Point-of-care beta-hydroxybutyrate (ketone) meters vs laboratory and clinical reference standards — 5 studies (documented N = 256 in the clinical cohorts, meter evaluations stated per card): sensitivity 98% / specificity 85% for diabetic ketoacidosis at 1.5 mmol/L, close tracking of pH and bicarbonate during paediatric treatment, capillary values substantially higher than venous on the same device, and 2 of 4 meters passing a multicentre accuracy evaluation
Shelf #9488 · Diabetes, glucose & metabolism
MEASUREMENT ACCURACY — at the 1.5 mmol/L threshold the founding triage study found sensitivity 98% / specificity 85% (labelled in words) for diabetic ketoacidosis, and bedside values track the laboratory and acid-base course during treatment, but capillary samples read substantially higher than venous blood on the same meter and only 2 of 4 ketone meters passed formal accuracy evaluation — accurate enough for triage and monitoring, not interchangeable across sample types or brands
Stamped: 2026-11-28.
Evidence published: 2006–2021.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do bedside capillary/whole-blood beta-hydroxybutyrate meters measure ketosis against laboratory reference methods and clinical ketoacidosis criteria — at the diabetic ketoacidosis triage threshold, during paediatric ketoacidosis treatment, across blood sample types, and between meter brands?
Balance of the evidence
MEASUREMENT ACCURACY — Composition: 5 studies (documented N = 256 across the four clinical cohorts; the multicentre meter evaluation is stated per card). Naunheim 2006 (160 emergency patients with glucose above 250 mg/dL, PMID 16690813): on the Precision Xtra meter (Abbott Laboratories, North Chicago, IL — as named in the record), beta-hydroxybutyrate at the manufacturer-suggested 1.5 mmol/L level (≈15.6 mg/dL) gave sensitivity 98% and specificity 85% (both labelled in words) for a clinical diagnosis of diabetic ketoacidosis made in 57 of 160 patients. Rewers 2006 (68 children with diabetic ketoacidosis, PMID 17109599): bedside meter values correlated closely with duplicate laboratory serum values and with pH (r = −0.63), bicarbonate (r = −0.74) and pCO2 (r = −0.55) across treatment. Norgren 2020 (15 volunteers, PMID 31821040): venous whole-blood values on the point-of-care meter agreed well with plasma values, but capillary values on the same device ran substantially higher (regression slope 0.64–0.78 against venous), so sample type changes the number. Moore 2021 (13 participants, PMID 33413456): a lower-cost meter agreed excellently with the Precision Xtra for ketones (intraclass correlation 0.968; area under the curve 0.913 for detecting nutritional ketosis at 0.5 mmol/L), with the plain-words reliability context that 0.70 suffices for comparing groups and 0.90 is the ideal for individual-patient decisions — 0.968 clears both. Guimont 2015 (multicentre evaluation, PMID 26232287): of four ketone meters formally evaluated, only two were considered acceptable. PROPOSED MEASUREMENT ACCURACY: bedside beta-hydroxybutyrate is accurate enough for ketoacidosis triage at the studied threshold and for following treatment, provided the sample type and the specific meter are held constant — the numbers are not interchangeable across capillary versus venous sampling or across brands.
MeSH: 3-Hydroxybutyric Acid · Point-of-Care Systems · Diabetic Ketoacidosis · Sensitivity and Specificity · Reproducibility of Results
Cards (5)
Card 1 — Triage performance: at 1.5 mmol/L, sensitivity 98% / specificity 85% for diabetic ketoacidosis (Naunheim 2006 — 160 emergency patients)
Study: Diagnostic accuracy study (prospective, blinded clinical reference standard; n = 160)
Population: One hundred sixty emergency-department patients presenting with blood glucose above 250 mg/dL (13.9 mmol/L) consented to point-of-care beta-hydroxybutyrate testing with the Precision Xtra meter (Abbott Laboratories, North Chicago, IL) at triage; the diagnosis of diabetic ketoacidosis was made by clinicians using standard clinical criteria (glucose above 250 mg/dL, anion gap at least 15 mmol/L, carbon dioxide at most 18 mmol/L, other causes excluded) without knowledge of the beta-hydroxybutyrate result, and was made in 57 of 160 patients. N = 160, printed on the face of this card.
Summary: Triage performance: at 1.5 mmol/L, sensitivity 98% / specificity 85% for diabetic ketoacidosis (Naunheim 2006 — 160 emergency patients)
Result: At the manufacturer-suggested beta-hydroxybutyrate level of 1.5 mmol/L (approximately 15.6 mg/dL), cross-classification against the clinical diagnosis yielded sensitivity 98% (95% CI 91–100%) and specificity 85% (95% CI 78–91%) — sensitivity and specificity labelled in words at first occurrence here and throughout this shelf — with a positive likelihood ratio of 6.7 and a negative likelihood ratio of 0.021. Beta-hydroxybutyrate correlated with anion gap (r = 0.66) and carbon dioxide (r = −0.69), both P < 0.001. The authors conclude the point-of-care test was as sensitive as more established indicators and more useful than glucose alone, offering immediate diagnosis at triage.
Watch out: The reference standard is a clinical criteria diagnosis, not a laboratory beta-hydroxybutyrate assay; this card measures agreement with the ketoacidosis syndrome, and the meter-versus-laboratory agreement evidence is on cards 2 and 3 of this shelf.
Source: Naunheim R, Jang TJ, Banet G et al. Point-of-care test identifies diabetic ketoacidosis at triage. Academic emergency medicine : official journal of the Society for Academic Emergency Medicine. 2006. PMID: 16690813. Registry ID: none in the record. MeSH: 3-Hydroxybutyric Acid · Acid-Base Equilibrium · Adolescent · Adult · Aged · Aged, 80 and over · Biomarkers · Blood Glucose · Carbon Dioxide · Diabetic Ketoacidosis · Female · Humans · Male · Middle Aged · Point-of-Care Systems · Prospective Studies · Reproducibility of Results · Sensitivity and Specificity · Triage
Card 2 — During paediatric ketoacidosis treatment, bedside values track the laboratory method and the acid-base course (Rewers 2006 — 68 children)
Study: Agreement/monitoring study (repeated measures during treatment; n = 68)
Population: Sixty-eight children treated for diabetic ketoacidosis in an emergency department under a standard protocol with monitoring of venous pH, pCO2, bicarbonate, glucose, urea nitrogen and electrolytes at two to ten time points per patient; venous beta-hydroxybutyrate was measured at the bedside with the Precision Xtra meter (MediSense/Abbott Diabetes Care, Abbott Park, IL — as named in the record) and on duplicate batched serum samples by a reference laboratory method (Cobas Mira Plus; Roche Diagnostics, Indianapolis, IN). N = 68, printed on the face of this card.
Summary: During paediatric ketoacidosis treatment, bedside values track the laboratory method and the acid-base course (Rewers 2006 — 68 children)
Result: Bedside meter beta-hydroxybutyrate levels were significantly correlated with pH (r = −0.63), bicarbonate (r = −0.74) and pCO2 (r = −0.55), all P < 0.0001, at all measurement points during treatment, and time-series models confirmed bedside levels correlated very closely with the laboratory and acid-base parameters (the record's abstract is truncated at that sentence in the fetched record; no further coefficient is claimed). The authors' stated purpose — evaluating whether bedside monitoring can simplify management by eliminating superfluous laboratory monitoring — is the use this card supports: following a treatment course within a patient.
Watch out: Repeated measures within 68 children are not 68 independent accuracy estimates at a single threshold; this card supports monitoring use, and the single-threshold diagnostic claim rests on card 1.
Source: Rewers A, McFann K, Chase HP et al. Bedside monitoring of blood beta-hydroxybutyrate levels in the management of diabetic ketoacidosis in children. Diabetes technology & therapeutics. 2006. PMID: 17109599. Registry ID: none in the record. MeSH: 3-Hydroxybutyric Acid · Adolescent · Child · Diabetic Ketoacidosis · Electrochemistry · Female · Humans · Male · Point-of-Care Systems
Card 3 — Sample type changes the number: capillary blood reads substantially higher than venous on the same meter (Norgren 2020 — 15 volunteers)
Study: Agreement study (three methods, simultaneous sampling; n = 15)
Population: Fifteen healthy older volunteers in whom ketosis in the range 0–1.5 mmol/L was induced by medium-chain fatty-acid intake after a 12-hour fast; beta-hydroxybutyrate was measured on one point-of-care meter in venous whole blood (12 time points over 4 hours) and capillary blood (3 time points), with a laboratory plasma measure of total ketones (beta-hydroxybutyrate plus acetoacetate) at 4 time points; 180 cases had simultaneous data. N = 15, printed on the face of this card.
Summary: Sample type changes the number: capillary blood reads substantially higher than venous on the same meter (Norgren 2020 — 15 volunteers)
Result: Total ketones correlated with plasma beta-hydroxybutyrate (r = 0.99), venous meter values (r = 0.91) and capillary meter values (r = 0.91), all P < 0.0001. Venous meter values and plasma values had good agreement in absolute values; however, the slope between capillary and venous values on the same device was 0.64–0.78 across regression models, indicating substantially higher concentrations in capillary than venous blood. The authors conclude all three methods detect relative changes in ketosis, but cut-offs may need adjusting — for ketoacidosis management and for comparing dietary interventions — by method.
Watch out: Fifteen volunteers — N printed on the face of this card, and the ketosis range studied (0–1.5 mmol/L) is nutritional, below ketoacidosis values; the sample-type warning is carried into the stamp as a direction of bias, not a quantified correction for high ranges.
Source: Norgren J, Sindi S, Sandebring-Matton A et al. Capillary blood tests may overestimate ketosis: triangulation between three different measures of β-hydroxybutyrate. American journal of physiology. Endocrinology and metabolism. 2020. PMID: 31821040. Registry ID: none in the record. MeSH: 3-Hydroxybutyric Acid · Adult · Capillaries · Diet, Ketogenic · Fatty Acids · Female · Healthy Volunteers · Hematologic Tests · Humans · Ketones · Ketosis · Male · Reproducibility of Results · Sensitivity and Specificity
Card 4 — Brand comparison: a lower-cost meter agrees excellently with the established meter for ketones (Moore 2021 — 13 participants)
Study: Device agreement and diagnostic performance study (randomised crossover; n = 13)
Population: Thirteen participants (7 women, 6 men; 21.6 ± 3.0 years) attended the laboratory three times in a randomised double-blind crossover design, with ketone and glucose measured twice before and twice after ingesting a racemic ketone, a natural ketone or a maltodextrin supplement, on Meter 1 (Precision Xtra, previously validated) and Meter 2 (Keto-Mojo, not previously validated — both as named in the record). N = 13, printed on the face of this card.
Summary: Brand comparison: a lower-cost meter agrees excellently with the established meter for ketones (Moore 2021 — 13 participants)
Result: Inter-meter reliability for ketones was excellent (intraclass correlation coefficient 0.968, 95% CI 0.942–0.981) and good for glucose (0.809, 0.642–0.893). For context in plain words: a reliability of 0.70 suffices for comparing groups, and 0.90 is the ideal — a measure is acceptable for individual-patient decisions when it reaches the 90% threshold — so 0.968 clears both benchmarks, while the glucose value clears only the group benchmark. Area under the receiver-operating curve for Meter 2 detecting nutritional ketosis at a 0.5 mmol/L threshold (as identified by Meter 1) was 0.913 (0.828–0.998). The authors conclude Meter 2 is an effective, more affordable alternative for ketone measurement.
Watch out: Thirteen participants — N printed on the face of this card. The reference here is another meter, not a laboratory assay, and the population is athletes in nutritional ketosis; this card supports brand comparability, not ketoacidosis diagnosis.
Source: Moore AR, Holland-Winkler AM, Ansley JK et al. Reliability and diagnostic performance of a new blood ketone and glucose meter in humans. Journal of the International Society of Sports Nutrition. 2021. PMID: 33413456. Registry ID: none in the record. MeSH: 3-Hydroxybutyric Acid · Area Under Curve · Athletes · Blood Glucose · Cross-Over Studies · Diagnostic Equipment · Diet, Ketogenic · Dietary Supplements · Double-Blind Method · Female · Humans · Ketone Bodies · Ketones · Ketosis · Male · Polysaccharides · Reproducibility of Results · Young Adult
Card 5 — The formal evaluation: of four ketone meters tested to the ISO standard, only two were acceptable (Guimont 2015 — multicentre evaluation)
Study: Device evaluation study (multicentre, standard-based; patient N not documented in the record)
Population: A multicentre evaluation within the Assistance Publique-Hôpitaux de Paris of eight glucose meters and four ketone meters, connected or not to laboratory software, assessing accuracy, precision, correlation with central-laboratory plasma values and haematocrit interference for glucose, and accuracy, precision and recovery for ketone meters measuring beta-hydroxybutyrate, applying the ISO 15197:2003 norm. The record does not state a patient N for this evaluation, and none is claimed on this card.
Summary: The formal evaluation: of four ketone meters tested to the ISO standard, only two were acceptable (Guimont 2015 — multicentre evaluation)
Result: Among the glucose meters, seven were considered acceptable for glucose measurement; among the four ketone meters, two were considered acceptable for beta-hydroxybutyrate measurement. The authors conclude that, since not all meters meet clinically relevant criteria, meter performance has to be evaluated before use in clinical practice — the brand-level caution that, with card 3's sample-type caution, bounds this shelf's otherwise favourable triage evidence.
Watch out: No patient N is documented in the record — this card contributes no number to the shelf's documented clinical total (256). The record does not name which two ketone meters passed, and no brand is named as passing or failing on this card.
Source: Guimont MC, Desjobert H, Fonfrède M et al. Multicentric evaluation of eight glucose and four ketone blood meters. Clinical biochemistry. 2015. PMID: 26232287. Registry ID: none in the record. MeSH: Adult · Automation, Laboratory · Blood Glucose · Diabetes Mellitus · Guidelines as Topic · Hematocrit · Humans · Hydroxybutyrates · Hyperglycemia · Ketones · Point-of-Care Systems · Prospective Studies · Software