Non-invasive estimation of central aortic blood pressure versus invasive catheter measurement — 4 validation studies (documented N = 191): transfer-function devices correlate strongly (r about 0.90–0.91) but underestimate central systolic pressure — by 13.3 mmHg with cuff calibration in one study and 15 mmHg in another — the error is mostly calibration, not the transfer function (invasive calibration: −2.3 ± 5.6 mmHg), and the cuff-based XCEL device did not meet the ARTERY Society accuracy criterion (mean difference ± SD within 5 ± 8 mmHg) in severe aortic stenosis
Shelf #9291 · Heart, vessels, blood pressure & clotting
MEASUREMENT ACCURACY — NOT ESTABLISHED for individual central systolic pressure: across 4 catheter-validation studies (191 subjects) estimates correlate strongly with invasive central pressure (r about 0.90–0.91) but systematically underestimate systolic pressure by 2–15 mmHg depending on device and calibration, with individual spread (SD 13–17 mmHg in severe aortic stenosis) far beyond the ARTERY Society's acceptable mean difference ± SD of 5 ± 8 mmHg; the transfer function itself is accurate when calibrated invasively (−2.3 ± 5.6 mmHg), so the dominant error is the non-invasive brachial calibration
Stamped: 2026-11-28.
Evidence published: 1979–2021.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do non-invasive devices that estimate central aortic blood pressure from a radial or brachial waveform (SphygmoCor tonometry, Omron HEM9000-AI, SphygmoCor XCEL cuff — named as published) reproduce central pressure measured by an invasive aortic catheter — what are the bias and spread, and how much of the error comes from the brachial-cuff calibration rather than the transfer function?
Balance of the evidence
MEASUREMENT ACCURACY — NOT ESTABLISHED for individual central systolic pressure — Composition: 4 simultaneous invasive-validation studies (191 subjects). Cloud 2003 (30 catheterisation patients): the SphygmoCor estimate (named as published), calibrated with non-invasive brachial cuff pressure, underestimated central systolic pressure by 13.3 mmHg on average and overestimated diastolic by 11.5 mmHg; cuff brachial pressure alone was a closer estimate of central pressure than the device output. Ding 2011 (33 patients, 99 paired measurements): central systolic estimates correlated with invasive values at r = 0.91 (SphygmoCor) and r = 0.90 (Omron HEM9000-AI — named as published) but underestimated by −15 mmHg (95% CI −17 to −13) and −2 mmHg (−4 to 0) respectively; the Omron's own brachial oscillometric systolic read −19 mmHg against the brachial catheter. Laugesen 2014 (34 patients with type 2 diabetes): with invasive calibration the transfer function's error was only −2.3 ± 5.6 mmHg systolic and +1.0 ± 0.9 diastolic; with oscillometric calibration, systolic error ranged from −9.6 ± 8.1 to +1.9 ± 12.2 mmHg by strategy while diastolic error stayed about +14 mmHg — the calibration, not the transfer function, carries the error. De la Torre Hernández 2021 (94 patients with severe aortic stenosis): the cuff-based SphygmoCor XCEL underestimated central systolic pressure by −3 ± 15 mmHg before and −9 ± 13 mmHg after valve replacement and did not meet the ARTERY Society criterion (acceptable mean difference ± SD of 5 ± 8 mmHg).
MeSH: Blood Pressure Determination · Aorta · Arterial Pressure · Calibration · Pulse Wave Analysis
Cards (4)
Card 1 — Cuff-calibrated tonometry misses both ways: central systolic underestimated by 13.3 mmHg, diastolic overestimated by 11.5 mmHg (Cloud 2003, Clinical Science — 30 catheterisation subjects)
Study: Invasive validation study (n = 30)
Population: 30 subjects (age 27–84 years, half aged 65 or over) undergoing routine diagnostic cardiac catheterisation; central aortic pressure measured directly, and estimated simultaneously by SphygmoCor applanation tonometry (named as published) of the left radial artery using the radial transfer function, calibrated with non-invasive brachial cuff pressure.
Summary: Cuff-calibrated tonometry misses both ways: central systolic underestimated by 13.3 mmHg, diastolic overestimated by 11.5 mmHg (Cloud 2003, Clinical Science — 30 catheterisation subjects)
Result: On average the SphygmoCor system underestimated systolic central arterial pressure by 13.3 mmHg and overestimated diastolic pressure by 11.5 mmHg, with similar results under and over age 65. Non-invasively measured brachial pressures gave an overall closer estimate of central arterial pressure than the SphygmoCor system. The authors note the transfer function had been validated from invasively measured arterial pressures, and that using non-invasive measures may explain the discrepancies.
Watch out: Thirty subjects — N printed on the face of this card. A 2003 study of an early device generation and cuff calibration: it identifies the calibration problem that the later cards on this shelf quantify, rather than fixing the error of current devices.
Source: Cloud GC et al. Estimation of central aortic pressure by SphygmoCor requires intra-arterial peripheral pressures. Clinical science (London, England : 1979). 2003. PMID: 12710885. Registry ID: none in the record. MeSH: Adult · Age Factors · Aged · Aged, 80 and over · Aorta · Blood Pressure Determination · Brachial Artery · Cardiac Catheterization · Humans · Middle Aged · Reproducibility of Results
Card 2 — Strong correlation, systematic underestimation: SphygmoCor −15 mmHg and Omron −2 mmHg against the aortic catheter (Ding 2011, American Journal of Hypertension — 33 patients, 99 paired measurements)
Study: Invasive validation study (n = 33; 99 paired measurements)
Population: 33 patients in whom radial pulse was recorded sequentially with the Omron HEM9000-AI and SphygmoCor devices (named as published), calibrated with brachial oscillometric pressure, while central pressure was measured simultaneously by a catheter-based fluid-filled manometer in the ascending aorta; the procedure was repeated three times for 99 paired measurements.
Summary: Strong correlation, systematic underestimation: SphygmoCor −15 mmHg and Omron −2 mmHg against the aortic catheter (Ding 2011, American Journal of Hypertension — 33 patients, 99 paired measurements)
Result: Non-invasive central systolic estimates correlated significantly with the invasive measurement (correlation coefficient 0.91 for SphygmoCor, 0.90 for Omron; P less than 0.001), but both underestimated central systolic pressure: by −15 mmHg (95% CI −17 to −13; P less than 0.001) for SphygmoCor and −2 mmHg (−4 to 0; P less than 0.05) for Omron. Against the brachial artery catheter, the Omron oscillometric device underestimated brachial systolic pressure by −19 mmHg (−23 to −15). The authors conclude the estimations can be used properly only if device-specific diagnostic thresholds become available.
Watch out: Ninety-nine measurements come from 33 patients measured three times — the effective sample is 33 people, and the tight confidence intervals reflect repeated measures, not 99 independent subjects.
Source: Ding FH et al. Validation of the noninvasive assessment of central blood pressure by the SphygmoCor and Omron devices against the invasive catheter measurement. American journal of hypertension. 2011. PMID: 21976274. Registry ID: none in the record. MeSH: Aged · Aged, 80 and over · Antihypertensive Agents · Blood Pressure Determination · Calibration · Catheters · Coronary Artery Disease · Female · Humans · Hydroxymethylglutaryl-CoA Reductase Inhibitors · Hypertension · Male · Middle Aged · Oscillometry · Radial Artery · Reproducibility of Results · Sphygmomanometers
Card 3 — The transfer function is fine; the calibration is the error: invasive calibration −2.3 ± 5.6 mmHg, oscillometric calibration up to −9.6 ± 8.1 systolic and +14 diastolic (Laugesen 2014, American Journal of Hypertension — 34 patients with type 2 diabetes)
Study: Invasive validation study (calibration strategies; n = 34)
Population: 34 patients with type 2 diabetes; ascending aortic pressure estimated with the SphygmoCor device (named as published) and recorded invasively, testing the transfer function calibrated with invasive diastolic and mean pressure, and non-invasive strategies calibrated with brachial oscillometric systolic+diastolic versus diastolic+mean pressure (form factors 0.33 and 0.40).
Summary: The transfer function is fine; the calibration is the error: invasive calibration −2.3 ± 5.6 mmHg, oscillometric calibration up to −9.6 ± 8.1 systolic and +14 diastolic (Laugesen 2014, American Journal of Hypertension — 34 patients with type 2 diabetes)
Result: Calibrated with invasive pressure, estimated minus invasive ascending aortic pressure was −2.3 ± 5.6 mmHg systolic and +1.0 ± 0.9 mmHg diastolic — the transfer function itself is valid in this group. Calibrated with oscillometric brachial pressure, differences were −9.6 ± 8.1 / +14.1 ± 6.2 mmHg (systolic+diastolic calibration), −8.3 ± 11.7 / +13.9 ± 6.1 (diastolic+mean, form factor 0.33) and +1.9 ± 12.2 / +14.1 ± 6.2 (diastolic+mean, form factor 0.40); averaging three brachial measurements did not improve accuracy. The authors conclude non-invasive calibration enables accurate group-level mean systolic estimation, but wide limits of agreement indicate limited accuracy in the individual patient.
Watch out: Thirty-four subjects — N printed on the face of this card. The diastolic overestimation of about +14 mmHg under every oscillometric strategy is as large as the systolic problem and is easy to miss if only systolic accuracy is quoted.
Source: Laugesen E et al. Assessment of central blood pressure in patients with type 2 diabetes: a comparison between SphygmoCor and invasively measured values. American journal of hypertension. 2014. PMID: 24304654. Registry ID: NCT01538290 (in the record). MeSH: Aged · Arterial Pressure · Blood Pressure · Blood Pressure Determination · Calibration · Diabetes Mellitus, Type 2 · Female · Humans · Male · Middle Aged · Pulse Wave Analysis
Card 4 — Cuff-based estimation in severe aortic stenosis fails the accuracy criterion: −3 ± 15 mmHg before and −9 ± 13 mmHg after valve replacement (De la Torre Hernández 2021, Journal of Clinical Hypertension — 94 patients)
Study: Invasive validation study (severe aortic stenosis; n = 94)
Population: 94 patients with severe aortic stenosis undergoing transcatheter aortic valve replacement, with simultaneous non-invasive (SphygmoCor XCEL cuff device — named as published) and invasive central blood pressure measurements before and after the procedure; agreement judged against the ARTERY Society recommendation that acceptable error is a mean difference ± SD within 5 ± 8 mmHg (in plain words: on average within 5 mmHg of the catheter value, with the spread of individual errors within 8 mmHg).
Summary: Cuff-based estimation in severe aortic stenosis fails the accuracy criterion: −3 ± 15 mmHg before and −9 ± 13 mmHg after valve replacement (De la Torre Hernández 2021, Journal of Clinical Hypertension — 94 patients)
Result: Before valve replacement, central systolic pressure was on average slightly underestimated with the default brachial-cuff systolic calibration (mean difference ± SD −3 ± 15 mmHg); after replacement the underestimation increased (−9 ± 13 mmHg). Agreement was somewhat better in low-gradient than high-gradient stenosis (−2 ± 11 vs −4 ± 17 mmHg; P = 0.3). Cuff-brachial systolic pressure itself agreed less well than the device estimate. Despite strong correlation, the device did not meet the ARTERY Society accuracy criteria for non-invasive central systolic measurement in this population.
Watch out: Severe aortic stenosis distorts the arterial waveform itself, so this is the hardest test case on the shelf, not the expected error in general clinic populations — the shelf's general claim rests on the first three cards, with this card showing the criterion failure where central pressure matters most.
Source: De la Torre Hernández JM et al. Validation study to determine the accuracy of central blood pressure measurement using the SphygmoCor XCEL cuff device in patients with severe aortic stenosis undergoing transcatheter aortic valve replacement. Journal of clinical hypertension (Greenwich, Conn.). 2021. PMID: 33942487. Registry ID: none in the record. MeSH: Aortic Valve Stenosis · Blood Pressure · Blood Pressure Determination · Humans · Hypertension · Transcatheter Aortic Valve Replacement