Point-of-care haemoglobin by HemoCue versus laboratory haemoglobin — 4 paired-measurement studies (documented N = 344 subjects): average bias is small (−0.18 to +0.3 g/dL where reported) but the 95% limits of agreement span roughly −1.6 to +2.3 g/dL, and in the transfusion-relevant 6–10 g/dL range 5.3% of HemoCue readings fell outside a 10% error zone
Shelf #9288 · Critical care & sepsis
MEASUREMENT ACCURACY — close on average, NOT ESTABLISHED as interchangeable with laboratory haemoglobin for individual decisions: across 4 studies (344 subjects) HemoCue bias was −0.18 to +0.3 g/dL, but 95% limits of agreement reached about −1.6 to +2.3 g/dL — a spread wide enough to cross transfusion thresholds — and one surgical study found all point-of-care devices, HemoCue included, underestimated the laboratory value and could not be used interchangeably with it, even though HemoCue was the most precise of the devices tested
Stamped: 2026-11-28.
Evidence published: 2012–2020.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately does the HemoCue point-of-care haemoglobinometer (HemoCue Hb 201+ / Hb-201, named as published) measure haemoglobin against a laboratory haematology analyser — what are its bias and limits of agreement in intensive care, emergency, outpatient and surgical settings, and is it interchangeable with the laboratory value for transfusion decisions?
Balance of the evidence
MEASUREMENT ACCURACY — close on average, NOT ESTABLISHED as interchangeable — Composition: 4 paired-measurement studies (344 subjects; 48–292 specimens). Chutipongtanate 2020 (35 surgical-ICU patients, 48 specimens): HemoCue Hb201+ (named as published) versus central laboratory — correlation r = 0.922, bias −0.18 g/dL, 95% limits of agreement −1.63 to +1.28 g/dL, better than non-invasive SpHb (limits −3.12 to +3.39) and cheaper per test than the laboratory. Zatloukal 2016 (99 emergency patients, 292 measurements): HemoCue 201+ mean 115 ± 33 g/L versus laboratory 111 ± 30 g/L, r² = 0.93, bias −3.7 g/L with limits −20.9 to +13.5 g/L; 94% of values in the acceptable error-grid zone, and the authors call absolute accuracy low despite the high correlation. Raikhel 2012 (152 outpatients): HemoCue 201+ bias +0.3 ± 1.0 g/dL, limits −1.7 to +2.3 g/dL, similar to non-invasive SpHb. Johnson 2020 (58 cardiothoracic patients, 70 laboratory measurements): all point-of-care devices underestimated laboratory haemoglobin; HemoCue had the narrowest limits and 5.3% of readings beyond the 10% error zone in the 6–10 g/dL range versus 19.0% and 16.0% for two blood-gas analysers.
MeSH: Hemoglobinometry · Point-of-Care Systems · Hemoglobins · Reproducibility of Results · Blood Transfusion
Cards (4)
Card 1 — Surgical ICU: HemoCue bias −0.18 g/dL with limits of agreement −1.63 to +1.28 g/dL — the best sole-purpose device of four point-of-care methods (Chutipongtanate 2020, BMC Anesthesiology — 35 patients, 48 specimens)
Study: Method-comparison study (point-of-care versus laboratory; n = 35 patients, 48 specimens)
Population: 35 patients in a surgical intensive care unit (48 blood specimens) in Thailand, each specimen measured by four point-of-care methods — capillary haematocrit, HemoCue Hb201+, iSTAT with CG8+ cartridge, and SpHb from the Radical-7 pulse co-oximeter (all named as published) — against central laboratory haemoglobin (mean 11.2 ± 1.9 g/dL).
Summary: Surgical ICU: HemoCue bias −0.18 g/dL with limits of agreement −1.63 to +1.28 g/dL — the best sole-purpose device of four point-of-care methods (Chutipongtanate 2020, BMC Anesthesiology — 35 patients, 48 specimens)
Result: Ranked by correlation, bias and limits of agreement: iSTAT first (r = 0.941; bias +0.15 g/dL; limits −1.41 to +1.12), then HemoCue (r = 0.922; bias −0.18 g/dL; limits −1.63 to +1.28), then capillary haematocrit (r = 0.905; bias +0.46; limits −1.16 to +2.08), then SpHb (r = 0.670; bias +0.13; limits −3.12 to +3.39). All point-of-care methods had less workload and turnaround time than the laboratory; only HemoCue and capillary haematocrit had lower unit cost. The authors identify HemoCue as the suitable point-of-care method for the sole purpose of haemoglobin measurement in this setting.
Watch out: Only 35 patients and 48 specimens from one surgical ICU — the ranking is a single-centre result, and even the best device's limits of agreement span nearly 3 g/dL.
Source: Chutipongtanate A et al. Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: a cross-sectional method comparison study. BMC anesthesiology. 2020. PMID: 32321425. Registry ID: none in the record. MeSH: Aged · Cross-Sectional Studies · Female · Hematocrit · Hemoglobins · Humans · Intensive Care Units · Laboratories · Male · Middle Aged · Oximetry · Point-of-Care Systems · Reproducibility of Results · Thailand · Time Factors
Card 2 — Emergency department: correlation r² = 0.93 but bias −3.7 g/L with limits −20.9 to +13.5 g/L — 'absolute accuracy low' (Zatloukal 2016, Journal of Clinical Monitoring and Computing — 99 patients, 292 measurements)
Study: Method-comparison study (emergency department; n = 99 patients, 292 measurements)
Population: 99 patients admitted to an emergency department, their blood samples analysed by laboratory reference (Beckman Coulter LH 750 — named as published), HemoCue 201+ and Gem Premier 3000, with transfusion-decision reliability assessed by a three-zone error grid.
Summary: Emergency department: correlation r² = 0.93 but bias −3.7 g/L with limits −20.9 to +13.5 g/L — 'absolute accuracy low' (Zatloukal 2016, Journal of Clinical Monitoring and Computing — 99 patients, 292 measurements)
Result: Mean haemoglobin was 115 ± 33 g/L (HemoCue), 110 ± 28 g/L (Gem Premier) and 111 ± 30 g/L (laboratory). Correlations with the laboratory were high (HemoCue r² = 0.93; Gem r² = 0.86; both P less than 0.001), but Bland–Altman bias and limits were −3.7 g/L (limits −20.9 to +13.5) for HemoCue and +2.5 g/L (limits −18.6 to +23.5) for the Gem device. On the error grid, 94% (HemoCue) and 91% (Gem) of values fell in the acceptable zone and 0% and 1% in the unacceptable zone. The authors conclude the absolute accuracy of both point-of-care devices was low despite the high correlation, with unfavourable error-grid results.
Watch out: Values are in g/L as the record prints them (115 g/L = 11.5 g/dL): the limits of agreement of about ±20 g/L are about ±2 g/dL — the same individual-reading spread as the other cards, in different units.
Source: Zatloukal J et al. Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room. Journal of clinical monitoring and computing. 2016. PMID: 26507548. Registry ID: none in the record. MeSH: Adult · Aged · Automation · Blood Transfusion · Emergency Medicine · Emergency Service, Hospital · Female · Hematology · Hemoglobinometry · Hemoglobins · Hemorrhage · Humans · Male · Middle Aged · Point-of-Care Testing · Prospective Studies · Reproducibility of Results
Card 3 — Outpatients: HemoCue bias +0.3 ± 1.0 g/dL, limits −1.7 to +2.3 g/dL — no tighter than non-invasive pulse co-oximetry in this study (Raikhel 2012, Postgraduate Medicine — 152 subjects)
Study: Method-comparison study (outpatient; n = 152)
Population: 152 adult outpatients (average age 46 years; 69% female) at a research clinic, each tested three ways: non-invasive pulse co-oximetry (Pronto SpHb, Masimo Corp. — named as published), finger-stick HemoCue 201+ (HemoCue, Inc. — named as published), and venous laboratory haematology analyser as reference.
Summary: Outpatients: HemoCue bias +0.3 ± 1.0 g/dL, limits −1.7 to +2.3 g/dL — no tighter than non-invasive pulse co-oximetry in this study (Raikhel 2012, Postgraduate Medicine — 152 subjects)
Result: Bias ± standard deviation versus the laboratory was −0.5 ± 1.0 g/dL for SpHb and +0.3 ± 1.0 g/dL for HemoCue 201+. Bland–Altman limits of agreement were −2.5 to +1.5 g/dL for SpHb and −1.7 to +2.3 g/dL for HemoCue. A non-invasive measurement could not be obtained in 4 subjects (2.5% failure rate); HemoCue measurements were obtained for all subjects. The author concludes non-invasive testing had bias and spread similar to HemoCue.
Watch out: A single-author study in stable outpatients with mostly normal haemoglobin values: agreement at the low values where transfusion is decided is extrapolated, not demonstrated, in this population.
Source: Raikhel M et al. Accuracy of noninvasive and invasive point-of-care total blood hemoglobin measurement in an outpatient setting. Postgraduate medicine. 2012. PMID: 22913913. Registry ID: none in the record. MeSH: Adult · Ambulatory Care · Female · Hematologic Tests · Hemoglobinometry · Hemoglobins · Humans · Male · Oximetry · Point-of-Care Systems · Prospective Studies · Spectrophotometry
Card 4 — At transfusion range, HemoCue errs least but still errs: 5.3% of readings beyond the 10% error zone at 6–10 g/dL, and every device underestimated the laboratory (Johnson 2020, Anesthesia & Analgesia — 58 surgical patients)
Study: Method-comparison study (transfusion-decision range; n = 58 patients, 70 laboratory measurements)
Population: 58 adult cardiothoracic-surgery patients; 70 central-laboratory measurements (Siemens Advia 2120 — named as published; range 5.8–16.7 g/dL, 36% below 10.0 g/dL) compared with two Ilex GEM Premier 3500 blood-gas analysers and a HemoCue Hb-201 device; a between-method difference over 10% in the 6–10 g/dL range was prespecified as likely to influence transfusion decisions erroneously.
Summary: At transfusion range, HemoCue errs least but still errs: 5.3% of readings beyond the 10% error zone at 6–10 g/dL, and every device underestimated the laboratory (Johnson 2020, Anesthesia & Analgesia — 58 surgical patients)
Result: All three point-of-care devices showed statistically significant bias with predominant underestimation of laboratory haemoglobin, and the authors state none can be used interchangeably with standard laboratory measurement. HemoCue had narrower limits of agreement and a narrower mountain plot (greater precision) than either blood-gas analyser. In the 6–10 g/dL range, 5.3% of HemoCue measurements were beyond the permissible 10% error zone, versus 19.0% and 16.0% for the blood-gas analysers; erroneous decisions generally erred toward unnecessary transfusion, and chance-corrected agreement on transfusion was better for HemoCue.
Watch out: The 10% error zone and its transfusion consequence are the authors' prespecified judgement, labelled as such here; the study's own bottom line — not interchangeable with the laboratory — is the shelf's headline, not the device ranking.
Source: Johnson M et al. Comparison of Hemoglobin Measurements by 3 Point-of-Care Devices With Standard Laboratory Values and Reliability Regarding Decisions for Blood Transfusion. Anesthesia and analgesia. 2020. PMID: 31764158. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Blood Transfusion · Female · Hematologic Tests · Hemoglobinometry · Hemoglobins · Humans · Male · Middle Aged · Point-of-Care Testing · Prospective Studies · Reference Standards · Reproducibility of Results