Peak flow meter accuracy against the pneumotachograph — 4 evaluation studies (documented human N = 182, plus bench testing): meters disagree with the reference and with each other by make — one meter underestimated by a mean of 27 L/min and another overestimated by 45 L/min, the American and European Mini-Wright scales differ by 15 L/min in the same volunteers, and portable devices read peak cough flow about 50 L/min low with limits of agreement spanning about 150 L/min
Shelf #9290 · Lungs, breathing & sleep
MEASUREMENT ACCURACY — device- and scale-dependent, NOT ESTABLISHED as interchangeable: in 4 evaluation studies (182 human subjects plus bench series) the best adult meters agreed closely with the pneumotachograph but others carried mean biases of −27 to +45 L/min, the two official Mini-Wright scales (American Thoracic Society versus European Union — named as published) read 541 versus 526 L/min in the same volunteers and are stated by the authors not to be interchangeable, and for peak cough flow portable devices read about 50 L/min low with 95% limits of agreement spanning about 150 L/min
Stamped: 2026-11-28.
Evidence published: 1990–2015.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do portable peak expiratory flow meters (the named makes and scales as published) measure peak flow against a pneumotachograph reference — how large are the between-meter and between-scale differences, and are readings from different meters interchangeable?
Balance of the evidence
MEASUREMENT ACCURACY — device- and scale-dependent, NOT ESTABLISHED as interchangeable — Composition: 4 evaluation studies (50 adults + 25 children; 57 volunteers; 20 volunteers; 30 patients with airflow obstruction; plus bench testing of meter units). Folgering 1998 (11 meters — 7 adult, 4 low-range — in series with a Fleisch #4 pneumotachograph, named as published): substantial differences in quality among adult meters; Personal Best and Micro Medical (named as published) agreed most closely with the pneumotachograph in adults and in low-range versions. Pretto 1990 (30 subjects): against the pneumotachograph, the Breath-Taker meter underestimated by a mean of 27 L/min, the mini-Wright overestimated by a mean of 45 L/min, and the Wright meter was not significantly different; bench coefficients of variation were 4.8% (Wright), 8.4% (Breath-Taker) and 7.5% (mini-Wright). Pesola 2010 (57 volunteers): American-scale Mini-Wright read 541 ± 114 L/min versus European-scale 526 ± 112 L/min (mean bias 15 ± 15 L/min; concordance correlation 0.98), reading +3.0% and −2.0% versus the pneumotachometer on the bench — not interchangeable, per the authors. Kulnik 2015 (20 volunteers, peak cough flow): portable devices read on average about 50 L/min lower than the pneumotachograph, with 95% limits of agreement spanning about 150 L/min.
MeSH: Peak Expiratory Flow Rate · Flowmeters · Respiratory Function Tests · Spirometry
Cards (4)
Card 1 — Eleven meters ranked against a pneumotachograph: substantial quality differences; Personal Best and Micro Medical closest in adults and low-range versions (Folgering 1998, European Respiratory Journal — 50 adults and 25 children, plus bench series)
Study: Device evaluation study (11 meters versus pneumotachograph; n = 75 human subjects plus bench series)
Population: Seven adult peak flow meters (Miniwright, Personal Best, Wright Pocket, Vitalograph, Assess, Pocket Peak flow meter, Truzone — makers as published) and four low-range meters, tested in series downstream of a Fleisch #4 pneumotachograph (named as published): one subject performed 50 partial forced expiratory manoeuvres, and 50 adult patients and 25 healthy children performed sequential maximal manoeuvres on each meter and the pneumotachograph.
Summary: Eleven meters ranked against a pneumotachograph: substantial quality differences; Personal Best and Micro Medical closest in adults and low-range versions (Folgering 1998, European Respiratory Journal — 50 adults and 25 children, plus bench series)
Result: Substantial differences in the quality of the adult meters were found under a ranking system for agreement with the pneumotachograph. The adult meters with the closest agreement were Personal Best and Micro Medical; among low-range meters the lowest differences were seen in the LR Personal Best and LR Micro Medical.
Watch out: The record reports the ranking without printing each meter's numeric bias, so no per-meter error is carded — the card establishes that make matters, and the size of the disagreement comes from the other cards on this shelf.
Source: Folgering H et al. Eleven peak flow meters: a clinical evaluation. The European respiratory journal. 1998. PMID: 9543292. Registry ID: none in the record. MeSH: Adult · Child · Evaluation Studies as Topic · Humans · Peak Expiratory Flow Rate · Rheology
Card 2 — Same lungs, three meters, three answers: Breath-Taker −27 L/min, mini-Wright +45 L/min, Wright not significantly different from the pneumotachograph (Pretto 1990, Medical Journal of Australia — 30 subjects)
Study: Device evaluation study (bench plus 30 subjects)
Population: 30 subjects with various degrees of irreversible airflow obstruction, each performing at least three reproducible peak-flow manoeuvres on a computerised pneumotachograph system and on each of three meter types (Wright, mini-Wright, Breath-Taker — named as published) in randomised order; bench reproducibility of five units of each type was measured with an explosive decompression device across six flow rates (97–622 L/min).
Summary: Same lungs, three meters, three answers: Breath-Taker −27 L/min, mini-Wright +45 L/min, Wright not significantly different from the pneumotachograph (Pretto 1990, Medical Journal of Australia — 30 subjects)
Result: Between-unit reproducibility on the bench: mean coefficient of variation 4.8% for Wright meters (falling with flow), 8.4% for Breath-Taker (falling with flow) and 7.5% for mini-Wright (rising with flow); the Breath-Taker had less inter-unit variability than the mini-Wright above 200 L/min. In subjects, against the pneumotachograph, the Breath-Taker underestimated peak flow by a mean of 27 L/min and the mini-Wright overestimated by a mean of 45 L/min, while the Wright meter was not significantly different; the Breath-Taker's error was independent of flow rate, so a scale offset would correct it.
Watch out: A 1990 evaluation of meter versions then on the market: the specific biases belong to those versions, but the structural finding — fixed, make-specific offsets of tens of litres per minute — is what the shelf carries forward.
Source: Pretto JJ et al. An evaluation of the Breath-Taker peak flow meter. The Medical journal of Australia. 1990. PMID: 2093803. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Evaluation Studies as Topic · Humans · Middle Aged · Peak Expiratory Flow Rate · Respiratory Function Tests
Card 3 — Two official scales, one meter family, a 15 L/min gap: American-scale Mini-Wright 541 versus European-scale 526 L/min in the same volunteers — 'not interchangeable' (Pesola 2010, Journal of Asthma — 57 volunteers)
Study: Device comparison study (two scales, bench plus 57 volunteers)
Population: 57 nonsmoking adult volunteers (average age 37 ± 12 years) performing three standing peak flows on each of the American Thoracic Society-scale and European Union-scale Mini Wright meters (named as published) in random order, best of three analysed; bench accuracy of each meter against a pneumotachometer in series.
Summary: Two official scales, one meter family, a 15 L/min gap: American-scale Mini-Wright 541 versus European-scale 526 L/min in the same volunteers — 'not interchangeable' (Pesola 2010, Journal of Asthma — 57 volunteers)
Result: Average peak flows differed: 541 ± 114 L/min (American scale) versus 526 ± 112 L/min (European scale; P less than 0.01), and both differed from spirometrically predicted values (483 ± 86 L/min). Concordance correlation was 0.98 and the Bland–Altman mean bias was 15 ± 15 L/min with no increase in scatter across the range. On the bench, the American meter read 3.0 ± 2.1% above and the European meter 2.0 ± 1.5% below the pneumotachometer. The authors conclude the American meter reads 2.8% higher across flows and the two meters are not interchangeable.
Watch out: Healthy volunteers with high flows: a fixed 15 L/min scale gap is a smaller proportion at 540 L/min than at the 150–250 L/min flows typical of obstructed patients, where the same gap is proportionally larger.
Source: Pesola GR et al. Comparison of the ATS versus EU Mini Wright peak flow meter in normal volunteers. The Journal of asthma : official journal of the Association for the Care of Asthma. 2010. PMID: 21039207. Registry ID: none in the record. MeSH: Adult · Female · Flowmeters · Humans · Male · Random Allocation · Respiratory Function Tests · Spirometry · Statistics, Nonparametric
Card 4 — Peak cough flow reads low on portables: about 50 L/min below the pneumotachograph, limits of agreement spanning about 150 L/min (Kulnik 2015, Physiological Measurement — 20 healthy volunteers)
Study: Device accuracy study (peak cough flow; n = 20)
Population: 20 healthy volunteers (mean age 45 ± 16 years) coughing through a calibrated laboratory pneumotachograph connected in series with each of four portable devices in turn (Mini-Wright and Assess peak flow meters, SpiroUSB and Microlab spirometers — named as published), plus mechanically generated constant-peak-flow waves.
Summary: Peak cough flow reads low on portables: about 50 L/min below the pneumotachograph, limits of agreement spanning about 150 L/min (Kulnik 2015, Physiological Measurement — 20 healthy volunteers)
Result: Agreement between portable-device and pneumotachograph peak cough flow readings was poor: readings were on average lower by approximately 50 L/min on the portable devices, and the 95% limits of agreement spanned approximately 150 L/min. The authors conclude absolute peak cough flow values reported with different instruments may not be directly comparable.
Watch out: Twenty subjects — N printed on the face of this card. Peak cough flow is a different manoeuvre from peak expiratory flow (a cough, not a forced expiration): the card extends the shelf's warning to cough-flow thresholds used in neuromuscular care, it does not re-measure the expiratory meters.
Source: Kulnik ST et al. Accuracy of portable devices in measuring peak cough flow. Physiological measurement. 2015. PMID: 25582526. Registry ID: none in the record. MeSH: Cough · Female · Flowmeters · Forced Expiratory Flow Rates · Healthy Volunteers · Humans · Male · Middle Aged · Peak Expiratory Flow Rate · Spirometry