Capillary blood gas vs arterial blood gas — 1 meta-analysis (29 studies; 664 earlobe, 222 fingertip, 880 carbon-dioxide and 587 pH paired samples) and 3 primary studies (documented N = 147): earlobe sampling predicts arterial carbon dioxide and pH closely (bias 1.9 mmHg and 0.02) and arterial oxygen with a residual error of 6 mmHg, while fingertip sampling does not predict arterial oxygen (bias 11.5 mmHg)

Shelf #9340 · Emergency & first hours

MEASUREMENT ACCURACY — pH and carbon dioxide established from the earlobe, oxygen NOT ESTABLISHED from the fingertip: in a 29-study meta-analysis earlobe sampling predicted arterial carbon dioxide (mean bias 1.9 mmHg) and pH (bias 0.02) closely and arterial oxygen with mean bias 3.8 mmHg but a residual standard error of 6 mmHg, while fingertip sampling did not predict arterial oxygen (bias 11.5 mmHg); primary studies agree for pH and carbon dioxide, and a capillary carbon dioxide of 51.7 mmHg detected hypercarbia with sensitivity 100% and specificity 95.12% (labelled in words)

Stamped: 2026-11-28.

Evidence published: 1994–2019.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do capillary blood gases (arterialised earlobe or fingertip) reproduce arterial blood gas values for pH, carbon dioxide and oxygen — and where does the substitution fail?

Balance of the evidence

MEASUREMENT ACCURACY — pH and carbon dioxide established from the earlobe, oxygen NOT ESTABLISHED from the fingertip — Composition: 1 meta-analysis of 29 studies (664 earlobe and 222 fingertip paired samples for oxygen; 880 paired samples for carbon dioxide; 587 for pH) and 3 primary studies (147 patients). Zavorsky 2007: earlobe predicted arterial oxygen (adjusted r-squared 0.88, mean bias 3.8 mmHg, residual standard error 6 mmHg) while fingertip did not (0.48, bias 11.5 mmHg); both sites predicted arterial carbon dioxide (bias 1.9 and 2.2 mmHg) and pH (bias 0.02). Pitkin 1994 (40 patients): earlobe–arterial mean differences were 0.21 kPa for carbon dioxide and −0.17 kPa for oxygen, with close agreement especially below an arterial oxygen tension of 8 kPa. Kongstad 2019 (62 patients): non-arterialised fingertip limits of agreement were clinically acceptable for carbon dioxide, pH and bicarbonate but too wide for oxygen, base excess, lactate and oxygen saturation. Eizadi-Mood 2009 (45 patients): pH correlated satisfactorily (r-squared 0.91) and a capillary carbon dioxide of 51.7 mmHg had sensitivity 100% and specificity 95.12% (labelled in words) for detecting hypercarbia.

MeSH: Blood Gas Analysis · Capillaries · Carbon Dioxide · Oxygen · Hydrogen-Ion Concentration

Cards (4)

Card 1 — The meta-analysis: earlobe predicts arterial oxygen (bias 3.8 mmHg), fingertip does not (bias 11.5 mmHg); both predict carbon dioxide and pH (Zavorsky 2007, Respiratory Physiology & Neurobiology — 29 studies)

Study: Meta-analysis (paired-sample accuracy; 29 studies)

Population: Meta-analysis of 29 studies with paired arterial and capillary samples — 664 earlobe and 222 fingertip paired samples for oxygen, 880 total paired samples for carbon dioxide (arterial range 10–114 mmHg) and 587 total paired samples for pH (range 6.77–7.74) — analysed with a mixed-effects model.

Summary: The meta-analysis: earlobe predicts arterial oxygen (bias 3.8 mmHg), fingertip does not (bias 11.5 mmHg); both predict carbon dioxide and pH (Zavorsky 2007, Respiratory Physiology & Neurobiology — 29 studies)

Result: Earlobe sampling showed about 2.5 times better precision than fingertip sampling for arterial oxygen over a wide range (21–155 mmHg), and the lower the arterial oxygen, the more accurate any capillary prediction. Earlobe sampling predicted arterial oxygen (adjusted r-squared 0.88, mean bias 3.8 mmHg) but fingertip sampling did not (adjusted r-squared 0.48, mean bias 11.5 mmHg); the residual standard error when predicting arterial oxygen from an earlobe sample was 6 mmHg. For carbon dioxide, earlobe (adjusted r-squared 0.94, mean bias 1.9 mmHg) and fingertip (0.95, bias 2.2 mmHg) both closely reflected arterial values. For pH both sites accurately reflected arterial pH (adjusted r-squared 0.90–0.94, mean bias 0.02). The authors conclude earlobe (preferably) or fingertip sampling accurately reflects arterial carbon dioxide and pH, and earlobe — never fingertip — sampling may replace arterial oxygen unless precision is required.

Watch out: Sample counts are paired samples, not patients, and are stated as such; they are not added to the primary-study patient total. The 6 mmHg residual error on earlobe oxygen is the reason the stamp stops short of calling oxygen established even from the earlobe.

Source: Zavorsky GS et al. Arterial versus capillary blood gases: a meta-analysis. Respiratory physiology & neurobiology. 2007. PMID: 16919507. Registry ID: none in the record. MeSH: Arteries · Blood Gas Analysis · Blood Specimen Collection · Capillaries · Carbon Dioxide · Humans · Hydrogen-Ion Concentration · Linear Models · Oxygen · Oxyhemoglobins · Reproducibility of Results · Venules

Estimate type: adjusted estimate, as stated in the card.

Card 2 — Earlobe vs arterial in chronic lung disease: mean differences 0.21 kPa (carbon dioxide) and −0.17 kPa (oxygen) — yet only 18% of UK centres used the technique (Pitkin 1994, Thorax — 40 patients)

Study: Primary accuracy study (paired earlobe–arterial samples; n = 40)

Population: Forty patients with chronic lung disease and a wide range of arterial blood gas values, with simultaneous earlobe (arterialised) and arterial samples at rest analysed in the same blood gas analyser, plus a telephone survey of respiratory function laboratory staff in 50 UK hospitals.

Summary: Earlobe vs arterial in chronic lung disease: mean differences 0.21 kPa (carbon dioxide) and −0.17 kPa (oxygen) — yet only 18% of UK centres used the technique (Pitkin 1994, Thorax — 40 patients)

Result: Earlobe and arterial blood gas tensions agreed closely over a wide range of arterial pH, carbon dioxide (mean difference 0.21, 95% confidence intervals −0.24 to +0.67 kPa) and oxygen (mean difference −0.17, 95% confidence intervals −1.09 to +0.75 kPa), especially at arterial oxygen values lower than 8 kPa. Of 50 UK centres surveyed, 18% used the arterialised earlobe technique and 4% planned to introduce it; reasons for not using it were lack of knowledge in 64%, no blood gas analyser in 6%, considered inaccurate in 4%, and insufficient staff in 4%. The authors conclude earlobe analysis is sufficiently accurate to be reliably substituted for arterial sampling in routine clinical practice.

Watch out: Forty patients — N printed on the face of this card. Values are in kPa as printed in the record (0.21 kPa is about 1.6 mmHg); the survey half of the paper is about adoption, not accuracy, and is carded as context only.

Source: Pitkin AD et al. Arterialised earlobe blood gas analysis: an underused technique. Thorax. 1994. PMID: 8202909. Registry ID: none in the record. MeSH: Blood Gas Analysis · Carbon Dioxide · Ear, External · Humans · Hydrogen-Ion Concentration · Lung Diseases · Oxygen · Sensitivity and Specificity

Card 3 — No arterialisation, fingertip only: acceptable limits for carbon dioxide, pH and bicarbonate — too wide for oxygen, lactate and saturation (Kongstad 2019, European Clinical Respiratory Journal — 62 patients with COPD)

Study: Primary accuracy study (paired fingertip–arterial samples; n = 62)

Population: 62 patients with stable-phase chronic obstructive pulmonary disease with consecutive arterial punctures and non-arterialised fingertip capillary samples, compared by Bland–Altman agreement; this record carries no MeSH terms — MeSH: none in the record.

Summary: No arterialisation, fingertip only: acceptable limits for carbon dioxide, pH and bicarbonate — too wide for oxygen, lactate and saturation (Kongstad 2019, European Clinical Respiratory Journal — 62 patients with COPD)

Result: Limits of agreement for oxygen (limits of agreement −1.27 to 4.45 kPa), base excess (−1.35 to 0.55), lactate (−0.77 to 0.20 mmol/L, about −6.9 to 1.8 mg/dL) and oxygen saturation (−0.02 to 0.06) were wider than what would be applicable for clinical use. Clinically acceptable limits of agreement were obtained for carbon dioxide (−0.64 to 0.38 kPa), pH (−0.02 to 0.03) and bicarbonate. The authors conclude measurement of carbon dioxide, pH and bicarbonate in non-arterialised capillary blood may be useful in clinical practice in an acute setting, while the other parameters cannot be substituted.

Watch out: This record carries no MeSH terms — stated plainly on the card. The lactate limits are printed in mmol/L with their mg/dL equivalents (about −6.9 to 1.8 mg/dL) as this shelf's unit rule requires; bicarbonate limits are described qualitatively because the record's own acceptability judgement, not a converted number, is the finding.

Source: Kongstad HK et al. Agreement between arterial and non-arterialised fingertip capillary blood gas and acid-base values. European clinical respiratory journal. 2019. PMID: 31489152. Registry ID: none in the record. MeSH: none in the record

Card 4 — In comatose poisoned patients, capillary pH substitutes and a capillary carbon dioxide of 51.7 mmHg flags hypercarbia — sensitivity 100%, specificity 95.12% (labelled in words) (Eizadi-Mood 2009, Human & Experimental Toxicology — 45 patients)

Study: Primary accuracy study (paired samples, emergency setting; n = 45)

Population: 45 adult patients with coma because of poisoning and stable vital signs, with simultaneously obtained arterial and capillary blood gas values compared prospectively in a poisoning emergency department.

Summary: In comatose poisoned patients, capillary pH substitutes and a capillary carbon dioxide of 51.7 mmHg flags hypercarbia — sensitivity 100%, specificity 95.12% (labelled in words) (Eizadi-Mood 2009, Human & Experimental Toxicology — 45 patients)

Result: For pH, arterial and capillary values correlated satisfactorily (r-squared 0.91) with acceptable limits of agreement (−0.04 to 0.06). Base excess correlated well (r-squared 0.85) but its 95% limits of agreement seemed too wide to allow substitution (−4.4 to 2.7). Carbon dioxide (r-squared 0.61), bicarbonate (0.71) and oxygen (0.53) correlated less reliably. A capillary carbon dioxide of 51.7 mmHg had a sensitivity of 100% and a specificity of 95.12% (sensitivity and specificity labelled in words) for detecting hypercarbia (area under the curve 0.99). The authors conclude capillary analysis for pH may be a reliable substitute for arterial analysis in the initial evaluation of comatose patients with stable vital signs, with arterial sampling required when capillary carbon dioxide exceeds 51.7 mmHg.

Watch out: Forty-five patients — N printed on the face of this card, all with stable vital signs; the record does not support extending the substitution to unstable patients. The sensitivity/specificity pair is for the hypercarbia threshold test, not for capillary sampling in general.

Source: Eizadi-Mood N et al. Comparison of arterial and capillary blood gas values in poisoning department assessment. Human & experimental toxicology. 2009. PMID: 19744970. Registry ID: none in the record. MeSH: Adult · Arteries · Blood Gas Analysis · Capillaries · Carbon Dioxide · Coma · Female · Humans · Hydrogen-Ion Concentration · Male · Oxygen · Poisoning · Prospective Studies · ROC Curve