Hepatitis C virus self-testing accuracy vs laboratory and trained-user reference standards — 3 field studies (documented N = 855 participants): self-read oral-fluid testing found sensitivity 88.4% where trained staff reading the same tests found 97.7% (labelled in words: sensitivity and specificity are reported per study below), 95% of lay users completed saliva self-testing successfully but positive agreement with a trained user's rapid test was only 70%, and a blood-based self-test in an Indian slum cohort reported sensitivity 97.4% / specificity 99.6% on a 35-person ELISA-positive panel
Shelf #9541 · Liver & gut
MEASUREMENT ACCURACY — NOT ESTABLISHED for unsupervised self-reading: across 3 field studies (documented N = 855) the act of self-reading cost about 9 points of sensitivity on oral fluid (88.4% self-read vs 97.7% staff-read on the same self-collected tests), positive percent agreement with a trained user's second rapid test was 70% in the Karachi study, and the one high pair (sensitivity 97.4% / specificity 99.6%) rests on a 35-person positive panel
Stamped: 2026-11-28.
Evidence published: 2019–2026.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurate is hepatitis C virus self-testing — self-collected, self-performed and self-read antibody tests — against laboratory reference testing and against the same or a second rapid test performed and read by trained staff?
Balance of the evidence
MEASUREMENT ACCURACY — NOT ESTABLISHED for unsupervised self-reading — Composition: 3 field studies (documented N = 855; 95 + 295 + 465). Kimble 2019 (95 participants, oral-fluid OraQuick HCV test — OraSure Technologies, as named in the record — self-collected and self-interpreted, with a researcher reading the same test independently and true status from records or venipuncture): sensitivity was 88.4% (95% CI 74.9–96.1) and specificity 100% (93–100) when participants read their own tests, against sensitivity 97.7% (88–99.9) and specificity 98% (89.6–100) when trained staff read them — the self-reading step, not the self-collection, is where sensitivity is lost. Mazzilli 2024 (295 participants in a Karachi informal settlement, saliva-based self-test): 95% of tests were completed successfully, but only 13% of participants managed without any verbal assistance, inter-reader concordance was 96%, and positive percent agreement with a second rapid test performed by a trained user was only 70%. Bhadoria 2026 (465 participants in northern-India urban slums performing blood-based self-testing under observation, plus 35 ELISA-positive individuals recruited for the accuracy panel): over 90% performed critical steps correctly, 96% interpreted results accurately, and the kit's sensitivity was 97.4% and specificity 99.6% — on a positive panel of 35. The studies agree that lay users can perform the tests; they do not yet agree that self-read results equal trained results. MEASUREMENT ACCURACY — NOT ESTABLISHED for unsupervised self-reading: hepatitis C self-testing is performable and acceptable in the field, but self-interpretation measurably lowers sensitivity on oral fluid, and the accuracy base is three field studies, one with a very small positive panel.
MeSH: Hepatitis C Antibodies · Self Testing · Sensitivity and Specificity · Mass Screening
Cards (3)
Card 1 — Self-reading cost 9 points of sensitivity: 88.4% when participants read their own oral-fluid tests vs 97.7% when staff read the same tests (Kimble 2019, N = 95)
Study: Primary accuracy study (self-collected oral fluid, self-read vs staff-read; n = 95)
Population: Ninety-five participants (48 men, 47 women) used the OraQuick HCV Rapid Antibody Test (OraSure Technologies, Inc., Bethlehem, PA — named in the record) on self-collected oral fluid, following the kit instructions and interpreting their own result; a researcher interpreted each test simultaneously and independently. True antibody status was determined from medical records or a venipuncture blood sample.
Summary: Self-reading cost 9 points of sensitivity: 88.4% when participants read their own oral-fluid tests vs 97.7% when staff read the same tests (Kimble 2019, N = 95)
Result: When participants interpreted their own tests, sensitivity was 88.4% (95% CI 74.9–96.1) and specificity 100% (95% CI 93–100). When trained staff interpreted the same self-collected tests, sensitivity was 97.7% (95% CI 88–99.9) and specificity 98% (95% CI 89.6–100). Agreement between participant and staff reading was κ = 0.89 (95% CI 0.80–0.98). The authors concluded the rapid HCV test showed good performance for self-testing of oral fluid specimens.
Watch out: The sensitivity gap is a reading effect on identical specimens — self-collection is exonerated by this design, self-interpretation is not. The confidence interval on self-read sensitivity runs to 74.9%, and the setting was a supervised study, not the home.
Source: Kimble MM et al. Clinical evaluation of a hepatitis C antibody rapid immunoassay on self-collected oral fluid specimens. Diagnostic microbiology and infectious disease. 2019. PMID: 31204109. Registry ID: none in the record. MeSH: Adult · Aged · Aged, 80 and over · Female · Hepacivirus · Hepatitis C · Hepatitis C Antibodies · Humans · Immunoassay · Male · Middle Aged · Reagent Kits, Diagnostic · Reproducibility of Results · Saliva · Self Care · Sensitivity and Specificity · Specimen Handling · Young Adult
Card 2 — 95% of lay users completed saliva self-testing successfully, but positive agreement with a trained user's rapid test was only 70% (Mazzilli 2024, N = 295)
Study: Primary accuracy and usability study (saliva self-test vs trained-user rapid test; n = 295)
Population: In a hepatitis C clinic in a high-prevalence informal settlement in Karachi, Pakistan, 295 participants (97, or 32%, illiterate) performed a saliva-based HCV self-test (OraSure Technologies, USA — named in the record) following pictorial instructions, with a study member scoring performance on a standardised checklist and giving verbal support when a step could not be completed. Self-test results were compared with the study member's reading of the same test and with a second rapid HCV test (Abbott Diagnostics Korea Inc — named in the record) performed by a trained user.
Summary: 95% of lay users completed saliva self-testing successfully, but positive agreement with a trained user's rapid test was only 70% (Mazzilli 2024, N = 295)
Result: In total 280 of 295 self-tests (95%, 95% CI 92–97%) were successful, but only 38 participants (13%) completed the test without verbal assistance, 67 (23%) needed help with one step and 190 (64%) with two or more steps, most often managing the buffer and reading times. Inter-reader concordance was 96% and inter-operator concordance 93%; inter-reader and inter-operator positive percent agreement were 84% and 70% respectively. All participants said they would use the self-test again and recommend it.
Watch out: Positive percent agreement of 70% is not sensitivity against a laboratory reference — it is agreement with a second rapid test, printed as the record defines it. Most participants needed verbal help, so this is assisted self-testing, not the unsupervised home use the accuracy question ultimately concerns.
Source: Mazzilli S et al. Usability and acceptability of self-testing for hepatitis C virus exposure in a high-prevalence urban informal settlement in Karachi, Pakistan. BMC infectious diseases. 2024. PMID: 39333922. Registry ID: none in the record. MeSH: Humans · Pakistan · Male · Female · Cross-Sectional Studies · Hepatitis C · Adult · Middle Aged · Prevalence · Self-Testing · Hepacivirus · Mass Screening · Hepatitis C Antibodies · Patient Acceptance of Health Care · Urban Population · Surveys and Questionnaires · Young Adult · Saliva
Card 3 — Blood-based self-testing under observation: 96% interpreted results accurately; kit sensitivity 97.4% / specificity 99.6% on a small recruited positive panel (Bhadoria 2026, N = 465 + 35)
Study: Primary accuracy and usability study (observed blood-based self-testing; n = 465, accuracy panel including 35 ELISA-positive recruits)
Population: In urban slums of Haridwar, northern India, 465 participants recruited from community and urban primary-health-centre settings performed blood-based HCV self-testing under observation (nested in a larger cluster randomised trial, registration CTRI/2023/07/054590 as given in the record), with errors tracked on standardised checklists and inter-reader concordance assessed against trained personnel. Separately, 35 ELISA-positive individuals were recruited to assess the kit's sensitivity and specificity.
Summary: Blood-based self-testing under observation: 96% interpreted results accurately; kit sensitivity 97.4% / specificity 99.6% on a small recruited positive panel (Bhadoria 2026, N = 465 + 35)
Result: Over 90% of participants correctly performed critical steps such as finger pricking, and 96% interpreted results accurately (98.8% with education beyond high school vs 94.4% without, P = 0.02). Inter-reader concordance was 95.9% (Cohen's κ = 0.25, 95% CI 0.07–0.43). The kit's sensitivity was 97.4% and specificity 99.6% on the recruited panel. Satisfaction was 94.2% and 95.4% trusted the test's accuracy; challenges included blood collection, instructions and the absence of counselling.
Watch out: The headline sensitivity rests on 35 recruited ELISA-positive individuals — printed on the face of this card — and testing was observed, not unsupervised. The low κ (0.25) alongside 95.9% concordance reflects the low prevalence in the usability cohort, as the record's own figures show.
Source: Bhadoria AS et al. Assessment of Usability and Acceptability of Blood-based Hepatitis C Virus Self-testing Among Residents of Urban Slums in Northern India: A Cross-sectional Study. Journal of clinical and experimental hepatology. 2026. PMID: 41536884. Registry ID: CTRI/2023/07/054590 (parent cluster trial, as given in the record). MeSH: