Computerised ECG interpretation vs expert reference reading — 1 meta-analysis and 3 primary studies (documented N = 55,376 ECGs in the synthesis; 5,000, 117 and 69 ECGs or patients in the primaries): automated software detects atrial fibrillation at sensitivity 89% / specificity 99% in the pool (labelled in words at first occurrence) but misclassified atrial fibrillation in 30% of flagged ECGs in the largest primary, and automated ST-segment measurement disagrees with cardiologists beyond clinically acceptable limits

Shelf #9542 · Heart, vessels, blood pressure & clotting

MEASUREMENT ACCURACY — NOT ESTABLISHED as a standalone replacement for expert over-read: pooled automated sensitivity for atrial fibrillation was 89% with specificity 99% across 55,376 ECGs, but in the largest primary the computer misclassified 30% of its atrial-fibrillation-related ECGs (sensitivity 78% vs the cardiologists' 87%), automated STEMI statements were wrong in 26.5% of confirmed STEMI cases, and automated ST-deviation measurement showed clinically inadequate limits of agreement with a cardiologist

Stamped: 2026-11-28.

Evidence published: 2005–2025.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurate is computerised (automated) electrocardiogram interpretation against cardiologist or expert reference reading — for atrial fibrillation, for ST-elevation myocardial infarction statements, and for the underlying ST-segment measurements?

Balance of the evidence

MEASUREMENT ACCURACY — NOT ESTABLISHED as a standalone replacement for expert over-read — Composition: 1 systematic review and meta-analysis plus 3 primary studies (55,376 participant ECGs in the synthesis; 5,000 + 117 + 69 in the primaries — overlapping evidence, counted as reported). Taggar 2015 (10 studies, 16 interpretation methods, 55,376 ECGs): automated software detected atrial fibrillation at pooled sensitivity 0.89 (95% CI 0.82–0.93) and specificity 0.99 (0.99–0.99), with healthcare professionals showing similar sensitivity (0.92) but lower specificity (0.93) in indirect comparison. Escabí-Mendoza 2025 (5,000 ECGs from a veterans' health system, all computer-interpreted, with blinded review by two cardiologists): of 803 ECGs the computer flagged as atrial-fibrillation-related, 70% were true AF, 11% false positives and 19% false negatives by the study's accounting — computer sensitivity 78% and specificity 97% against cardiologist sensitivity 87% and specificity 99%; ventricular pacing, undetermined rhythm and tachycardia independently predicted misclassification, and 41% of computer errors went unrecognised by treating providers, with delayed or inappropriate anticoagulation documented. Kim 2022 (117 patients with confirmed STEMI): the automated interpretation misread 26.5% (31/117) of STEMI ECGs, more often with single-vessel than multi-vessel occlusion (31.3% vs 12.9%, P = 0.045). Eskola 2005 (69 patients with suspected acute coronary syndrome): agreement between computer and cardiologist on ST elevation ≥0.2 mV was only moderate in lead V2 (κ = 0.72), and Bland–Altman limits of agreement for ST-segment deviation were clinically inadequate, although a lowered computer cut-off of 0.115 mV detected cardiologist-defined elevation with sensitivity 89% and specificity 98%. MEASUREMENT ACCURACY — NOT ESTABLISHED as a standalone replacement for expert over-read: computerised ECG interpretation is a good screening reader — highly specific for atrial fibrillation in pooled data — whose errors concentrate in exactly the high-stakes patterns (fibrillation with pacing or tachycardia, single-vessel STEMI, borderline ST deviation) where a wrong statement changes treatment.

MeSH: Electrocardiography · Diagnosis, Computer-Assisted · Atrial Fibrillation · ST Elevation Myocardial Infarction

Cards (4)

Card 1 — Pooled automated software: sensitivity 0.89 / specificity 0.99 for atrial fibrillation on the 12-lead ECG (Taggar 2015, meta-analysis of 55,376 ECGs)

Study: Systematic review and meta-analysis (diagnostic accuracy of ECG interpretation methods; 55,376 ECGs)

Population: A systematic review and meta-analysis (searches to March 2014) of methods for diagnosing atrial fibrillation from 12-lead ECGs: 10 studies investigating 16 interpretation methods in 55,376 participant ECGs, appraised with QUADAS-2, pooled with a bivariate hierarchical random-effects model. Sensitivity and specificity are labelled in words here at first occurrence on this shelf, per the shelf's labelling rule.

Summary: Pooled automated software: sensitivity 0.89 / specificity 0.99 for atrial fibrillation on the 12-lead ECG (Taggar 2015, meta-analysis of 55,376 ECGs)

Result: Automated software (8 studies, 9 methods) achieved sensitivity 0.89 (95% CI 0.82–0.93) and specificity 0.99 (95% CI 0.99–0.99), positive likelihood ratio 96.6 and negative likelihood ratio 0.11. Healthcare professionals (5 studies, 7 methods) in indirect comparison had similar sensitivity (0.92, 95% CI 0.81–0.97) but lower specificity (0.93, 95% CI 0.76–0.98); among primary-care professionals, GPs were more specific (0.96) than nurses. The authors concluded the best interpretation method was not established by direct comparison.

Watch out: The software-versus-human comparison is indirect — different studies, not head-to-head readings of the same ECGs — and the pool predates modern machine-learning interpreters; it estimates conventional automated interpretation, printed as published.

Source: Taggar JS et al. Accuracy of methods for diagnosing atrial fibrillation using 12-lead ECG: A systematic review and meta-analysis. International journal of cardiology. 2015. PMID: 25705010. Registry ID: none in the record. MeSH: Atrial Fibrillation · Electrocardiography · Humans · Software

Card 2 — In 5,000 ECGs the computer misclassified atrial fibrillation in 30% of flagged cases; cardiologists were more accurate on every metric (Escabí-Mendoza 2025, N = 5,000 ECGs)

Study: Primary accuracy study (automated interpretation vs blinded cardiologist review; 5,000 ECGs)

Population: A retrospective study of 5,000 ECGs from the Veterans Affairs Caribbean Healthcare System (September–November 2017), all initially interpreted by automated software; a subset of 905 had been over-read at the time of care, and two board-certified cardiologists independently re-reviewed all 5,000 blinded, as the reference. Clinical records were reviewed for treatment consequences of errors.

Summary: In 5,000 ECGs the computer misclassified atrial fibrillation in 30% of flagged cases; cardiologists were more accurate on every metric (Escabí-Mendoza 2025, N = 5,000 ECGs)

Result: The computer identified 803 (16%) atrial-fibrillation-related ECGs; of these, 70% were true AF, 11% false positives and 19% false negatives. Computer accuracy was sensitivity 78%, specificity 97%, positive predictive value 86% and negative predictive value 97%, against cardiologist sensitivity 87%, specificity 99%, positive predictive value 98% and negative predictive value 99% (all P < 0.01). Ventricular pacing (odds ratio 10.2), undetermined rhythm (13.4) and tachycardia (1.8) independently predicted misclassification. Among misclassified ECGs, 41% of errors were unrecognised by providers, leading to delayed anticoagulation (18%), inappropriate initiation (12%) and three adverse events.

Watch out: A single health system's 2017 software and a veterans' population — device- and era-specific, printed as published. The 70/11/19 split is the study's accounting of its flagged set; the sensitivity and specificity pair is computed against the cardiologist reference across all 5,000 ECGs.

Source: Escabí-Mendoza J et al. Diagnostic accuracy of atrial fibrillation by computerized electrocardiogram analysis versus cardiologist interpretation. Journal of electrocardiology. 2025. PMID: 41172878. Registry ID: none in the record. MeSH: Humans · Atrial Fibrillation · Electrocardiography · Retrospective Studies · Male · Female · Diagnosis, Computer-Assisted · Sensitivity and Specificity · Cardiologists · Diagnostic Errors · Aged · Middle Aged · Reproducibility of Results

Card 3 — Automated interpretation misread 26.5% of confirmed STEMI ECGs (Kim 2022, N = 117)

Study: Primary accuracy study (automated interpretation in confirmed STEMI; n = 117 patients)

Population: A single-centre retrospective cohort of 117 adults with ST-elevation myocardial infarction who attended an emergency department between January 2017 and December 2020; the automated ECG interpretation was compared with the confirmed diagnosis, and misinterpretation was analysed by culprit artery and by the leads showing ST change.

Summary: Automated interpretation misread 26.5% of confirmed STEMI ECGs (Kim 2022, N = 117)

Result: The automated ECG misinterpretation rate in confirmed STEMI was 26.5% (31 of 117 patients). There was no significant association between misinterpretation and ST change in the anteroseptal, lateral, inferior or aVR lead groups, and no association with which culprit artery was occluded (P = 0.132); single-vessel occlusion was misread significantly more often than multi-vessel occlusion (27 of 86, 31.3%, vs 4 of 31, 12.9%; P = 0.045).

Watch out: A STEMI-only cohort: this measures the false-negative direction (missed or mislabelled STEMI statements) and says nothing about false-positive STEMI statements in non-STEMI ECGs, which this study did not sample.

Source: Kim S et al. Analysis of the accuracy of automatic electrocardiogram interpretation in ST-segment elevation myocardial infarction. Clinical and experimental emergency medicine. 2022. PMID: 35354230. Registry ID: none in the record. MeSH:

Card 4 — Computer ST-segment measurement agreed only moderately with a cardiologist in V2, with clinically inadequate limits of agreement for ST deviation (Eskola 2005, N = 69)

Study: Primary accuracy study (computer vs manual ECG measurement in suspected ACS; n = 69)

Population: Sixty-nine consecutive patients with suspected acute coronary syndrome had lead-specific computer-based and manual (cardiologist) measurements of J-point, ST-segment and T-wave deviations on standard 12-lead ECGs (excluding aVR), compared by kappa statistics and Bland–Altman analysis against cardiologist-defined ST elevation thresholds.

Summary: Computer ST-segment measurement agreed only moderately with a cardiologist in V2, with clinically inadequate limits of agreement for ST deviation (Eskola 2005, N = 69)

Result: Interobserver reliability between human readers for ST deviation ≥0.2 mV was very good (κ = 0.94 in V2, 0.93 in V3), but agreement between cardiologist and computer on ST elevation ≥0.2 mV was only moderate in V2 (κ = 0.72) and very good in V3 (κ = 0.85); for ST depression or elevation ≥0.05 mV in lead III, agreement was good and moderate (κ = 0.79 and 0.51). Bland–Altman analysis showed clinically acceptable limits of agreement for J-point and T-wave measurements but clinically inadequate limits for ST-segment deviation, with automatically measured deviations systematically smaller; a computer cut-off of 0.115 mV detected cardiologist-defined ≥0.2 mV elevation at sensitivity 89% / specificity 98%.

Watch out: Sixty-nine patients — N printed on the face of this card — and 2005-era algorithms. This is the measurement layer beneath the statement layer: a computer can misstate STEMI because it first mis-measures the ST segment, systematically reading deviations smaller than a cardiologist does.

Source: Eskola MJ et al. Comparative accuracy of manual versus computerized electrocardiographic measurement of J-, ST- and T-wave deviations in patients with acute coronary syndrome. The American journal of cardiology. 2005. PMID: 16310446. Registry ID: none in the record. MeSH: Acute Disease · Coronary Disease · Electrocardiography · Electronic Data Processing · Humans · Observer Variation · Reproducibility of Results · Retrospective Studies · Syndrome