Completely non-invasive cardiac output monitoring vs bolus thermodilution — 1 meta-analysis (37 studies, 1,543 patients) plus 2 primary comparisons (n = 21 and n = 60): pooled percentage error was 47%, well above the 30% interchangeability criterion; in a head-to-head study the bioreactance device (NICOM) differed from the pulmonary-artery catheter by a bias of −0.71 L/min with a percentage error of 47%
Shelf #9239 · Critical care & sepsis
MEASUREMENT ACCURACY — NOT ESTABLISHED (interchangeability with thermodilution): the meta-analysis of 37 studies (1,543 patients) pooled a bias of −0.13 L/min with limits of agreement of −2.38 to +2.12 L/min and a percentage error of 47%, against the 30% criterion for accepting a new method; primary head-to-head data agree in kind (bioreactance vs pulmonary-artery catheter percentage error 47%), while one paediatric comparison against echocardiography — a different, itself imperfect reference — showed much tighter agreement in children without heart defects
Stamped: 2026-11-28.
Evidence published: 2015–2018.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In perioperative and critically ill patients, how accurately do completely non-invasive cardiac output technologies (thoracic bioimpedance/bioreactance, pulse-wave transit time, non-invasive pulse contour, carbon-dioxide rebreathing) measure cardiac output against bolus thermodilution — are they interchangeable with the reference method within the accepted 30% percentage-error criterion?
Balance of the evidence
MEASUREMENT ACCURACY — NOT ESTABLISHED (interchangeability with thermodilution) — Composition: 1 systematic review and meta-analysis (37 studies, 1,543 patients) plus 2 primary comparison studies (n = 21 adults, n = 60 children). Joosten 2017: across pulse-wave transit time, non-invasive pulse contour, bioimpedance/bioreactance and carbon-dioxide rebreathing devices, the pooled bias was −0.13 L/min, limits of agreement −2.38 to +2.12 L/min, and percentage error 47% — above the 30% threshold the field uses to judge a new method interchangeable with bolus thermodilution; only 6 of 37 studies assessed the repeatability of the tested device. Lamia 2018 (21 cardiac-surgery patients): the bioreactance device (NICOM — named as published) differed from the pulmonary-artery catheter by a bias of −0.71 L/min with limits of agreement of ±2.70 L/min and a percentage error of 47%, although the direction of changes tracked across devices. Sun 2015 (60 children): NICOM against echocardiography showed a bias of 0.03 L/min in children with normal hearts but 0.31 L/min and wider limits in children with ventricular septal defects.
MeSH: Cardiac Output · Humans · Monitoring, Physiologic · Perioperative Care · Reproducibility of Results
Cards (4)
Card 1 — Meta-analysis verdict: pooled percentage error 47% against a 30% acceptability criterion (Joosten 2017, British Journal of Anaesthesia — systematic review and meta-analysis, 37 studies, 1,543 patients)
Study: Systematic review and meta-analysis (37 studies; 1,543 patients)
Population: 37 studies (1,543 patients) comparing cardiac output by bolus thermodilution with commercially available completely non-invasive technologies: pulse-wave transit time, non-invasive pulse contour analysis, thoracic electrical bioimpedance/bioreactance, and carbon-dioxide rebreathing. Device families are described by technology class in the record; no single brand is the subject of the pooled claim.
Summary: Meta-analysis verdict: pooled percentage error 47% against a 30% acceptability criterion (Joosten 2017, British Journal of Anaesthesia — systematic review and meta-analysis, 37 studies, 1,543 patients)
Result: Mean cardiac output by both methods was 4.78 L/min. The overall random-effects pooled bias was −0.13 L/min with limits of agreement of −2.38 to +2.12 L/min, and the pooled percentage error was 47%, against the previously recommended acceptability threshold of less than 30%. Inter-study heterogeneity was high (I² = 83%, P less than 0.001), and only six studies assessed the random error (repeatability) of the tested device. The authors conclude that, with so wide a percentage error, completely non-invasive cardiac output devices are not interchangeable with bolus thermodilution.
Watch out: The 30% criterion itself assumes the reference method's own precision; thermodilution is an imperfect standard, which cuts both ways — it can inflate the apparent error of a test device, but the pooled 47% exceeds the criterion by a margin too wide to be explained by reference noise alone, and heterogeneity between studies was high.
Source: Joosten A et al. Accuracy and precision of non-invasive cardiac output monitoring devices in perioperative medicine: a systematic review and meta-analysis†. British journal of anaesthesia. 2017. PMID: 28203792. Registry ID: none in the record. MeSH: Cardiac Output · Humans · Monitoring, Physiologic · Perioperative Care · Reproducibility of Results
Card 2 — Head-to-head in cardiac surgery: bioreactance (NICOM) vs pulmonary-artery catheter — bias −0.71 L/min, percentage error 47% (Lamia 2018, Journal of Clinical Monitoring and Computing — 21 patients)
Study: Prospective device comparison study (simultaneous multi-device measurements; n = 21)
Population: 21 cardiac-surgery patients in the first 2 hours after surgery, with simultaneous cardiac-output estimates from a bolus thermodilution pulmonary-artery catheter, three arterial pulse-contour systems (LiDCOplus, FloTrac and PiCCOplus — named as published) and a bioreactance device (NICOM — named as published); 21 simultaneous measurement sets were made before and after therapeutic interventions.
Summary: Head-to-head in cardiac surgery: bioreactance (NICOM) vs pulmonary-artery catheter — bias −0.71 L/min, percentage error 47% (Lamia 2018, Journal of Clinical Monitoring and Computing — 21 patients)
Result: Mean pulmonary-artery catheter cardiac output was 5.7 ± 1.5 L/min and NICOM output 5.3 ± 1.0 L/min. The bias between the catheter and NICOM was −0.71 L/min with limits of agreement of ±2.70 L/min and a percentage error of 47%; pairwise percentage errors among all devices ranged from 33% to 59%. Dynamic changes in cardiac output showed good cross-correlations between devices. The authors conclude all devices recorded similar mean values that changed in similar directions, but with markedly different bias and precision, so results from one device cannot validate another.
Watch out: Twenty-one patients — N printed on the face of this card. Trending (direction of change) performed better than absolute agreement; the card's accuracy figures concern absolute values, which is the basis of the interchangeability claim being tested.
Source: Lamia B et al. Cross-comparisons of trending accuracies of continuous cardiac-output measurements: pulse contour analysis, bioreactance, and pulmonary-artery catheter. Journal of clinical monitoring and computing. 2018. PMID: 28188408. Registry ID: none in the record. MeSH: Aged · Algorithms · Analysis of Variance · Cardiac Output · Cardiac Surgical Procedures · Catheterization, Swan-Ganz · Female · Heart Rate · Humans · Linear Models · Male · Middle Aged · Monitoring, Physiologic · Postoperative Period · Pulmonary Artery · Reproducibility of Results · Thermodilution · Time Factors
Card 3 — The paediatric exception: NICOM vs echocardiography agreed tightly in children with normal hearts, less well with a septal defect (Sun 2015, Paediatric Anaesthesia — 60 children)
Study: Prospective device comparison study (bioreactance vs echocardiography; n = 60)
Population: 60 children during anaesthesia induction: 28 with normal heart anatomy and 32 with isolated ventricular septal defects; cardiac output was measured minute by minute simultaneously by the bioreactance device (NICOM — named as published) and by echocardiography (Simpson's rule).
Summary: The paediatric exception: NICOM vs echocardiography agreed tightly in children with normal hearts, less well with a septal defect (Sun 2015, Paediatric Anaesthesia — 60 children)
Result: Measurements by the two techniques were highly correlated in children with normal hearts (r = 0.96, P less than 0.005) and in those with septal defects (r = 0.84, P less than 0.005). Mean bias (echocardiography minus NICOM) was 0.03 L/min with limits of agreement of −0.29 to +0.35 L/min in the normal-heart group, and 0.31 L/min with limits of −0.44 to +1.05 L/min in the septal-defect group; median percent errors were significantly lower at all time points in the normal-heart group. The authors conclude NICOM agrees well with echocardiography in children without heart defects but underestimates it in children with a ventricular septal defect.
Watch out: The reference here is echocardiography, not thermodilution, and echocardiographic cardiac output carries its own error; tight agreement with echo in anaesthetised children does not establish agreement with the adult reference standard tested in the other two cards.
Source: Sun Y et al. Noninvasive cardiac output monitoring using bioreactance-based technique in pediatric patients with or without ventricular septal defect during anesthesia: in comparison with echocardiography. Paediatric anaesthesia. 2015. PMID: 25039420. Registry ID: none in the record. MeSH: Anesthesia · Cardiac Output · Child, Preschool · Echocardiography · Female · Heart Septal Defects, Ventricular · Humans · Infant · Male · Monitoring, Physiologic · Reproducibility of Results · Thermodilution
Card 4 — Synthesis — MEASUREMENT ACCURACY — NOT ESTABLISHED for absolute values: small average bias, wide individual disagreement
Study: Synthesis
Population: 1 meta-analysis (37 studies, 1,543 patients) plus 2 primary comparisons (n = 21 against thermodilution; n = 60 against echocardiography).
Summary: Synthesis — MEASUREMENT ACCURACY — NOT ESTABLISHED for absolute values: small average bias, wide individual disagreement
Result: The pooled bias against thermodilution is small (−0.13 L/min) only because errors cancel: the limits of agreement span more than 4 L/min and the percentage error is 47%, against the 30% interchangeability criterion. The one tight agreement on the shelf uses a different reference (echocardiography), a different population (anaesthetised children), and deteriorates with an intracardiac shunt. Direction-of-change tracking is consistently better than absolute accuracy in every report. A non-invasive cardiac output number should therefore be read as a trend, not as a thermodilution-equivalent value.
Watch out: No sensitivity/specificity pair is reported by any carded record, so none is quoted; the accuracy currency of this shelf is bias, limits of agreement and percentage error. Invasive and minimally invasive (pulse-contour calibrated) monitoring are different questions and are not covered by this shelf.
Source: Synthesis of the reports carded above; no additional source.