Breath alcohol analysers vs blood alcohol concentration — 4 evaluation studies (documented N = 83 volunteers plus 714 offence records; 1 simulator study states no participant N): evidential breath analysers agree closely with venous blood in the elimination phase, read high just after drinking, and consumer smartphone-paired devices all underestimated blood alcohol by more than 0.01%

Shelf #9540 · Emergency & first hours

MEASUREMENT ACCURACY — established for evidential instruments used by trained operators in the elimination phase, NOT ESTABLISHED for consumer smartphone-paired devices: two evidential breath instruments agreed with each other to a mean bias of 0.003 mg/L (limits of agreement −0.016 to +0.023 mg/L), breath read higher than venous blood 15 minutes after drinking and lower thereafter, and all six consumer devices tested underestimated blood alcohol by more than 0.01%

Stamped: 2026-11-28.

Evidence published: 2005–2021.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do breath alcohol analysers — evidential police instruments, hand-held roadside screeners and consumer smartphone-paired devices — estimate blood alcohol concentration measured from venous blood?

Balance of the evidence

MEASUREMENT ACCURACY — established for evidential instruments used by trained operators in the elimination phase, NOT ESTABLISHED for consumer smartphone-paired devices — Composition: 4 evaluation studies (documented N = 83 volunteers — 63 + 20 — plus 714 retrospectively analysed offence records; 1 simulator study states no participant N). Jurič 2018: back-calculation of blood results in 714 offence records found no significant difference between calculated blood alcohol and the roadside breath result in the elimination phase, and in a controlled drinking study (63 volunteers; named Dräger 6810 hand-held analyser) breath and blood again did not differ significantly. Fransson 2005 (20 volunteers, ethanol 0.4 g/kg): the new Evidenzer evidential analyser and the established Intoxilyzer 5000S agreed with a mean bias of only 0.003 mg/L and 95% limits of agreement of −0.016 to +0.023 mg/L, with breath measurement standard deviation 0.006 mg/L (coefficient of variation 4.7%); against venous blood, both breath instruments read higher at 15 minutes after drinking and lower at all later times, and breath–blood correlation was r = 0.95. Delgado 2021 (20 volunteers dosed toward a peak blood alcohol of 0.10%): all seven devices — a police-grade Intoxilyzer 240 and six consumer smartphone-paired devices — underestimated blood alcohol by more than 0.01% at peak, with two consumer brands (Drinkmate and Evoc, as named in the record) consistently less accurate than the police device and the best consumer device. Razatos 2005 (simulator evaluation, no participants): the portable Intoxilyzer 8000 was evaluated against the stationary Intoxilyzer 5000 across ethanol concentrations from 0.04 to 0.55 g/100 mL using wet-bath simulator controls, with attention to the 0.08 and 0.16 legal levels. MEASUREMENT ACCURACY — established for evidential instruments used by trained operators in the elimination phase, NOT ESTABLISHED for consumer smartphone-paired devices: breath alcohol is a trustworthy estimate of blood alcohol under evidential conditions and timing, and an systematically low one from consumer devices at peak concentration.

MeSH: Breath Tests · Ethanol · Alcoholic Intoxication · Sensitivity and Specificity

Cards (4)

Card 1 — Roadside breath results matched back-calculated blood alcohol in 714 offence records, and breath matched blood in a 63-volunteer drinking study (Jurič 2018)

Study: Primary accuracy study (records analysis plus controlled drinking study; 714 records, 63 volunteers)

Population: Two limbs: a retrospective analysis of 714 offence records from 2011, in which blood alcohol measured 0–303 minutes after roadside breath screening was back-calculated to the screening time (all offenders were in the alcohol elimination phase); and a controlled drinking study in 63 healthy volunteers, in whom breath alcohol was measured immediately after drinking with a hand-held analyser (named in the record as the Dräger model 6810) and 29 participants gave blood and urine 1 hour later for gas-chromatography alcohol measurement.

Summary: Roadside breath results matched back-calculated blood alcohol in 714 offence records, and breath matched blood in a 63-volunteer drinking study (Jurič 2018)

Result: In the offence records, the analyses did not reveal any significant differences between back-calculated blood alcohol and the breath screening result. In the drinking study, breath alcohol measured immediately after consumption and blood alcohol measured 1 hour later again showed no significant differences. The authors concluded the hand-held breath analysers approved for roadside screening are highly reliable when police officers perform measurements according to the manufacturer's instructions.

Watch out: 'No significant difference' is a group-level statement; the record does not report limits of agreement for individual pairs, so individual-level interchangeability is not established by this study.

Source: Jurič A et al. Evaluation of breath alcohol analysers by comparison of breath and blood alcohol concentrations. Arhiv za higijenu rada i toksikologiju. 2018. PMID: 29604198. Registry ID: none in the record. MeSH: Adult · Alcoholic Intoxication · Blood Alcohol Content · Breath Tests · Ethanol · Female · Healthy Volunteers · Humans · Male · Middle Aged · Reproducibility of Results · Retrospective Studies · Time Factors

Card 2 — Two evidential breath analysers agreed to a bias of 0.003 mg/L; both read above venous blood at 15 minutes after drinking and below it later (Fransson 2005, N = 20)

Study: Primary accuracy study (two evidential analysers vs venous blood, controlled dosing; n = 20)

Population: Twenty healthy volunteers (10 men, 10 women) drank ethanol 0.4 g/kg in 15 minutes, 2–3 hours after their last meal; venous blood and breath were sampled every 15–30 minutes for up to 4 hours. Breath was measured with the new Evidenzer evidential analyser and the established Intoxilyzer 5000S, and analytical precision was also assessed on simulator-generated ethanol vapour at 0.50 mg/L.

Summary: Two evidential breath analysers agreed to a bias of 0.003 mg/L; both read above venous blood at 15 minutes after drinking and below it later (Fransson 2005, N = 20)

Result: The two breath instruments agreed closely: mean bias 0.003 mg/L with 95% limits of agreement of −0.016 to +0.023 mg/L, and measurement standard deviation about 0.006 mg/L for both (coefficient of variation 4.7%; 0.7% on simulator vapour). Breath and venous blood alcohol were highly correlated (r = 0.95), but both breath instruments gave results higher than venous blood at 15 minutes after the end of drinking, whereas at all later times venous blood alcohol was higher — the breath-to-blood relationship reverses between the absorption and elimination phases.

Watch out: Twenty volunteers at a modest dose (0.4 g/kg) — N printed on the face of this card. The 210-to-1 style fixed breath:blood conversion is not what this study validates; it validates the instruments against each other and shows the phase-dependence of the blood comparison.

Source: Fransson M et al. Laboratory evaluation of a new evidential breath-alcohol analyser designed for mobile testing--the Evidenzer. Medicine, science, and the law. 2005. PMID: 15745276. Registry ID: none in the record. MeSH: Adult · Alcoholic Intoxication · Automobile Driving · Breath Tests · Female · Humans · Male · Time Factors

Card 3 — All six consumer smartphone-paired breath devices underestimated peak blood alcohol by more than 0.01% (Delgado 2021, N = 20)

Study: Primary accuracy study (consumer devices vs police device and phlebotomy blood alcohol; n = 20)

Population: Twenty healthy moderate drinkers (mean age 27) received weight-based ethanol doses targeting a peak blood alcohol concentration of 0.10%. Breath alcohol was measured every 20 minutes with a police-grade device (named in the record as the Intoxilyzer 240) and with two randomly ordered series of three consumer smartphone-paired devices (six devices in total, each series tested by 10 participants), alongside peak phlebotomy blood alcohol.

Summary: All six consumer smartphone-paired breath devices underestimated peak blood alcohol by more than 0.01% (Delgado 2021, N = 20)

Result: Peak blood alcohol ranged from 0.06% to 0.14%. All seven devices underestimated blood alcohol by more than 0.01% at peak, though the BACtrack Mobile Pro and the police-grade device were consistently more accurate than the Drinkmate and Evoc devices (brands as named in the record). The authors concluded the accuracy of consumer smartphone-paired devices had not been established by this laboratory validation.

Watch out: Twenty young moderate drinkers in a laboratory — N printed on the face of this card. Underestimation at peak is the safety-relevant direction for a device marketed to decide whether one is fit to drive.

Source: Delgado MK et al. Accuracy of Consumer-marketed smartphone-paired alcohol breath testing devices: A laboratory validation study. Alcoholism, clinical and experimental research. 2021. PMID: 33966283. Registry ID: none in the record. MeSH: Adult · Blood Alcohol Content · Breath Tests · Direct-To-Consumer Screening and Testing · Female · Humans · Male · Reproducibility of Results · Smartphone · Young Adult

Card 4 — A portable evidential analyser (Intoxilyzer 8000) was evaluated against the stationary Intoxilyzer 5000 on simulator standards from 0.04 to 0.55 g/100 mL (Razatos 2005; simulator study, no participants)

Study: Device evaluation study (simulator comparison of portable vs stationary evidential analysers; no participants)

Population: A regulatory laboratory evaluation for a state implied-consent programme: the portable Intoxilyzer 8000's precision and accuracy were assessed against the approved stationary Intoxilyzer 5000 using in-house ethanol controls in a wet-bath simulator, at concentrations ranging from 0.04 to 0.55 g/100 mL in blood-equivalent terms, with special attention to the per-se level of 0.08 g/100 mL and the aggravated level of 0.16 g/dL stated in the record.

Summary: A portable evidential analyser (Intoxilyzer 8000) was evaluated against the stationary Intoxilyzer 5000 on simulator standards from 0.04 to 0.55 g/100 mL (Razatos 2005; simulator study, no participants)

Result: The record describes the evaluation design and its regulatory purpose — establishing whether the portable instrument's performance supports replacing stationary testing in a geographically large, rural state — and reports precision and accuracy testing across the simulator range. Numerical coefficients of variation from the evaluation are not stated in the abstract record, and no human blood comparison was performed in this study; the human breath–blood evidence on this shelf comes from the Jurič, Fransson and Delgado studies.

Watch out: A simulator study — printed on the face of this card. Simulator agreement shows an instrument measures ethanol vapour precisely; it does not by itself establish agreement with blood alcohol in drinking humans.

Source: Razatos G et al. Evaluation of a portable evidential breath alcohol analyzer. Forensic science international. 2005. PMID: 15935582. Registry ID: none in the record. MeSH: Altitude · Breath Tests · Central Nervous System Depressants · Ethanol · Forensic Medicine · Humans · New Mexico · Substance Abuse Detection · Temperature