SourceMed · Shelf #9489 in the ask interface
Lactose hydrogen breath test accuracy vs genotype and alternative criteria — 1 meta-analysis (17 breath-test studies) plus 3 primary studies (documented N = 124, one study's N not documented in its record): pooled sensitivity 88% / specificity 85% against the lactase genotype, conventional criteria missing a quarter of malabsorbers that late-sampling criteria catch, and breath hydrogen outperforming blood and urine markers head-to-head
Shelf #9489 · Liver & gut
MEASUREMENT ACCURACY — against the North European lactase genotype, pooled sensitivity 88% / specificity 85% (labelled in words) across 17 breath-test studies, with accuracy strongly criterion-dependent: newer late-sampling criteria reclassified a quarter of conventionally negative patients as malabsorbers, and breath hydrogen (area under the curve 1.00 in a 40-woman head-to-head) beat plasma glucose (0.75) and urinary galactose (0.73) against genotype
Stamped: 2026-11-28.
Evidence published: 2004–2020.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately does the lactose hydrogen breath test identify lactose malabsorption / lactase non-persistence against reference standards — the lactase C/T-13910 genotype, alternative breath-test criteria and sampling schedules, and competing blood and urine markers?
Balance of the evidence
MEASUREMENT ACCURACY — Composition: 1 meta-analysis plus 3 primary studies (documented N = 124 across two of them; the third study's record states no subject number). Marton 2012 (PMID 22211845): across 17 breath-hydrogen studies against the North European lactase polymorphism C/T-13910, pooled sensitivity was 88% and specificity 85% (labelled in words), while the lactose tolerance test pooled at sensitivity 94% / specificity 90%. Di Stefano 2004 (84 patients, PMID 15614254): the conventional breath-hydrogen criterion classified 51% as malabsorbers, but a 6th-hour value above 6 parts per million classified 76% and a summed 5–7-hour criterion 80% — diagnosing malabsorption in 24% and 27% respectively of those negative by the conventional criterion. Shrestha 2020 (40 women, PMID 32600320): against genotype as reference after a 50 g lactose challenge, breath hydrogen discriminated lactase non-persistence with area under the curve 1.00, against 0.75 for plasma glucose and 0.73 for urinary galactose/creatinine. Nagy 2009 (PMID 19156157): in Hungarian children and controls, the C/C-13910 genotype was closely associated with a positive breath test in symptomatic children and persistent genotypes with a negative test in controls — qualitative concordance, with the study's subject number not stated in its record. PROPOSED MEASUREMENT ACCURACY: the breath test is a moderately accurate, criterion-bound test of lactase non-persistence — which criterion and sampling schedule are used moves the answer by tens of percentage points, so a negative result under conventional early criteria does not exclude malabsorption.
MeSH: Breath Tests · Hydrogen · Lactose Intolerance · Lactase · Sensitivity and Specificity
Cards (4)
Card 1 — The pooled accuracy against genotype: sensitivity 88% / specificity 85% for the breath test (Marton 2012 — meta-analysis, 17 breath studies)
Study: Meta-analysis of diagnostic accuracy studies (19 studies; 17 breath-test)
Population: Nineteen studies comparing indirect lactose-digestion tests with the North European lactase gene polymorphism C/T-13910: 17 evaluating breath hydrogen and 5 the lactose tolerance test (3 overlapping), analysed with random-effects (breath) and fixed-effect (tolerance) models, exploring lactose dose, inclusion of children and study latitude. Participant totals are stated per study in the review and are not summed on this card.
Summary: The pooled accuracy against genotype: sensitivity 88% / specificity 85% for the breath test (Marton 2012 — meta-analysis, 17 breath studies)
Result: For the lactose breath hydrogen test, overall sensitivity was 88% (95% CI 85–90%) and specificity 85% (95% CI 82–87%) — sensitivity and specificity labelled in words at first occurrence here and throughout this shelf. The lactose tolerance test showed sensitivity 94% (90–97%) and specificity 90% (84–95%) with non-significant heterogeneity. The authors set out to determine the tests' ability to predict geno/phenotype relationships; the breath test's pooled pair is therefore accuracy against a genetic reference for lactase non-persistence, not against symptoms.
Watch out: Synthesis card: the reference standard is genotype, which defines lactase non-persistence; a person can be genotype-positive without symptoms, so this accuracy does not translate directly into accuracy for symptom causation — stated here rather than in the stamp's pooled pair.
Source: Marton A, Xue X, Szilagyi A et al. Meta-analysis: the diagnostic accuracy of lactose breath hydrogen or lactose tolerance tests for predicting the North European lactase polymorphism C/T-13910. Alimentary pharmacology & therapeutics. 2012. PMID: 22211845. Registry ID: none in the record. MeSH: Breath Tests · Genotype · Humans · Hydrogen · Lactase · Lactose Intolerance · Lactose Tolerance Test · Phenotype · Polymorphism, Single Nucleotide · Predictive Value of Tests · White People
Card 2 — Criterion choice moves the answer: conventional criteria call 51% malabsorbers, late-sampling criteria 76–80% (Di Stefano 2004 — 84 patients)
Study: Diagnostic criteria comparison study (three-day protocol with false-result controls; n = 84)
Population: Eighty-four consecutive patients with functional or organic gastrointestinal disease each underwent, on 3 separate days, breath hydrogen measurement after lactose (testing for malabsorption), after lactulose (to rule out false negatives from hydrogen-nonproducer status) and after a nonabsorbable electrolyte solution (to rule out false positives from release of preformed hydrogen). N = 84, printed on the face of this card.
Summary: Criterion choice moves the answer: conventional criteria call 51% malabsorbers, late-sampling criteria 76–80% (Di Stefano 2004 — 84 patients)
Result: By the conventional criterion, 51% of subjects were lactose malabsorbers. Positive results were obtained in 76% using the 6th-hour criterion (breath hydrogen above 6 parts per million 6 hours after the load) and in 80% using the sum of the 5th, 6th and 7th hour values (above 15 parts per million) — both P < 0.05 versus conventional — making it possible to diagnose malabsorption in 24% and 27% respectively of patients who tested negative by the conventional criterion. Ten of 21 and 11 of 25 patients in the reclassified groups, respectively, showed intolerance symptoms. The authors conclude the conventional criterion must be abandoned in favour of these later-sampling criteria.
Watch out: Breath hydrogen is measured in parts per million, a concentration with no mg/dL conversion; none is offered on this card. There is no independent reference standard here — the comparison is between criteria, with the lactulose and electrolyte days controlling the known false-result mechanisms.
Source: Di Stefano M, Missanelli A, Miceli E et al. Hydrogen breath test in the diagnosis of lactose malabsorption: accuracy of new versus conventional criteria. The Journal of laboratory and clinical medicine. 2004. PMID: 15614254. Registry ID: none in the record. MeSH: Adult · Breath Tests · Dietary Carbohydrates · Fasting · Female · Humans · Hydrogen · Lactose Intolerance · Male · Middle Aged · Sensitivity and Specificity
Card 3 — Head-to-head against genotype: breath hydrogen area under the curve 1.00, plasma glucose 0.75, urinary galactose 0.73 (Shrestha 2020 — 40 women)
Study: Diagnostic accuracy study (randomised crossover challenge, genotype reference; n = 40)
Population: Forty healthy young women challenged with 50 g lactose and then randomised in crossover to 750 mL of conventional milk (37.5 g lactose) and A1-beta-casein-free milk on separate visits, with blood, breath and urine collected before and up to 3 hours after each challenge; C/T-13910 and G/A-22018 genotyping was the diagnostic reference for lactase non-persistence, identifying 14 of 40 women as non-persistent. N = 40, printed on the face of this card.
Summary: Head-to-head against genotype: breath hydrogen area under the curve 1.00, plasma glucose 0.75, urinary galactose 0.73 (Shrestha 2020 — 40 women)
Result: After the lactose challenge, all three malabsorption markers discriminated lactase persistence from non-persistence: breath hydrogen area under the receiver-operating curve 1.00, plasma glucose 0.75 and urinary galactose/creatinine 0.73. The record further reports plasma glucose and urinary galactose/creatinine were unreliable under at least one milk condition (the fetched abstract truncates mid-sentence at that point; no value beyond those printed here is claimed). The authors' design conclusion carried on this card is that breath hydrogen was the only marker with perfect discrimination in this sample.
Watch out: Forty women — N printed on the face of this card, with only 14 genotype-positive participants; an area under the curve of 1.00 in 40 people is a small-sample result and is not read as proof of perfect accuracy in practice.
Source: Shrestha A, Barnett MPG, Perry JK et al. Evaluation of breath, plasma, and urinary markers of lactose malabsorption to diagnose lactase non-persistence following lactose or milk ingestion. BMC gastroenterology. 2020. PMID: 32600320. Registry ID: none in the record. MeSH: Adult · Animals · Breath Tests · Eating · Female · Humans · Hydrogen · Lactase · Lactose · Lactose Intolerance · Milk
Card 4 — Population concordance: genotype and breath test agree in symptomatic children and in controls (Nagy 2009 — Hungarian cohort)
Study: Genotype-phenotype concordance study (subject number not documented in the record)
Population: Hungarian children and adults genotyped for the C/T(-13910) polymorphism upstream of the lactase gene and tested with the lactose hydrogen breath test, examining the prevalence of adult-type hypolactasia, its age of onset, and the applicability of the breath test against genetic screening. The record does not state the total subject number, and none is claimed on this card.
Summary: Population concordance: genotype and breath test agree in symptomatic children and in controls (Nagy 2009 — Hungarian cohort)
Result: The prevalence of the C/C(-13910) (lactase non-persistent) genotype among adults was 37%. Hypolactasia starts to appear at around 5 years of age, and over the age of 12 years almost all of those with a C/C(-13910) genotype have lactase non-persistence. The C/C genotype was closely associated with a positive lactose hydrogen breath test in symptomatic children, whereas lactase-persistent genotypes correlated better with a negative breath test in a control group. The authors conclude the breath and genotyping tests together furnish the clinical diagnosis — qualitative concordance evidence, printed without a sensitivity/specificity pair because the record prints none.
Watch out: No subject number and no sensitivity/specificity pair are documented in this record, so this card contributes no number to the shelf's documented total (124) and no pair to its stamp; it is carded for the age-of-onset and concordance pattern only.
Source: Nagy D, Bogácsi-Szabó E, Várkonyi A et al. Prevalence of adult-type hypolactasia as diagnosed with genetic and lactose hydrogen breath tests in Hungarians. European journal of clinical nutrition. 2009. PMID: 19156157. Registry ID: none in the record. MeSH: Adolescent · Adult · Biopsy · Breath Tests · Celiac Disease · Child · Child, Preschool · Diagnosis, Differential · Female · Gastrointestinal Diseases · Genotype · Humans · Hungary · Hydrogen · Intestine, Small · Lactase · Lactose Intolerance · Male · Phenotype · Polymorphism, Single Nucleotide · Predictive Value of Tests · Prevalence · Young Adult