello J et al. Reduced burden of bacterial airway colonization with a novel silver-coated endotracheal tube in a randomized multiple-center feasibility study. Critical care medicine. 2006. PMID: 16957639. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Coated Materials, Biocompatible* · Colony Count, Microbial · Cross Infection/etiology/prevention & control · Equipment Contamination/prevention & control · Feasibility Studies · Female · Humans · Intubation, Intratracheal/adverse effects/instrumentation · Male · Middle Aged · Pneumonia, Aspiration/etiology/prevention & control · Respiration, Artificial/adverse effects/methods · Silver* · Single-Blind Method · Spain · United States

Card 3 — Synthesis — POSSIBLE: one large trial shows fewer pneumonias; nothing shows patients live longer or leave sooner

Study: Synthesis

Population: 2 randomised trials: Kollef 2008 NASCENT (2,003 patients randomised) and Rello 2006 (149 patients). Documented N = 2,152.

Summary: Synthesis — POSSIBLE: one large trial shows fewer pneumonias; nothing shows patients live longer or leave sooner

Result: The evidence chain is coherent but single-stranded: the coating delays and reduces bacterial colonisation (Rello 2006), and in the one trial powered for the clinical outcome, confirmed ventilator-associated pneumonia fell from 7.5% to 4.8% (NASCENT) — an absolute reduction of 2.7 percentage points — while mortality, intubation duration and length of stay showed no difference. A two-trial meta-analysis of these same studies exists, but it pools the same patients and adds no independent evidence. Tier C: the benefit claim rests on a single outcome trial.

Watch out: Both trials predate modern ventilator-care bundles, which have lowered baseline pneumonia rates since 2008; at today's lower rates the same relative effect would prevent fewer pneumonias per hundred tubes, and neither record names the product or its maker.

Source: Synthesis of the reports carded above; no additional source.