Wrist actigraphy vs polysomnography for sleep–wake detection — 4 validation studies (documented N = 348): sensitivity for sleep is high (sensitivity 89–96.5%, labelled in words) but specificity for wake is low (specificity 32.9–69%), total sleep time agrees well while wake after sleep onset is poorly identified (intraclass correlation 0.00–0.10)
Shelf #9339 · Lungs, breathing & sleep
MEASUREMENT ACCURACY — sleep detection established, wake detection NOT ESTABLISHED: across 4 validation studies (348 participants) actigraphy sensitivity for sleep was 89–96.5% but specificity for wake was only 32.9–69% (sensitivity and specificity labelled in words); total sleep time and sleep efficiency agree closely with polysomnography, while wake after sleep onset agreement is poor (intraclass correlation 0.00–0.10)
Stamped: 2026-11-28.
Evidence published: 2004–2020.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately does wrist actigraphy detect sleep and wake against polysomnography — what are its sensitivity and specificity (labelled in words), and which summary measures (total sleep time, wake after sleep onset) does it get right or wrong?
Balance of the evidence
MEASUREMENT ACCURACY — sleep detection established, wake detection NOT ESTABLISHED — Composition: 4 validation studies against polysomnography (348 participants). Hedner 2004 (228 subjects including sleep apnoea patients): overall sensitivity 89% and specificity 69% (labelled in words) for identifying sleep, with agreement 80–86% across apnoea severity and tight agreement on sleep efficiency and total sleep time. Marino 2013 (77 participants, 232,849 epochs): sensitivity 0.965 and specificity 0.329 — that is, sensitivity 96.5% and specificity 32.9% — with accuracy 0.863, and actigraphy underestimated wake after sleep onset (36.8 vs 49.1 minutes). Quante 2018 (35 volunteers): both devices' accuracy was 0.81–0.86, algorithm choice moved sensitivity and specificity in opposite directions, and agreement for wake after sleep onset was poor (intraclass correlation 0.00–0.10; as a plain-words guide, a reliability of 0.70 suffices for comparing groups and 0.90 is the ideal for individual-patient decisions — these values fall far below both). Devine 2020 (8 participants): a wrist watch device showed accuracy and sensitivity above 90% but specificity of about 26% against polysomnography.
MeSH: Actigraphy · Polysomnography · Sensitivity and Specificity · Sleep · Wakefulness
Cards (4)
Card 1 — In sleep apnoea patients too: sensitivity 89% and specificity 69% (labelled in words) for sleep, agreement 80–86% by severity (Hedner 2004, Sleep — 228 subjects)
Study: Validation study (actigraphy algorithm vs polysomnography; n = 228)
Population: 228 subjects from 3 sleep centres (normal subjects and patients with mild, moderate and severe obstructive sleep apnoea) with simultaneous polysomnography and a Watch_PAT100 ambulatory device containing a built-in actigraph — device named as published — scored by a novel automatic algorithm on an epoch-by-epoch basis.
Summary: In sleep apnoea patients too: sensitivity 89% and specificity 69% (labelled in words) for sleep, agreement 80–86% by severity (Hedner 2004, Sleep — 228 subjects)
Result: The overall sensitivity and specificity to identify sleep was 89% and 69% respectively (sensitivity and specificity labelled in words). Agreement ranged from 86% in normal subjects to 86%, 84% and 80% in patients with mild, moderate and severe obstructive sleep apnoea. There was tight agreement between actigraphy and polysomnography in determining sleep efficiency (78.4 ± 9.9% vs 78.8 ± 13.4%), total sleep time (690 ± 152 vs 690 ± 154 epochs) and sleep latency (56.8 ± 31.4 vs 43.3 ± 45.4 epochs), although for some individuals the disagreement was substantial. The authors conclude the algorithm provides reasonably accurate epoch-by-epoch estimation of sleep and wakefulness in normal subjects and apnoea patients.
Watch out: This is the most favourable specificity on the shelf (69%) and it comes from an algorithm designed for apnoea patients' movement patterns; the general-population studies below read wake much worse.
Source: Hedner J et al. A novel adaptive wrist actigraphy algorithm for sleep-wake assessment in sleep apnea patients. Sleep. 2004. PMID: 15683148. Registry ID: none in the record. MeSH: Adaptation, Physiological · Adult · Algorithms · Cohort Studies · Electromyography · Electronic Data Processing · Equipment Design · Female · Humans · Male · Middle Aged · Polysomnography · Prospective Studies · Sensitivity and Specificity · Severity of Illness Index · Sleep Apnea Syndromes · Sleep Disorders, Circadian Rhythm · Wrist
Card 2 — The definitive split: sensitivity 96.5% for sleep, specificity only 32.9% for wake (labelled in words), over 232,849 epochs (Marino 2013, Sleep — 77 participants)
Study: Validation study (epoch-by-epoch vs polysomnography; n = 77)
Population: 77 young and older adults — healthy volunteers, chronic primary insomnia patients, and 23 night-workers studied in daytime sleep — with simultaneous actigraphy and polysomnography during sleep-laboratory admissions of 8.5 hours in bed (mean 3.2 nights per participant), scored in 30-second epochs (n = 232,849).
Summary: The definitive split: sensitivity 96.5% for sleep, specificity only 32.9% for wake (labelled in words), over 232,849 epochs (Marino 2013, Sleep — 77 participants)
Result: Overall, sensitivity was 0.965 and accuracy was 0.863, whereas specificity was 0.329 — that is, sensitivity 96.5% and specificity 32.9% (labelled in words). Each was only slightly modified by gender, insomnia, or day/night sleep timing (magnitude of change under 0.04), and increasing age slightly reduced specificity. Mean wake after sleep onset was 49.1 minutes by polysomnography compared with 36.8 minutes per night by actigraphy, unbiased when wake after sleep onset was under 30 minutes and overestimated when it was over 30 minutes. Participant-specific accuracy was above 80% for most participants. The authors conclude actigraphy is a useful means of estimating total sleep time and wake after sleep onset, with limitations in specificity.
Watch out: The record prints sensitivity and specificity as proportions (0.965 / 0.329); they are carded both ways so the pair is never a bare unlabelled pair. A specificity of 32.9% means about two-thirds of true wake epochs are scored as sleep.
Source: Marino M et al. Measuring sleep: accuracy, sensitivity, and specificity of wrist actigraphy compared to polysomnography. Sleep. 2013. PMID: 24179309. Registry ID: none in the record. MeSH: Actigraphy · Adult · Age Factors · Female · Humans · Male · Middle Aged · Polysomnography · Reproducibility of Results · Sensitivity and Specificity · Sleep · Wakefulness · Wrist
Card 3 — Device and algorithm both move the answer: accuracy 0.81–0.86, but wake-after-sleep-onset agreement near zero (intraclass correlation 0.00–0.10) (Quante 2018, Nature and Science of Sleep — 35 volunteers)
Study: Validation study (two devices, two algorithms vs polysomnography; n = 35)
Population: 35 healthy volunteers (13 school children and 22 adults) comparing two wrist actigraphs — the GT3X+ and the Actiwatch Spectrum, both named as published — against in-home polysomnography, with the GT3X+ scored by both the Sadeh and the Cole-Kripke algorithms; this record carries no MeSH terms — MeSH: none in the record.
Summary: Device and algorithm both move the answer: accuracy 0.81–0.86, but wake-after-sleep-onset agreement near zero (intraclass correlation 0.00–0.10) (Quante 2018, Nature and Science of Sleep — 35 volunteers)
Result: Each actigraph showed comparable accuracy (0.81–0.86) for sleep–wake estimation compared with polysomnography. For the GT3X+, the Cole-Kripke algorithm was more sensitive (sensitivity 0.88–0.96, labelled in words) to detect sleep but less specific (specificity 0.35–0.64) to detect wake than the Sadeh algorithm (sensitivity 0.82–0.91, specificity 0.47–0.68). Total sleep time from the GT3X+ with both algorithms was similar to polysomnography (intraclass correlation 0.64–0.88), while agreement for wake after sleep onset was poor (intraclass correlation 0.00–0.10). In adults the Cole-Kripke GT3X+ data were comparable to the Actiwatch Spectrum (mean bias 3.7 ± 19.7 minutes for total sleep time).
Watch out: This record carries no MeSH terms — stated plainly on the card. The intraclass correlations are agreement statistics: in plain words, a reliability of 0.70 suffices for comparing groups and 0.90 is the ideal for individual-patient decisions — total sleep time (0.64–0.88) straddles the group threshold, wake after sleep onset (0.00–0.10) reaches neither.
Source: Quante M et al. Actigraphy-based sleep estimation in adolescents and adults: a comparison with polysomnography using two scoring algorithms. Nature and science of sleep. 2018. PMID: 29403321. Registry ID: none in the record. MeSH: none in the record
Card 4 — A purpose-built watch: accuracy and sensitivity above 90%, specificity about 26% against polysomnography (Devine 2020, Sensors — 8 healthy young adults)
Study: Validation study (commercial watch vs polysomnography; n = 8)
Population: Eight healthy young adult participants who wore a Zulu watch and an Actiwatch (both named as published) simultaneously over a 3-day laboratory polysomnography sleep study, with epoch-by-epoch accuracy, sensitivity and specificity tested against polysomnography and actigraphy.
Summary: A purpose-built watch: accuracy and sensitivity above 90%, specificity about 26% against polysomnography (Devine 2020, Sensors — 8 healthy young adults)
Result: Compared with either polysomnography or actigraphy, both the accuracy and the sensitivity for the Zulu watch's sleep–wake determination were above 90% (labelled in words), while the specificity was low (about 26% vs polysomnography and about 33% vs actigraphy — specificity labelled in words). Accuracy for sleep scoring against polysomnography was about 87% for interrupted sleep, about 52% for light sleep and about 49% for deep sleep. The watch performed well for total sleep time, sleep efficiency, sleep onset and final awakening compared with polysomnography or actigraphy.
Watch out: Eight participants — N printed on the face of this card. Sleep-depth (light/deep) scoring at about 50% accuracy is a separate, weaker claim than sleep–wake detection and is not part of the shelf's stamp.
Source: Devine JK et al. Validation of Zulu Watch against Polysomnography and Actigraphy for On-Wrist Sleep-Wake Determination and Sleep-Depth Estimation. Sensors (Basel, Switzerland). 2020. PMID: 33375557. Registry ID: none in the record. MeSH: Actigraphy · Humans · Polysomnography · Reproducibility of Results · Sleep · Wrist · Young Adult