SourceMed · Shelf #9240 in the ask interface

Handheld spirometry (FEV1/FEV6 devices, PiKo-6) for detecting airflow obstruction — 3 primary validation studies (documented N = 297, 2,397 screened with 267 verified, and 241): at a ratio cut-off below 0.75 in primary-care case finding, sensitivity was 81% and specificity 71%; in a smoker screening programme at a cut-off below 0.70, sensitivity was 97.8% and specificity 93.8% in the subgroup recalled for full spirometry; a disposable handheld device classified obstruction severity with 87.1% accuracy overall

Shelf #9240 · Lungs, breathing & sleep

PREDICTIVE PERFORMANCE — adequate for case finding and triage to full spirometry, cut-off- and spectrum-dependent: in primary-care case finding (n = 297) the PiKo-6 at an FEV1/FEV6 cut-off below 0.75 gave sensitivity 81% / specificity 71% (labelled in words at first occurrence throughout this shelf), in a smoker screening programme the same device family at a cut-off below 0.70 gave sensitivity 97.8% / specificity 93.8% in the verified subgroup, and a disposable handheld device classified obstruction with 87.1% accuracy (93.7% in subjects able to exhale for 6 seconds) — performance good enough to decide who needs diagnostic spirometry, not to replace it

Stamped: 2026-11-28.

Evidence published: 2009–2018.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults at risk of chronic obstructive pulmonary disease, how accurately does handheld spirometry using the FEV1/FEV6 ratio (the PiKo-6 device and a disposable pneumotachograph device) detect airflow obstruction against conventional diagnostic spirometry (post-bronchodilator FEV1/FVC) — what sensitivity and specificity does it achieve, and at what cut-off?

Balance of the evidence

PREDICTIVE PERFORMANCE — adequate for case finding and triage to full spirometry, cut-off- and spectrum-dependent — Composition: 3 prospective primary validation studies (n = 297; n = 2,397 screened with 267 verified against full spirometry; n = 241). Frith 2011: in smokers aged 50 or over, the PiKo-6 FEV1/FEV6 ratio gave areas under the receiver operating characteristic curve of 0.85 (case finding) and 0.88 (differential diagnosis); at a cut-off below 0.75, sensitivity was 81% and specificity 71% for case finding, and sensitivity 86% / specificity 67% against an asthma-diagnosis comparison group. Duong-Quy 2009: at a cut-off below 0.70, sensitivity was 97.8% and specificity 93.8% — but only 267 of 2,397 screened smokers were recalled for the reference test, so those figures carry verification bias. Chen 2018: a disposable handheld pneumotachograph device classified obstruction severity with 87.1% accuracy in all 241 subjects, rising to 93.7% in the 159 able to exhale for at least 6 seconds at an FEV1/FEV6 cut-off of 74%.

MeSH: Aged · Early Diagnosis · Equipment Design · Female · Forced Expiratory Volume · Humans · Male · Mass Screening · Middle Aged · New South Wales · Prevalence · Primary Health Care · Prospective Studies · Pulmonary Disease, Chronic Obstructive · Sensitivity and Specificity · Severity of Illness Index · Spirometry · Surveys and Questionnaires

Cards (4)

Card 1 — Primary-care validation: at FEV1/FEV6 below 0.75, sensitivity 81% and specificity 71% for case finding (Frith 2011, Primary Care Respiratory Journal — 297 subjects)

Study: Prospective validation study (handheld device vs diagnostic spirometry; n = 297)

Population: 297 current and former smokers aged 50 years or over in primary care: 204 with no previous respiratory diagnosis (case finding) and 93 with an asthma diagnosis (differential diagnosis); all had pre-bronchodilator FEV1/FEV6 by the handheld PiKo-6 (nSpire Health, Inc. — named as published) and post-bronchodilator FEV1/FVC by conventional spirometry as the reference.

Summary: Primary-care validation: at FEV1/FEV6 below 0.75, sensitivity 81% and specificity 71% for case finding (Frith 2011, Primary Care Respiratory Journal — 297 subjects)

Result: The PiKo-6 FEV1/FEV6 showed sensitivity and specificity summarised by areas under the receiver operating characteristic curves of 0.85 (95% CI 0.79 to 0.90) for case finding and 0.88 (95% CI 0.80 to 0.96) for differential diagnosis, exceeding the accuracy of validated questionnaires. An FEV1/FEV6 cut-off below 0.75 provided optimal sensitivity (case finding 81%; differential diagnosis 86%) and specificity (case finding 71%; differential diagnosis 67%) for COPD screening. The authors conclude the device allows simple and reliable screening that could optimise early referral for spirometry.

Watch out: The reference diagnosis uses post-bronchodilator spirometry while the handheld test is pre-bronchodilator, so part of the disagreement is built into the design; a specificity of 71% means roughly 3 in 10 unobstructed smokers screen positive and need the full test anyway.

Source: Frith P et al. Simplified COPD screening: validation of the PiKo-6® in primary care. Primary care respiratory journal : journal of the General Practice Airways Group. 2011. PMID: 21597667. Registry ID: none in the record. MeSH: Aged · Early Diagnosis · Equipment Design · Female · Forced Expiratory Volume · Humans · Male · Mass Screening · Middle Aged · New South Wales · Prevalence · Primary Health Care · Prospective Studies · Pulmonary Disease, Chronic Obstructive · Sensitivity and Specificity · Severity of Illness Index · Spirometry · Surveys and Questionnaires

Card 2 — Smoker screening programme: sensitivity 97.8% and specificity 93.8% at a cut-off below 0.70 — in the 267 of 2,397 recalled for verification (Duong-Quy 2009, Revue des Maladies Respiratoires — 2,397 screened)

Study: Screening programme validation (partial verification; 2,397 screened, 267 verified)

Population: 2,397 smokers over 40 years of age with more than 10 years of smoking (2,130 active smokers, 267 ex-smokers; mean age 52 ± 13 years; mean 24 ± 13 pack-years) screened with the Piko-6 device (named as published); smokers in the lowest ratio group and a sample of the other groups (267 in total) were recalled for full spirometric assessment as the reference.

Summary: Smoker screening programme: sensitivity 97.8% and specificity 93.8% at a cut-off below 0.70 — in the 267 of 2,397 recalled for verification (Duong-Quy 2009, Revue des Maladies Respiratoires — 2,397 screened)

Result: The prevalence of COPD detected by the Piko-6 in the study population was 13.5%. For the threshold FEV1/FEV6 below 0.70 and with the detected prevalence, the Piko-6 had a sensitivity of 97.8%, a specificity of 93.8%, a positive predictive value of 71% and a negative predictive value of 99.6% (95% confidence intervals as reported). The authors conclude the Piko-6 is a useful tool for early screening for COPD in smokers in a developing-country setting.

Watch out: Only 267 of 2,397 screened smokers received the reference test, and they were selected by their screening result — a classic verification-bias design that inflates sensitivity and specificity relative to the fully verified primary-care study on this shelf. The article is published in French; its record's abstract is in English and every figure carded appears in that abstract.

Source: Duong-Quy S et al. [Early detection of smoking related chronic obstructive pulmonary disease in Vietnam]. Revue des maladies respiratoires. 2009. PMID: 19367200. Registry ID: none in the record. MeSH: Adult · Aged · Early Diagnosis · Feasibility Studies · Female · Forced Expiratory Volume · Humans · Male · Mass Screening · Middle Aged · Predictive Value of Tests · Pulmonary Disease, Chronic Obstructive · Sensitivity and Specificity · Smoking · Spirometry · Vietnam

Card 3 — A disposable handheld device: 87.1% classification accuracy overall, 93.7% in subjects who could exhale for 6 seconds (Chen 2018, International Journal of COPD — 241 subjects)

Study: Prospective device agreement and classification study (n = 241)

Population: 241 adult Chinese subjects ranging from healthy to mixed levels of pulmonary disease, performing spirometry on a conventional body plethysmograph (the reference) and on a handheld disposable pneumotachograph device (D-PNEU — described as published; the record names no manufacturer) in randomised order; the 159 subjects able to exhale for at least 6 seconds were analysed separately.

Summary: A disposable handheld device: 87.1% classification accuracy overall, 93.7% in subjects who could exhale for 6 seconds (Chen 2018, International Journal of COPD — 241 subjects)

Result: Bland–Altman analysis showed between-device agreement for key airflow metrics within clinically acceptable limits. Using FEV1/FVC below 70% as the reference definition of obstruction, the handheld device had 87.1% accuracy in classifying the severity of obstruction in all 241 subjects; in the 159 subjects able to exhale for at least 6 seconds, accuracy was 93.7% at an FEV1/FEV6 cut-off point of 74%. The authors conclude the disposable handheld device can accurately identify and quantify airway obstruction in at-risk patients, while cautioning that all available brands should be tested.

Watch out: This record reports overall classification accuracy, not a sensitivity/specificity pair, and none is derived on this shelf; accuracy also depends on the 6-second exhalation the FEV6 manoeuvre requires, which the most obstructed patients are least able to perform.

Source: Chen G et al. The accuracy of a handheld "disposable pneumotachograph device" in the spirometric diagnosis of airway obstruction in a Chinese population. International journal of chronic obstructive pulmonary disease. 2018. PMID: 30122915. Registry ID: none in the record. MeSH: Adult · Airway Obstruction · China · Disposable Equipment · Forced Expiratory Volume · Humans · Plethysmography · Pulmonary Disease, Chronic Obstructive · ROC Curve · Random Allocation · Sensitivity and Specificity · Spirometry · Vital Capacity

Card 4 — Synthesis — PREDICTIVE PERFORMANCE: a triage test whose numbers move with cut-off, spectrum and verification

Study: Synthesis

Population: 3 primary validation studies: Frith 2011 (n = 297, fully verified), Duong-Quy 2009 (2,397 screened, 267 verified), Chen 2018 (n = 241).

Summary: Synthesis — PREDICTIVE PERFORMANCE: a triage test whose numbers move with cut-off, spectrum and verification

Result: Fully verified primary-care data give the conservative estimate: sensitivity 81% / specificity 71% at an FEV1/FEV6 cut-off below 0.75. Loosening the design (partial verification, a screening spectrum of heavy smokers) raises the apparent figures to sensitivity 97.8% / specificity 93.8%, and a second device family classifies obstruction with 87.1–93.7% accuracy. Every report positions the handheld result as a route to confirmatory spirometry, not a diagnosis. Read the shelf's headline performance as the Frith figures; treat the higher pair as an upper bound set by verification bias.

Watch out: All evidence is for FEV1/FEV6 handheld devices in smoker or at-risk spectra; the figures do not transfer to general-population screening, to children, or to restrictive disease, and device brands other than those tested should not borrow these numbers.

Source: Synthesis of the reports carded above; no additional source.