Rebound and non-contact tonometry vs Goldmann applanation tonometry for intraocular pressure — 1 meta-analysis (6 studies, 672 eyes) and 3 primary studies (336 and 125 patients and 1,000 eyes): mean differences are small (−0.14 to +1.01 mmHg) but agreement is only moderate (intraclass correlation 0.73 in 1,000 eyes), one large study found a consistent 4.2 mmHg underestimate, and rebound readings run low at Goldmann pressures of 23 mmHg or more
Shelf #9388 · Eyes, skin, nose & allergy
MEASUREMENT ACCURACY — approximate at group level, NOT ESTABLISHED as interchangeable with Goldmann for an individual eye: mean differences between Icare rebound and Goldmann readings are small in most studies (−0.14 mmHg in a 6-study meta-analysis; −0.22 mmHg in 336 patients) but one 1,000-eye study found a consistent 4.2 mmHg underestimate with an intraclass correlation of 0.73 — below the 0.90 ideal for individual-patient decisions — and rebound readings run significantly low when Goldmann pressure is 23 mmHg or more; non-contact tonometry read 0.49 mmHg below Goldmann in healthy subjects
Stamped: 2026-11-28.
Evidence published: 2017–2023.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do rebound tonometry (Icare devices, named as published) and non-contact tonometry measure intraocular pressure against Goldmann applanation tonometry, the reference standard — at group level, for an individual eye, and at high pressures?
Balance of the evidence
MEASUREMENT ACCURACY — approximate at group level, NOT ESTABLISHED as interchangeable with Goldmann for an individual eye — Composition: 1 meta-analysis (6 studies, 672 eyes) and 3 primary studies (336 patients, 125 healthy individuals, and 1,000 eyes; eye and patient counts are stated per card and not summed). Rödter 2020: the pooled Icare PRO minus Goldmann difference was −0.14 mmHg (95% CI −0.43 to 0.15), non-significant, with strong heterogeneity (I-squared 77%) and two studies significantly different in opposite directions. Subramaniam 2021 (1,000 eyes): the ic100 read lower than Goldmann in every stratum (mean 12.1 vs 16.2 mmHg; mean difference 4.2 mmHg) with an intraclass correlation coefficient of 0.73. Gao 2017 (336 patients): mean readings were close (18.30 ± 5.10 vs 18.52 ± 4.46 mmHg) but the rebound device read 1.66 ± 3.87 mmHg lower when Goldmann pressure was 23 mmHg or more. Duran 2023 (125 healthy individuals): non-contact tonometry read 0.49 ± 1.89 mmHg below Goldmann and Icare Pro 1.01 ± 1.90 mmHg above it, with inter-device intraclass correlations above 0.8. For reliability coefficients on this shelf: 0.70 suffices for comparing groups, and 0.90 is the ideal — a measure is acceptable for individual-patient decisions when it reaches the 90% threshold.
MeSH: Tonometry, Ocular · Intraocular Pressure · Glaucoma · Reproducibility of Results
Cards (4)
Card 1 — The meta-analysis: pooled difference −0.14 mmHg, but studies disagree in opposite directions (Rödter 2020, European Journal of Ophthalmology — 6 studies, 672 eyes)
Study: Meta-analysis (paired device agreement; 6 studies, 672 eyes)
Population: Systematic review and meta-analysis of paired method-comparison studies of the Icare PRO rebound tonometer against Goldmann applanation tonometry in adult glaucoma patients, searching PubMed and the Cochrane Library for January 2010 to March 2018; 6 studies (672 eyes) were included from 147 publications.
Summary: The meta-analysis: pooled difference −0.14 mmHg, but studies disagree in opposite directions (Rödter 2020, European Journal of Ophthalmology — 6 studies, 672 eyes)
Result: Four studies showed an underestimation of intraocular pressure by the Icare PRO and two an overestimation. The overall meta-difference (Icare PRO minus Goldmann) was −0.14 mmHg (95% CI −0.43 to 0.15), a non-significant, clinically irrelevant mean deviation (P = 0.335). Heterogeneity between studies was strong (I-squared 77%): two studies presented a significant difference, in opposite directions. The authors conclude the review could neither confirm nor deny considerable agreement between the devices in glaucomatous patients, despite the manufacturer's 'superior accuracy' claim for the Icare PRO (device named as published).
Watch out: Eye counts, not patient counts, as printed in the record. The near-zero pooled mean hides study-level disagreement in both directions — the reason the stamp stops at approximate.
Source: Rödter TH et al. Meta-analysis of the concordance of Icare® PRO-based rebound and Goldmann applanation tonometry in glaucoma patients. European journal of ophthalmology. 2020. PMID: 31466475. Registry ID: none in the record. MeSH: Glaucoma · Humans · Intraocular Pressure · Prospective Studies · Reproducibility of Results · Tonometry, Ocular
Card 2 — The discordant large study: ic100 underestimates by 4.2 mmHg on average, intraclass correlation 0.73 (Subramaniam 2021, Ophthalmic Research — 1,000 eyes)
Study: Primary accuracy study (paired device comparison; 1,000 eyes)
Population: Cross-sectional study measuring intraocular pressure in 1,000 eyes (500 left, 500 right) with the Icare ic100 rebound tonometer and Goldmann applanation tonometry, with central corneal thickness by hand-held pachymeter, analysed overall and by pressure stratum, corneal thickness group, medication count and diagnosis.
Summary: The discordant large study: ic100 underestimates by 4.2 mmHg on average, intraclass correlation 0.73 (Subramaniam 2021, Ophthalmic Research — 1,000 eyes)
Result: Agreement was moderate (intraclass correlation coefficient 0.73). Mean pressure was significantly lower by ic100 than by Goldmann (12.1 vs 16.2 mmHg, P < 0.0001), and ic100 means were lower within every pressure stratum (≤12: 7.9 vs 9.7; 13–21: 12.1 vs 16.6; ≥22: 18.4 vs 25.2 mmHg; all P < 0.0001) and every subanalysis. The authors conclude the Icare ic100 (named as published) consistently underestimated pressure against Goldmann regardless of corneal thickness, and the mean difference of 4.2 mmHg can have significant clinical implications in glaucoma management. For reliability coefficients on this shelf: 0.70 suffices for comparing groups, and 0.90 is the ideal — a measure is acceptable for individual-patient decisions when it reaches the 90% threshold. An intraclass correlation of 0.73 clears the group-comparison level but not the 0.90 individual-decision ideal.
Watch out: The record counts eyes (1,000), not patients; the number of patients is not stated in the record and no patient total is claimed. This study's 4.2 mmHg gap is far larger than the meta-analysis pooled estimate — device generation and population differ, and the shelf does not average them away.
Source: Subramaniam AG et al. Comparison of the Icare ic100 Rebound Tonometer and the Goldmann Applanation Tonometer in 1,000 Eyes. Ophthalmic research. 2021. PMID: 32906125. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Cross-Sectional Studies · Equipment Design · Female · Glaucoma · Humans · Intraocular Pressure · Male · Middle Aged · Ocular Hypertension · Prospective Studies · Reproducibility of Results · Tonometry, Ocular · Young Adult
Card 3 — Close on average at low-to-moderate pressures, significantly low at Goldmann 23 mmHg or more (Gao 2017, Experimental and Therapeutic Medicine — 336 patients, 672 eyes)
Study: Primary accuracy study (paired device comparison; n = 336 patients)
Population: Three hundred and thirty-six patients (672 eyes) whose intraocular pressure was measured by the iCare rebound tonometer and Goldmann applanation tonometer, divided by the Goldmann reading into groups A (7–15 mmHg, n = 74), B (16–22 mmHg, n = 218) and C (23–50 mmHg, n = 44), with pachymetry and slit-lamp inspection; this record carries no MeSH terms — MeSH: none in the record.
Summary: Close on average at low-to-moderate pressures, significantly low at Goldmann 23 mmHg or more (Gao 2017, Experimental and Therapeutic Medicine — 336 patients, 672 eyes)
Result: Mean pressures were 18.30 ± 5.10 mmHg (rebound) and 18.52 ± 4.46 mmHg (Goldmann), not significantly different (P = 0.19), and the readings correlated well (r = 0.806, P = 0.001); the 95% confidence interval of the differences between devices was −5.80 to 6.24 mmHg, as printed. In group C (Goldmann ≥23 mmHg) rebound measurements were significantly lower (−1.66 ± 3.87 mmHg; P = 0.007). Both devices correlated positively with central corneal thickness, the rebound device more strongly (r = 0.390 vs 0.191). The measurement was well tolerated with no corneal epithelial defects. The authors conclude the rebound tonometer (named as published) is a reliable alternative in the low-to-moderate range but not at high pressures.
Watch out: Three hundred and thirty-six patients — N printed on the face of this card. The difference interval is printed in the record as a '95% confidence interval of the differences'; it is reproduced as printed and spans about 12 mmHg, which is itself the caution.
Source: Gao F et al. Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer. Experimental and therapeutic medicine. 2017. PMID: 28565785. Registry ID: none in the record. MeSH: none in the record
Card 4 — Three devices in healthy eyes: non-contact 0.49 mmHg below Goldmann, rebound 1.01 mmHg above (Duran 2023, Journal Français d'Ophtalmologie — 125 healthy individuals)
Study: Primary accuracy study (three-device paired comparison; n = 125)
Population: One hundred and twenty-five healthy individuals measured by non-contact tonometry, then iCare Pro rebound tonometry, then (after 5 minutes, under topical anaesthesia) three Goldmann applanation readings averaged, with central corneal thickness by topography. N = 125, printed on the face of this card.
Summary: Three devices in healthy eyes: non-contact 0.49 mmHg below Goldmann, rebound 1.01 mmHg above (Duran 2023, Journal Français d'Ophtalmologie — 125 healthy individuals)
Result: Mean pressures were 15.97 ± 2.99 mmHg (non-contact), 17.47 ± 2.86 mmHg (rebound) and 16.46 ± 2.68 mmHg (Goldmann). Mean differences were −0.49 ± 1.89 mmHg (non-contact minus Goldmann), +1.01 ± 1.90 mmHg (rebound minus Goldmann) and −1.50 ± 2.02 mmHg (non-contact minus rebound). Inter-device agreement exceeded an intraclass correlation of 0.8 for each pairing, and all three devices correlated significantly with central corneal thickness, non-contact most affected. For reliability coefficients on this shelf: 0.70 suffices for comparing groups, and 0.90 is the ideal — a measure is acceptable for individual-patient decisions when it reaches the 90% threshold. Agreement above 0.8 still falls short of the 0.90 individual-decision ideal. The authors conclude all three suit routine inspection and screening, but follow-up in advanced glaucoma benefits from Goldmann measurement.
Watch out: Healthy eyes only — N printed on the face of this card. Agreement in healthy eyes does not settle performance in glaucoma, which the other cards on this shelf address.
Source: Duran M et al. Comparison of intraocular pressure measurements obtained by icare pro tonometer, non-contact tonometer and Goldmann applanation tonometer in healthy individuals. Journal francais d'ophtalmologie. 2023. PMID: 37666735. Registry ID: none in the record. MeSH: Humans · Intraocular Pressure · Prospective Studies · Reproducibility of Results · Glaucoma · Tonometry, Ocular