Pulse CO-oximetry (SpCO) vs blood carboxyhaemoglobin — 1 meta-analysis (21 studies; 2,089 subjects in the accuracy pool) plus 3 primary studies (documented N = 5,701): pooled sensitivity 77% / specificity 83%, mean bias 0.75% with limits of agreement −7.08% to +8.57%, bias 2.99% in 1,578 unselected emergency patients, and close low-range agreement (bias 0.1%) in 50 children with sickle cell disease

Shelf #9490 · Critical care & sepsis

MEASUREMENT ACCURACY — NOT ESTABLISHED for confirming or excluding carbon monoxide poisoning: pooled sensitivity 77% / specificity 83% (labelled in words) with a mean bias of 0.75% but limits of agreement from −7.08% to +8.57% carboxyhaemoglobin, and the largest unselected emergency study found bias 2.99% with limits from −3.55% to +9.53% — agreement tight enough in low-range paediatric measurements (bias 0.1%, precision ±2.1%) does not extend to poisoning-range decisions

Stamped: 2026-11-28.

Evidence published: 2011–2023.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately does non-invasive pulse CO-oximetry (SpCO) estimate blood carboxyhaemoglobin against laboratory co-oximetry / blood gas analysis — its bias, limits of agreement and sensitivity/specificity for carbon monoxide poisoning, in unselected emergency patients, in suspected-poisoning cohorts and in children?

Balance of the evidence

MEASUREMENT ACCURACY — NOT ESTABLISHED for confirming or excluding carbon monoxide poisoning. Composition: 1 systematic review/meta-analysis plus 3 primary studies (documented N = 5,701). Papin 2023 (PMID 37171830): 21 studies in the review, 11 in the quantitative accuracy analysis (2,089 subjects, 3,381 observations) — mean sensitivity 77% and specificity 83% (labelled in words), area under the summary curve 0.86; for agreement (2,794 subjects, 4,646 observations) mean bias was 0.75% with limits of agreement −7.08% to +8.57%. Roth 2011 (1,578 unselected emergency patients, PMID 21459480): bias 2.99% carboxyhaemoglobin (1.50% in smokers, 4.33% in non-smokers) with limits of agreement −3.55% to +9.53%; only 17 patients (1.1%) had carbon monoxide poisoning. Koyuncu 2020 (4,073 emergency patients with suspected poisoning, PMID 31416641): 106 (2.6%) were diagnosed with carbon monoxide poisoning and the authors report non-invasive and arterial blood gas measurements as compatible, without a printed bias figure in the record. Caboot 2012 (50 children with sickle cell disease, PMID 22328189): at low carboxyhaemoglobin levels (0.4–6.1%) bias was 0.1% with precision ±2.1%. MEASUREMENT ACCURACY — NOT ESTABLISHED as a stand-alone diagnostic: the average bias looks small, but the limits of agreement span roughly ±8 percentage points of carboxyhaemoglobin — wider than the range that decides treatment — so the meta-analysis authors conclude current devices should probably not be used to confirm or exclude poisoning with certainty.

MeSH: Carboxyhemoglobin · Oximetry · Carbon Monoxide Poisoning · Blood Gas Analysis · Reproducibility of Results

Cards (4)

Card 1 — The pooled verdict: sensitivity 77% / specificity 83%, bias 0.75% but limits of agreement −7.08% to +8.57% (Papin 2023 — meta-analysis, 21 studies)

Study: Systematic review and meta-analysis of diagnostic test accuracy studies (21 studies)

Population: Twenty-one diagnostic accuracy studies of non-invasive pulse CO-oximetry (SpCO) against blood carboxyhaemoglobin in human volunteers or ill patients including children, from four databases searched to August 2022 (PROSPERO CRD42020177940, as printed in the record); 11 studies (2,089 subjects, 3,381 observations) entered the primary accuracy analysis and 18 (2,794 subjects, 4,646 observations) the agreement analysis. Study-level totals are stated in the review and are not added to this shelf's primary total.

Summary: The pooled verdict: sensitivity 77% / specificity 83%, bias 0.75% but limits of agreement −7.08% to +8.57% (Papin 2023 — meta-analysis, 21 studies)

Result: The area under the summary receiver-operating curve was 0.86. Mean sensitivity was 77% (95% CI 66–85%) and mean specificity 83% (95% CI 74–89%) — sensitivity and specificity labelled in words at first occurrence here and throughout this shelf. Mean bias was 0.75% carboxyhaemoglobin, with limits of agreement from −7.08% to +8.57% (95% CI −8.89 to +10.38). No publication bias was found. The authors conclude non-invasive measurement with current pulse CO-oximeters does not seem highly accurate for estimating blood carboxyhaemoglobin and should probably not be used to confirm (rule in) or exclude (rule out) carbon monoxide poisoning with certainty.

Watch out: Synthesis card: its subjects are not added to this shelf's documented primary total (5,701). The stamp follows this review's own conclusion rather than its pooled sensitivity/specificity pair, because the limits of agreement are the decision-relevant number.

Source: Papin M, Latour C, Leclère B et al. Accuracy of pulse CO-oximetry to evaluate blood carboxyhemoglobin level: a systematic review and meta-analysis of diagnostic test accuracy studies. European journal of emergency medicine : official journal of the European Society for Emergency Medicine. 2023. PMID: 37171830. Registry ID: none in the record. MeSH: Child · Humans · Carboxyhemoglobin · Reproducibility of Results · Oximetry · Carbon Monoxide Poisoning · Diagnostic Tests, Routine

Card 2 — In unselected emergency patients: bias 2.99% (4.33% in non-smokers), limits of agreement −3.55% to +9.53% (Roth 2011 — 1,578 patients)

Study: Diagnostic accuracy study (prospective, unselected cohort; n = 1,578)

Population: A prospective STARD diagnostic accuracy study in a tertiary emergency department including all patients over 1 year for whom both non-invasive multiwave pulse oximetry and invasive blood gas carboxyhaemoglobin measurement within 60 minutes were available, regardless of presenting complaint; 1,578 subjects were studied, of whom 17 (1.1%) were diagnosed with carbon monoxide poisoning. N = 1,578, printed on the face of this card.

Summary: In unselected emergency patients: bias 2.99% (4.33% in non-smokers), limits of agreement −3.55% to +9.53% (Roth 2011 — 1,578 patients)

Result: Bias was 2.99% carboxyhaemoglobin overall (1.50% for smokers, 4.33% for non-smokers) with precision 3.27% (2.90% smokers, 2.98% non-smokers) and limits of agreement from −3.55% to +9.53% (−4.30% to +7.30% in smokers; −1.63% to +10.29% in non-smokers). An upper-limit-of-normal cut-off of 6.6% carboxyhaemoglobin had the highest sensitivity in screening for carbon monoxide poisoning (the record prints no numeric sensitivity/specificity pair for that cut-off in its abstract, and none is claimed here). Smoking status and carboxyhaemoglobin level most influenced the deviation between measurements. The authors suggest the device can be used to screen, within the bias and precision they report.

Watch out: Only 17 poisoning cases in 1,578 patients: this is predominantly a low-range agreement study in unselected patients, and its screening suggestion rests on very few true cases — stated on the face, and the reason the shelf's stamp follows the meta-analysis conclusion instead.

Source: Roth D, Herkner H, Schreiber W et al. Accuracy of noninvasive multiwave pulse oximetry compared with carboxyhemoglobin from blood gas analysis in unselected emergency department patients. Annals of emergency medicine. 2011. PMID: 21459480. Registry ID: none in the record. MeSH: Blood Gas Analysis · Carbon Monoxide Poisoning · Carboxyhemoglobin · Emergency Service, Hospital · False Positive Reactions · Female · Humans · Male · Middle Aged · Oximetry · Prospective Studies · ROC Curve · Reproducibility of Results · Sensitivity and Specificity · Smoking

Card 3 — In a suspected-poisoning triage cohort, 2.6% had occult poisoning and the two methods were reported compatible (Koyuncu 2020 — 4,073 patients)

Study: Cross-sectional triage study (n = 4,073)

Population: A cross-sectional study in the emergency department of Kayseri Training and Research Hospital (November 2012 to May 2013) in which patients presenting with signs of carbon monoxide poisoning had carboxyhaemoglobin measured non-invasively by pulse CO-oximetry at triage, with patients at elevated levels investigated further including arterial blood gas measurement; 4,073 patients were enrolled and 106 (2.6%) were diagnosed with carbon monoxide poisoning, headache being the most common presenting complaint in occult cases. N = 4,073, printed on the face of this card.

Summary: In a suspected-poisoning triage cohort, 2.6% had occult poisoning and the two methods were reported compatible (Koyuncu 2020 — 4,073 patients)

Result: Further evaluation found the non-invasive pulse CO-oximeter and arterial blood gas measurements were compatible (the record's abstract prints no bias, limits of agreement or sensitivity/specificity pair, and none is claimed on this card). The authors conclude non-invasive measurement at triage might reduce morbidity and mortality from occult poisoning and prevent unnecessary diagnostic testing. This card is carried as the large-cohort counterweight on the face of the shelf: a favourable qualitative report at scale, without the quantitative agreement figures that cards 1 and 2 supply.

Watch out: The absence of printed agreement statistics in this record is stated on the card itself; a qualitative 'compatible' in 4,073 patients does not override the quantified limits of agreement on cards 1 and 2, and the stamp does not treat it as doing so.

Source: Koyuncu S, Bol O, Ertan T et al. The detection of occult CO poisoning through noninvasive measurement of carboxyhemoglobin: A cross-sectional study. The American journal of emergency medicine. 2020. PMID: 31416641. Registry ID: none in the record. MeSH: Adult · Blood Gas Analysis · Carbon Monoxide Poisoning · Carboxyhemoglobin · Cross-Sectional Studies · Emergency Service, Hospital · Female · Humans · Male · Morbidity · Retrospective Studies · Survival Rate · Triage · Turkey

Card 4 — At low levels in children, agreement is close: bias 0.1%, precision ±2.1% over a 0.4–6.1% range (Caboot 2012 — 50 children)

Study: Agreement study (paired measurements, low range; n = 50)

Population: Fifty children with sickle cell disease (SS genotype) underwent non-invasive pulse co-oximetry and invasive whole-blood co-oximetry while breathing room air, comparing non-invasive carboxyhaemoglobin (SpCO) and methaemoglobin (SpMet) readings with the corresponding blood measurements. N = 50, printed on the face of this card.

Summary: At low levels in children, agreement is close: bias 0.1%, precision ±2.1% over a 0.4–6.1% range (Caboot 2012 — 50 children)

Result: The pulse co-oximeter bias was 0.1% for carboxyhaemoglobin and −0.22% for methaemoglobin. Precision of SpCO was ±2.1% within a carboxyhaemoglobin range of 0.4–6.1%, and precision of SpMet was ±0.33% within a methaemoglobin range of 0.1–1.1%. The authors conclude non-invasive pulse co-oximetry was useful for measuring carboxyhaemoglobin and methaemoglobin in children with sickle cell disease, with close agreement between the methods.

Watch out: Fifty children — N printed on the face of this card. The range studied (up to 6.1%) is far below poisoning levels; close agreement here does not extend the device's validity to the poisoning range, and the card is placed last on the shelf for exactly that reason.

Source: Caboot JB, Jawad AF, McDonough JM et al. Non-invasive measurements of carboxyhemoglobin and methemoglobin in children with sickle cell disease. Pediatric pulmonology. 2012. PMID: 22328189. Registry ID: none in the record. MeSH: Adolescent · Anemia, Sickle Cell · Blood Gas Analysis · Carboxyhemoglobin · Child · Child, Preschool · Female · Humans · Male · Methemoglobin · Oximetry · Reproducibility of Results · Spectrophotometry