SourceMed · Shelf #9439 in the ask interface
Electronic-nose breath analysis for diagnosis — 3 systematic reviews (52, 46 and 73 studies) and 1 primary device study (144 subjects): pooled cancer sensitivity 90% and specificity 87% (labelled in words) in feasibility studies, but only 7 of 46 studies validated externally and area under the curve fell by up to 0.23 when they did
Shelf #9439 · Cancer
MEASUREMENT ACCURACY — NOT ESTABLISHED for clinical diagnosis: pooled feasibility-study sensitivity 90% and specificity 87% for cancer (labelled in words) shrink on validation — only 7 of 46 human studies used an external validation set, where area under the curve fell by up to 0.23 — and the best device study's 78–88% accuracy is internal cross-validation in diagnosed cases
Stamped: 2026-11-28.
Evidence published: 2015–2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately does electronic-nose analysis of exhaled breath diagnose disease (cancer, asthma, COPD) against established reference diagnoses — and does the accuracy survive validation in independent samples?
Balance of the evidence
MEASUREMENT ACCURACY — NOT ESTABLISHED for clinical diagnosis — Composition: 3 systematic reviews (52, 46 and 73 studies) and 1 two-centre primary study (144 subjects); the reviews overlap in primary studies and their counts are stated per card, never summed. Scheepers 2022: across 52 cancer studies (3,677 patients with cancer), pooled sensitivity was 90% and specificity 87%, both labelled in words, with a mean area under the curve of 0.94 — but every study was a feasibility study with considerable heterogeneity and high risk of bias. Leopold 2015: of 46 human e-nose studies only 7 used external validation; performance fell by 0.10–0.15 of area under the curve on internal validation and by a further 0.09–0.23 on external validation in two of four re-analysed datasets. de Vries 2015: the SpiroNose device (named as published) was reproducible (kappa 0.81–0.94 between sensor arrays) and discriminated asthma, COPD, lung cancer and controls with cross-validation values of 78–88% — internal cross-validation only. Krilaviciute 2015: in 25% of 73 breath-cancer studies the classification model was built and validated on the same dataset. MEASUREMENT ACCURACY — NOT ESTABLISHED: electronic-nose breathprints separate cases from controls impressively inside feasibility datasets, but the accuracy that survives independent external validation — the accuracy a clinical diagnosis would need — has not been established.
MeSH: Electronic Nose · Breath Tests · Neoplasms · ROC Curve · Reproducibility of Results
Cards (4)
Card 1 — The pooled cancer estimate looks excellent — sensitivity 90%, specificity 87% — and every included study was a feasibility study (Scheepers 2022, JAMA Network Open — 52 articles, 3,677 patients with cancer)
Study: Systematic review and meta-analysis (diagnostic accuracy; 52 studies)
Population: Systematic review and meta-analysis of electronic-nose technology analysing exhaled breath for cancer detection in adults (PubMed and Embase, January 2000 to July 2021): 52 articles with a total of 3,677 patients with cancer, quality-assessed with the QUADAS-2 tool. Counts are as printed in the record.
Summary: The pooled cancer estimate looks excellent — sensitivity 90%, specificity 87% — and every included study was a feasibility study (Scheepers 2022, JAMA Network Open — 52 articles, 3,677 patients with cancer)
Result: Across studies, the sensitivity of electronic noses ranged from 48.3% to 95.8% and the specificity from 10.0% to 100.0% — sensitivity and specificity labelled in words at first occurrence, as throughout this shelf. Pooled analysis gave a mean area under the receiver operating characteristic curve of 94% (95% CI 92%–96%), a sensitivity of 90% (95% CI 88%–92%) and a specificity of 87% (95% CI 81%–92%). Considerable heterogeneity existed between studies from differences in patient selection, endogenous and exogenous factors, and breath collection. All studies were feasibility studies. The authors conclude electronic noses have high diagnostic accuracy for cancer in exhaled breath, but most studies were small feasibility studies with a lack of standardisation and a high risk of bias, and the lack of standardisation and reproducibility should be addressed in future research.
Watch out: The pooled pair (sensitivity 90% / specificity 87%) pools case-control feasibility studies that typically compare known cancer patients with healthy controls — a design that inflates accuracy versus a real diagnostic population. The stamp follows the validation evidence on cards 2 and 4, not this pooled number.
Source: Scheepers MHMC et al. Diagnostic Performance of Electronic Noses in Cancer Diagnoses Using Exhaled Breath: A Systematic Review and Meta-analysis. JAMA network open. 2022. PMID: 35767259. Registry ID: none in the record. MeSH: Adult · Breath Tests · Electronic Nose · Exhalation · Humans · Neoplasms · Reproducibility of Results
Card 2 — The validation audit: only 7 of 46 studies validated externally, and performance fell on every step towards independence (Leopold 2015, Journal of Breath Research — 46 studies, 4 datasets re-analysed)
Study: Systematic / methodological review with dataset re-analysis (46 studies)
Population: Methodological review of all human electronic-nose breath studies to the end of 2014 (46 studies, quality assessed with a tailored QUADAS-2 tool), with re-analysis of the original datasets of externally validated studies — 4 of the 7 such datasets were shared by their authors. Counts are as printed in the record.
Summary: The validation audit: only 7 of 46 studies validated externally, and performance fell on every step towards independence (Leopold 2015, Journal of Breath Research — 46 studies, 4 datasets re-analysed)
Result: Risk of bias was high in most studies because of non-random selection of patients. Only 7 of the 46 studies applied external validation in an independent dataset, mostly with case-control designs. In the re-analysed datasets, internal validation (leave-one-out cross-validation) decreased the area under the receiver operating characteristics curve versus the training-set performance by 0.15, 0.14, 0.10 and 0.11 in datasets 1–4 respectively; external validation lowered it further in dataset 1 (−0.23) and dataset 3 (−0.09), and did not decrease it in datasets 2 (+0.07) and 4 (+0.04). No single combination of dimension-reduction and classification methods gave consistent results between internal and external validation. The authors conclude it is not meaningful to estimate diagnostic performance on a training set alone, even after internal validation, and recommend an external validation set in all future electronic-nose projects in medicine.
Watch out: This is the load-bearing card for the stamp: the drops are in area-under-the-curve points, reproduced exactly as printed. Four datasets only were available for re-analysis, so the size of the external-validation penalty is itself imprecisely known — the direction (down or flat, never up in the training sense) is the finding.
Source: Leopold JH et al. Comparison of classification methods in breath analysis by electronic nose. Journal of breath research. 2015. PMID: 26669708. Registry ID: none in the record. MeSH: Area Under Curve · Breath Tests · Electronic Nose · Humans · ROC Curve · Reproducibility of Results
Card 3 — The best-engineered device study: SpiroNose breathprints are reproducible (kappa 0.81–0.94) and separate asthma, COPD, lung cancer and controls at 78–88% cross-validated accuracy (de Vries 2015 — 144 subjects)
Study: Primary accuracy / reproducibility study (case-control; n = 144)
Population: Two-centre case-control study integrating an electronic nose with routine spirometry (the SpiroNose, built on five identical metal-oxide semiconductor sensor arrays, named as published): technical validation in 60 healthy subjects (age 43 ± 11 years), then a cross-sectional study of 37 patients with asthma, 31 with COPD, 31 with lung cancer and 45 healthy controls in day-to-day clinical care. N = 144 in the diagnostic study, printed on the face of this card.
Summary: The best-engineered device study: SpiroNose breathprints are reproducible (kappa 0.81–0.94) and separate asthma, COPD, lung cancer and controls at 78–88% cross-validated accuracy (de Vries 2015 — 144 subjects)
Result: Expiratory flow correlated linearly with raw sensor deflections (R-squared 0.84); exhaled volume, humidity and temperature did not correlate with sensor readings. After environment correction, exhaled data were highly reproducible for each sensor array (Cohen's kappa 0.81–0.94). For reliability coefficients: 0.70 suffices for comparing groups, and 0.90 is the ideal — a measure is acceptable for individual-patient decisions when it reaches the 90% threshold; these kappas straddle that ideal, applying to breathprint repeatability, not to diagnostic correctness. The device distinguished controls, asthma, COPD and lung cancer patients with cross-validation values ranging between 78% and 88%. The authors conclude the standardised integration provides repeatable, interchangeable results and promising accuracy, paving the route towards point-of-care breath diagnostics.
Watch out: One hundred and forty-four subjects — N printed on the face of this card. The 78–88% figures are cross-validation values within the same case-control sample (the design card 2 shows overstates performance), in patients with already-established diagnoses — not accuracy in undiagnosed presenters.
Source: de Vries R et al. Integration of electronic nose technology with spirometry: validation of a new approach for exhaled breath analysis. Journal of breath research. 2015. PMID: 26469298. Registry ID: none in the record. MeSH: Adult · Aged · Asthma · Breath Tests · Case-Control Studies · Electronic Nose · Exhalation · Female · Humans · Lung Neoplasms · Male · Middle Aged · Principal Component Analysis · Pulmonary Disease, Chronic Obstructive · Reproducibility of Results · Spirometry
Card 4 — The wider breath-cancer literature: areas under the curve of 0.9 or more are common — and a quarter of studies validated on their own training data (Krilaviciute 2015, Oncotarget — 73 studies)
Study: Systematic review (breath analysis for cancer; 73 studies)
Population: Systematic review of exhaled breath analysis for cancer detection using standard analytical techniques and electronic nose (PubMed and Web of Science to 30 April 2015): 73 studies, two-thirds of them on lung cancer, extracting test performance (sensitivity and specificity) and the volatile compounds used for discrimination. Counts are as printed in the record.
Summary: The wider breath-cancer literature: areas under the curve of 0.9 or more are common — and a quarter of studies validated on their own training data (Krilaviciute 2015, Oncotarget — 73 studies)
Result: Good discrimination usually required a combination of multiple biomarkers, and area under the receiver operating characteristic curve or accuracy reached levels of 0.9 or higher in multiple studies. In 25% of the reported studies, classification models were built and validated on the same datasets. Huge variability was seen between studies in breath sampling and other methods. The authors conclude breath analysis yielded promising results, but standardisation of breath collection, sample storage and data handling remain critical issues, and larger studies in true screening settings with validation in independent population samples are needed before implementation into practice.
Watch out: This review mixes electronic-nose studies with chemical-analyser (mass spectrometry) studies; only its electronic-nose subset belongs to this shelf's question, and the card does not attribute the pooled impression to electronic noses alone. No single sensitivity/specificity pair is printed for electronic noses in this record, and none is claimed.
Source: Krilaviciute A et al. Detection of cancer through exhaled breath: a systematic review. Oncotarget. 2015. PMID: 26440312. Registry ID: none in the record. MeSH: Biomarkers, Tumor · Breath Tests · Exhalation · Humans · Lung Neoplasms