SourceMed · Shelf #9491 in the ask interface
Methylated SEPT9 blood test (Epi proColon) for colorectal cancer — 1 meta-analysis (19 studies) plus 3 primary studies (documented N = 1,221): pooled sensitivity 69% / specificity 92%, second-generation assay sensitivity 74.8% / specificity 87.4% with only 27.4% of advanced adenomas detected, and a prospective head-to-head with the faecal immunochemical test (sensitivity 73.3% versus 68.0%; specificity 81.5% versus 97.4%)
Shelf #9491 · Cancer
MEASUREMENT ACCURACY — pooled sensitivity 69% / specificity 92% (labelled in words) for colorectal cancer across 19 studies: specific but only moderately sensitive, detecting 66.7% of stage I cancers in the second-generation assay study and only 27.4% of advanced adenomas, with head-to-head specificity well below the faecal immunochemical test (81.5% versus 97.4%) — NOT ESTABLISHED as a stand-alone screening replacement, a possible option for people declining stool testing
Stamped: 2026-11-28.
Evidence published: 2011–2020.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately does the plasma methylated SEPT9 blood test (Epi proColon) detect colorectal cancer against colonoscopy-verified status — its pooled sensitivity/specificity, its performance by cancer stage and for advanced adenomas, and how it compares head-to-head with the faecal immunochemical test?
Balance of the evidence
MEASUREMENT ACCURACY — Composition: 1 meta-analysis plus 3 primary studies (documented N = 1,221). Hariharan & Jenkins 2020 (PMID 32128229): across 19 studies, pooled sensitivity was 69% and specificity 92% (labelled in words) for colorectal cancer, with area under the curve 0.89; in an average-risk population (0.3% prevalence) the positive predictive value was 2.6% and negative predictive value 99.9%. Warren 2011 (PMID 22168215): the improved assay showed sensitivity 90% / specificity 88% in case-control samples (50 cancers, 94 controls) and detected 12% of adenomas in a 300-person colonoscopy cohort. Jin 2015 (476 colonoscopy-verified subjects, PMID 25471329): the second-generation Epi proColon 2.0 test (Epigenomics — the maker as named in the record's test name) gave sensitivity 74.8% / specificity 87.4% for cancer, rising by stage (66.7% stage I to 100% stage IV), but sensitivity for advanced adenomas was only 27.4%. Johnson 2014 (301 subjects, PMID 24901436): prospectively, Epi proColon sensitivity was 73.3% versus 68.0% for the faecal immunochemical test, but specificity was 81.5% versus 97.4%; combining both tests detected 88.7% at specificity 78.8%. PROPOSED MEASUREMENT ACCURACY: the blood test is a high-specificity, moderate-sensitivity cancer detector whose weakness is precancerous lesions — it misses roughly three-quarters of advanced adenomas — so the pooled evidence supports it, at most, as an option for people who decline stool-based screening, not as an equivalent replacement.
MeSH: Septins · Colorectal Neoplasms · Early Detection of Cancer · DNA Methylation · Sensitivity and Specificity
Cards (4)
Card 1 — The pooled pair: sensitivity 69% / specificity 92%, area under the curve 0.89, positive predictive value 2.6% at screening prevalence (Hariharan & Jenkins 2020 — meta-analysis, 19 studies)
Study: Meta-analysis of diagnostic test accuracy studies (19 studies)
Population: Nineteen studies of the methylated SEPT9 test for early diagnosis of colorectal cancer, identified from Medline and Embase, pooled for sensitivity, specificity and diagnostic accuracy. Participant totals are stated per study in the review and are not summed on this card.
Summary: The pooled pair: sensitivity 69% / specificity 92%, area under the curve 0.89, positive predictive value 2.6% at screening prevalence (Hariharan & Jenkins 2020 — meta-analysis, 19 studies)
Result: Pooled sensitivity was 69% (95% CI 62–75%) and pooled specificity 92% (95% CI 89–95%) — sensitivity and specificity labelled in words at first occurrence here and throughout this shelf — with positive likelihood ratio 9.1, negative likelihood ratio 0.34, diagnostic odds ratio 27 and area under the curve 0.89 (0.86–0.91). At an average-risk prevalence of 0.3%, the positive predictive value was 2.6% and the negative predictive value 99.9%; at a high-risk prevalence of 1.2%, 9.5% and 99.6%. The authors conclude the test has high specificity and moderate sensitivity, is a potential alternative for people declining the faecal immunochemical test, but is limited by poor performance for precancerous lesions, relatively high cost and unknown acceptability.
Watch out: Synthesis card: its participants are not added to this shelf's documented primary total (1,221). The 2.6% positive predictive value at screening prevalence is carried on the face — at population prevalence, most positive blood tests are not cancer.
Source: Hariharan R, Jenkins M et al. Utility of the methylated SEPT9 test for the early detection of colorectal cancer: a systematic review and meta-analysis of diagnostic test accuracy. BMJ open gastroenterology. 2020. PMID: 32128229. Registry ID: none in the record. MeSH: Colorectal Neoplasms · Diagnostic Tests, Routine · Early Detection of Cancer · Humans · Occult Blood · Septins
Card 2 — The optimistic early assay: sensitivity 90% / specificity 88% in case-control samples, 12% of adenomas prospectively (Warren 2011 — 444 subjects)
Study: Diagnostic accuracy study (case-control plus small prospective cohort; n = 444)
Population: Plasma samples from 50 untreated colorectal cancer patients at 3 institutions and 94 control samples from 4 US institutions, plus plasma collected from 300 colonoscopy patients at 1 US clinic before endoscopy, tested for methylated SEPT9 DNA concentration with the improved-sensitivity assay. N = 444, printed on the face of this card.
Summary: The optimistic early assay: sensitivity 90% / specificity 88% in case-control samples, 12% of adenomas prospectively (Warren 2011 — 444 subjects)
Result: The test's overall sensitivity for colorectal cancer was 90% (95% CI 77.4–96.3%) and specificity 88% (95% CI 79.6–93.7%), with sensitivity 87% for stage I–II cancer, detection from all colonic regions including the proximal colon, and a 12% false-positive rate in the case-control set. In the small prospective colonoscopy study, the test detected 12% of adenomas with a false-positive rate of 3%. The authors conclude the blood test specifically detects a majority of cancers of all stages and locations and could be offered to average-risk individuals who decline other screening.
Watch out: Case-control sampling (known cancers versus controls) characteristically flatters accuracy relative to screening use; the pooled estimate on card 1 (sensitivity 69%) is the shelf's headline for that reason, and this card's higher pair is labelled with its design on the face.
Source: Warren JD, Xiong W, Bunker AM et al. Septin 9 methylated DNA is a sensitive and specific blood test for colorectal cancer. BMC medicine. 2011. PMID: 22168215. Registry ID: none in the record. MeSH: Adult · Aged · Biomarkers, Tumor · Case-Control Studies · Colorectal Neoplasms · DNA Methylation · DNA, Neoplasm · Early Detection of Cancer · Humans · Mass Screening · Middle Aged · Prospective Studies · Real-Time Polymerase Chain Reaction · Sensitivity and Specificity · Septins
Card 3 — Second-generation assay, colonoscopy-verified: sensitivity 74.8% / specificity 87.4% for cancer, 27.4% for advanced adenomas (Jin 2015 — 476 subjects)
Study: Diagnostic accuracy study (colonoscopy-verified, four-group; n = 476)
Population: A total of 476 subjects whose clinical status was verified by colonoscopy — 135 with colorectal cancer, 169 with adenomatous polyps, 81 with hyperplastic polyps and 91 healthy controls — all tested with the second-generation methylated SEPT9 assay (Epi proColon 2.0); 177 patients also had a faecal immunochemical test. N = 476, printed on the face of this card.
Summary: Second-generation assay, colonoscopy-verified: sensitivity 74.8% / specificity 87.4% for cancer, 27.4% for advanced adenomas (Jin 2015 — 476 subjects)
Result: Sensitivity of SEPT9 for colorectal cancer was 74.8% (95% CI 67.0–81.6%) and specificity 87.4% (against non-cancer, 95% CI 83.5–90.6%). Sensitivity by stage was 66.7% (stage I), 82.6% (stage II), 84.1% (stage III) and 100% (stage IV). Sensitivity for advanced adenomas was 27.4% (95% CI 18.7–37.6%). The faecal immunochemical test in the paired subset showed sensitivity 58.0% / specificity 82.4% for cancer. The authors conclude the second-generation assay showed better cancer detection than the faecal test in this sample, with similar performance for advanced adenomas.
Watch out: The specificity denominator here mixes polyp patients and healthy controls, and the paired faecal-test subset (177) differs from the prospective head-to-head on card 4, whose specificity ordering reverses — both designs are labelled on their faces rather than merged.
Source: Jin P, Kang Q, Wang X et al. Performance of a second-generation methylated SEPT9 test in detecting colorectal neoplasm. Journal of gastroenterology and hepatology. 2015. PMID: 25471329. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Asian People · Biomarkers, Tumor · Colorectal Neoplasms · Early Diagnosis · Female · Humans · Male · Middle Aged · Neoplasm Staging · Occult Blood · Reagent Kits, Diagnostic · Sensitivity and Specificity · Septins · Young Adult
Card 4 — The prospective screening head-to-head: sensitivity 73.3% vs 68.0% for the faecal test, specificity 81.5% vs 97.4% (Johnson 2014 — 301 subjects)
Study: Prospective comparative accuracy study (n = 301)
Population: A prospective multicentre study enrolling 102 patients with screening-identified colorectal cancer who provided blood and stool samples before surgery (for sensitivity) and 199 prospectively enrolled patients who provided samples before bowel preparation for screening colonoscopy (for specificity), analysed with the Epi proColon test and the OC Fit-Check faecal immunochemical test (both as named in the record). N = 301, printed on the face of this card.
Summary: The prospective screening head-to-head: sensitivity 73.3% vs 68.0% for the faecal test, specificity 81.5% vs 97.4% (Johnson 2014 — 301 subjects)
Result: Sensitivity for cancer detection was 73.3% (95% CI 63.9–80.9%) for Septin9 and 68.0% (95% CI 58.2–76.5%) for the faecal immunochemical test; specificity was 81.5% (95% CI 75.5–86.3%) for Epi proColon compared with 97.4% (95% CI 94.1–98.9%) for the faecal test. In paired samples, Septin9 sensitivity (72.2%) was statistically non-inferior to the faecal test; combining both tests detected 88.7% of cancers at a specificity of 78.8%. The authors conclude that at a sensitivity of 72%, the Epi proColon test is non-inferior to the faecal test for cancer detection.
Watch out: Sensitivity and specificity were estimated in two separately enrolled groups (cancer patients for one, screening patients for the other), not in a single screening cohort — the design is stated on the face, and it is why this shelf does not present a single-cohort screening accuracy for the blood test.
Source: Johnson DA, Barclay RL, Mergener K et al. Plasma Septin9 versus fecal immunochemical testing for colorectal cancer screening: a prospective multicenter study. PloS one. 2014. PMID: 24901436. Registry ID: none in the record. MeSH: Aged · Aged, 80 and over · Colonoscopy · Colorectal Neoplasms · Early Detection of Cancer · Female · Humans · Male · Mass Screening · Middle Aged · Neoplasm Staging · Occult Blood · Prospective Studies · Reproducibility of Results · Sensitivity and Specificity · Septins